ENGLISH Cavitron® Select™ SPS™ Ultrasonic Scaler Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System Directions For Use
Please read carefully and completely before operating unit.
Introduction Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System DENTSPLY® Professional is an ISO9001 certified company. The Cavitron® Select™ System is classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical hazards only in accordance UL 2601-1 1 and Can/CSA C22.2 NO. 601.1. The Scaling System operates by converting ordinary AC house current into high frequency current. The ultrasonic system consists of two parts: an insert, and the SPS™ (Sustained Performance System) electronic system. The system incorporates two closed loops. One loop provides automatic tuning (operating frequency is adjusted to be at resonance for each insert), the second loop automatically controls the tip stroke over different working conditions. The scaler produces 30,000 microscopically small strokes per second at the insert’s working tip. This combined with acoustic effects of the coolant water, produces a synergistic action that literally “powers away” the heaviest calculus deposits while providing exceptional operator and patient comfort. The Pump/Reservoir accessory for the Cavitron® Select™ SPS™ is an optional lavage dispenser. Footswitch plug, handpiece plug, water tube and power supply plug are removed from the Scaler before mating with the Pump/Reservoir. The Scaler is snapped together with the Pump/Reservoir. The footswitch plug, handpiece plug and power supply plug are inserted into mating receptacles at the rear of the Pump/Reservoir. Ultrasonic signal, footswitch and DC power connections between the units are made via an edgecard connector at the rear of the Scaler and the front of the Pump
module. The Reservoir has a capacity of 450 ml. Fluid from the Reservoir is propelled from the Reservoir Bottle with 15–19 psig compressed air generated by an internal diaphragm pump. ON/OFF of the lavage is controlled through an internal solenoid valve. Fluid to the Handpiece passes directly from the Pump/ Reservoir to the Handpiece. No fluid passes through the Cavitron® Select™ SPS™ when used with this accessory. Flow through any insert in the Handpiece will be greater than 40 cc/m water when adjusted to maximum.
Technical Support For technical support and repair assistance in the U.S., call 1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other areas, contact your local DENTSPLY® representative.
Supplies & Replacement Parts To order supplies or replacement parts in the U.S., contact your local DENTSPLY® Distributor or call 1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other areas, contact your local DENTSPLY® representative. Caution: U.S. Federal Law restricts this device to sale by or on the order of a Dentist.
ENGLISH • 1
Table of Contents Section Number
1 2
3
4
5
6
7
8
9
10
11
12 13 14
Section Title Description of Contents
Page Number
Indications for use... 3 Contraindications and Warnings... 3 2.1 Contraindications 2.2 Warnings Precautions... 3 3.1 Precautions for All Systems 3.2 Precautions for Ultrasonic Prophylaxis Procedures 3.3 Precautions for Cavitron® Select™ Reservoir System Infection Control... 4 4.1 General Infection Control Recommendations 4.2 Water Supply Recommendations Installation Instructions for Cavitron® Select™ SPS™ Ultrasonic Scaler... 4–5 5.1 General Information 5.2 Water Line Requirements 5.3 Electrical Requirements 5.4 Unpacking the System 5.5 Power Cord/Power Connection 5.6 Foot Control Cable Assembly Connection 5.7 Water Supply Line Connection 5.8 Lavage Control Detachable Handpiece Cable Assembly Connection Installation Instructions for Cavitron® Select™ SPS™ Scaler with Cavitron® Select™ Reservoir System... 6–7 6.1 General Information 6.2 Electrical Requirements 6.3 Unpacking the System 6.4 Assembling the System 6.5 Power Cord/Power Connection 6.6 Foot Control Cable Assembly Connection 6.7 Lavage Control Detachable Handpiece Cable Assembly Connection Cavitron® Select™ SPS™ Ultrasonic Scaler Description... 8–10 7.1 System Controls for Cavitron® SPS™ Select™ Ultrasonic Scaler 7.2 System Controls for Cavitron® Select™ SPS™ Scaler with Cavitron® Select™ Reservoir System 7.3 Handpiece 7.4 Cavitron® 30KTM Ultrasonic Inserts 7.5 Foot Control Information & Operation Accessories... 11 8.1 Accessories List 8.2 Medicaments Techniques for Use... 11 9.1 Patient Positioning 9.2 Performing Ultrasonic Scaling Procedures 9.3 Patient Comfort Considerations System Care...12–13 10.1 Daily Maintenance - Start-Up Procedures, Between Patients, Shut-Down Procedures 10.2 Weekly Maintenance 10.3 Water Line Filter Maintenance 10.4 Pump Lavage Filter Maintenance Troubleshooting... 14 11.1 Troubleshooting Guide 11.2 Technical Support and Repairs Specifications... 15 Classifications... 15 Disposal of Unit... 15
ENGLISH • 2
Section 1: Indications for Use Ultrasonic procedures: • All general supra and subgingival scaling applications. • Periodontal debridement for all types of periodontal diseases. • Endodontic procedures. • For patients with a history of sensitivity to ultrasonics.
