CooperSurgical Inc
Fetal and Vascular Dopplers
Tria II TR150 and TR250 series Directions for Use Rev A March 2013
Directions for Use
20 Pages
Preview
Page 1
Explanation of Symbols REF
Reorder Number
SN
Serial Number Latex Free ATTENTION: See instructions for use.
®
Keep Dry Type B Applied Part
Class II Equipment
Fetal and Vascular Dopplers User Manual for TR150 Series Units (Non-Display) and TR250 Series Units (Display)
Manufacturer Date of Manufacture Symbol indicates that the device should be sent to the special agencies according to local regulations for separate collection after its useful life. Product conforms to the Medical Device Directive 93/42/EEC Authorized Representative in the European Community.
TRIA®II and MedaSonics® are registered trademarks of CooperSurgical, Inc. Clorox® is a registered trademark of The Clorox Company. CooperSurgical is a registered trademark of CooperSurgical, Inc. © 2013 CooperSurgical, Inc.
Made in the USA 95 Corporate Drive • Trumbull, CT 06611 USA Phone: (800) 243-2974 Fax: (800) 262-0105 www.coopersurgical.com
Non-Display Models with Probe
Display Models with Probe
T334 T335 T336 T337 T337-S T345
T334-D T335-D T336-D T337-D T337-DS T345-D
International Phone: (203) 601-9818 Fax: (203) 601-4747
MAN0035-DFU • Rev. A • 3/13
MAN0035-DFU • Rev. A • 3/13
Thank you for choosing TRIA® II from CooperSurgical, Inc. MedaSonics® handheld Dopplers have been a standard of care for 30 years. Your total satisfaction is our highest priority as we strive to continually improve our products and services. Please contact us with any suggestions. We look forward to enjoying a long-term relationship with you! CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611 USA
Year of manufacture located on the device.
Warranty and Servicing Policy This warranty is in lieu of all other warranties, expressed or implied. CooperSurgical warrants this TRIA® II Doppler against defects in material or workmanship as follows: Parts CooperSurgical will repair or replace parts in exchange for defective parts for the specified life of the device of five (5) years after date of original purchase with no charge for the parts. CooperSurgical reserves the right to determine whether a part should be repaired or replaced. Labor CooperSurgical will repair this instrument with no labor charge for a period of one (1) year after the date of original shipment when it is shipped to the CooperSurgical Service Center. Transportation and insurance to the Service Center are to be prepaid by the purchaser.
Here’s how you can reach us… Phone: 1-800-243-2974 (203) 601-5200 Fax: (203) 601-4740 e-mail us at: [email protected] visit our website at: www.coopersurgical.com
Table of Contents: Page Intended Use/Contraindications/Warnings ... 1 Description of Product ... 3 Operation and Installation ... 5 Obtaining Doppler Signals... 8 Maintenance and Cleaning... 10 Replacing Batteries ... 11 Troubleshooting ... 12 Clinical References ... 13 Accessories ... 13 Specifications... 14 Warranty and Servicing Policy ... 17 Explanation of Symbols ...Back Cover Please read the manual carefully and become familiar with the operation, features and maintenance of your Doppler prior to using the device or accessories.
Limitations - This product is to be used only for the intended purpose and labeled indications presented in the literature accompanying the product. This warranty shall not apply to any products repaired or altered by anyone not authorized in writing by CooperSurgical. CooperSurgical makes no warranty of the results to be obtained. CooperSurgical’s sole responsibility shall be to replace or repair this product under the terms stated above and does not cover loss or damage from external causes such as, but not limited to, weather, theft or abuse. This warranty does not apply to batteries, shipping case or other accessories, or to damage from shipping, tampering, misuse or negligence. Preventive maintenance, the refinishing or replacement of any cosmetic defect or deterioration, or the replacement of batteries unless damaged by a component failure, is not covered by this warranty. CooperSurgical will not be responsible for any loss, damage or injury resulting from delay in rendering service under this warranty. COOPERSURGICAL SHALL NOT BE RESPONSIBLE FOR ANY SECONDARY CHARGES OR CONSEQUENTIAL DAMAGES FROM ANY BREACH OF ANY WARRANTY, EXPRESSED OR IMPLIED. Since some states do not allow the exclusion or limitation of consequential damages, some of this limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state. To return products to CooperSurgical: 1. Call CooperSurgical to obtain a Return Authorization and to receive any final instructions prior to shipping 2. Clean the product prior to shipping 3. Ensure the device is well-packaged and suitable for shipment 4. Send the product to: Service Center CooperSurgical, Inc. 95 Corporate Drive
Trumbull, CT 06611 USA If you should have questions, please call 1-800-243-2974 for Customer Service, 1-800-444-8456 for Technical Support or (203) 601-9818 for International calls.
i
MAN0035-DFU • Rev. A • 3/13
17
MAN0035-DFU • Rev. A • 3/13
Obstetrical Probe Information:
ISATA (max) (mW/cm2) Po (mW) Effective Radiating Area (cm2) Ultrasound Frequency (MHz) Pulse Duration Repetition Freq. ISATA ISPTA.3 ISPPA.3 MI Pr.3 Wo fc zsp x-6, y-6 EBD
Intended Use Model Number TSP-2 TSP-3 TSP-2W 19.6 17.0 48.0 24.4 2.45 1.57 2.1 MHz 3.2 MHz CW CW CW CW
the spatial-average temporal-average intensity (mwatts per cm2). the derated spatial-peak temporal-average intensity (mwatts per cm2). the derated spatial-peak pulse-average intensity (watts per cm2). the Mechanical Index. the peak rarefactional pressure (megapascals) associated with the transmit pattern giving rise to the value reported for MI. the total time-average ultrasonic power (mwatts). the probe center frequency (MHz). the axial distance at which the reported parameter is measured (cm). are the –6dB beam dim. in the x-y plane where zsp is found (cm). the entrance beam dimensions (cm). These dimensions are the same as the dimensions of the transmit crystal.
