DENTAURUM IMPLANTS
Adhesive residue remover, wheel
Instructions for the reprocessing of pliers and instruments
16 Pages
Preview
Page 1
EN
Instructions for the reprocessing of pliers and instruments These instructions apply to all Dentaurum pliers, instruments and rotary instruments that are approved for use on patients. These instructions comply with the requirements of the latest version of DIN EN ISO 17664 and the Robert Koch Institute guidelines.
General requirements Pliers and instruments must be cleaned, disinfected and if necessary, sterilized before each use. This also applies to initial use following delivery. Pliers and instruments are normally delivered non-sterile (clean and disinfect following removal of the protective transport packaging; sterilize after packing). Efficient cleaning and disinfecting is essential for effective sterilization. This includes rotary instruments (drills), hereafter referred to as instruments. During use, ensure that contaminated instruments are collected separately and are not placed back in the instrument tray to avoid greater contamination of the loaded instrument tray. Clean / disinfect the contaminated instruments. Then replace them in the instrument tray and sterilize the fully loaded instrument tray if necessary. Should the dental professional and / or the patient become aware of serious problems arising from the use of the product, it is important that the manufacturer and the competent authority in the country in which the dental professional and / or the patient are resident is informed accordingly.
Manual reprocessing Surface contamination must be removed from pliers and instruments immediately after use, at the latest within one hour.
1h
Rinse pliers and instruments under water and immerse in a disinfectant solution. Use only a soft brush (nylon brush) or a clean soft cloth for manual removal of contamination. Do not use metal brushes or steel wool. Disinfectants should be aldehyde-free (otherwise fixation of blood residue), have a certified efficacy (e.g. DGHM [German Society for Hygiene and Microbiology] or FDA approval or CE marking), be suitable for disinfecting instruments and compatible with the instruments.
2
Machine-based cleaning and disinfection
When using a disinfector, make sure: the efficacy of the disinfector has been certified (e.g. DGHM or FDA approved or CE marking according to DIN EN ISO 15883), a certified program for thermal disinfection (minimum 10 mins at 93 °C / 199 °F or an A0 > 3000) is used (with chemical disinfection there is the risk of disinfectant residue on the instruments), it uses only water that is sterile or has a low bacteria count (max. 10 bacteria / ml) and is low in endotoxins (max. 0.25 endotoxin units / ml) (e.g. purified water / highly purified water) for rinsing, and t he air used for drying is filtered.
When choosing a cleaning agent system, make sure: it is suitable for cleaning metal and plastic instruments, a n additional disinfectant with certified efficacy (e.g. DGHM or FDA approved or CE marking) is used – provided that thermal sterilization is not used – and that it is compatible with the cleaning agent used, and the chemicals used are compatible with the instruments (see chapter Material resistance), Adhere to the concentrations given by the manufacturer of the cleaning agent and disinfectant.
3
Procedure
D ismantle the pliers and instruments as far as possible. Place the disassembled, half-opened (hand-width) pliers and instruments in the disinfector. Pliers and instruments should not come into contact with one another.
Start the program. R emove the pliers and instruments from the disinfector at the end of the program.
Check and pack the instruments, immediately after removal if possible.
If the disinfector does not have an automatic drying program, leave the door of the unit slightly open to dry the instruments. Proof of basic suitability for effective machine-based cleaning and disinfecting was provided by an independent, accredited test laboratory using a G 7836 GD disinfector (thermal disinfection, Miele & Cie. GmbH & Co., Gütersloh, Germany) and the cleaning agent neodisher® Medizym (Dr. Weigert GmbH & Co. KG, Hamburg, Germany). The procedure described above was taken into account during the tests.
Manual cleaning and disinfection. Observe the following points when choosing a cleaning agent and disinfectant: m etal and / or plastic pliers and instruments must be suitable for cleaning and disinfecting. the cleaning agent must be suitable for ultrasonic cleaning. u se only disinfectants with certified efficacy (e.g. DGHM or FDA approval or CE marking). Disinfectants should be compatible with the cleaning agent used. Combined cleaning agents / disinfectants should not be used if possible. Combined cleaning / disinfecting agents can only be used if there is minimal contamination (no visible contamination). The concentrations and reaction times given by the cleaning agent and disinfectant manufacturer should be strictly adhered to. Use only freshly mixed solutions, only sterile or low-germ (max. 10 germs / ml) as well as low-endotoxin (max. 0.25 endotoxin units / ml) water (e.g. purified water / highly purified water) and only filtered air for drying.