Section 2: Contraindications and Warnings 2.2 Warnings
2.1 Contraindications • Ultrasonic Systems should not be used for restorative dental procedures involving the condensation of amalgam.
Section 3: Precautions 3.1 Precautions for All Systems • Cavitron Inserts manufactured by DENTSPLY Professional should be used for optimal performance. ®
®
• Do not place the System on or next to a radiator or other heat source. Excessive heat may damage the System’s electronics. Place the System where air is free to circulate on all sides and beneath it. • The System is portable, but must be handled with care when moving. • Equipment flushing and dental water supply system maintenance are strongly recommended. See Water Supply under Section 10: System Care. • Close the water shut-off valve to the System every night before leaving the office. • Never operate scaler without fluid flowing through handpiece.
3.2 Precautions for Ultrasonic Prophylaxis Procedures • Like a toothbrush, ultrasonic inserts “wear out” with use. Inserts with just 2 mm of wear lose about 50% of their scaling efficiency. In general, it is recommended that ultrasonic inserts be discarded and replaced after 2 mm of wear to maintain optimal efficiency and avoid breakage.
• Persons fitted with cardiac pacemakers, defibrillators and other active implanted medical devices have been cautioned that some types of electronic equipment might interfere with the operation of the device. Although no instance of interference has ever been reported to DENTSPLY®, we recommend that the handpiece and cables be kept at least 6 to 9 inches (15 to 23 cm) away from any device and their leads during use. • There are a variety of pacemakers and other medically implanted devices on the market. Clinicians should contact the device manufacturer or the patient’s physician for detailed information about the device.
• Retract the lips, cheeks and tongue to prevent contact with the insert tip whenever it is placed in the patient’s mouth. • The DENTSPLY® power supply shipped with this unit is approved for use with devices and must be used to operate this system.
3.3 Precautions for Cavitron® Select™ Reservoir System • Do not place the System on or next to a radiator or other heat source. Excessive heat may damage the System’s electronics. Place the System where air is free to circulate on all sides and beneath it. • The System is portable, but must be handled with care when moving. • Equipment flushing and dental water supply system maintenance are strongly recommended. See Section 10: System Care. • Pour liquid into the cap opening of the reservoir bottle. To prevent potential leaking at the cap, do not overfill bottle with liquid into the neck of the bottle. Ensure cap is tightly sealed. • Turn unit off before refilling reservoir bottle to prevent potential leaking of fluid. Turn cap a 1/4 revolution to release air pressure and then remove the cap completely.
• If excessive wear is noted, or the insert has been bent, reshaped or otherwise damaged, discard the insert immediately.
ENGLISH • 3
Section 4: Infection Control 4.1 General Infection Control Recommendations • As with all dental procedures, use standard personal protection equipment (i.e., wear face mask, eyewear, or face shield, gloves and protective gown). • For maximal operator and patient safety, carefully follow the Infection Control Information procedures detailed in the Cavitron Infection Control Procedures Booklet (Form No. 81158).
4.2 Water Supply Recommendations It is highly recommended that all dental water supply systems conform to applicable CDC (Centers for Disease Control and Prevention) and ADA (American Dental Association) standards, and that all recommendations be followed in terms of flushing, chemical flushing, and general infection control procedures.
• As with high speed handpieces, and other dental devices, the combination of water and ultrasonic vibration from your Cavitron® Select™ SPS™ Scaler will create aerosols. With proper technique, much of the Cavitron® Select™ SPS™ Scaler’s aerosol dispersion can be effectively controlled and minimized. Please carefully follow the procedural guidelines in this manual regarding the use of your System.
Section 5: Installation Instructions for Cavitron® Select™ SPS™ Ultrasonic Scaler 5.1 General Information
5.3 Electrical Requirements
If the installation of your Cavitron® Select™ SPS™ Scaler is performed by someone other than trained DENTSPLY® Distributor personnel, care should be taken to observe the following requirements and recommendations.
Refer to Section 12: Specifications.
5.2 Water Line Requirements • Incoming water supply line pressure to the Scaler must be 20 psi (138 kPa minimum) to 40 psi (276 kPa maximum). If your dental water system’s supply line pressure is above 40 psi, install a water pressure regulator on the water supply line to your Cavitron® Select™ SPS™ Scaler. • A manual shut-off valve on the dental water system supply line should be used so that the water can be completely shut off when the office is unoccupied. • After the above installations are completed on the dental water supply system, the dental office water line should be thoroughly flushed prior to connection to the Scaler.
