Operators Manual
189 Pages
Preview
Page 1
Passport 2/Passport 2 LT Operating Instructions Chapter 5 - Accessories
1-1
Table of Contents
Foreword ... vii Warnings, Precautions And Notes ... vii Warnings ... viii Precautions ... xi Notes...xiii Unpacking ... xiii Symbols and Descriptions... xiv General Description... 1 - 1 Controls, Indicators and Connectors... 2 - 1 Front Panel ... 2 - 2 Display ... 2 - 6 Menus ... 2 - 8 Patient... 2 - 8 Monitor Setup ... 2 - 10 Printer Setup... 2 - 13 Parameters Menu... 2 - 14 Functions Menu... 2 - 15 Left Side Panel ... 2 - 16 Right Side Panel ... 2 - 18 Rear Panel ... 2 - 19 Remote Color Display (Passport 2 only) ... 2 - 20 Gas Module (Optional) - Front Panel... 2 - 21 Gas Module (Optional) - Rear Panel... 2 - 22 Comm-Port (optional) ... 2 - 23 Operation... 3 - 1 Abbreviated Operating Check List... 3 - 1 Detailed Operating Instructions ... 3 - 3 Setting-Up / Turning Power On ... 3 - 4 Setting Up Your Patient ... 3 - 4 Settings ... 3 - 5 Transferring Monitor Default Settings... 3 - 5 Installation and Use of “Extended Trend” Feature ... 3 - 5 Installation Menu... 3 - 6 Installation Mode ... 3 - 6 System Information Screen... 3 - 9 Non-Invasive Blood Pressure Measurements (NIBP)... 3 - 11 The NIBP Menu ... 3 - 11 Manual NIBP Measurements ... 3 - 11 Automatic NIBP Measurements ... 3 - 13 Automatic Adjustment in the Interval Mode ... 3 - 14 Suspension of NIBP Measurements ... 3 - 14 NIBP Pressure Limit Fail Safe... 3 - 14 Cuff Inflation Time ... 3 - 14 START and STOP Functions... 3 - 15 NIBP Auto Time Out Functions ... 3 - 15 ECG Measurements... 3 - 16
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
i
Table of Contents
The ECG Menu ... 3 - 17 3 Lead or 5 Lead ECG Measurements ... 3 - 17 “ECG Lead Fault” Messages ... 3 - 18 Adult Electrode Placement... 3 - 19 Neonatal Electrode Placement ... 3 - 21 12-Lead ECG Monitoring (optional Passport 2) ... 3 - 21 12-lead ECG Interpretation (Optional Passport 2) ... 3 - 22 Invasive Blood Pressure (IBP1, IBP2) (optional) ... 3 - 24 SpO2 Pulse Oximetry ... 3 - 25 SpO2 Menu... 3 - 25 SpO2 Measurements... 3 - 25 Performance Considerations ... 3 - 27 Calibration ... 3 - 27 Auto Scaling ... 3 - 28 MASIMO® Sensors and Patient Cable... 3 - 28 MASIMO® Sensors And Accessories ... 3 - 28 Selecting a Sensor ... 3 - 28 Cleaning and Re-use ... 3 - 29 NELLCOR® Sensors and Patient Cable... 3 - 29 Selecting a Sensor ... 3 - 30 Cleaning and Re-use ... 3 - 30 ST Monitoring (optional) ... 3 - 31 ST Setup... 3 - 33 Arrhythmia Analysis (optional) ... 3 - 34 Arrhythmia Pattern Definitions... 3 - 35 Temperature Menu ... 3 - 37 List Trends... 3 - 38 Modification of Parameters Displayed ... 3 - 39 Modification of Trend Entry Conditions ... 3 - 39 Filtering of List Trend Data Displayed ... 3 - 39 Printing List Trend Data ... 3 - 40 Transferring List Trend Data Between Different Passport 2 Monitors ... 3 - 40 Transfer Notes... 3 - 40 Clearing Trend Data... 3 - 40 Removing the List Trend Display... 3 - 40 Graph Trends (Passport 2 Only) ... 3 - 41 Modification of Parameters Displayed ... 3 - 42 Modification of Trend Entry Conditions ... 3 - 42 Printing Graph Trend Data... 3 - 42 Transferring Graph Trend Data Between Different Passport 2 Monitors ... 3 - 42 Clearing Trend Data... 3 - 43 Removing the Graph Trend Display ... 3 - 43 OXY-CRG Display Menu... 3 - 44 Parameters Displayed... 3 - 44 Printing OXY-CRG Data ... 3 - 44 Transferring OXY-CRG Data Between Different Passport 2/Passport 2 LT Monitors ... 3 - 45 Clearing Trend Data... 3 - 45 Removing the OXY-CRG Display ... 3 - 45