Measurement Uncertainties:
Power: Pressure: Intensity (Ispta): Frequency:
+34, -42% +11, -16% +23, -26% ±5%
Acoustic Output Parameters are measured in water. Derated values, denoted by the subscript “.3”, take into account a conservative level of attenuation that would be encountered in the human body. The derated intensity values (I.3) are obtained from water values of intensity (Iw) at a depth of z calculated by: I.3 = exp(-0.23*0.3*f*z)*Iw (where f is the probe frequency in MHz and z is the depth in centimeters) The derated peak rarefactional pressure is calculated from the value of measure water (pr) by: Pr.3 = exp(-0.115*0.3*f*z)*pr (where pressure is given in megapascals) Additional Output Reporting Information for IEC 61157 5 MHz: Iob < 51 mW/cm2 8 MHz: Iob < 47 mW/cm2 The 2 MHz, 2 MHz WP and 3 MHz obstetrical probes are exempt from the declaration requirements of IEC61157. These probes meet the conditions: Iob < 20 mW/cm2, Ispta < 100 mW/cm2, and Pr < 1 MPa. Iob is output power divided by beam area. Note that parameter Zsp in the probe reporting tables is the same parameter as Ip in IEC 61157.
16
MAN0035-DFU • Rev. A • 3/13
Obstetric (2 and 3 MHz Probes) This product will be used to detect fetal heart beats as an aid for determining fetal viability. Vascular (5 and 8 MHz Probes) This product will be used to detect blood flow in veins and arteries for assisting in the detection of peripheral vascular disease. Caution • U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner.
Contraindications Warnings • The vascular probes (5 and 8 MHz) are not for fetal use. • The ultrasound probes are not to be used on or near the eyes. • The device is for use only on intact skin. • Do not plug any part of this device into a telephone or modem system. • This device is not intended for use with HF surgical equipment. If there are questions or concerns regarding these warnings or contraindications, please do not hesitate to contact CooperSurgical for further clarification. Caution • Dropping the TRIA® II, probe or accessories may cause damage to the housing or electronics. In order to preserve, protect and improve the quality of the environment, protect human health and utilize natural resources prudently and rationally – do not dispose of waste electrical or electronic equipment (WEEE) as unsorted municipal waste. Contact local WEEE disposal sites. Safety of Ultrasound The TRIA® II was designed with physician and patient safety in mind. In early design phases all potential hazards were eliminated or reduced to As Low As Reasonably Achievable (ALARA) by adhering to good design practices and industry wide safety standards. Ultrasound procedures should be performed with the ALARA principle in mind when delivering ultrasound energy into the body.
1
MAN0035-DFU • Rev. A • 3/13
The following official statements from the American Institute of Ultrasound Medicine (AIUM) are provided for your general information regarding the safe use of ultrasound.
Transducer Model: TSP-5 Application(s): Peripheral Vascular
Clinical Safety Approved March 1997, October 1982 Diagnostic ultrasound has been in use since the late 1950s. Given its known benefits and recognized efficacy for medical diagnosis, including use during human pregnancy, the American Institute of Ultrasound in Medicine herein addresses the clinical safety of such use:
Global Maximum Value Pr.3
ACOUSTIC OUTPUT
Associated Acoustic Parameter
MI (Mpa)
0.04 0.09
ISPTA.3 2 (mW/cm ) 223
ISPPA.3 2 (W/cm ) 0.22
12.8
0.013
wo
(mW)
fc
(MHz)
5.3
5.3
5.3
zsp
(cm)
0.85
0.85
0.85
Beam Dimensions
There are no confirmed biological effects on patients or instrument operators caused by exposures from present diagnostic ultrasound instruments. Although the possibility exists that such biological effects may be identified in the future, current data indicate that the benefits to patients of the prudent use of diagnostic ultrasound outweigh the risks, if any, that may be present. Prudent Use Approved May 1999 The AIUM advocates the responsible use of diagnostic ultrasound. The AIUM strongly discourages the non-medical use of ultrasound for psychosocial or entertainment purposes. The use of either two-dimensional (2D) or three-dimensional (3D) ultrasound to only view the fetus, obtain a picture of the fetus or determine the fetal gender without a medical indication is inappropriate and contrary to responsible medical practice. Although there are no confirmed biological effects on patients caused by exposures from present diagnostic ultrasound instruments, the possibility exists that such biological effects may be identified in the future. Thus ultrasound should be used in a prudent manner to provide medical benefit to the patient.
Operating Mode: Continuous-Wave (cw)
EBD
x-6
(cm)
0.4
0.4
y-6
(cm)
0.6
0.6
Az (cm)
0.4
Ele. (cm)
0.8
Transducer Model: TSP-8 Operating Mode: Continuous-Wave (cw) Application(s): Peripheral Vascular
ACOUSTIC OUTPUT Global Maximum Value Pr.3 Associated Acoustic Parameter
MI (Mpa)
0.03 0.09
ISPTA.3 2 (mW/cm ) 229
ISPPA.3 2 (W/cm ) 0.23
13.9
0.014
wo
(mW)
fc
(MHz)
8.0
8.0
8.0
zsp
(cm)
0.66
0.66
0.66
Beam Dimensions EBD
x-6
(cm)
0.2
0.2
y-6
(cm)
0.4
0.4
Az (cm)
0.3
Ele. (cm)
0.6
Safety in Training and Research Approved March 1997, March 1983 Diagnostic ultrasound has been in use since the late 1950s. There are no confirmed adverse biological effects on patients resulting from this usage. Although no hazard has been identified that would preclude the prudent and conservative use of diagnostic ultrasound in education and research, experience from normal diagnostic practice may or may not be relevant to extended exposure times and altered exposure conditions. It is therefore considered appropriate to make the following recommendation: In those special situations in which examinations are to be carried out for purposes other than direct medical benefit to the individual being examined, the subject should be informed of the anticipated exposure conditions, and of how these compare with conditions for normal diagnostic practice.