ENGLISH • 4
5.4 Unpacking the System Carefully unpack your Cavitron® Select™ SPS™ Scaler and verify that all components and accessories are included: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10.
Cavitron® Select™ SPS™ Scaler. Detachable AC Power Cord set. Steri-Mate Detachable Sterilizable Handpiece. Steri-Mate Flow Control Detachable Handpiece Cable Assembly with Swivel. Dual Position Foot Control Assembly. Hose / Filter / Quick Disconnect Assembly. Additional Water Line Filter. Additional Pump Lavage Filters. Medically Approved Power Supply. Literature Packet.
Section 5: Installation Instructions for Cavitron® Select™ SPS™ Ultrasonic Scaler, cont’d
5.5 Power Cord/Power Connection
5.7 Water Supply Line Connection
• Verify the Power Control knob is set to the OFF position before proceeding.
• Push the blue hose into the orange water receptacle until the hose cannot be pushed in any further.
• Plug the detachable AC cord into the back of the Medically Approved Power Supply.
• Connect the free end of the system’s water supply line to the office water supply line.
• Plug the power supply outlet into the back of the System.
• Inspect all connections to make certain there are no leaks.
• Plug the 3 prong grounded plug into an approved outlet.
• To remove water line from Cavitron® Select™ SPS™ Scaler, turn off or disconnect water supply line FIRST, then push on outer orange ring of the outlet and gently pull out water hose.
5.8 Lavage Control Detachable Handpiece Cable Assembly Connection • Plug the Handpiece outlet into the receptacle on the back of the System. • Holding the Handpiece Cable Assembly seat male connector into appropriate female connector. • Once seated, snap male connector into female connector by pushing on the end of the male connector. • Rotate Lavage Control on the handpiece to middle setting. • Turn on the on/off switch (power adjustment).
5.6 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated.
ENGLISH • 5
Section 6: Installation Instructions for Cavitron® Select™ SPS™ Scaler with Cavitron® Select™ Reservoir System 6.1 General Information If the installation of your Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System is performed by someone other than trained DENTSPLY® Distributor personnel, care should be taken to observe the following requirements and recommendations.
• Connect Cavitron® Select™ SPS™ Scaler to Cavitron® Select™ Reservoir System. Insert male edge card connector on Cavitron® Select™ Reservoir Pump into Cavitron® SPS™ Select™ Scaler opening on the back of the Cavitron® Select™ SPS™ Scaler. Press down on the top of the Cavitron® Select™ SPS™ Scaler (directly above power setting knob) until the black striker knob located on the Reservoir Bottle locks into place.
6.2 Electrical Requirements Refer to Section 12: Specifications.
6.3 Unpacking the System Carefully unpack your Cavitron® Select™ SPS™ Scaler and verify that all components and accessories are included: 1. 2. 3. 4. 5. 6.
Cavitron® Select™ SPS™ Scaler. Cavitron® Select™ Reservoir Pump. Cavitron® Select™ Reservoir Bottle. Detachable AC Power Cord set. Steri-Mate Detachable Sterilizable Handpiece. Steri-Mate Flow Control Detachable Handpiece Cable Assembly with Swivel. 7. Dual Position Foot Control Assembly. 8. Hose / Filter / Quick Disconnect Assembly.
9. 10. 11. 12.
• To detach Cavitron® Select™ SPS™ Scaler from Cavitron® Select™ Reservoir Pump and Bottle, lift up on the front of the Cavitron® Select™ SPS™ Scaler (below power setting knob and Power/Boost light). After the striker knob is unsnapped, pull the Cavitron® Select™ SPS™ Scaler away from the Cavitron® Select™ Reservoir Pump.
Additional Water Line Filter. Additional Pump Lavage Filters. Medically Approved Power Supply. Literature Packet.
6.4 Assembling the System • Remove protective sticker from edge card connector on the scaler. • Rinse Cavitron® Select™ Reservoir Bottle before using. • Insert Cavitron® Select™ Reservoir Bottle into Cavitron® Select™ Reservoir Pump and rotate clockwise until bottle locks into place.
ENGLISH • 6
NOTE: Always disengage striker knob before pulling the Cavitron® Select™ SPS™ Scaler away from the Cavitron® Select™ Reservoir Pump.
Section 6: Installation Instructions for Cavitron® Select Scaler with Cavitron® Select™ SPS™ Reservoir System cont'd.
6.5 Power Cord/Power Connection • Verify the Power Control knob is set to the OFF position before proceeding. • Plug the detachable AC cord into the back of the Medically Approved Power Supply. • Plug the power supply outlet into the back of the Cavitron® Select™ Reservoir Pump. • Plug the 3 prong grounded plug into an approved outlet.