ii
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Table of Contents
Respiration Monitoring ... 3 - 46 Respiration Menu ... 3 - 46 Thoracic Impedance ... 3 - 46 CO2 Waveform ... 3 - 47 Respiration Monitoring on the Passport 2/Passport 2 LT ... 3 - 47 Thoracic Impedance ... 3 - 47 CO2 Waveform (requires optional Microstream® CO2 or Gas Module) ... 3 - 47 Microstream® CO2 Gas Monitoring (Optional for Passport 2 only) ... 3 - 48 Microstream® CO2 ... 3 - 48 Microstream® CO2 Calibration ... 3 - 50 Gas Module (Optional)... 3 - 52 Sequence for Monitoring Anesthetic Gases, O2, N2O and/or CO2... 3 - 52 Gas Monitor Calibration ... 3 - 54 Alarms ... 3 - 56 Setting Parameter Alarm Limits ... 3 - 56 Alarm Limits ... 3 - 57 Auto-Set Alarms ... 3 - 59 Alarm Violations... 3 - 59 Battery Indicators... 3 - 61 Beep Tones... 3 - 62 How to Set the Clock/Date... 3 - 63 Recorder (optional) ... 3 - 64 Print Setup Menu ... 3 - 64 Operation of Recorder... 3 - 64 Printer Formats... 3 - 65 Laser Printing 12 Lead ECG (Optional - Passport 2 only) ... 3 - 69 Printing 12 Lead to the Laser printer ... 3 - 69 Laser Printing 12 Lead Format: (Passport 2 Only)... 3 - 70 Status Messages... 3 - 71 NIBP Measurement Messages ... 3 - 71 SpO2 Messages ... 3 - 72 Recorder Messages (only units equipped with recorder)... 3 - 73 CO2 Messages (only units equipped with CO2) ... 3 - 73 Passport 2/Passport 2 LT Gas Module Messages (only observed when Gas Module is installed)... 3 - 74 Cooling Fan Message... 3 - 78 Monitor Problem Solving... 3 - 79 Using the Passport 2/Passport 2 LT with the VISA or PatientNet® Central Stations ... 3 - 82 User Maintenance... 4 - 1 Introduction ... 4 - 1 Care and Cleaning Of Monitor... 4 - 1 Care and Cleaning Of SpO2 Sensors... 4 - 2 Sterilization and Cleaning of Reusable Cuffs ... 4 - 2 Battery Replacement and Maintenance... 4 - 3 Battery Replacement ... 4 - 3 Battery Maintenance ... 4 - 3 Recorder Paper Replacement ... 4 - 4
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
iii
Table of Contents
Care and Storage Of Thermal Chart Paper... 4 - 5 Care and Cleaning Of Gas Module ... 4 - 6 Care and Cleaning of 3 Lead and 5 Lead ECG Cables and Leadwires... 4 - 7 Care and Cleaning of View™ 12 Lead ECG Analysis Module... 4 - 7 Accessories... 5 - 1 Standard Destination Kits... 5 - 1 SpO2 Kits ... 5 - 4 NIBP / ECG Kits... 5 - 4 CO2 Kits... 5 - 5 Optional Accessories... 5 - 6 NIBP Accessories ... 5 - 6 SpO2 Accessories... 5 - 7 Oridion CO2 Accessories ... 5 - 8 Gas Module Accessory Kits... 5 - 9 Reusable Temperature Probes... 5 - 11 Disposable Temperature Probes... 5 - 12 ECG Accessories... 5 - 12 ECG Cables ... 5 - 12 ECG Leadwires ... 5 - 13 12 Lead ECG Kits ... 5 - 14 Hewlett Packard Accessories ... 5 - 14 Electrodes ... 5 - 15 IBP Accessories ... 5 - 15 Comm-Port Accessories... 5 - 15 Miscellaneous Accessories ... 5 - 17 Mounting Kits and Accessories ... 5 - 17 Upgrade Kits ... 5 - 18 Central Station Accessories ... 5 - 18 Appendix ... 6 - 1 Safety Designations... 6 - 1 Safety designations per IEC 601-1 Standard: ... 6 - 1 Performance Specifications... 6 - 2 ECG... 6 - 2 ECG Performance Requirements ... 6 - 2 Analog Output Specifications ... 6 - 5 3 Lead and 5 Lead ECG* ... 6 - 5 Arterial Blood Pressure ... 6 - 5 Sync Pulse for Cardioversion ... 6 - 5 Systole Detector and Heart Rate Meter... 6 - 6 ECG Derived Heart Rate Meter Performance Requirements ... 6 - 6 IBP Derived Heart Rate Meter Performance ... 6 - 7 SpO2 Derived Heart Rate Meter Performance ... 6 - 7 NIBP Derived Heart Rate Meter ... 6 - 8 S-T Segment Analysis... 6 - 8 S-T Segment Analysis Performance Requirements ... 6 - 8 Arrhythmia Analysis ... 6 - 9
