DePuy Synthes
Synthes Air Pen Drive System Ref J8803A Users Manual June 2010
Users Manual
65 Pages
Preview
Page 1
Air Pen Drive. Compact system for a wide range of applications.
User’s Manual
Synthes Air Pen Drive System
Table of Contents
Introduction
Operating Instructions
Product Information
Indications
2
Specifications
3
Drive Unit
13
Hand Switch
14
Foot Switch
20
Irrigation
21
Drilling Speed Attachments
24
Screw Insertion Attachments
28
K-Wire Attachment
29
Sawing Attachments
30
Burring Speed Attachments
35
Hudson Coupling Attachment
40
Craniotomy Attachments
41
Drill/Burr Attachment, for 2.35 mm Shafts
43
Adaptor for Intra-Coupling
45
Sterilization
46
Troubleshooting
47
Set Lists
49
Cutting Tools
53
Synthes
Indications
The Air Pen Drive is indicated for screw insertion, pin and wire placement, cutting of bone and metal, drilling, reaming, decorticating, shaping and smoothing of bones and teeth in a wide variety of surgical procedures, including general orthopaedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngological, reconstructive, and spine surgery.
2
Synthes Air Pen Drive User’s Manual
General traumatology
Spine surgery
Maxillofacial surgery
Hand surgery
Foot surgery
Neurosurgery and ENT
Specifications - System
Environmental conditions Operation
Transportation and storage
Temperature
10 °– 40°C 50 °– 104°F
-20 °– 50°C -4 °– 122°F
Relative humidity
30 – 75%
10 – 75%
Atmospheric pressure
500 – 1060 hPA 0.5 – 1.06 bar
500 – 1060 hPA 0.5 – 1.06 bar
Duty cycles Intermittent operation with 6.5 bar
Xmin on
Ymin off
Cycles
Drilling Attachment
5
3
unlimited
Burr Attachment
unlimited –
unlimited
Hudson Coupling Attachment
5
3
unlimited
Reciprocating Saw Attachment
3
2
unlimited
Oscillating Saw Attachment
1
2
unlimited
Sagittal Saw Attachment
1
2
unlimited
Intermittent operation with 12 bar*
Xmin on
Ymin off
Cycles
Burr Attachment
10
10
2
Hudson Coupling Attachment
3
5
unlimited
Caution – The Air Pen Drive must never be operated with oxygen, due to danger of explosion. The Air Pen Drive should also not be stored or operated in an explosive atmosphere. – Dirt inside adaptors, air hoses, angled coupling, and pen can cause loss of power. – Heavy side loads and long operating periods may cause an attachment to overheat, making it uncomfortable to hold. – Never place an overheated Drive Unit on the patient, or on patient draping during surgery. – Mitigate overheating of an attachment by discontinuing use until it cools completely, or by using the attachment intermittently. – Never immerse an attachment or the Drive Unit, or wrap with a moist sterile towel, as this will attract excessive moisture to the internal components.
* For 10 minutes maximum.
Synthes
3
Specifications - Air Pen Drive (05.001.080)
– Weight: 169 g (5.96 oz) – Length: 144 mm (5.7 in) – Recommended pressure: 6 – 8 bar – Recommended air flow rate: 200 liters/minute at 6.5 bar For maximum 10 minutes (continuous operation) it is possible to run the Air Pen Drive up to 12 bar with the following attachments: 05.001.045 – 05.001.050, 05.001.055, 05.001.063, 05.001.059, 05.001.123, 05.001.177 – Continuously variable speed: 0 – 60,000 rpm at 6.5 bar 0–80,000 rpm at 12 bar (only with 05.001.045–05.001.050, 05.001.055).
3
2 4
1 5
7
6
Note: Use a regulator to verify that the pressure of the air supply is within the recommended pressure range for the system. Mode Switch To avoid an unintentional change of the operating mode, the mode switch locks automatically. To move the mode switch, the lock has to be pushed backward. Release the lock to secure the mode switch in the desired position. By turning the mode switch to the HAND position , the Drive Unit can be used with the Hand Switch. In the FOOT position , only the Foot Switch can be used. If the FOOT position is set and no Foot Switch is attached, the Drive Unit will run with full speed. Either a Hand Switch or a Foot Switch can be used for speed control. The LOCK position is used for safety when changing attachments and tools. This prevents accidental start-up of the unit. For instructions on mounting the attachments, see “Operating Instructions” (pages 18 and 19). Adjusting the Drive Unit
7
Mode switch Release ring for attachment Coupling for hoses LOCK position HAND position FOOT position Locking mechanism for mode switch
4
Synthes Air Pen Drive User’s Manual
1 2 3 4 5 6
Lock position
Hand switch position
Foot switch position
Specifications - Angled Coupling (05.001.085) and Hand Switch (05.001.082)
Angled Coupling The Angled Coupling (05.001.085) connects the Air Hose to the Drive Unit and is used to guide the Air Hose away from the Drive Unit at a 45° angle. It allows 360° rotation.
