da Vinci Xi Manual Reprocessing Xi ENDOSCOPE Inspect
7 Flush Flush all ports with pressurized cold water.
6 x 20 sec
Connector Cover (if present) 2 Clean Wipe thoroughly using disinfectant, dry, and attach cover.
8 Spray Submerge and spray gap and base.
2 x 20 sec
Solution 3 Prepare Prepare cleaning solution (pH 7-11).
9 Brush Brush submerged or under running cold water.
60 sec
4 Prime Prime button ports, input discs, and base with cleaning solution.
Rinse 10 Final Rinse with high purity water.
60 sec
5 Soak Soak endoscope and cable.
11 Dry Empty water and dry all surfaces with a soft cloth.
6 Rinse Rinse under running cold water.
Dirty Side Inspect 12 Final Inspect entire endoscope for soil.
1
Inspect entire endoscope for damage.
60 sec
Chemical Disinfection (optional) 13 Apply per manufacturer’s instructions and rinse.
da Vinci Universal Reprocessing INSTRUMENTS AND STAPLER
8 Rinse Rinse under running cold water.
60 sec
Solution 2 Prepare Prepare cleaning solution (pH 7-11).
Cleaning 9 Ultrasonic Prime and clean in ultrasonic bath.
15 min
3 Prime Prime primary port with cleaning solution.
10 Flush Flush ports with pressurized cold water.
20 sec
4 Soak Soak the device.
30 min
Rinse 11 Final Rinse with high purity water.
60 sec
20 sec
12 Inspect Inspect entire device for soil.
6 Spray Submerge and spray tip.
30 sec
Thermal Disinfection 13 Disinfect per hospital policy.
7 Brush Brush submerged or under running cold water.
60 sec
1
Remove Accessories Remove all accessories before cleaning.
5 Flush Flush all ports with pressurized cold water.
ACCESSORIES GROUP 1 Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, etc.
Ultrasonic Cleaning (optional) 6 Clean in ultrasonic bath.
15 min
Solution 2 Prepare Prepare cleaning solution (pH 7-11).
Rinse 7 Final Rinse with high purity water.
10 sec
3 Soak Soak and actuate moving parts.
10 min
8 Inspect Inspect entire device for soil.
4 Rinse Rinse under running cold water.
20 sec
Disinfection 9 Thermal Disinfect per hospital policy.
5 Brush Brush submerged or under running cold water.
60 sec
1
Remove Accessories Remove all accessories before cleaning.
ACCESSORIES GROUP 2 Cautery Cord, Stapler Cable, PK Cords, Light Guide Cable, etc.
1
Ultrasonic Cleaning (optional) 5 Clean in ultrasonic bath.
Disconnect Cords or Cables Disconnect all cords and cables from device.
Solution 2 Prepare Prepare cleaning solution (pH 7-11).
Rinse 6 Final Rinse with high purity water.
3 Wipe Wipe thoroughly using cleaning solution.
7 Inspect Inspect entire accessory for soil.
4 Brush Brush submerged or under running cold water.
Disinfection 8 Thermal Disinfect per hospital policy.
60 sec
Copyright © 2018 Intuitive Surgical, Inc. All rights reserved. Intuitive Surgical, da Vinci, da Vinci Xi, da Vinci S, da Vinci Si, EndoWrist, and Single-Site are trademarks or registered trademarks of Intuitive Surgical, Inc. Product names are trademarks or registered trademarks of their respective holders. PN 554162-01 Rev. A
60 sec
For Intuitive Surgical da Vinci Xi® Surgical System Instruments, Stapler, and Endoscopes. This chart is not intended to replace an in-service or a manual. Refer to da Vinci Xi Reprocessing Instructions and da Vinci Xi Stapler Reprocessing Instructions.