Preparation and cleaning of pliers and instruments.
2a Use a suitable immersion disinfectant.
2a Clean under running water and with a soft nylon brush.
1 Pliers are to be used only for the indications intended.
2b
5
3
Dry the pliers, especially the joint, with oil-free compressed air.
Rinse the pliers and especially the joint with distilled water.
7
Apply protection and maintenance oil (REF 055-201-00) to the joint and hard metal inserts.
Autoclave pliers in the fully-opened position.
Turnstr. 31 I 75228 Ispringen I Germany I Phone + 49 72 31 / 803 - 0 I Fax + 49 72 31 / 803 - 295 www.dentaurum.com I [email protected]
989-504-60
6
Printed by Dentaurum
Germany
Discontinue use of damaged or corroded instruments. Dull pliers need to be reground.
4
09/19/C/R1-2
Clean the pliers in an instrument washer.