iv
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Table of Contents
12-Lead ECG Interpretation ... 6 - 9 NIBP Sub-System Performance Characteristics ... 6 - 9 Systolic Pressure Measurements ... 6 - 9 Diastolic Pressure Measurements ... 6 - 9 Pressure Measurement System ... 6 - 10 Pulse Rate ... 6 - 10 Maximum Cuff Pressure ... 6 - 10 Cuff Inflation ... 6 - 11 Maximum Leakage ... 6 - 11 Vent Rate ... 6 - 11 NIBP Sub-System Functional Requirements ... 6 - 11 Initial Conditions ... 6 - 11 NIBP Starting Pressure Settings and Ranges ... 6 - 11 NIBP Measurement Cycle ... 6 - 12 IBP Parameter Sub-System Performance Characteristics... 6 - 12 IBP Performance Requirements ... 6 - 12 IBP Connector Type ... 6 - 12 IBP Transducer Performance ... 6 - 12 IBP Heart Rate Meter ... 6 - 12 Temperature Parameter Performance Characteristics... 6 - 13 Connector Type ... 6 - 13 Temperature Performance Requirements ... 6 - 13 Respiration ... 6 - 14 ECG Respiration Performance Requirements ... 6 - 14 CO2 Respiration Performance Requirements ... 6 - 14 SpO2 ... 6 - 15 Masimo® SpO2 Performance Requirements ... 6 - 15 Masimo® Pulse Rate Performance ... 6 - 16 Nellcor SpO2 Performance Requirements ... 6 - 17 Nellcor SpO2 Pulse Rate Performance Requirements ... 6 - 17 CO2... 6 - 17 Microstream® CO2 (Passport 2 Only) ... 6 - 17 Physical Characteristics ... 6 - 18 Printer... 6 - 18 Comm-Port... 6 - 18 Physical Characteristics ... 6 - 19 Communication Characteristics ... 6 - 19 Normal Operating Noise ... 6 - 20 Battery ... 6 - 20 AC Power ... 6 - 21 Real Time Clock ... 6 - 21 Power Selection... 6 - 21 Fan Control ... 6 - 21 Trend Storage... 6 - 21 Transferring Monitor Default Settings... 6 - 21 Installation and Use of “Extended Trend” feature ... 6 - 22 Display ... 6 - 22 Gas Module ... 6 - 23
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
v
Table of Contents
Environmental Conditions... 6 - 25 Agency Compliance ... 6 - 28 Electromagnetic Compatibility ... 6 - 29 Indirect Blood Pressure Measurements And Associated Errors ... 6 - 30 Precautions While Making Automatically Cycled Blood Pressure Measurements ... 6 - 30 Cuff Size ... 6 - 30 Other Factors... 6 - 30 User Verification Of Passport 2/Passport 2 LT Measurements ... 6 - 31 Newborn NIBP Technique ... 6 - 31 How To Get Help ... 6 - 32 Warranty... 6 - 33 USA, Canada, Mexico, and Puerto Rico... 6 - 33 International (excluding North America)... 6 - 33 Datascope’s Responsibility ... 6 - 35 Extended Warranty... 6 - 35 Trademark Information ... 6 - 35
vi
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Foreword
Introduction
Foreword The Passport 2/Passport 2 LT Operating Instructions (0070-00-0440 Rev R, October 7, 2002) is intended to provide information for proper operation. General knowledge of monitoring and an understanding of the features and functions of the Datascope Passport 2/Passport 2 LT Monitor are prerequisites for its proper use. Do not operate this monitor before reading these instructions. CAUTION:
Information for servicing this instrument is contained in the Passport 2 Monitor Service Manual, Part No. 0070-00-00513-02. For additional information or assistance, please contact an authorized Datascope representative in your area.