4
Cleaning procedure
D ismantle the pliers and instruments as far as possible. P lace the disassembled pliers and instruments half-opened (hand-width) into the cleaning solution for the prescribed reaction time; ensure that the pliers and instruments are adequately covered. Pliers and instruments should not come into contact with one another. Rinse all hollow sections in the instruments using a single-use syringe (minimum volume 5 ml) before and after the reaction time. Special holders, e.g. bur blocks, should be used if necessary.
Remove the pliers and instruments from the cleaning solution and rinse thoroughly with water. Check that the pliers and instruments have been properly cleaned.
Disinfection procedure
P lace the disassembled pliers and instruments in the disinfectant solution according to the prescribed reaction time. Ensure that the pliers and instruments are covered and are not in contact with one another. Rinse all hollow sections in the instruments using a single-use syringe (minimum volume 5 ml) before and after the reaction time.
Remove the pliers and instruments from the disinfectant solution and rinse thoroughly at least three times with water.
Do not pack the pliers and instruments until they are dry. Pack the dry pliers and instruments immediately. Proof of basic suitability for effective manual cleaning and disinfecting was provided by an independent, accredited test laboratory using Bodedex® forte cleaning agent and Korsolex® plus disinfectant (Bode Chemie, Hamburg, Germany). The procedure described above was taken into account during the tests.
Check Check all pliers and instruments after cleaning or cleaning / disinfecting for corrosion, damaged surfaces, chipped areas or contamination and separate damaged pliers and instruments. Pliers and instruments that are still contaminated must be cleaned and disinfected again.
5
Pliers and instrument care
General: Instrument oil should not be used if possible. If it has to be used, however, ensure that only instrument oils (white oil) that are approved for steam sterilization and have certified biocompatibility are used – taking into account the maximum sterilization temperature. Pliers: Following drying and before sterilization, plier jaws, locks, sliding surfaces for springs and other critical surfaces must be conditioned with medical white oil, which is steam permeable and temperature resistant. The oil protection on the extra-hard plier inserts reduces the risk of corrosion. We recommend cleaning / protective oil (white oil) for pliers and instruments (REF 055-201-00).
Packaging
Pack the pliers and instruments or sterilization tray in a single-use sterilization pack (single or double packaging) and / or sterilization container, which comply with the following requirements: DIN EN ISO/ANSI AAMI ISO 11607-1/2 (formerly: DIN EN 868/ANSI AAMI ISO 11607) s uitable for steam sterilization (temperature resistant to min. 134 °C / 273 °F and adequate steam permeability) adequate protection of pliers and instruments and sterilization packaging against mechanical damage regular servicing according to manufacturer’s instructions (sterilization container)
6
Sterilization
Only the following sterilization procedures should be used for sterilization. Other sterilization procedures are unsuitable. Flash sterilization is not permitted. Other procedures that should not be used are: hot-air sterilization, X-ray sterilization, formaldehyde or ethylene oxide sterilization or plasma sterilization.
Steam sterilization f ractional vacuum or gravitation process* (with adequate product drying) s team sterilizer according to DIN EN 13060-2004 and DIN EN 285 validated (valid commissioning and product-specific performance evaluation) in accordance with DIN EN ISO/ ANSI AAMI ISO 17665 (formerly: DIN EN 554/ANSI AAMI ISO 11134) m aximum sterilization temperature 134 °C / 273 °F (plus tolerance in accordance with DIN EN ISO / ANSI AAMI ISO 17665 (formerly: DIN EN 554 / ANSI AAMI ISO 11134)) s terilization time (exposure time at the sterilization temperature) minimum 5 mins at 134 °C / 273 °F Proof of basic suitability for effective steam sterilization was provided by an independent, accredited test laboratory using a EuroSelectomat steam sterilizer (MMM Münchener Medizin Mechanik GmbH, Planegg, Germany) and the vacuum process as well as a Systec V-150 steam sterilizer (Systec GmbH Labor-Systemtechnik, Wettenberg, Germany) and the gravitation process. The procedure described above was taken into account during the tests.