2
MAN0035-DFU • Rev. A • 3/13
15
MAN0035-DFU • Rev. A • 3/13
Specifications
Description of Product
Degree of protection against electric shock:
The MedaSonics TRIA® II Doppler includes user interchangeable ultrasound transducers to meet your specific needs. Other features include:
Type B Applied part Class II Equipment Degree of protection against ingress of water: 2 MHz Waterproof: IPX7 – entire probe and cable excluding connector All other probes: IPX4 – extending 2.5 cm from tip IPX1 – entire probe 2.5 cm from tip, excluding connector Designed and tested to meet: IEC601-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-37, EN5011-A Dimensions (h x w x l): Weight:
140 x 70 x 35 mm 320 grams
Operating temperature: 10 to 40 °C Operating humidity: 30 to 75 % Transport/Storage temperature: –20 to 50 °C Transport/Storage humidity: 5 to 90%, non-condensing Battery type and voltage
3 – AA Alkaline 1.5 volt (for non-rechargeable units) 3 – AA NiMH 1.5 volt (for rechargeable units) (as applicable)
Battery life:
Batteries provided with the TRIA® II: 1250, 1-minute exams
Audio bandwidth and power:
350 Hz – 2 KHz, 0.5 W
Main Unit The main handheld unit is ergonomically designed to fit the palm of your hand with comfort and allow easy access to each control feature. Each unit is individually tested and inspected to ensure the highest quality standards. SSQ – Superior Sound Quality. Every TRIA® II Doppler is designed with a state of the art sound system that produces excellent sound quality and longterm reliability. LCD Display – The LCD display (TR250 only) allows you to view the fetal heart rate in large, easy to read digits, monitor battery life, observe signal strength indicators, and view multiple diagnostic indicators to ensure your unit is functioning at peak performance levels. Non-display units (TR150 models) incorporate LED indicators for battery status.
Heart rate calculation accuracy: ±3 BPM over range 50 to 220 BPM Operating Conditions: There are no user controls which affect the ultrasound output. Attention: Consult Accompanying Documents
14
MAN0035-DFU • Rev. A • 3/13
3
MAN0035-DFU • Rev. A • 3/13
TRIA II ultrasound transducers were designed to meet your specific applications needs. Each probe has been ergonomically designed for comfort while providing excellent maneuverability for locating the fetus or vascular target.
Radio Frequency Interference The TRIA® II was tested for immunity to electromagnetic interference at a level of 3V/meter. Interference during normal operation may occur in the presence of fields stronger than 3V/meter. If this occurs, try to increase the distance between the TRIA® II and the source of interference. Contact CooperSurgical for more information.
Each probe is carefully measured and tested to ensure it meets exacting performance standards.
Diagnostic Codes – Contact CooperSurgical’s Service Dept. at 1-800-444-8456
2 MHz – Late term obstetrical examination. This probe frequency is typically used during the last trimester for deep fetal positions associated with larger women. Waterproof versions of the 2 MHz probe are available.
1 – Temperature too low 2 – Temperature too high 3 – Reference Voltage too low 4 – Reference Voltage too high
Probes ®
3 MHz – Early and general-purpose obstetrical examination. This probe frequency is a general use model ideal for most stages of fetal examination. Fetal heart sounds can be heard as early as 12 weeks and sometimes sooner depending on the position and size of the fetus. 5 MHz – This standard “pencil” style probe is an excellent vascular tool for locating deep specific vessels in the peripheral vascular system. The narrow grip and small face of the probe make it ideal for maneuvering for maximizing the signal. 8 MHz – This standard “pencil” style probe is an excellent vascular tool for locating shallow specific vessels in the peripheral vascular system. The narrow grip and small face of the probe make it ideal for maneuvering for maximizing the signal. Sterilizable versions of the 8 MHz probe are available.
5 – 5 Volt Supply too low 6 – 5 Volt Supply too high 7 – Battery Voltage too low 8 – Battery Voltage too high
Reference materials for Obstetrical and Peripheral Vascular testing: Handbook of Fetal Heart Rate Monitoring; Julian T. Parer, 1997 Doppler Ultrasound and Its Use In Clinical Measurement; Peter Atkinson and John P. Woodcock, 1982 Noninvasive Diagnosis of Peripheral Vascular Disease; W. Robert Felix, Jr., 1988 Current Noninvasive Vascular Diagnosis; Ali F. Aburahma, Edward B. Diethrich, 1988
Accessories To order accessories (gel, batteries, probe cleaner, cables), contact CooperSurgical’s Service Department at 1-800-645-3760 or (203) 601-5200 to order by phone.
4
MAN0035-DFU • Rev. A • 3/13
13
MAN0035-DFU • Rev. A • 3/13
Troubleshooting
Operation and Installation
Warning • Use alternate equipment in case of unit failure. Call CooperSurgical’s Service Department if the probe or main unit malfunctions at 1-800-444-8456.
Probe Clip On/Off Button
Caution • Do not drop or mishandle the TRIA® II unit or probes. Damage to sensitive electrical components and transducers is likely to occur. Doppler Cable Connector
Poor sound quality Inadequate gel use Try and relocate the probe for a better signal – refer to Signal Strength Indicator Improper choice of probe Frequency Interference from other equipment Probe interconnect cable or battery contacts may be intermittent Debris in the speaker may cause poor sound Device damage from dropping the TRIA® II, probes or accessories
Head Phone Connector
Heart Rate inaccurate Try and relocate the probe for a better signal – refer to Signal Strength Indicator For OB, ensure maternal sounds are not mixing with fetal sounds Ensure by manual counting that the rate is between 50 and 220 BPM Battery indicator flashing Consult the Battery Monitoring section; replace batteries as described in the Replacing Batteries section.
Battery Compartment Door
Probe frequency does not match the connected probe Check probe that is attached to ensure it is the correct one, or check to see if probe is attached. If the correct probe is being used, contact CooperSurgical Service Department at 1-800-444-8456.
Belt Clip
Error 5 or 7 Batteries are low and require replacement.
12
Volume Slider
MAN0035-DFU • Rev. A • 3/13
5
MAN0035-DFU • Rev. A • 3/13
Replacing Batteries Warning • The battery compartment only accepts AA size batteries. See the Accessories section for reordering of parts information. Open the battery compartment by depressing the tab and pulling outward on the battery door. Remove the existing drained batteries by pushing on the end of the battery that compresses the battery contact spring and lift upwards. It is acceptable to carefully use a simple tool, such as a pen, to assist in lifting the batteries out.
LCD Panel for TR250 Display Units Only
Turning Unit On/Off Turn the unit on by pressing the On/Off button. LED or LCD indicators (depending on the model) indicate power status.
Batteries slide into compartment
The TRIA® II automatically shuts itself off after 3 minutes if it is not being used. This complete power shutdown preserves the life of the batteries and ensures the unit will be ready for operation in case it was accidentally left on. Diagnostic Monitoring (LCD Display units only) Once the unit is on, the TRIA® II TR250 performs a series of diagnostic checks. The unit first checks and temporarily displays the frequency of the probe that is being used. This display will not reappear unless the probe is changed or the power is cycled, in which case the display will again temporarily confirm the frequency of probe that is connected. The unit then checks for proper internal operating temperature, battery voltage, reference voltage and power supply voltage levels. If any of these characteristics are out of range, the display will show the ERROR indicator and a failure code associated with the diagnostic error. Diagnostic functions are periodically checked while the unit is on to ensure the Doppler is operating at peak performance. Refer to the Troubleshooting section for a listing of failure codes.