CAUTION:
Federal Law restricts this device to sale by or on the order of a physician or other practitioner licensed by state law to use or order the use of this device.
•
Patents: This device is covered under one or more of the following U.S. Patents 4,621,643, 4,653,498, 4,700,708, 4,770,179, 4,869,254, 4,911,167, 4,928,692, 4,934,372, 5,078,136, 5,368,224, 5,482,036, 5,490,505, 5,632,272, 5,685,299, 5,758,644, 5,769,785, 6,002,952, 6,036,642, 6,067,462, 6,157,850, 6,206,830, 6,247,674 and foreign equivalents. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
Warnings, Precautions And Notes Please read and adhere to all warnings, precautions and notes listed here and in the appropriate areas throughout this manual. A WARNING is provided to alert the user to potential serious outcomes (death, injury, or serious adverse events) to the patient or the user. A CAUTION is provided to alert the user to use special care necessary for the safe and effective use of the device. They may include actions to be taken to avoid effects on patients or users that may not be potentially life threatening or result in serious injury, but about which the user should be aware. Cautions are also provided to alert the user to adverse effects on this device of use or misuse and the care necessary to avoid such effects. A NOTE is provided when additional general information is applicable.
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
vii
Introduction
Warnings
Warnings
viii
WARNING:
Possible Explosion Hazard - This instrument is not explosion proof in the presence of flammable anesthetics. Do not use in the presence of flammable gases.
WARNING:
Internal Electrical Shock Hazard - This unit does not contain any user-serviceable parts. Do not remove instrument covers. Refer Servicing to qualified personnel.
WARNING:
Trace Gas Hazard - When using the optional Gas Module, a health hazard exists when trace amounts of vaporized anesthetic agents are chronically inspired by operating room personnel. See Appendix A in NFPA 56A on Inhalation Anesthetics. During any procedure where such agents are employed, the Gas Module exhaust output should be connected to a medical gasscavenging system.
WARNING:
Do not use this monitor during MRI (Magnetic Resonance Imaging) scanning. Induced current could potentially cause burns. Accuracy of measurements on this unit and the MRI unit may be also be affected.
WARNING:
For continued protection against a fire hazard, replace all fuses with the specified type and rating. See the Passport 2 Service Manual, P/N 0070-00-0513-02.
WARNING:
Do not clean monitor while it is on and/or plugged in.
WARNING:
This unit uses a common isolation path for the ECG leads and the Invasive Pressure Channels. Ensure that conductive parts of the ECG electrodes do not contact other conductive parts including earth ground. Do not connect any nonisolated accessories to the Passport 2 or to the ECG or invasive pressure channel inputs when connected to a patient. Insure that the total chassis leakage currents of all connected units does not exceed 300mA. Use an IEC 601-1 approved isolation / separation transformer if required. Do not simultaneously touch the patient and any piece of electrical equipment if any cover has been removed from the equipment.
WARNING:
The AC line cord and interface cables (ie nonpatient cables) may utilize the same ground. Therefore, removal of the AC line cord does not necessarily isolate the Passport 2, if non-patient interface cables are attached.
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Warnings
Introduction
WARNING:
Observe extreme caution when a defibrillator is used on a patient. Do not touch any part of patient, table, or monitor when a defibrillator is in use.
WARNING:
Microstream waste material and CO2 filter should be treated as biohazard material.
