da Vinci Xi Accessories Reprocessing Instructions
117 Pages
Preview
Page 1
da Vinci Xi ACCESSORIES
Reprocessing Instructions
1
2
Cleaning, Disinfection, and Sterilization Information for Reusable Accessories that are used with da Vinci Xi Systems
3
The reprocessing instructions and parameters given represent the manufacturer’s recommendations according to ISO 17664.
4
These Reprocessing Instructions are intended to be used together with the Reprocessing Instructions Appendices (PN 554154), which is provided as a separate document.
Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 USA
Intuitive Surgical SAS 11 avenue de Canteranne 33600 Pessac, France
www.intuitive.com Customer Service 800.876.1310
Customer Service (Europe) +800.0821.2020
PN 554153-02 Rev. B
2460
Copyright © 2021 Intuitive Surgical Operations, Inc. All rights reserved. Trademarks Intuitive, Intuitive Surgical, da Vinci Xi, EndoWrist, and Single-Site are trademarks or registered trademarks of Intuitive Surgical. Product and brand names/logos are trademarks or registered trademarks of Intuitive Surgical or their respective owner. See www.intuitive.com/trademarks.
2
Reprocessing Manual for da Vinci Xi Accessories
Table of Contents INTRODUCTION
Symbols: Warnings, Cautions, and Notes General Warnings, Cautions, and Notes Text Glossary Symbol Glossary Overview of Accessory Groups
GROUP 1
7 7 8 10 11 13
15
Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, and others (see Reprocessing Instructions Appendices PN 554154 for complete list) Accessories Group 1 Overview Automated Cleaning Process Overview Manual Cleaning Process Overview
17 18 19
Operating Room (OR)...21 Remove Accessories 23 Wipe 23 Wet 24 Chemical Pre-Disinfection 25 Transport 26 Dirty Side, Automated Cleaning...27 Remove Accessories 29 Prepare Solution 30 Soak 31 Rinse 31 Brush 32 Final Rinse 32 Inspect 33 Automated Cleaning 33 Transfer 34
Reprocessing Manual for da Vinci Xi Accessories
3
Table of Contents (continued) Dirty Side, Manual Cleaning...35 Remove Accessories 37 Prepare Solution 38 Soak 39 Rinse 39 Brush 40 Ultrasonic Cleaning 41 Final Rinse 43 Inspect 43 Thermal Disinfection 44 Transfer 44 Clean Side...45 Dry 47 Final Inspection 47 Sterilization Warnings, Cautions, and Notes 49 Pack 50 Sterilization Process Parameters 51 Wrap or Prepare 52 Sterilize 53 Store 53
GROUP 2
55
Cords (see Reprocessing Instructions Appendices PN 554154 for complete list) Accessories Group 2 Overview Automated Process Overview Manual Process Overview
57 58 59
Operating Room (OR)...61 Disconnect Cords 62 Chemical Pre-Disinfection 63 Transport 64 Dirty Side, Automated Cleaning...65 Remove Accessories 67 Prepare Solution 67 Wipe 68 Inspect 69 Automated Cleaning 69 Transfer 70 4
Reprocessing Manual for da Vinci Xi Accessories
Table of Contents (continued) Dirty Side, Manual Cleaning...71 Remove Accessories 73 Prepare Solution 74 Wipe 75 Brush 75 Ultrasonic Cleaning 76 Final Rinse 78 Inspect 78 Thermal Disinfection 79 Transfer 79 Clean Side...81 Dry 83 Final Inspection 84 Sterilization Warnings, Cautions, and Notes 85 Pack 86 Sterilization Process Parameters 87 Wrap or Prepare 88 Sterilize 89 Store 89
GROUP 3
Xi Endoscope Trays Accessories Group 3 Overview Automated Process Overview Manual Process Overview
91 93 94 95
Operating Room (OR)...97 Wipe 99 Chemical Pre-Disinfection 99 Transport 100 Dirty Side, Automated Cleaning... 101 Wipe 103 Inspect 103 Pack 104
Reprocessing Manual for da Vinci Xi Accessories
5
Table of Contents (continued) Dirty Side, Manual Cleaning... 105 Prepare Solution 107 Wipe 108 Brush 108 Final Rinse 109 Inspect 109 Thermal Disinfection 110 Transfer 110 Clean Side... 111 Dry 113 Final Inspection 114 Sterilize 115
6
Reprocessing Manual for da Vinci Xi Accessories
Introduction
INTRODUCTION
These Instructions for Use contain the cleaning, disinfection, and sterilization instructions and parameters recommended by Intuitive Surgical. These instructions apply to da Vinci Xi® Accessories. These instructions contain essential information for reprocessing accessories. Read these instructions in their entirety before reprocessing the accessories. Reusable accessories are not supplied sterile and must be reprocessed prior to first use. If there are any questions or comments about any information in these instructions, or if a problem occurs while reprocessing that cannot be resolved, contact Intuitive Surgical.