6
MAN0035-DFU • Rev. A • 3/13
Replace the batteries by paying close attention to the polarity indicators on the battery and the polarity indicators on the battery holder in the compartment. Positive (+) aligns with positive (button) and negative (-) aligns with negative (spring). Insert the battery such that the spring contacts are loaded first and then press the battery firmly into place. After all three have been inserted, replace the battery door. Warning • If the batteries have been inserted incorrectly, the unit will not function but the TRIA® II will not be damaged.
11
MAN0035-DFU • Rev. A • 3/13
Maintenance and Cleaning
Battery Monitoring
Warnings • The TRIA® II is not designed for liquid immersion. Do not soak the Doppler main unit or probes in liquids. Use only spray or wipe cleaners and disinfectants. Do not use products containing bleach.
All TRIA® II Doppler units perform continuous battery monitoring and give a visual indication of battery level. Display units use a multiple level battery shaped icon that indicates the voltage level of the battery. The battery outline will flash when the battery level is very low indicating that the user should change the batteries soon after the current examination is complete.
•
The TRIA® II is not designed for sterilization processes such as autoclaving, gamma radiation or hydrogen gas.
•
The TRIA® II is not intended to be used on open skin. If there is evidence of open wound contamination, disinfect the probe before using again as described below.
Non-display units (TR150) use the On/Off LED as a battery indicator by flashing at a low rate (approximately once per second) when the battery level is low. Several exams can still be performed in this state of operation. The battery indicator will flash at a higher rate (approximately twice per second) when the battery level is very low. The user should change the batteries soon after the current exam is complete.
The TRIA® II Doppler requires very little maintenance. However, it is important to the continuing function of the unit and the health of the patients that the unit is cleaned and examined regularly per the following guideline: After every examination: Excess gel should be wiped off prior to docking the probe. Probes and main unit should be cleaned with a damp warm water cloth or presaturated isopropyl alcohol wipes or spray. In particular, pay close attention to clean the seams along the plastic lines at the probe face but do not allow water or spray to enter through the connectors or speaker grill. To disinfect unit, use commercially available spray or wipe disinfectants registered with the EPA. Clorox® Broad Spectrum Quaternary Disinfectant is the only disinfectant that is CooperSurgical approved for use with the TRIA® II products. Follow the manufacturer’s instructions and wipe unit until it is dry of solutions. Examiners should wash hands and change gloves after every exam. Refer to local and hospital policies for cleaning and disinfection policies. Store unit in a clean area free from dust and debris. Follow temperature and humidity guidelines as specified at the end of this manual. Warning • If the unit is to be stored for longer than 90 days without use, remove the batteries prior to storage. Periodically (at least annually): Inspect the main unit and probes for signs of cracks or breaks in the mechanical housing. Inspect cables and connectors for signs of wear or failure. The user should discontinue use of the unit with any sign of loss of housing integrity. Contact CooperSurgical for service. It is recommended that rechargeable batteries be replaced annually.
10
MAN0035-DFU • Rev. A • 3/13
7
MAN0035-DFU • Rev. A • 3/13
Obtaining Doppler Signals Caution • Doppler examinations should be performed only by trained individuals. For any Doppler examination, it is essential that an adequate supply of coupling gel is used to transmit the ultrasound energy from the probe to the surface of the skin. Re-apply more gel if it starts to dry out or spread so thinly that an air gap occurs between the probe and the skin. It is not necessary to cover the entire surface of the probe, only the probe face. Applying too much gel makes the unit difficult to clean and does not aid in the performance of the probe. Volume Slider The audio level can be adjusted using the Volume Slider. Moving the slider up will increase the volume, while moving it down will decrease it. Signal Quality Indicator (TR250 only) An inadequate signal can produce erroneous rates from the heart rate calculation. The signal level that is being obtained is shown on the Signal Quality Indicator bars. This indicator provides a visual aid in obtaining a strong audio signal by showing the pulsatile nature of the signal. A large difference between the highest and lowest signal bars that are lit confirms that the quality of the signal is good and thus ensures the heart rate calculation is operating at peak performance.
Many times when attempting to detect the fetal heart, the maternal vascular sounds are heard instead of (or in some cases, in addition to) the fetal sounds. These maternal sounds can come from one of the major arteries, the placenta or the umbilical cord. The maternal vascular sounds are typically higher in frequency at a lower rate. The heart rate calculation will display either the maternal rate or the fetal rate, whichever portion of the signal is stronger. If the fetal heart sounds cannot be located using the procedure as described above, a second exam should be performed using another commercially available fetal monitor as a repeated test. 5 and 8 MHz Vascular Peripheral arterial sounds are typically higher in frequency. For the best sounds, angle the probe approximately 45 degrees from the skin surface over the general location of the vessel. Slowly move the probe side to side and vary the angle of the probe until the vascular sounds are heard. Changing the angle of the probe has an effect on the frequency of the sound. The steeper the probe angle is, the higher the frequency of the sound. Peripheral venous sounds are not typically periodic and vary greatly depending on patient movement and breathing. These sounds are more like the wind at the ocean and vary in pitch as the patient moves or breathes.
The heart rate can be verified manually by counting the audible beats for 20 seconds and multiplying by 3, or for 15 seconds and then multiplying by 4. Counting for less than 15 seconds is not recommended due to a decrease in accuracy with the small sample size. Obstetrical Fetal heart sounds are quite different from peripheral vascular blood flow sounds. Fetal sounds are typically much lower in frequency and much higher in rate. For early term fetal detection, start the probe at the pubic bone and slowly move along the midline – rocking the probe slowly from side to side until a heart beat is heard. For mid to late term fetal detection the best chance of finding the heart sounds are to start on the fundus and move toward the navel and from one side of the abdomen to the other, slowly rocking the probe until the heart beat is heard. The fetal heart reminds many people of a galloping horse and can vary in tone from a distant swishing sound to a hard clopping sound depending on the position of the baby and probe.