WARNING:
Route cables neatly. Ensure cables, hoses, and wires are away from patient’s neck to avoid strangulation. Keep floors and walkways free of cables to reduce risk to hospital personnel, patients and visitors.
WARNING:
Thoracic respiration measurement may interfere with some pacemakers. Refer to pacemaker’s manual.
WARNING:
Do not incinerate battery, possible explosion may occur.
WARNING:
Do not put MPSO (Multiple Portable Socket Outlets ie. Multiple outlet extension cords) used with the Passport 2/Passport 2 LT or its accessories on the floor. Connect only Passport 2/Passport 2 LT accessories to the same MPSO as the Passport 2/Passport 2 LT. Do not overload the MPSO.
WARNING:
Compressed gasses are considered Dangerous Goods/Hazardous Materials per I.A.T.A. and D.O.T. regulations. It is a violation of federal and international law to offer any package or over pack of dangerous goods for transportation without the package being appropriately identified, packed, marked, classified, labeled and documented according to D.O.T. and I.A.T.A. regulations. Please refer to the applicable I.A.T.A. Dangerous Goods Regulations and/or the Code of Federal Regulations 49 (Transportation, Parts 171-180) for further information.
WARNING:
PACEMAKER PATIENTS: Rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. Keep pacemaker patients under close surveillance. See section 6.2 of this manual for disclosure of the pacemaker pulse rejection capability of this instrument.
WARNING:
Computerized ECG Analysis should be reviewed by qualified medical personnel. It should not be used exclusively for treatment or non-treatment of patients.
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
ix
Introduction
x
Warnings
WARNING:
ST segment measurements may be affected by one or more of the following ECG rhythm morphologies: wide complex QRS such as bundle branch blocks, ventricular pacemaker rhythm, left ventricular hypertrophy or Wolff-Parkinson-White Syndrome. Consult with qualified medical personnel prior to treatment or non-treatment.
WARNING:
The View 12™ ECG Analysis Module is not intended for use during electrosurgery. If the electrosurgical ground connection is not satisfactory, there exists a possibility of patient burns at the ECG electrode sites.
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Precautions
Introduction
Precautions CAUTION:
Only use the Abbreviated Operating Check List (0070-00-0493) if you are already familiar with this product. If not, please use the Detailed Operating Instructions.
CAUTION:
Always place the monitor on a rigid, flat surface or on approved mounts. Do not block the vents.
CAUTION:
Never place fluids on top of this monitor. In case of accidental wetting, wipe clean immediately and have the monitor serviced to ensure no hazard exists.
CAUTION:
Do not operate the Passport 2/Passport 2 LT with a frayed or damaged power cord.
CAUTION:
This unit must only be operated with Datascope approved software.
CAUTION:
Cuffs must be used with the correct Datascope hoses. See Chapter 5 for part numbers.
CAUTION:
Use only Datascope accessories with this product.
CAUTION:
When cleaning sensors, do not use excessive amounts of liquid. Wipe the sensor surface with a soft cloth, dampened with the cleaning solution. See section 4 for more details. Do not clean the monitor while it is on and/or plugged in.
CAUTION:
Dispose of single use items in accordance with hospital policy.
CAUTION:
Do not use a damaged or broken unit or accessory.
CAUTION:
Do not operate the Passport 2/Passport 2 LT with the ventilation or speaker vents obstructed.
CAUTION:
To prevent condensation, allow the Passport 2/ Passport 2 LT to warm up and dry if it is moved from a cold area to a warm one.
CAUTION:
This product contains natural rubber latex which may cause allergic reactions.
CAUTION:
Please consult a physician for interpretation of blood pressure measurements.
CAUTION:
A blood pressure measurement can be affected by the position of the patient, and his / her physiological condition as well as other factors, such as patient movement.
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
xi
Introduction
xii
Precautions
CAUTION:
Substitution of a component different from that supplied might result in measurement error.
CAUTION:
The Passport 2/Passport 2 LT may not meet its performance specifications if stored or operated outside of specified temperature and humidity ranges.
CAUTION:
Prior to use, be sure the rail supporting the bed rail mounting hook can support the weight of the monitor. Consult the bed manufacturer’s specifications if necessary. Datascope cannot be responsible for injury or damage resulting from improper or inadequate support of the monitor.