*U se of the less effective gravitation process is only permitted if a fractional vacuum process is not available.
7
Storage
Following sterilization, pliers and instruments should be stored dry and dust-free in the sterilization packaging.
Material resistance
When choosing the cleaning agent and disinfectant ensure that they do not contain the following components: o rganic, mineral or oxidizing acids (maximum permitted pH 9.5, neutral / enzymatic cleaner recommended) s trong alkali solutions o rganic solvents (e.g. alcohols, ethers, ketones, benzines) o xidation agents (e.g. hydrogen peroxides) h alogens (chlorine, iodine, bromine) a romatic / halogenated hydrocarbons s alts of heavy metals
Never clean pliers and instruments, sterilization and tomas® trays with metal brushes or steel wool. Do not expose pliers and instruments, sterilization and tomas® trays to temperatures above 134 °C / 273 °F !
8
Reusability
All rotary instruments can be reused up to 5 times. The operator bears responsibility for any further reuse or use of damaged and/ or contaminated rotary instruments.
Special instructions regarding Dentaurum pliers and instruments Pliers: o nly pretreat in half-opened (hand-width) position (open and close several times when conditioning) c lean d isinfect a fter each cleaning process, apply oil to plier jaws, locks, sliding surfaces for springs and other critical areas of the pliers using the cleaning / protective oil (white oil) for pliers and instruments (REF 055-201-00) p ack and sterilize Note: The oil protection on the extra-hard plier inserts reduces the risk of corrosion. Product group Instruments: The ligator ( and its individual parts) must always be cleaned and disinfected by machine.
9
Overview of pliers / instruments The overview contains all pliers and instruments approved for use on patients. The processing instructions given should be adhered to, unless directed otherwise. Further information is included in the respective Instructions for use; refer to www.dentaurum.com, leaflets and Orthodontics catalog.
Designation
Nance loop bending pliers, EQ-Line Needle holder Medium Clean and disinfect in the open position, sterilize in the first lock.
Needle holder Mini Clean and disinfect in the open position, sterilize in the first lock.
Needle holder Medium with finger ring Clean and disinfect in the open position, sterilize in the first lock.
Mosquito forceps without hook Clean and disinfect in the open position, sterilize in the first lock.
Mosquito forceps with hook Clean and disinfect in the open position, sterilize in the first lock.
Mosquito forceps, curved
000-001-00 000-030-00 000-031-00 000-035-00 000-730-00 000-731-00
Clean and disinfect in the open position, sterilize in the first lock.
000-732-00
Nance loop closing pliers, Premium-Line
001-002-00
Lingual bending pliers, Premium-Line
003-001-00
Band seater
10
REF
Do not use alkaline / acid cleaners.
003-006-00
How pliers, curved, Premium-Line
003-111-00
How pliers, straight, Premium-Line
003-112-00
Johnson contouring pliers, Premium-Line
003-114-00
Weingart universal pliers, Premium-Line
003-120-00
Weingart universal pliers Mini, Premium-Line
003-121-00
Jarabak Light Wire pliers, Premium-Line
003-125-00
Light Wire pliers, Premium-Line
003-130-00
Designation
REF
Angle wire bending pliers Standard, Premium-Line
003-139-00
Facebow bending pliers Magnum, Premium-Line
003-180-01
Lingual arch bending pliers, Premium-Line
003-185-00
Anderer 3-Prong pliers Extra-Mini, Premium-Line
003-205-00
Bracket removing pliers, Premium-Line Remove the insert for cleaning and disinfecting.
003-349-00
Hollow chop pliers Extra-Mini, Premium-Line
003-350-00
Loop forming pliers Mini, Premium-Line
003-351-00
Hammerhead pliers, Premium-Line
003-355-00
Angle / Tweed ribbon arch pliers, Premium-Line
003-442-00
Distal end cutter Maxi, Premium-Line
003-700-00
Distal end cutter Mini, Premium-Line
003-701-00
Distal end cutter Mini with longer handle, Premium-Line
003-702-00
Distal end cutter, lingual
003-703-00
Flush Cutter Mini with silicone O-ring, Premium-Line Remove silicone O-Ring for cleaning, disinfecting and sterilizing.
