Datex-Ohmeda
Aestiva 5 MRI Operation Manual Part 2 Software Rev 3.X Feb 2002
Aestiva 5 MRI Operation Manual Part 2 Software Rev 3.X Feb 2002
144 Pages
Preview
Page 1
Aestiva/5 MRI
Operation Manual – Part 2 Software Revision 3.X Setup, Cleaning and Sterilization, Maintenance and Troubleshooting
User Responsibility This Product will perform in conformity with the description thereof contained in this Operation manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Customer Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
w CAUTION
U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A., check local laws for any restriction that may apply. Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture, and a sequential unit number for identification. AAA F 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned; “D” = 2000, “E” = 2001, “F” = 2002, etc. “I” and “O” are not used.
Aestiva, S/5, Tec 5, Tec 6 and Tec 7 are registered trademarks of Datex-Ohmeda Inc. Other brand names or product names used in this manual are trademarks or registered trademarks of their respective holders.
Important
Important MRI Compatibility Information Any electronic medical device intended for use in the MRI suite requires special design considerations to ensure MRI compatibility. Care was taken during the design of the Aestiva/5 MRI to help ensure that the performance specifications were the same as the existing Aestiva/5 Anesthesia System. However, while the performance specifications have not changed, you as users must take special precautions when using the Aestiva/5 MRI in the MRI Suite. This section of the Aestiva/5 MRI Operation and Maintenance Manual is designed to provide you with the additional precautions you must take when using this device in the MRI suite.
Important
The Aestiva/5 MRI is MRI Compatible. It was tested with a 1.5 Tesla Active Shielded MRI unit in a magnetic fringe field ≤ 300 gauss. The Aestiva/5 MRI will function to specifications in an MRI environment that does not exceed 300 gauss. Do not store or operate the Aestiva/5 MRI within the 300 gauss line for your unit.
w WARNING
Use only MRI compatible cylinders Do not set up or service this machine in the MRI room. Do not add or remove accessories or components in the MRI room. Do not use or mount monitors or other accessories that are not MRI compatible at the 300 gauss line. Check with the manufacturer of the monitor or accessory equipment for MRI compatibility information. Do not store any items in the drawer that are not MRI compatible. The Aestiva/5 MRI does contain some ferromagnetic material and will be attracted to the MR Imager if positioned closer than 300 gauss. Always keep the Aestiva/5 MRI in a fringe field of 300 gauss or less. Set the brake during use or storage in MRI room.
1006-0839-000
i
Aestiva/5 MRI
ii
1006-0839-000
Table of Contents
Important MRI Compatibility Information... i
1/Introduction How to use this manual...1-2 Symbols used in the manual or on the equipment...1-3 Maintenance summary and schedule...1-6 Operator maintenance...1-6 Datex-Ohmeda approved service...1-7
2/Cleaning and Sterilization Summary...2-2 Patient path...2-2 Scavenging path...2-3 Clean and sterilize...2-4 To wash (by hand or machine)...2-4 To Autoclave...2-5 Special requirements...2-6 Assemble...2-6 Disassemble the patient path...2-7 Canister disassembly... 2-10 Disassemble the scavenging path... 2-12 How to clean and disinfect the flow sensors... 2-13 CIDEX sterilization... 2-13 Procedure (Figure 2-3)... 2-13 How to clean and sterilize the optional CO2 bypass assembly... 2-15
3/Setup and Connections MRI Compatibility...3-1 Breathing system setup...3-3 Canister Setup...3-7 Pneumatic and electrical connections...3-9 How to install gas cylinders (high pressure leak test)... 3-11 Cylinder yokes... 3-11 DIN connections... 3-12 Installation Notes... 3-13
1006-0839-000
iii
Aestiva/5 MRI
4/User maintenance Repair policy... 4-2 Manifold maintenance... 4-3 Expiratory valve maintenance... 4-6 Receiver maintenance (active gas scavenging only)... 4-8 Flow sensor maintenance... 4-9 Breathing circuit maintenance... 4-10 Bellows maintenance... 4-12 Bellows tests... 4-15 O2 sensor calibration - 21% O2... 4-18 O2 sensor calibration - 100% O2... 4-21 Flow sensor calibration... 4-24 How to prevent water build-up... 4-25