CAUTION:
To assure successful triggering of a Datascope Intra-Aortic balloon pump from the Passport 2/ Passport 2 LT monitor, set the “ECG Filter” to “Extended” and set “Pacer Enhancement” to “On”. Both of these settings are located in the ECG setup menu of the Passport 2/Passport 2 LT.
CAUTION:
The Analog Output on the Passport 2/Passport 2 LT supports triggering the IABP from 3-Lead or 5-Lead ECG only. Invasive Blood Pressure triggering is not supported.
CAUTION:
Use only Datascope supplied power cords or if a substitute is necessary use only hospital grade power cords.
CAUTION:
Removal of the View 12™ ECG Analysis Module without first disabling the 12-Lead ECG card may cause a temporary disruption in patient monitoring.
CAUTION:
The 2.4 GHz radio optionally used in this device must be at least 20 cm away from the user and/or patient during normal operating conditions.
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Introduction
xii
Precautions
CAUTION:
Substitution of a component different from that supplied might result in measurement error.
CAUTION:
The Passport 2/Passport 2 LT may not meet its performance specifications if stored or operated outside of specified temperature and humidity ranges.
CAUTION:
Prior to use, be sure the rail supporting the bed rail mounting hook can support the weight of the monitor. Consult the bed manufacturer’s specifications if necessary. Datascope cannot be responsible for injury or damage resulting from improper or inadequate support of the monitor.
CAUTION:
To assure successful triggering of a Datascope Intra-Aortic balloon pump from the Passport 2/ Passport 2 LT monitor, set the “ECG Filter” to “Extended” and set “Pacer Enhancement” to “On”. Both of these settings are located in the ECG setup menu of the Passport 2/Passport 2 LT.
CAUTION:
The Analog Output on the Passport 2/Passport 2 LT supports triggering the IABP from 3-Lead or 5-Lead ECG only. Invasive Blood Pressure triggering is not supported.
CAUTION:
Use only Datascope supplied power cords or if a substitute is necessary use only hospital grade power cords.
CAUTION:
Removal of the View 12™ ECG Analysis Module without first disabling the 12-Lead ECG card may cause a temporary disruption in patient monitoring.
CAUTION:
The 2.4 GHz radio optionally used in this device must be at least 20 cm away from the user and/or patient during normal operating conditions.
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
Notes
Introduction
Notes NOTE:
This unit is not designed to be used with a peripheral pulse sensor. SpO2 is a standard function in this monitor, and may be used to obtain a plethysmograph waveform and heart rate.
NOTE:
The comparison testing conducted via the ausculatory method used both Phase 4 and Phase 5 Korotkoff sounds. Reports of study findings for both the auscultatory method as well as the intraarterial methods are available by contacting Datascope Technical Support (201) 995-8116.
NOTE:
Potential hazards due to errors in software or hardware have been minimized by actions taken in accordance with IEC 60601-1-4.
NOTE:
This device has not been tested in a patient population having common arrhythmias such as ventricular premature beats or atrial fibrillation.
Unpacking Remove the instrument from the shipping carton and examine it for signs of shipping damage. Save all packing materials, invoice, and bill of lading. These may be required to process a claim with the carrier. Check all materials against the packing list. Contact the Datascope Service Department (201) 995-8116 for prompt assistance in resolving shipping problems.
Copyright © Datascope Corp., 1999. Printed in U.S.A. All rights reserved. Contents of this publication may not be reproduced in any form without permission of Datascope Corp.
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
xiii
Introduction
Symbols and Descriptions
Symbols and Descriptions
xiv
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
1.0
General Description
The Datascope Passport 2 is a vital signs monitor intended for intra-hospital use on human patients. The Passport 2 LT is limited to 3 traces with an option for 4. Both units have many features and functions, yet are very easy to use through an integrated keypad, NavigatorTM Control Knob and intuitive menu system. The patient parameters that can be monitored are: ECG, 3 Lead ST analysis with adjustable ISO and J points (optional), Invasive (optional) and Non-invasive Blood Pressure, Masimo SET® or Nellcor® OxiSmart® SpO2, Respiration Rate, Temperature, Arrhythmia (optional), Gases (optional), Microstream® CO2 (optional only with Passport 2), and 12-Lead ST analysis and ECG Interpretation (optional only with Passport 2).