6.7 Lavage Control Detachable Handpiece Cable Assembly Connection • Plug the Handpiece outlet into the receptacle on the back of the Cavitron® Select™ Reservoir Pump. • Holding the Handpiece Cable Assembly seat male connector into appropriate female connector. • Once seated, snap male connector into female connector by pushing on the end of the male connector. • Rotate Lavage Control on the handpiece to middle setting. Turn on the on/off switch (power adjustment). This energizes the system and the ultrasonic unit is ready for operation.
6.6 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the Cavitron® Select™ Reservoir Pump and push it in until firmly seated.
ENGLISH • 7
Section 7: Cavitron® Select™ SPS™ Ultrasonic Scaler Description 7.1 System Controls: Cavitron® Select™ SPS™ Ultrasonic Scaler Magnetic Handpiece Holder Holds the System's Handpiece when system is not being used. Magnet integrated into plastic housing.
Power & Boost Indicator Light Green when system is switched ON. Orange when “Boost” Mode is activated.
Power Control (ON/OFF Switch) Turn knob to select ultrasonic power level for operation: clockwise increases system power, counter clockwise decreases system power.
7.2 System Controls: Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System Reservoir Pump Lavage dispenser.
Handpiece Holder Holds the Handpiece when it is not being used. A magnet is integrated into the plastic housing.
Power Control (ON/OFF Switch) Turn knob to select ultrasonic power level for operation: clockwise increases system power, counter clockwise decreases system power.
Power & Boost Indicator Light Green when system is switched ON. Orange when “Boost” Mode is activated.
ENGLISH • 8
Refillable Cavitron® Select™ Water Reservoir Cap
Section 7: Cavitron® Select™ SPS™ Ultrasonic Scaler Description cont'd.
7.3 Handpiece The handpiece accepts all Cavitron® 30K Ultrasonic Inserts. TM
Handpiece Connector(Handpiece & Cable Assembly are Keyed) Cable Assembly Metal ring is integrated into plastic flow control knob.
Steri-Mate® Handpiece
Date Codes (MMYY) Prior to connecting, align Handpiece and Cable Assembly electrical connections. If Cable Assembly does not seat into Handpiece, gently rotate handpiece until contacts align.
Insert Port
DO NOT TWIST WHILE PUSHING CONNECTOR INTO HANDPIECE.
Autoclave Symbol- Sterilizable up to the temperature specified
30K
Blue Lavage Knob Lavage Control Adjustment Holding open end of handpiece upright, lavage can be increased by holding the handpiece and gently twisting the blue lavage adjustment knob clockwise.
ENGLISH • 9
Section 7: Cavitron® Select™ SPS™ Ultrasonic Scaler Description cont'd.
7.4 Cavitron® 30K Ultrasonic Inserts TM
The many styles of Cavitron® 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications. See enclosed booklet for specific information. TM
O-Ring Provides seal for handpiece coolant.
Insert Tip Shape and size of tip determines access and adaptation. Lavage directed to tip.
Connecting Body Transfers and amplifies mechanical motion of stack to insert tip.
Finger Grip Magnetostrictive Stack Converts energy provided by the handpiece into mechanical oscillations used to activate the insert tip.
Insert Marking Manufacturer, Date (YYMM), Frequency, Type, Tip Lot
Hold the handpiece in an upright position. Activate the Foot Control to bleed any air bubbles that might be trapped inside the handpiece. Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Fully seat insert with a gentle push-twist motion. DO NOT FORCE.
7.5 Foot Control Information & Operation
IPX1
The Foot Control is a two position momentary switch, which activates both ultrasonic energy and Lavage Flow at the insert tip. Foot Control released...
Foot Control depressed half way... (1st position)
Both ultrasonic activation and irrigating flow stop.
The ultrasonic insert tip is activated and irrigating water flows.
ENGLISH • 10
Foot Control fully depressed... (2nd position)
This activates the “Boost” Mode.
Section 8: Accessories 8.1 Accessories List 1. Cavitron Select™ SPS™ Lavage Control Detachable Handpiece Cable Assembly with Swivel 2. Steri-Mate Detachable Sterilizable Handpiece 3. Medically Approved Power Supply 4. AC Power Cord Set 5. Cavitron® Ultrasonic Inserts 6. Dual Position Foot Control Assembly 7. DualSelect Dispensing System 8. Cavitron® Select™ Reservoir Bottle 9. Cavitron® Select™ Pump ®
*For detailed information, contact your local DENTSPLY® Representative or authorized DENTSPLY® Distributor.