5/Troubleshooting About alarms... 5-2 Alphabetical list... 5-4 Breathing system problems (no alarm)... 5-16 Electrical problems (power failure, etc.)... 5-17 Pneumatic problems... 5-19
6/Illustrated Parts Main manifold... 6-2 Breathing system parts... 6-4 Exhalation valve... 6-7 Breathing circuit modules... 6-8 Bellows... 6-9 Test tools and system parts... 6-10
7/External Communications External communications... 7-2 Protocol Description (Datex-Ohmeda Com 1)... 7-3 Serial Communication Parameters... 7-3 DEVICE COMMANDS Sent By External Device... 7-4
iv
1006-0839-000
Table of Contents
8/Specifications and Theory of Operation Ventilator Theory...8-2 Modes...8-3 Breathing system schematic...8-5 Internal signal schematic...8-6 Pneumatic circuits...8-7 Breathing system... 8-10 Breathing system specifications... 8-12 Gas scavenging... 8-13 Pneumatic specifications... 8-14 Flow specifications... 8-15 Electrical power... 8-16 Power cord... 8-16 Battery Information... 8-16 Electro-magnetic Compatibility... 8-17 Gauss alarm... 8-17 Physical specifications... 8-18 Environmental requirements... 8-18 MRI environment... 8-18 Temperature... 8-18 Humidity... 8-18 Altitude... 8-18 Ventilation Operating Specifications... 8-19 Aestiva/5 MRI Ventilator Accuracy Data... 8-21
Index Warranty
1006-0839-000
v
1
Introduction
In this section
How to use this manual... 1-2 Symbols used in the manual or on the equipment... 1-3 Maintenance summary and schedule... 1-6 Operator maintenance... 1-6 Datex-Ohmeda approved service... 1-7
1006-0839-000
1-1
Aestiva/5 MRI
How to use this manual This is part two of the Aestiva™/5 MRI operation and maintenance manual. It tells you how to: • Remove and clean parts • Set up the system • Identify and replace worn or damaged parts • Calibrate the O2 sensor The second half of this chapter is a maintenance schedule. The last section, troubleshooting, tells you what causes each alarm and what you can do about it. Use this manual together with part 1, which includes the operating instructions and preoperative checkout.
w WARNING
1-2
If an alarm occurs, safeguard the patient first, before troubleshooting or repair procedures.
1006-0839-000
Introduction
Symbols used in the manual or on the equipment wWarnings and wCautions tell you about dangerous conditions that can occur if you do not follow all instructions in this manual. Warnings tell about a condition that can cause injury to the operator or the patient. Cautions tell about a condition that can cause damage to the equipment. Read and follow all warnings and cautions. Other symbols replace words on the equipment or in Datex-Ohmeda manuals. No one device or manual uses all of the symbols. These symbols include:
On (power) Off (power) Standby Standby or preparatory state for part of the equipment “ON” only for part of the equipment “OFF” only for part of the equipment Direct current
∏
Alternating current
x
Protective earth ground
r 1006-0839-000
Í m µ H w wW N N
l O o q p œ †
π y
Not autoclavable Type B equipment Type BF equipment Type CF equipment Caution, ISO 7000-0434 Attention, refer to product instructions, IEC 601-1 This way up
Dangerous Voltage
Earth ground
Electrical input
Electrical output
Pneumatic inlet
Pneumatic outlet
Frame or chassis ground
REF
Stock Number
1-3
Aestiva/5 MRI
å Y
Alarm silence button
SN
Equipotential XXXX
Serial Number Systems with this mark agree with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The xxxx is the certification number of the Notified Body used by DatexOhmeda’s Quality Systems.
t
Variability
Read top of float.
T
Variability in steps
Vacuum inlet
+
Plus, positive polarity
Suction bottle outlet
P
Minus, negative polarity Lamp, lighting, illumination
Cylinder
N
Movement in one direction
Isolation transformer
ˆ
Movement in two directions
Linkage system
z
Lock
Risk of Explosion.
Z
Unlock
Low pressure leak test
1-4
O2 +
O2 Flush button
1006-0839-000
Introduction
134°C C
R u q t
< 345 kPa
Autoclavable
Bag position/ manual ventilation Open drain (remove liquid) Inspiratory flow
r U Q
O2 sensor connection.
The primary regulator is set to pressure less than 345 kPa
Absorber on.
Mechanical ventilation Close drain Expiratory flow End case
< 414 kPa
The primary regulator is set to pressure less than 414 kPa.