8
MAN0035-DFU • Rev. A • 3/13
9
MAN0035-DFU • Rev. A • 3/13
Obtaining Doppler Signals Caution • Doppler examinations should be performed only by trained individuals. For any Doppler examination, it is essential that an adequate supply of coupling gel is used to transmit the ultrasound energy from the probe to the surface of the skin. Re-apply more gel if it starts to dry out or spread so thinly that an air gap occurs between the probe and the skin. It is not necessary to cover the entire surface of the probe, only the probe face. Applying too much gel makes the unit difficult to clean and does not aid in the performance of the probe. Volume Slider The audio level can be adjusted using the Volume Slider. Moving the slider up will increase the volume, while moving it down will decrease it. Signal Quality Indicator (TR250 only) An inadequate signal can produce erroneous rates from the heart rate calculation. The signal level that is being obtained is shown on the Signal Quality Indicator bars. This indicator provides a visual aid in obtaining a strong audio signal by showing the pulsatile nature of the signal. A large difference between the highest and lowest signal bars that are lit confirms that the quality of the signal is good and thus ensures the heart rate calculation is operating at peak performance.
Many times when attempting to detect the fetal heart, the maternal vascular sounds are heard instead of (or in some cases, in addition to) the fetal sounds. These maternal sounds can come from one of the major arteries, the placenta or the umbilical cord. The maternal vascular sounds are typically higher in frequency at a lower rate. The heart rate calculation will display either the maternal rate or the fetal rate, whichever portion of the signal is stronger. If the fetal heart sounds cannot be located using the procedure as described above, a second exam should be performed using another commercially available fetal monitor as a repeated test. 5 and 8 MHz Vascular Peripheral arterial sounds are typically higher in frequency. For the best sounds, angle the probe approximately 45 degrees from the skin surface over the general location of the vessel. Slowly move the probe side to side and vary the angle of the probe until the vascular sounds are heard. Changing the angle of the probe has an effect on the frequency of the sound. The steeper the probe angle is, the higher the frequency of the sound. Peripheral venous sounds are not typically periodic and vary greatly depending on patient movement and breathing. These sounds are more like the wind at the ocean and vary in pitch as the patient moves or breathes.
The heart rate can be verified manually by counting the audible beats for 20 seconds and multiplying by 3, or for 15 seconds and then multiplying by 4. Counting for less than 15 seconds is not recommended due to a decrease in accuracy with the small sample size. Obstetrical Fetal heart sounds are quite different from peripheral vascular blood flow sounds. Fetal sounds are typically much lower in frequency and much higher in rate. For early term fetal detection, start the probe at the pubic bone and slowly move along the midline – rocking the probe slowly from side to side until a heart beat is heard. For mid to late term fetal detection the best chance of finding the heart sounds are to start on the fundus and move toward the navel and from one side of the abdomen to the other, slowly rocking the probe until the heart beat is heard. The fetal heart reminds many people of a galloping horse and can vary in tone from a distant swishing sound to a hard clopping sound depending on the position of the baby and probe.
8
MAN0035-DFU • Rev. A • 3/13
9
MAN0035-DFU • Rev. A • 3/13
Maintenance and Cleaning
Battery Monitoring
Warnings • The TRIA® II is not designed for liquid immersion. Do not soak the Doppler main unit or probes in liquids. Use only spray or wipe cleaners and disinfectants. Do not use products containing bleach.
All TRIA® II Doppler units perform continuous battery monitoring and give a visual indication of battery level. Display units use a multiple level battery shaped icon that indicates the voltage level of the battery. The battery outline will flash when the battery level is very low indicating that the user should change the batteries soon after the current examination is complete.
•
The TRIA® II is not designed for sterilization processes such as autoclaving, gamma radiation or hydrogen gas.
•
The TRIA® II is not intended to be used on open skin. If there is evidence of open wound contamination, disinfect the probe before using again as described below.
Non-display units (TR150) use the On/Off LED as a battery indicator by flashing at a low rate (approximately once per second) when the battery level is low. Several exams can still be performed in this state of operation. The battery indicator will flash at a higher rate (approximately twice per second) when the battery level is very low. The user should change the batteries soon after the current exam is complete.
The TRIA® II Doppler requires very little maintenance. However, it is important to the continuing function of the unit and the health of the patients that the unit is cleaned and examined regularly per the following guideline: After every examination: Excess gel should be wiped off prior to docking the probe. Probes and main unit should be cleaned with a damp warm water cloth or presaturated isopropyl alcohol wipes or spray. In particular, pay close attention to clean the seams along the plastic lines at the probe face but do not allow water or spray to enter through the connectors or speaker grill. To disinfect unit, use commercially available spray or wipe disinfectants registered with the EPA. Clorox® Broad Spectrum Quaternary Disinfectant is the only disinfectant that is CooperSurgical approved for use with the TRIA® II products. Follow the manufacturer’s instructions and wipe unit until it is dry of solutions. Examiners should wash hands and change gloves after every exam. Refer to local and hospital policies for cleaning and disinfection policies. Store unit in a clean area free from dust and debris. Follow temperature and humidity guidelines as specified at the end of this manual. Warning • If the unit is to be stored for longer than 90 days without use, remove the batteries prior to storage. Periodically (at least annually): Inspect the main unit and probes for signs of cracks or breaks in the mechanical housing. Inspect cables and connectors for signs of wear or failure. The user should discontinue use of the unit with any sign of loss of housing integrity. Contact CooperSurgical for service. It is recommended that rechargeable batteries be replaced annually.
10
MAN0035-DFU • Rev. A • 3/13
7
MAN0035-DFU • Rev. A • 3/13
Replacing Batteries Warning • The battery compartment only accepts AA size batteries. See the Accessories section for reordering of parts information. Open the battery compartment by depressing the tab and pulling outward on the battery door. Remove the existing drained batteries by pushing on the end of the battery that compresses the battery contact spring and lift upwards. It is acceptable to carefully use a simple tool, such as a pen, to assist in lifting the batteries out.
LCD Panel for TR250 Display Units Only
Turning Unit On/Off Turn the unit on by pressing the On/Off button. LED or LCD indicators (depending on the model) indicate power status.