8.2 Medicaments The following medicaments are compatible with the Cavitron® Select™ SPS™ when used with the Cavitron® Select™ Reservoir Pump: • Chlorhexidine Gluconate 0.12% (Oris™ CHX Antimicrobial Irrigating Solution) • Cetylpyridinium Chloride (Cepacol®) • Essential Oils (Listerine) • Hydrogen Peroxide, 3% USP • Povidone Iodine, 10% Solution • Saline Solution • Sanguinaria Extract (Viadent®) • Sterile Water, USP (For irrigating and System maintenance purposes).
Section 9: Techniques for Use 9.1 Patient Positioning • For optimal access to both the upper and lower arches, the backrest of the chair should be adjusted as for other dental procedures. This assures patient comfort and clinician visibility.
• Hold the empty Handpiece in an upright position. Activate the Foot Control until fluid exits. Once fluid exits, deactivate foot controller. • Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Fully seat insert with a gentle push-twist motion. DO NOT FORCE. • Activate the System. Hold the handpiece over a sink or drain. Check spray temperature to verify fluid is reaching the working end of the insert tip. Adjust the Lavage Control knob to ensure adequate flow for the selected Power setting. Greater flow settings provide cooler irrigant. • It may be necessary to adjust lavage with the system in “Boost” mode (Foot Control fully depressed) so adequate fluid will be available to cool tip to tooth interface. • In general, it is suggested that a “feather-light-touch” be used both supra- and subgingivally. The motion of the activated tip and acoustic effects of the irrigating fluid, in most cases, is adequate to remove even the most tenacious calculus.
• Have the patient turn their head to the right or left. Also position their chin up or down depending upon the quadrant and surface being treated. Evacuate irrigant using either a saliva ejector or High Volume Evacuator (HVE).
9.2 Performing Ultrasonic Scaling Procedures • NOTE: Refer to the Infection Control Information card supplied with your System for general procedures to be followed at the beginning of each day and between patients. • The edges of Cavitron® Ultrasonic Inserts are intentionally rounded so there is little danger of tissue laceration with proper ultrasonic scaling technique. Whenever the insert tip is placed in the patient’s mouth, the lips, cheek and tongue should be retracted to prevent accidental prolonged contact with the activated tip. • Turn unit on.
• Periodically check the Cavitron® Ultrasonic Insert for wear with the DENTSPLY® Insert Efficiency Indicator. • The use of a saliva ejector or High Volume Evacuator (HVE) is recommended during all procedures. • Set the System’s Power Adjustment knob to the lowest power setting for the application and the selected insert.
9.3 Patient Comfort Considerations Reasons for sensitivity • Incorrect tip placement. Point should be directed away from root surfaces. • Not keeping tip in motion on tooth. Do not allow the insert to remain in a static position on any one area of the tooth. Change the insert’s path of motion. • Applying pressure. Use extremely light grasp and pressure, especially on exposed cementum. • If sensitivity persists, decrease power setting and/or move from the sensitive tooth to another and then return.
ENGLISH • 11
Section 10: System Care It is recommended that you perform the following maintenance procedures to help minimize bio-film formation in the lavage path of your Cavitron Select SPS System which could affect the lavage flow to the ultrasonic insert and scaling performance.
10.1 Daily Maintenance Start-Up Procedures at the beginning of the day (Cavitron® Select™ SPS™ Ultrasonic Scaler): 1. Open the manual shut-off valve on the dental office water supply system. 2. Turn the System ON using the Power Control knob (see illustration in Section 7). Verify the Power Indicator Light is lighted. 3. Connect a freshly sterilized handpiece to its matching cable connector. Set the Power Control knob to full power. 4. Set the Lavage Control on Handpiece to maximum. 5. Hold the Handpiece (without the Cavitron® Insert installed) upright over a sink or drain. Activate the Foot Control and flush the water line for at least 2 minutes. 6. Place a sterilized Cavitron® Insert into the Handpiece. Set the Power and Lavage Control knobs to your preferred operating position. Start-Up Procedures at the beginning of the day (Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System): 1. Carefully lock Cavitron® Select™ Reservoir Bottle into Cavitron® Select™ Reservoir Pump. 2. Fill Cavitron® Select™ Reservoir Bottle with tap, distilled water, or medicament. Replace Cavitron® Select™ Bottle cap. 3. Connect Cavitron® Select™ Scaler to Cavitron® Select™ Reservoir Pump and Bottle (see Section 6 for complete instructions). 4. Turn the System ON using the Power Control knob (see illustration in Section 7). Verify the Power Indicator Light is lighted. 5. Connect a freshly sterilized handpiece to its matching cable connector. Set the Power Control knob to full power. 6. Set the Lavage Control on the Handpiece to maximum. 7. Hold the Handpiece (without the Cavitron® Insert installed) upright over a sink or drain. Activate the Foot Control and flush the water line for at least 2 minutes. 8. Place a sterilized Cavitron® Insert into the Handpiece. Set the Lavage Control on the Handpiece and Power control knobs to your preferred operating position. Between Patients (Cavitron® Select™ SPS™ Ultrasonic Scaler): 1. Remove all ultrasonic inserts and handpiece used. Clean and sterilize inserts and handpiece and clean and disinfect all the surfaces of the cabinet, power cord, handpiece cable and foot control and cable assembly as instructed in the Cavitron Systems Infection Control Procedure booklet (#81158). To maintain sterility of the Steri-Mate® handpiece, store it in the steam sterilization pouch until needed. 2. Connect a freshly sterilized handpiece to its matching cable connector. Hold the Handpiece over a sink or drain and flush the water line of the unit as above at maximum water flow for 30 seconds. 3. Place a freshly sterilized insert into the handpiece.