Flush Cutter Maxi with silicone O-ring, Premium-Line
003-705-00
Remove silicone O-Ring for cleaning, disinfecting and sterilizing.
003-706-00
Crimping pliers for Herbst TS / SUS, Premium-Line
003-710-00
Crimping pliers, Premium-Line
003-711-00
Ortho-Cast M-Series cap removal pliers, Premium-Line Remove the blade for cleaning and disinfection; sterilize when the blade has been put back into place.
003-805-00
De La Rosa contouring pliers, Premium-Line
004-109-00
Begg Light Wire pliers, Premium-Line
004-130-00
Angle wire bending pliers, Premium-Line
004-139-00
Side cutter Maxi, Premium-Line
004-266-00
discovery® pearl debonding instrument
019-001-00
Bracket removing pliers, straight, Premium-Line
004-346-00
Bracket removing pliers, angled, Premium-Line
004-347-00
Bracket removing pliers, Premium-Line Remove the insert for cleaning and disinfecting.
004-349-00
11
Designation
REF
Special bracket debonding pliers, 45° angled, Premium-Line
004-350-00
Tweed loop forming pliers, Premium-Line
012-350-00
Tweed loop forming pliers with individually replaceable hard metal tip, Premium-Line Remove the insert for cleaning and disinfecting.
012-351-00
Ligature cutter Mini, Premium-Line
014-151-00
Ligature cutter 45°, Premium-Line
014-152-00
Ligature cutter, lingual
014-153-00
Ligature cutter
014-154-00
Double beak pliers, Premium-Line
015-157-00
Ligature forming pliers, Premium-Line
015-158-00
Separating pliers, Premium-Line
015-160-00
Side cutter Medium, Premium-Line
016-153-00
Separating strip holder Remove the separating strip before cleaning, disinfecting and sterilizing.
018-000-00
Distal end bender, double-ended
019-000-00
Ligature adapter, double-ended
021-273-00
Ligature tightener and adapter
022-274-00
LinguBall instrument
023-275-00
sl-instrument, gold
023-276-00
sl-instrument, silver
023-277-00
Positioning tool, lingual
023-278-00
Dental mirror with reference lines (assembled), lingual Fully disassemble for cleaning and disinfecting
023-279-00
Dental mirror with reference lines (without grip), lingual Ligator
023-279-10
Completely disassemble for cleaning and disinfecting; remove the wire from the guidance. The ligator and its individual parts must always be cleaned and disinfected by machine.
023-292-00
Ligator Completely disassemble for cleaning and disinfecting; remove the wire from the guidance. The ligator and its individual parts must always be cleaned and disinfected by machine.
023-288-00
Ligator
12
Completely disassemble for cleaning and disinfecting; remove the wire from the guidance. The ligator and its individual parts must always be cleaned and disinfected by machine.
023-284-00
Band pusher
024-300-01
Mershon band pusher
024-300-00
Designation
Crimping tweezers, with long placement and angulation aid Blow dry the joint gap after reprocessing (e.g. with compressed air gun).
Crimping tweezers, delicate design Blow dry the joint gap after reprocessing (e.g. with compressed air gun).
Crimping tweezers, for posterior teeth Blow dry the joint gap after reprocessing (e.g. with compressed air gun).
Crimping tweezers, delicate tweezer design Blow dry the joint gap after reprocessing (e.g. with compressed air gun).
Oliver-Jones band removing pliers, Premium Line Remove the insert for cleaning and disinfecting.
Molar band seater, with replaceable plastic head and hard metal triangular tip Do not use cleaning agents or disinfectants that contain solvents. Unscrew the head before cleaning and disinfecting.
Molar band seater, plastic
REF
025-276-00 025-277-00 025-279-00 025-280-00 026-347-00 026-354-00
Do not use disinfectants that contain solvents.
026-355-00
Scaler, double-ended
027-349-00
Bracket positioning gauge, with metal tips
030-390-00
Ruler
044-730-01
Angle / Tweed ribbon arch pliers, EQ-Line
045-000-00
Tweed loop forming pliers, EQ-Line
045-001-00
Angle wire bending pliers, EQ-Line
045-002-00
Weingart universal pliers, EQ-Line
045-003-00
Oliver-Jones band removing pliers, EQ-Line Remove the insert for cleaning and disinfecting.