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
1-1
General Description
The Passport 2/Passport 2 LT Monitor can be mounted on a rolling stand, a wall mount bracket, gas machine arm, Gas Module, Bedrail or operated as a tabletop instrument. The keypad contains dedicated primary functions. The menu buttons provide access to setting up patient information, waveforms, and parameters. The Passport 2 comes with a color TFT LCD or a monochrome display. The Passport 2 LT comes with a color display or a monochrome display. Digital displays are provided for Heart Rate, Invasive Blood Pressure (IBP1 and IBP2) (optional), Anesthetic Agents, O2 and N2O (optional), ST (optional), Non-invasive Blood Pressure (NIBP), Pulse Oximetry (SpO2), Respiration Rate, Temperature (T1), and CO2 (optional). The built-in recorder (optional) provides hard copies of all digital data and waveforms as well as Tabular & Graphic Trend Information. The View 12™ ECG Analysis Module for the Passport 2 enables 12-Lead Acquisition, Continuous 12-Lead ST Analysis and Arrhythmia Analysis with print capability. The View 12™ ECG Analysis Module consists of a PCMCIA card for insertion into the Passport 2 with 12 Lead software, a M-12 cable and a detachable leadwire set. The Passport 2/Passport 2 LT has the capability of interfacing with Datascope’s Central Station, DPD Defibrillator, Gas Module and all IABP Systems 90 through 98. The optional Gas Module can be used on an anesthesia cart or mounted on a rolling stand or wall mounted. The Passport 2/Passport 2 LT monitor is powered from an AC connection or internal batteries. Batteries can be purchased separately as optional equipment. See Chapter 5. The Passport 2/Passport 2 LT monitor can operate with either battery removed so that fresh batteries can be installed during monitor operation. Key features of the Passport 2/Passport 2 LT are: • 3 or 5 Lead (I, II, III, aVR, aVL, aVF, V) ECG • 12-Lead (I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6) ECG (optional Passport 2 only) • ECG Cascade • ESIS Capability (3 or 5 Lead ECG only). • 2 invasive blood pressure channels (optional) • 3 Lead ST Analysis with adjustable ISO and ST points (optional Passport 2/ Passport 2 LT) • 12 Lead ST Analysis with adjustable ISO and ST points (optional Passport 2 only) • 12-Lead ECG Interpretation (optional Passport 2) • Arrhythmia Analysis (optional) • Non-invasive Blood Pressure (NIBP) • Lead Selectable Impedance Respiration • Masimo SET® SpO2 • Nellcor® OxiSmart® SpO2 (optional) • Microstream® CO2 (optional with Passport 2 only) • Gas Module Connectivity (optional)
1-2
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions
General Description
• 1 YSI 400/700 temperature channel • Automatic sensor detection and waveform display • Automatic Heart Rate source selection • Auto-Set™ Alarms • Dual channel thermal array recorder (optional) • Color TFT LCD display or Monochrome display (Passport 2) • Passive Color display or Monochrome display (Passport 2 LT) • Battery operation (optional) • Tabular and Graphic trend displays, 120 entries (500 entries optional) • Extended Trend Display, 500 entries (optional) • OXY-CRG display, 6 minutes (12 hours optional) • 6 trace erase bar refresh - Passport 2 • 3 trace erase bare refresh - Passport 2 LT (optional 4th trace) • NavigatorTM Control Knob • Internal isolated power module • External Remote Color Display Available with Passport 2 Color TFT LCD Equipped Monitor (optional) • Communication with Datascope Central Station and the DPD Defibrillator • Communication with hospital CIS (Clinical Information Systems) through DIAP (Datascope Improved ASCII Protocol, manual P/N 0070-00-0307). • Inter-Monitor Patient Data Transfer (with optional accessories) • Inter-Monitor System Set-up Transfer (with optional accessories) • Mounting Kits (optional accessory) • Soft-Grip Handle • Comm-Port (optional) • Dual PCMCIA Interface
Passport 2/Passport 2 LT Operating Instructions
0070-00-0440
1-3
General Description
This page intentionally left blank.
1-4
0070-00-0440
Passport 2/Passport 2 LT Operating Instructions