ENGLISH • 12
* NOTE: Water-based disinfection solutions are preferred. Some alcohol-based disinfectant solutions may be harmful and may discolor plastic materials. Between Patients (Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System): 1. Follow steps 1–2 listed in Between Patients (Cavitron® Select™ SPS™ Ultrasonic Scaler). 2. With the power OFF, unscrew cap and fill Cavitron® Select™ Reservoir Bottle with tap, distilled water, or medicament. 3. Do not overfill Bottle with liquid into the neck of the Bottle. Securely screw cap back on Bottle. 4. Place a freshly sterilized insert into the handpiece. Shut-Down Procedures at the end of the day (Cavitron® Select™ SPS™ Ultrasonic Scaler): 1. Turn the System OFF. 2. Remove all ultrasonic inserts and handpiece used. Clean and sterilize inserts and handpiece and clean and disinfect all the surfaces of the cabinet, power cord, handpiece cable and foot control and cable assembly as instructed in the Cavitron Systems Infection Control Procedure booklet (#81158). To maintain sterility of the Steri-Mate® handpiece, store it in the steam sterilization pouch until needed. 3. Close the manual shut-off valve on the dental water supply system. Shut-Down Procedures at the end of the day (Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System): 1. Remove Cavitron® Insert from Handpiece. 2. Turn the System OFF. 3. Detach Cavitron® Select™ SPS™ Scaler from Reservoir Pump and Bottle. 4. Detach Cavitron® Select™ Reservoir Bottle from Pump (see Section 6 for complete instructions). 5. Empty entire contents of Cavitron® Select™ Reservoir Bottle. 6. Replace empty Cavitron® Select™ Reservoir Bottle to Pump. 7. Carefully lock Cavitron® Select™ Reservoir Bottle into Cavitron® Select™ Reservoir Pump. 8. Connect Cavitron® Select™ SPS™ Scaler to Cavitron® Select™ Reservoir System (see Section 6 for complete instructions). 9. Turn unit ON. 10. Hold the Handpiece (without the Cavitron® Insert installed) upright over a sink or drain. 11. To air purge the water line, activate foot control to operate the Cavitron® Select™ SPS™ Scaler for at least 30 seconds. (This will eliminate fluid to prevent it from settling in the Cavitron® Select™ SPS™ System for extended periods of time. This may also help decrease the chance of bio-film formations in the Cavitron® Select™ SPS™ water lines.) 12. Follow steps 2–3 listed in Shut-Down Procedures at the end of the day (Cavitron® Select™ SPS™ Ultrasonic Scaler).
Section 10: System Care cont'd
10.2 Weekly Maintenance: Cavitron® Select™ SPS™ Ultrasonic Scaler with Cavitron® Select™ Reservoir System
26. Detach Cavitron® Select™ Reservoir Bottle from Pump (see Section 6 for complete instructions). 27. Autoclave Cavitron® Select™ Reservoir Bottle.