Bracket removing pliers, EQ-Line
045-004-00
Remove the insert for cleaning and disinfecting.
045-005-00
How pliers, straight, EQ-Line
045-006-00
Jarabak Light Wire pliers, EQ-Line
045-009-00
Nance loop bending pliers, EQ-Line
045-010-00
Distal end cutter, EQ-Line
045-021-00
Side cutter Medium, EQ-Line
045-022-00
Eby band seater Fully disassemble for cleaning and disinfecting: unscrew the tip; unscrew the sleeve from the handle; remove the hammer and spring Only sterilize the tip.
051-023-00
Lip retractor, large Reprocessing: max. 50 x; cleaning/disinfecting: only use cleaning and disinfecting agents suitable for polycarbonate and observe the manufacturer´s instructions; sterilization: steam sterilization, 121 °C / 249.8 °F, holding time 15 mins
075-600-00
Lip retractor, small Reprocessing: max. 50 x; cleaning/disinfecting: only use cleaning and disinfecting agents suitable for polycarbonate and observe the manufacturer´s instructions; sterilization: steam sterilization, 121 °C / 249.8 °F, holding time 15 mins
tomas®-screwdriver
075-601-00
Fully disassemble for cleaning and disinfecting. Sterilize fully assembled.
302-004-10
tomas®-applicator
302-004-20
13
Designation
tomas®-applicator, long version
302-004-70
tomas®-wheel
302-004-30
tomas®-torque ratchet Fully disassemble for cleaning and disinfecting.
302-004-40
tomas®-driver
302-004-50
tomas®-driver, long version
302-004-60
Manually turned unit mod. acc. to Pauls Fully disassemble for cleaning and disinfecting. Sterilize fully assembled.
302-004-81
tomas®-stop screw
302-013-01
tomas®-tool tray
302-156-00
tomas® PI-reusable punch
303-210-10
tomas® PI-round drill
303-220-00
tomas® PI-lance drill
303-230-00
tomas® PI-pre drill 2.0
303-240-20
tomas® PI-final drill 3.4
303-204-34
tomas® PI-depth gauge
303-240-54
tomas® PI-driver 2.5
303-310-00
tomas® PI-applicator
303-320-00
tomas® PI-torque ratchet
14
REF
Fully disassemble for cleaning and disinfecting. Sterilize fully assembled.
303-330-00
tomas® PI-wheel
303-340-00
tomas® PI-lock key
303-350-00
tomas® PI-driver 1.3
303-360-00
tomas® PI-guide sleeve
303-500-10
tomas® PI-trepan
303-500-20
tomas® PI-tool tray
303-010-10
Hexagon socket key 0.9
607-129-00
Hexagon socket key 1.5
607-144-00
Designation
Expansion screw key, large
Rotating instruments
REF
611-116-00
REF
tomas®-drill SD 1.1, short pilot drill
302-103-00
Carbide burs, coarse
123-603-00 123-603-30
Carbide burs, fine
123-604-00 123-604-30
Adhesive residue remover, conical shaped
123-700-00
Adhesive residue remover, wheel
123-701-00
Adhesive residue remover, pear shaped, large
123-702-00
Adhesive residue remover, pear shaped, small
123-703-00
References
Further tips on the correct reprocessing of pliers and instruments in the practice and laboratory are provided by the “Instrument Reprocessing Working Group” in its "Yellow Brochure". www.a-k-i.org
15
Dentaurum Germany I Benelux I España I France I Italia I Switzerland I Australia I Canada I USA and in more than 130 countries worldwide.
DENTAURUM QUALITY WORLDWIDE UNIQUE
989-590-20
Printed by Dentaurum
Germany
2023-04/C/R1-15
www.dentaurum.com
Date of information: 2023-04 Subject to modifications
Turnstr. 31 I 75228 Ispringen I Germany I Phone + 49 72 31 / 803 - 0 I Fax + 49 72 31 / 803 - 295 www.dentaurum.com I [email protected]