Batteries slide into compartment
The TRIA® II automatically shuts itself off after 3 minutes if it is not being used. This complete power shutdown preserves the life of the batteries and ensures the unit will be ready for operation in case it was accidentally left on. Diagnostic Monitoring (LCD Display units only) Once the unit is on, the TRIA® II TR250 performs a series of diagnostic checks. The unit first checks and temporarily displays the frequency of the probe that is being used. This display will not reappear unless the probe is changed or the power is cycled, in which case the display will again temporarily confirm the frequency of probe that is connected. The unit then checks for proper internal operating temperature, battery voltage, reference voltage and power supply voltage levels. If any of these characteristics are out of range, the display will show the ERROR indicator and a failure code associated with the diagnostic error. Diagnostic functions are periodically checked while the unit is on to ensure the Doppler is operating at peak performance. Refer to the Troubleshooting section for a listing of failure codes.
6
MAN0035-DFU • Rev. A • 3/13
Replace the batteries by paying close attention to the polarity indicators on the battery and the polarity indicators on the battery holder in the compartment. Positive (+) aligns with positive (button) and negative (-) aligns with negative (spring). Insert the battery such that the spring contacts are loaded first and then press the battery firmly into place. After all three have been inserted, replace the battery door. Warning • If the batteries have been inserted incorrectly, the unit will not function but the TRIA® II will not be damaged.
11
MAN0035-DFU • Rev. A • 3/13
Troubleshooting
Operation and Installation
Warning • Use alternate equipment in case of unit failure. Call CooperSurgical’s Service Department if the probe or main unit malfunctions at 1-800-444-8456.
Probe Clip On/Off Button
Caution • Do not drop or mishandle the TRIA® II unit or probes. Damage to sensitive electrical components and transducers is likely to occur. Doppler Cable Connector
Poor sound quality Inadequate gel use Try and relocate the probe for a better signal – refer to Signal Strength Indicator Improper choice of probe Frequency Interference from other equipment Probe interconnect cable or battery contacts may be intermittent Debris in the speaker may cause poor sound Device damage from dropping the TRIA® II, probes or accessories
Head Phone Connector
Heart Rate inaccurate Try and relocate the probe for a better signal – refer to Signal Strength Indicator For OB, ensure maternal sounds are not mixing with fetal sounds Ensure by manual counting that the rate is between 50 and 220 BPM Battery indicator flashing Consult the Battery Monitoring section; replace batteries as described in the Replacing Batteries section.
Battery Compartment Door
Probe frequency does not match the connected probe Check probe that is attached to ensure it is the correct one, or check to see if probe is attached. If the correct probe is being used, contact CooperSurgical Service Department at 1-800-444-8456.
Belt Clip
Error 5 or 7 Batteries are low and require replacement.
12
Volume Slider
MAN0035-DFU • Rev. A • 3/13
5
MAN0035-DFU • Rev. A • 3/13
TRIA II ultrasound transducers were designed to meet your specific applications needs. Each probe has been ergonomically designed for comfort while providing excellent maneuverability for locating the fetus or vascular target.
Radio Frequency Interference The TRIA® II was tested for immunity to electromagnetic interference at a level of 3V/meter. Interference during normal operation may occur in the presence of fields stronger than 3V/meter. If this occurs, try to increase the distance between the TRIA® II and the source of interference. Contact CooperSurgical for more information.
Each probe is carefully measured and tested to ensure it meets exacting performance standards.
Diagnostic Codes – Contact CooperSurgical’s Service Dept. at 1-800-444-8456
2 MHz – Late term obstetrical examination. This probe frequency is typically used during the last trimester for deep fetal positions associated with larger women. Waterproof versions of the 2 MHz probe are available.
1 – Temperature too low 2 – Temperature too high 3 – Reference Voltage too low 4 – Reference Voltage too high
Probes ®
3 MHz – Early and general-purpose obstetrical examination. This probe frequency is a general use model ideal for most stages of fetal examination. Fetal heart sounds can be heard as early as 12 weeks and sometimes sooner depending on the position and size of the fetus. 5 MHz – This standard “pencil” style probe is an excellent vascular tool for locating deep specific vessels in the peripheral vascular system. The narrow grip and small face of the probe make it ideal for maneuvering for maximizing the signal. 8 MHz – This standard “pencil” style probe is an excellent vascular tool for locating shallow specific vessels in the peripheral vascular system. The narrow grip and small face of the probe make it ideal for maneuvering for maximizing the signal. Sterilizable versions of the 8 MHz probe are available.
5 – 5 Volt Supply too low 6 – 5 Volt Supply too high 7 – Battery Voltage too low 8 – Battery Voltage too high
Reference materials for Obstetrical and Peripheral Vascular testing: Handbook of Fetal Heart Rate Monitoring; Julian T. Parer, 1997 Doppler Ultrasound and Its Use In Clinical Measurement; Peter Atkinson and John P. Woodcock, 1982 Noninvasive Diagnosis of Peripheral Vascular Disease; W. Robert Felix, Jr., 1988 Current Noninvasive Vascular Diagnosis; Ali F. Aburahma, Edward B. Diethrich, 1988
Accessories To order accessories (gel, batteries, probe cleaner, cables), contact CooperSurgical’s Service Department at 1-800-645-3760 or (203) 601-5200 to order by phone.
4
MAN0035-DFU • Rev. A • 3/13
13
MAN0035-DFU • Rev. A • 3/13
Specifications
Description of Product
Degree of protection against electric shock:
The MedaSonics TRIA® II Doppler includes user interchangeable ultrasound transducers to meet your specific needs. Other features include:
Type B Applied part Class II Equipment Degree of protection against ingress of water: 2 MHz Waterproof: IPX7 – entire probe and cable excluding connector All other probes: IPX4 – extending 2.5 cm from tip IPX1 – entire probe 2.5 cm from tip, excluding connector Designed and tested to meet: IEC601-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-37, EN5011-A Dimensions (h x w x l): Weight:
140 x 70 x 35 mm 320 grams
Operating temperature: 10 to 40 °C Operating humidity: 30 to 75 % Transport/Storage temperature: –20 to 50 °C Transport/Storage humidity: 5 to 90%, non-condensing Battery type and voltage
3 – AA Alkaline 1.5 volt (for non-rechargeable units) 3 – AA NiMH 1.5 volt (for rechargeable units) (as applicable)
Battery life:
Batteries provided with the TRIA® II: 1250, 1-minute exams
Audio bandwidth and power:
350 Hz – 2 KHz, 0.5 W
Main Unit The main handheld unit is ergonomically designed to fit the palm of your hand with comfort and allow easy access to each control feature. Each unit is individually tested and inspected to ensure the highest quality standards. SSQ – Superior Sound Quality. Every TRIA® II Doppler is designed with a state of the art sound system that produces excellent sound quality and longterm reliability. LCD Display – The LCD display (TR250 only) allows you to view the fetal heart rate in large, easy to read digits, monitor battery life, observe signal strength indicators, and view multiple diagnostic indicators to ensure your unit is functioning at peak performance levels. Non-display units (TR150 models) incorporate LED indicators for battery status.