1. Remove Cavitron® Insert from Handpiece. 2. Turn the System OFF. 3. Detach Cavitron® Select™ SPS™ Scaler from Reservoir System. 4. Detach Cavitron® Select™ Reservoir Bottle from Pump (see Section 6 for complete instructions). 5. Empty entire contents of Cavitron® Select™ Reservoir Bottle. 6. Replace empty Cavitron® Select™ Reservoir Bottle to Pump. 7. Carefully lock Cavitron® Select™ Reservoir Bottle into Cavitron® Select™ Reservoir Pump. 8. Connect Cavitron® Select™ SPS™ Scaler to Cavitron® Select™ Reservoir System (see Section 6 for complete instructions). 9. Turn the System ON. 10. Hold the Handpiece (without the Cavitron® Insert installed) upright over a sink or drain. 11. To air purge the water line, operate the Cavitron® Select™ SPS™ Scaler for at least 30 seconds. (This will eliminate fluid to prevent it from settling in the Cavitron® Select™ SPS™ Scaler for extended periods of time. This may also help decrease the chance of bio-film formations in the Cavitron® Select™ SPS™ water lines.) 12. Turn the System OFF. 13. Prepare a 1:10 hypochlorite (NaOCL) solution by mixing 1 part 5.25% sodium hypochlorite with 10 parts tap water. 14. Remove cap and pour this 1:10 solution into the Cavitron® Select™ SPS™ Reservoir Bottle. Do not overfill with solution (see Section 3.3 for complete instructions). 15. Securely replace cap on Bottle. 16. Turn the System ON. 17. Hold the Handpiece (without the Cavitron® Insert installed) upright over a sink or drain. 18. Operate the Cavitron® Select™ SPS™ Scaler for at least 10 seconds. 19. Turn the System OFF. 20. Leave the unit undisturbed for at least 10 minutes, BUT FOR NO MORE THAN 30 MINUTES to allow sodium hypochlorite solution to disinfect the system. As a suggestion, place a sign which states that the System is being disinfected and should not be used. 21. Unscrew cap and discard disinfectant solution. 22. Fill Cavitron® Select™ Reservoir Bottle with tap or distilled water. Do not overfill with liquid (see Section 3.3 for complete instructions). 23. Turn the System ON. 24. Operate the Cavitron® Select™ SPS™ Scaler for at least 60 seconds. 25. Turn off system. Detach Cavitron® Select™ SPS™ scaler from Select™ Reservoir System.
NOTE: ALL CHEMICALS MUST BE FLUSHED FROM THE SYSTEM BEFORE IT IS READY FOR PATIENT USE.
10.3 Water Line Filter Maintenance When the water line filter becomes discolored, the filter should be replaced to prevent reduced water flow to the Cavitron® Select™ SPS™ System. A 10-pack of replacement filters is available by ordering Part Number 90158 from your local DENTSPLY® distributor. 1. Disconnect the water supply hose from the water source. If a quick-disconnect connector is attached to the end of the hose, relieve the water pressure by pressing the tip of the connector in an appropriate container and drain the water. 2. Grasp the fittings on either side of the filter disk and twist counterclockwise. Remove the filter section from either side of the water hose. 3. Install the replacement filter onto the water hose fittings. The filter should be positioned to match up with the correct hose fitting. 4. Hand tighten the two hose fittings in a clockwise direction. Reconnect the water supply hose, operate the unit to bleed the air and test for leaks.
10.4 Pump Lavage Filter Maintenan ce When the pump lavage filter becomes discolored, the filter should be replaced to prevent reduced water flow to the Cavitron® Select™ SPS™ System. A 10-pack of replacement filters is available by ordering Part Number 90191 from your local DENTSPLY® distributor. 1. Empty the contents of the Cavitron® Select™ Reservoir bottle and disconnect it from the pump. 2. Remove the old filter and discard. 3. Insert the small end of the new filter until filter is seated.
10.5 O-Ring Maintenance Before placing Cavitron® Inserts into the handpiece, performing the following steps will increase the O-ring life on Cavitron® Inserts: • Fill entire handpiece with water. • Lubricate O-ring (black or green rubber material) with water. • Gently twist the insert down into the handpiece until the insert is fully seated in place.
ENGLISH • 13
Section 11: Troubleshooting Although service and repair of the Cavitron® Select™ SPS™ should be performed by DENTSPLY® personnel, the following are some basic trouble shooting procedures that will help avoid unnecessary service calls. Generally, check all lines and connections to and from the System, a loose plug or connection will often create problems. Check the settings on the System’s knobs.
11.1 Troubleshooting Guide System will not operate: (Power Indicator Light is not lighted.) 1. Check that the Power Control knob is in the ON position, and that the detachable Power Cord is fully seated in the receptacle on back of the Cavitron® Select™ SPS™ System. 2. Check that the System’s three-prong plug is fully seated in an appropriate AC receptacle, and that AC current is present. 3. Check that the DC power plug is properly inserted into the DC input located on the back of the System. (Power Indicator Light is lighted.) 1. Check that the Foot Control Connector is fully seated in the Foot Control Receptacle on the back of the System. 2. Check that the Handpiece Cable Assembly is fully seated in the Handpiece Receptacle.
ENGLISH • 14
System operates: (No lavage flow at insert tip.) 1. Assure that Lavage control is properly adjusted. 2. Check that irrigant supply control valve/s (dental office water supply) are open. 3. Check that the Handpiece Cable Assembly is fully seated in the Handpiece Receptacle. 4. Check to see if Water Line or Pump Lavage Filter is discolored-if so, replace. 5. Check to assure that liquid is available in reservoir bottle.