Symbols: Warnings, Cautions, and Notes
The most frequently used symbols in these instructions are provided below. See Appendix A for a complete list of symbols.
WARNING Alerts the reader about a situation which, if not avoided, could result in death or serious injury. CAUTION Alerts the reader about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. It may also be used to alert against unsafe practices. This includes the special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse. NOTE Highlights important information.
Reprocessing Manual for da Vinci Xi Accessories
7
INTRODUCTION
General Warnings, Cautions, and Notes WARNING • Devices are to be handled, reprocessed, and operated by qualified and trained personnel only. Inadequate training may result in device damage or patient/operator harm. • Cleaning processes described in these instructions have been validated by Intuitive Surgical. Alternative cleaning processes may result in incomplete reprocessing, which may lead to patient harm. • Do not reuse or reprocess single-use devices after being removed from package, even if unused, as this may result in incomplete sterilization and/or device damage, which may lead to patient harm. • Universal Precautions should be observed by all hospital personnel that work with contaminated or potentially contaminated devices. Personal Protective Equipment (PPE) should be worn when handling or working with contaminated or potentially contaminated devices, to prevent operator injury. PPE includes: gloves, gown, mask, goggles or face shield and shoe covers. • Use only the validated sterilization machines, parameters, and cycles, and follow instructions in entirety. Any other cycles and parameters have not been validated and may damage the device, or may result in incomplete sterilization, which may lead to patient harm. • Exercise care while handling devices with sharp points or cutting edges. Failure to follow Universal Precautions could result in injury. • Use care in handling chemicals and contaminated solutions, as skin contact, inhalation, or ingestion may cause injury. • Prior to use, after use, or during reprocessing, always inspect devices for any defects or damage. Failure to inspect may result in patient harm or additional device damage.
8
Reprocessing Manual for da Vinci Xi Accessories
INTRODUCTION
General Warnings, Cautions, and Notes (continued) CAUTION • Do not use chemical disinfection for devices, unless otherwise noted. Chemical disinfection may result in device damage. • Remove all accessories before cleaning. Failure to remove accessories prior to reprocessing may result in inadequate cleaning and sterilization, which may lead to patient harm. • In jurisdictions where the practice of chemical pre-disinfection is required, do not expose instruments, endoscopes, and accessories to chemical pre-disinfectants containing peroxidebased, strong alkaline or bleach-based chemicals. Doing so may result in device damage, which may lead to procedure delay. • Exposure of instruments, endoscopes, and accessories to chemical pre-disinfectants should be limited to the minimum contact time recommended by the pre-disinfectant manufacturer to achieve the level of pre-disinfection required by the hospital. Longer exposure times could result in device damage, which may lead to procedure delay.
Reprocessing Manual for da Vinci Xi Accessories
9
INTRODUCTION
General Warnings, Cautions, and Notes (continued) NOTE • Reusable devices, including cautery cords, gage pins, and endoscope trays, are not supplied sterile and must be cleaned and sterilized before use. • Do not use equipment (for example, steam cleaning devices) or procedures that employ pre-cleaning temperatures greater than 60 °C (140 °F). High temperatures may fix residual soil to the surface, and make cleaning more difficult. • Always handle devices with care to avoid mechanical shock or stress, which may result in device damage. • Use only process chemicals that conform to legal and regulatory requirements for safety as required by the jurisdiction in question. • When disposing of Intuitive Surgical instruments, accessories, or components, follow institution biohazard protocol and all applicable national and local laws and guidelines. • Follow ISO 17665-1 and the specifications of the sterilization device manufacturer. • Refer to the sterilization system manufacturers' instructions for proper use of the system and recommendations for sterilization. • Any reference in the instruction stating that reprocessing must start within 60 minutes after the procedure is based on our validation. A hospital's internal process validation may support a longer reprocessing start time. • Do not use the Instrument/Stapler Release Kit (IRK/SRK) on the instrument except during emergency or intraoperative scenarios. If kit is used, do not use device. • Avoid prolonged exposure to either ultrasonic cleaning or cleaning agents.