Heart rate calculation accuracy: ±3 BPM over range 50 to 220 BPM Operating Conditions: There are no user controls which affect the ultrasound output. Attention: Consult Accompanying Documents
14
MAN0035-DFU • Rev. A • 3/13
3
MAN0035-DFU • Rev. A • 3/13
The following official statements from the American Institute of Ultrasound Medicine (AIUM) are provided for your general information regarding the safe use of ultrasound.
Transducer Model: TSP-5 Application(s): Peripheral Vascular
Clinical Safety Approved March 1997, October 1982 Diagnostic ultrasound has been in use since the late 1950s. Given its known benefits and recognized efficacy for medical diagnosis, including use during human pregnancy, the American Institute of Ultrasound in Medicine herein addresses the clinical safety of such use:
Global Maximum Value Pr.3
ACOUSTIC OUTPUT
Associated Acoustic Parameter
MI (Mpa)
0.04 0.09
ISPTA.3 2 (mW/cm ) 223
ISPPA.3 2 (W/cm ) 0.22
12.8
0.013
wo
(mW)
fc
(MHz)
5.3
5.3
5.3
zsp
(cm)
0.85
0.85
0.85
Beam Dimensions
There are no confirmed biological effects on patients or instrument operators caused by exposures from present diagnostic ultrasound instruments. Although the possibility exists that such biological effects may be identified in the future, current data indicate that the benefits to patients of the prudent use of diagnostic ultrasound outweigh the risks, if any, that may be present. Prudent Use Approved May 1999 The AIUM advocates the responsible use of diagnostic ultrasound. The AIUM strongly discourages the non-medical use of ultrasound for psychosocial or entertainment purposes. The use of either two-dimensional (2D) or three-dimensional (3D) ultrasound to only view the fetus, obtain a picture of the fetus or determine the fetal gender without a medical indication is inappropriate and contrary to responsible medical practice. Although there are no confirmed biological effects on patients caused by exposures from present diagnostic ultrasound instruments, the possibility exists that such biological effects may be identified in the future. Thus ultrasound should be used in a prudent manner to provide medical benefit to the patient.
Operating Mode: Continuous-Wave (cw)
EBD
x-6
(cm)
0.4
0.4
y-6
(cm)
0.6
0.6
Az (cm)
0.4
Ele. (cm)
0.8
Transducer Model: TSP-8 Operating Mode: Continuous-Wave (cw) Application(s): Peripheral Vascular
ACOUSTIC OUTPUT Global Maximum Value Pr.3 Associated Acoustic Parameter
MI (Mpa)
0.03 0.09
ISPTA.3 2 (mW/cm ) 229
ISPPA.3 2 (W/cm ) 0.23
13.9
0.014
wo
(mW)
fc
(MHz)
8.0
8.0
8.0
zsp
(cm)
0.66
0.66
0.66
Beam Dimensions EBD
x-6
(cm)
0.2
0.2
y-6
(cm)
0.4
0.4
Az (cm)
0.3
Ele. (cm)
0.6
Safety in Training and Research Approved March 1997, March 1983 Diagnostic ultrasound has been in use since the late 1950s. There are no confirmed adverse biological effects on patients resulting from this usage. Although no hazard has been identified that would preclude the prudent and conservative use of diagnostic ultrasound in education and research, experience from normal diagnostic practice may or may not be relevant to extended exposure times and altered exposure conditions. It is therefore considered appropriate to make the following recommendation: In those special situations in which examinations are to be carried out for purposes other than direct medical benefit to the individual being examined, the subject should be informed of the anticipated exposure conditions, and of how these compare with conditions for normal diagnostic practice.
2
MAN0035-DFU • Rev. A • 3/13
15
MAN0035-DFU • Rev. A • 3/13
Obstetrical Probe Information:
ISATA (max) (mW/cm2) Po (mW) Effective Radiating Area (cm2) Ultrasound Frequency (MHz) Pulse Duration Repetition Freq. ISATA ISPTA.3 ISPPA.3 MI Pr.3 Wo fc zsp x-6, y-6 EBD
Intended Use Model Number TSP-2 TSP-3 TSP-2W 19.6 17.0 48.0 24.4 2.45 1.57 2.1 MHz 3.2 MHz CW CW CW CW
the spatial-average temporal-average intensity (mwatts per cm2). the derated spatial-peak temporal-average intensity (mwatts per cm2). the derated spatial-peak pulse-average intensity (watts per cm2). the Mechanical Index. the peak rarefactional pressure (megapascals) associated with the transmit pattern giving rise to the value reported for MI. the total time-average ultrasonic power (mwatts). the probe center frequency (MHz). the axial distance at which the reported parameter is measured (cm). are the –6dB beam dim. in the x-y plane where zsp is found (cm). the entrance beam dimensions (cm). These dimensions are the same as the dimensions of the transmit crystal.
Measurement Uncertainties:
Power: Pressure: Intensity (Ispta): Frequency:
+34, -42% +11, -16% +23, -26% ±5%
Acoustic Output Parameters are measured in water. Derated values, denoted by the subscript “.3”, take into account a conservative level of attenuation that would be encountered in the human body. The derated intensity values (I.3) are obtained from water values of intensity (Iw) at a depth of z calculated by: I.3 = exp(-0.23*0.3*f*z)*Iw (where f is the probe frequency in MHz and z is the depth in centimeters) The derated peak rarefactional pressure is calculated from the value of measure water (pr) by: Pr.3 = exp(-0.115*0.3*f*z)*pr (where pressure is given in megapascals) Additional Output Reporting Information for IEC 61157 5 MHz: Iob < 51 mW/cm2 8 MHz: Iob < 47 mW/cm2 The 2 MHz, 2 MHz WP and 3 MHz obstetrical probes are exempt from the declaration requirements of IEC61157. These probes meet the conditions: Iob < 20 mW/cm2, Ispta < 100 mW/cm2, and Pr < 1 MPa. Iob is output power divided by beam area. Note that parameter Zsp in the probe reporting tables is the same parameter as Ip in IEC 61157.