Absorber off (CO2 Bypass active).
European Union Representative
1006-0839-000
1-5
Aestiva/5 MRI
Maintenance summary and schedule These schedules show the minimum frequency. You will have to service the equipment more frequently if you use it: • In unusual conditions (dirty gas supplies, high temperature, high humidity, etc.) • More frequently than normal.
Operator maintenance
Examine all components and do the maintenance procedures more frequently if necessary. Minimum Frequency
Maintenance
Daily
• Clean the external surfaces. • 21% O2 calibration (circuit O2 sensor). • Flow sensor calibration
Two weeks
• Drain the vaporizers and discard the agent.
Monthly
• 100% O2 calibration (cIrcuit O2 sensor). • Put Krytox (or a lubricant approved for use with 100% O2) on all tee handle threads.
During cleaning • Inspect the parts for damage. Replace or repair as and setup necessary Annually
• Replace the external o-rings on the vaporizer ports.
As necessary
• Install new cylinder gaskets on cylinder yokes. • Replace the absorbent in the absorber. • Drain the breathing system. • Drain the breathing circuit module.1 • Drain the overflow trap on the optional suction regulator. • Replace the circuit O2 sensor. • Replace the disposable flow sensors (plastic) 2 • Replace the receiver filter (active gas scavenging only).
1. This is included in the preoperative test procedure. 2. Under typical use the sensor meets specifications for 3 months
1-6
1006-0839-000
Introduction
Datex-Ohmeda approved service Minimum Frequency
6 months
1006-0839-000
Maintenance
Have an approved service person do the service tests and scheduled service maintenance.
1-7
Aestiva/5 MRI
1-8
1006-0839-000
Cleaning and sterilization w WARNING Obey applicable safety precautions: • Read the material data sheet for each cleaning agent. • Read the operation and maintenance manual for all sterilization equipment. • Wear gloves and safety glasses. A damaged O2 sensor can leak and cause burns (contains potassium hydroxide). • Do not breathe the fumes.
w CAUTION To prevent damage: • Refer to the manufacturer’s data if you have questions about a cleaning agent. • Do not use organic, halogenated, or petroleum based solvents, anesthetic agents, glass cleaners, acetone, or other harsh cleaning agents. • Do not use abrasive cleaning agents (such as steel wool, silver polish or cleanser). • Keep all electronic parts away from liquids. • Do not permit liquid to go into the equipment housings. • Do not soak synthetic rubber parts for more than 15 minutes. Swelling or faster aging can occur. • Only autoclave parts that are marked 134°C.
In this section
Summary... 2-2 Clean and sterilize... 2-4 Disassemble the patient path... 2-7 Canister disassembly... 2-10 Disassemble the scavenging path... 2-12 How to clean and disinfect the flow sensors... 2-13 How to clean and sterilize the optional CO2 bypass assembly... 2-15
1006-0839-000
2-1
Aestiva/5 MRI
Summary The parts in Figure 2-1 send exhaled gas back to the patient. They may require more frequent cleaning/sterilization than parts in Figure 2-2. Refer to your hospital’s infection control policy.
Autoclave (134˚C) or wash (mild detergent pH <10.5)
AB.23.003
Patient path
*
Refer to cleaning/disinfection procedure.
AB.23.079
Wipe with a damp cloth.
*
Figure 2-1 • These parts return exhaled gas to the patient
2-2
1006-0839-000
Basic cleaning and sterilization
Autoclave (134˚C) or wash (mild detergent pH <10.5)
AB.23.080
AB.23.003
Scavenging path
Figure 2-2 • These parts do not send gas back to the patient
1006-0839-000
2-3
Aestiva/5 MRI
Clean and sterilize The Disassembly part of this section tells you how to remove parts for cleaning
To wash (by hand or machine)
AB.23.085
Patient path
AB.23.083
Scavenging path
Use a mild detergent (pH <10.5). Then, rinse and dry completely. All parts except the O2 sensor, and flow sensors can be washed. User maintenance tells you how to disassemble parts and clean inside them if necessary.
2-4
1006-0839-000
Basic cleaning and sterilization
To Autoclave
AB.23.087
Patient path
AB.23.086
Shown upside down
AB.23.083
Scavenging path
Autoclave at 134°C. Inspect the parts for deterioration. The user maintenance section tells you how to do this.
1006-0839-000
2-5