11.2 Technical Support and Repairs For technical support and assistance call 1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other areas, contact your local DENTSPLY® representative.
Section 12: Specifications Cavitron® Select™: 100-240 VAC Systems Electrical Voltage 30 VDC +/- 5% Current 2 amps DC Max Frequency 50/60 Hertz Water Pressure 20 – 40 psig Flow Rate Minimum setting (CCW) 0 – 15 ml/min Maximum setting (CW) >45 ml/min Weight 425 g / 15 oz. Dimensions Height: 2.4 in. Width: 3-1/8 in. Depth: 7 1/2 in. Handpiece Cable: 72 in. Foot Control Cable: 72 in. Power Cord: 2m Water Supply Line: 96 in. Cavitron® Select™ Power Supply (674022001): Electrical Voltage 100 – 240 VAC Current 1,6 amps Max @ 100 – 240 VAC Insulation Requirement Reinforce Frequency 47 – 63 Hertz Weight 550 g / 19.4 oz. Max Dimensions Height: 14,3 cm / 5.7 in. Width: 7,6 cm / 3.0 in. Depth: 14,3 cm / 5.7 in. Power Cord: 213 cm / 84 in. Footswitch Not for operating theatres Protection Class IPX1 Cavitron® Select™ Reservoir Pump: Electrical Voltage 30 VDC +/- 5% Pump Pressure 15 – 19 psig Flow Rate 40 ml/min Weight 513 g / 18 oz. Dimensions Height: 10,2 cm / 4.0 in. Width: 8,4 cm / 3.3 in. Length: 30 cm / 11.8 in.
Symbology DIRECT CURRENT
TYPE B EQUIPMENT
FOOTSWITCH
REFER TO INSTRUCTION MANUAL FOR INFORMATION
Operating Environment: Temperature: 15 to 40 Deg. Celsius Relative Humidity: 30% to 75% (non-condensing) Transport and Storage Conditions: Temperature: 0 to 70 Deg. Celsius Relative Humidity: 10% to 95% (non-condensing) Atmospheric Pressure: 500 to 1060 hPa
Section 13: Classifications • Type of protection against electric shock: Class 1 • Degree of protection against electric shock: Type B • Degree of protection against the harmful ingress of water: Ordinary • Mode of operation: Continuous • Degree of safety of application in the presence of a flammable Equipment not suitable for use in the presence anaesthetic mixture with air or with oxygen or nitrous oxide: of flammable anesthetics or oxygen. • According to medical device directive: IIA (rule 9)
Section 14: Disposal of Unit • Accordance with local and state laws.
ENGLISH • 15
ENGLISH • 16
DEUTSCH • 32
FRANÇAIS • 48
ITALIANO • 64
Worldwide Service Centers Centres de service après-vente mondiaux Centros de servicio mundiales Service-Zentren im Ausland Centri di assistenza internazionali
U. S. A.
DENTSPLY Professional Technical Service and Repair Department 1301 Smile Way York, PA 17404 Phone: (800) 989-8826
Deutschland
France
Australia
DENTSPLY DeTrey GmbH De-Trey-Strasse 1 D-78467 Konstanz Germany Phone: 7531 583 0
DENTSPLY DeTrey 17 Michael FARADAY 78380 Montigny Le Bretonneux France Phone: (1) 30 14 77 77
DENTSPLY (Australia) Pty. Ltd. 204-206 Gipps Street Abbotsford, Victoria 3067 Australia Phone: (3) 9417 1666 (3) 9417 1777
United Kingdom
Italia
Canada
DENTSPLY Ltd. Hamm Moor Lane Addlestone, Weybridge Surrey KT15 2SE England Phone: (0) 1932 853422
DENTSPLY DeTrey Italia s.r.l. Via A. Cavaglieri, 26 I-00173 Roma Italia Phone: (06) 723 3626
DENTSPLY Canada 161 Vinyl Court Woodbridge, Ontario L4L 4A3 Canada Phone: (905) 851-6060
Swiss Representative DENTSPLY DeTrey Sàrl Baar Office Oberdorfstrasse 11 6342 Baar Switzerland
Printed in U.S.A. Form #90295 Rev B 10/03
Manufactured by DENTSPLY Professional DENTSPLY International 1301 Smile Way York, PA 17404
ESPAÑOL • 80
EC
REP
DENTSPLY DeTrey GmbH De-Trey-Str.1 78467 Konstanz, Germany
Distributed by DENTSPLY Canada Woodbridge, Ontario L4L 4A3