10
Reprocessing Manual for da Vinci Xi Accessories
INTRODUCTION
Text Glossary Accessories: Components that are used in conjunction with instruments. Automated Cleaning: Cleaning process in which a washer disinfector is used as part of the process. Clean Side: An area in hospitals and other healthcare facilities where medical devices that have already been cleaned in the "Dirty Side" are inspected, packaged, and sterilized. Clean Side is sometimes referred to as the "Prep and Pack." Cold Water: Unless otherwise specified, the term “cold water” refers to tap water that is 1025 °C (50-77 °F) and potable or better. CSSD: (Central Sterile Supply Department), also called sterile processing department (SPD), is a department in hospitals and other healthcare facilities that performs cleaning, sterilization and other actions on medical devices and equipment. Dirty Side: An area in hospitals and other healthcare facilities where medical devices are cleaned and disinfected. Dirty Side is sometimes referred to as the "Decon." Final Rinse Step: The final rinsing step of a cleaning process. High purity water: Examples of high purity water include deionized (DI) water, reverse osmosis (RO), and distilled water. High purity water is a type of water defined by ISO 15883-1. High purity water is extensively treated (usually by a multi-step treatment process that could include carbon bed, softening, DI, and RO or distillation) to ensure that microorganisms and the inorganic and organic material are removed from the water. Operating Room (OR): Also called Operating Theater, is a room where surgery is often performed. Soil: Also called "residual soil." Body fluids or tissue that may remain on the device after reprocessing. Sterile Processing Department (SPD): Also called Central Sterile Supply Department (CSSD) is a department in hospitals and other healthcare facilities that performs cleaning, sterilization and other actions on medical devices and equipment. Washer Disinfector: A machine intended to clean and disinfect medical devices and other articles used in the context of medical practice as defined in ISO 15883-1.
Reprocessing Manual for da Vinci Xi Accessories
11
INTRODUCTION
Symbol Glossary CHEMICAL (BOTTLE)
CHEMICAL DISINFECTANT
SPRAY BOTTLE
SYRINGE (LUER SLIP TIP)
FLUSH PORT ICON
SOFT, LINT-FREE CLOTH
REFER TO MANUAL
NYLON BRUSH
TIME: SECONDS
CLEAN, DRY AIR
TIME: MINUTES
MAGNIFIER
STAINLESS STEEL STERILIZATION TRAY
SOAK BIN
PLASTIC STERILIZATION TRAY
FLUSH PORT ADAPTER
seconds
minutes
12
Reprocessing Manual for da Vinci Xi Accessories
INTRODUCTION
Symbol Glossary (continued) STERILIZATION CONTAINER
PEEL PACK
WASHER DISINFECTOR MACHINE
STERILIZATION MACHINE
STERILIZATION WRAP
HIGH PURITY WATER
Reprocessing Manual for da Vinci Xi Accessories
13
INTRODUCTION
Overview of Accessory Groups
For a complete list refer to Reprocessing Instructions Appendices PN 554154.
GROUP 1
Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Single-Site Suction Irrigator Kit, and others.