16
MAN0035-DFU • Rev. A • 3/13
Obstetric (2 and 3 MHz Probes) This product will be used to detect fetal heart beats as an aid for determining fetal viability. Vascular (5 and 8 MHz Probes) This product will be used to detect blood flow in veins and arteries for assisting in the detection of peripheral vascular disease. Caution • U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner.
Contraindications Warnings • The vascular probes (5 and 8 MHz) are not for fetal use. • The ultrasound probes are not to be used on or near the eyes. • The device is for use only on intact skin. • Do not plug any part of this device into a telephone or modem system. • This device is not intended for use with HF surgical equipment. If there are questions or concerns regarding these warnings or contraindications, please do not hesitate to contact CooperSurgical for further clarification. Caution • Dropping the TRIA® II, probe or accessories may cause damage to the housing or electronics. In order to preserve, protect and improve the quality of the environment, protect human health and utilize natural resources prudently and rationally – do not dispose of waste electrical or electronic equipment (WEEE) as unsorted municipal waste. Contact local WEEE disposal sites. Safety of Ultrasound The TRIA® II was designed with physician and patient safety in mind. In early design phases all potential hazards were eliminated or reduced to As Low As Reasonably Achievable (ALARA) by adhering to good design practices and industry wide safety standards. Ultrasound procedures should be performed with the ALARA principle in mind when delivering ultrasound energy into the body.
1
MAN0035-DFU • Rev. A • 3/13
Thank you for choosing TRIA® II from CooperSurgical, Inc. MedaSonics® handheld Dopplers have been a standard of care for 30 years. Your total satisfaction is our highest priority as we strive to continually improve our products and services. Please contact us with any suggestions. We look forward to enjoying a long-term relationship with you! CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611 USA
Year of manufacture located on the device.
Warranty and Servicing Policy This warranty is in lieu of all other warranties, expressed or implied. CooperSurgical warrants this TRIA® II Doppler against defects in material or workmanship as follows: Parts CooperSurgical will repair or replace parts in exchange for defective parts for the specified life of the device of five (5) years after date of original purchase with no charge for the parts. CooperSurgical reserves the right to determine whether a part should be repaired or replaced. Labor CooperSurgical will repair this instrument with no labor charge for a period of one (1) year after the date of original shipment when it is shipped to the CooperSurgical Service Center. Transportation and insurance to the Service Center are to be prepaid by the purchaser.
Here’s how you can reach us… Phone: 1-800-243-2974 (203) 601-5200 Fax: (203) 601-4740 e-mail us at: [email protected] visit our website at: www.coopersurgical.com
Table of Contents: Page Intended Use/Contraindications/Warnings ... 1 Description of Product ... 3 Operation and Installation ... 5 Obtaining Doppler Signals... 8 Maintenance and Cleaning... 10 Replacing Batteries ... 11 Troubleshooting ... 12 Clinical References ... 13 Accessories ... 13 Specifications... 14 Warranty and Servicing Policy ... 17 Explanation of Symbols ...Back Cover Please read the manual carefully and become familiar with the operation, features and maintenance of your Doppler prior to using the device or accessories.
Limitations - This product is to be used only for the intended purpose and labeled indications presented in the literature accompanying the product. This warranty shall not apply to any products repaired or altered by anyone not authorized in writing by CooperSurgical. CooperSurgical makes no warranty of the results to be obtained. CooperSurgical’s sole responsibility shall be to replace or repair this product under the terms stated above and does not cover loss or damage from external causes such as, but not limited to, weather, theft or abuse. This warranty does not apply to batteries, shipping case or other accessories, or to damage from shipping, tampering, misuse or negligence. Preventive maintenance, the refinishing or replacement of any cosmetic defect or deterioration, or the replacement of batteries unless damaged by a component failure, is not covered by this warranty. CooperSurgical will not be responsible for any loss, damage or injury resulting from delay in rendering service under this warranty. COOPERSURGICAL SHALL NOT BE RESPONSIBLE FOR ANY SECONDARY CHARGES OR CONSEQUENTIAL DAMAGES FROM ANY BREACH OF ANY WARRANTY, EXPRESSED OR IMPLIED. Since some states do not allow the exclusion or limitation of consequential damages, some of this limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state. To return products to CooperSurgical: 1. Call CooperSurgical to obtain a Return Authorization and to receive any final instructions prior to shipping 2. Clean the product prior to shipping 3. Ensure the device is well-packaged and suitable for shipment 4. Send the product to: Service Center CooperSurgical, Inc. 95 Corporate Drive
Trumbull, CT 06611 USA If you should have questions, please call 1-800-243-2974 for Customer Service, 1-800-444-8456 for Technical Support or (203) 601-9818 for International calls.
i
MAN0035-DFU • Rev. A • 3/13
17
MAN0035-DFU • Rev. A • 3/13
Explanation of Symbols REF
Reorder Number
SN
Serial Number Latex Free ATTENTION: See instructions for use.
®
Keep Dry Type B Applied Part
Class II Equipment
Fetal and Vascular Dopplers User Manual for TR150 Series Units (Non-Display) and TR250 Series Units (Display)
Manufacturer Date of Manufacture Symbol indicates that the device should be sent to the special agencies according to local regulations for separate collection after its useful life. Product conforms to the Medical Device Directive 93/42/EEC Authorized Representative in the European Community.
TRIA®II and MedaSonics® are registered trademarks of CooperSurgical, Inc. Clorox® is a registered trademark of The Clorox Company. CooperSurgical is a registered trademark of CooperSurgical, Inc. © 2013 CooperSurgical, Inc.
Made in the USA 95 Corporate Drive • Trumbull, CT 06611 USA Phone: (800) 243-2974 Fax: (800) 262-0105 www.coopersurgical.com
Non-Display Models with Probe
Display Models with Probe
T334 T335 T336 T337 T337-S T345
T334-D T335-D T336-D T337-D T337-DS T345-D
International Phone: (203) 601-9818 Fax: (203) 601-4747
MAN0035-DFU • Rev. A • 3/13
MAN0035-DFU • Rev. A • 3/13