Cannulae
Obturators
1
1
2
2
3
3
4
4
Release Kits Suction Irrigator Tube Adapter
Hasson Cone
Introducer
Suction Irrigator Tubing
Gage Pin
Single-Site Suction Irrigator Kit
GROUP 2 Cords
Monopolar Cautery Cord
GROUP 3
Trays
Plastic Sterilization Tray
14
Bipolar Cautery Cord
Stainless Steel Sterilization Tray
Reprocessing Manual for da Vinci Xi Accessories
Xi GROUP 1
ACCESSORIES
GROUP 1
Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, and others (see Reprocessing Instructions Appendices PN 554154 for complete list)
Cleaning, Disinfection, and Sterilization Information for Reusable Accessories that are used with da Vinci Xi Systems
1
2
3
4
Reprocessing Manual for da Vinci Xi Accessories
15
THIS PAGE LEFT INTENTIONALLY BLANK
16
Reprocessing Manual for da Vinci Xi Accessoriese
GROUP 1: INSTRUMENT ACCESSORIES
Accessories Group 1 Overview
Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Single-Site Suction Irrigator Kit, and other. For a complete list refer to Reprocessing Instructions Appendices PN 554154. NOTE • Do not use the Instrument/Stapler Release Kit (IRK/SRK) on the instrument except during emergency or intraoperative scenarios. If kit is used, do not use device. • Single-Site Suction Irrigator Kit is to be reprocessed with the Single-Site Suction Irrigator instrument.
Cannulae
Obturators
1
1
2
2
3
3
4
4
Release Kits Suction Irrigator Tube Adapter
Hasson Cone
Introducer
Gage Pin
Reprocessing Manual for da Vinci Xi Accessories
Suction Irrigator Tubing
Single-Site Suction Irrigator Kit
17
GROUP 1: INSTRUMENT ACCESSORIES
AUTOMATED CLEANING PROCESS
Automated Cleaning Process Overview
Reprocessing the accessories requires cleaning, disinfection, and sterilization. The 'Wet' step for the accessories should begin immediately after the procedure to prevent drying. If 'Wet' step in the operating room (OR) is not possible within 60 minutes, start the soak in dirty side within 60 minutes. See the diagram below for an overview of the reprocessing steps that take place in the OR, dirty side, and clean side.
OR (page 21) OR Preparation
Dirty Side, Automated Cleaning (page 27)
Step 1
Remove Accessories
Dirty Side Preparation
Step 2
Wipe
Step 1
Remove Accessories
Step 3
Wet
Step 2
Prepare Solution
Step 4
Chemical Pre-Disinfection (optional)¹
Prepare for Automated Cleaning
Step 5
Transport
Step 3
Soak
Step 4
Rinse
Step 5
Brush
Step 6
Final Rinse
Step 7
Inspect
18
Sterilization Preparation Step 1
Dry
Step 2
Final Inspection
Pack and Sterilize
Conduct Automated Cleaning and Disinfection
1
Clean Side (page 45)
Step 8
Automated Cleaning
Step 9
Transfer
Step 3
Pack
Step 4
Wrap or Prepare
Step 5
Sterilize
Store Step 6
Store
The practice of Chemical Pre-Disinfection may be required in some jurisdictions.
Reprocessing Manual for da Vinci Xi Accessories
GROUP 1: INSTRUMENT ACCESSORIES
MANUAL CLEANING PROCESS
Manual Cleaning Process Overview
Reprocessing the accessories requires cleaning, disinfection (optional), and sterilization. The 'Wet' step for the accessories should begin immediately after the procedure to prevent drying. If 'Wet' step in the operating room (OR) is not possible within 60 minutes, start the soak in dirty side within 60 minutes. See the diagram below for an overview of the reprocessing steps that take place in the OR, dirty side, and clean side.
OR (page 21) OR Preparation
Dirty Side, Manual Cleaning (page 35)
Step 1
Remove Accessories
Dirty Side Preparation
Step 2
Wipe
Step 1
Remove Accessories
Step 3
Wet
Step 2
Prepare Solution
Step 4
Chemical Pre-Disinfection (optional)¹
Conduct Cleaning
Step 5
Transport
Step 3
Soak
Step 4
Rinse
Step 5
Brush
Step 6
Ultrasonic Cleaning (optional)
Step 7
Final Rinse
Step 8
Inspect
Step 9
Thermal Disinfection²
Step 10
Transfer
1
The practice of Chemical Pre-Disinfection may be required in some jurisdictions.
2
Thermal Disinfection may be optional in some jurisdictions.
Reprocessing Manual for da Vinci Xi Accessories
Clean Side (page 45) Sterilization Preparation Step 1
Dry
Step 2
Final Inspection
Pack and Sterilize Step 3
Pack
Step 4
Wrap or Prepare
Step 5
Sterilize
Store Step 6
Store
19
THIS PAGE LEFT INTENTIONALLY BLANK
20
Reprocessing Manual for da Vinci Xi Accessories