Hand Switch – May be removed without disconnecting the Air Hose from the Drive Unit – Allows variable speed control of the Drive Unit – Includes a safety switch and pull-out finger rest 1 2 3 4 5
2
5
Positioning arrow Pull-out finger rest Positioning arrow Guide groove Locking switch
1
3
4
Synthes
5
Specifications - Foot Switch (05.001.081)
Foot Switch – Dimensions: 267 mm x 160 mm x 47 mm (bar included: 151 mm) 10.5 in x 6.3 in x 1.9 in (bar included: 5.9 in) – Allows variable speed control of the Drive Unit – Includes an irrigation ON / OFF switch
1
2
3
5
Pedal Irrigation ON/OFF button Plug for Double Air Hose Plug for Irrigation Flow Control Unit Plug for Air Hose
6
Synthes Air Pen Drive User’s Manual
1 2 3 4
4 5
Specifications - Attachments for Air Pen Drive
Drilling Attachments The drilling speed attachments are geared for a maximum speed of 1,800 rpm. 05.001.030 05.001.031 05.001.032 05.001.110
Drill Attachment, mini quick coupling Drill Attachment, J-latch coupling Drill Attachment, AO coupling Drill Attachment, hex coupling
Function: – Accept drill bits and burrs with mini quick, J-latch, hex and AO couplings
45° Drill Attachments 05.001.042 45° Drill Attachment, mini quick coupling 05.001.043 45° Drill Attachment, J-latch coupling 05.001.044 45° Drill Attachment, AO coupling Function: – Accept drill bits and burrs with mini quick, J-latch and AO couplings
90° Drill Attachments 05.001.035 90° Drill Attachment, short, mini quick coupling 05.001.036 90° Drill Attachment, long, mini quick coupling Functions: – Improve visibility during operations with narrow access – Accept drill bits and burrs with mini quick coupling
Synthes
7
Specifications - Attachments for Air Pen Drive
45° Drill Attachment, cannulated 05.001.120 Drill Attachment, with Jacobs Chuck Speed: – 0 to 1,800 rpm Function: – 1.6 mm cannulation permits the use of this attachment for drilling and reaming over K-wires (e.g., for cannulated screws and for cup and cone technique)
45° Oscillating Drill Attachment 05.001.033 45° Oscillating Drill Attachment, mini quick coupling Frequency: – 3,200 oscillations / minute Function: – Accepts drill bits and burrs with mini quick coupling – Prevents tissue and nerves from wrapping around the drill bit or burr
Screw Insertion Attachments 05.001.028 Screw Insertion Attachment, AO coupling 05.001.029 Screw Insertion Attachment, hex coupling 05.001.034 Screw Insertion Attachment, mini quick coupling Speed: – 0 to 400 rpm Function: – Accept mini quick, hex, or AO coupling screwdriver shafts
K-Wire Attachment 05.001.037 K-Wire Attachment Speed: – 0 to 2,700 rpm Function: – To insert/remove Kirschner wires 0.6 mm to 1.6 mm (0.02 to 0.06 in) in diameter, in any length
8
Synthes Air Pen Drive User’s Manual
Sawing Attachments
05.001.038
Oscillating Saw Attachment
Frequency: – 16,000 oscillations/minute Function: – Accepts Synthes crescentic and mandibular saw blades for Pen Drive
05.001.039
Sagittal Saw Attachment
Frequency: – 22,000 oscillations/minute Function: – Accepts Synthes sagittal saw blades for Pen Drive
05.001.040
Reciprocating Saw Attachment
Frequency: – 18,000 oscillations/minute Function: – Accepts Synthes reciprocating saw blades and Synthes rasps for Pen Drive
05.001.121
Guide for K-wires
Function: – Attaches to oscillating saw attachment – Accepts wires up to 1.6 mm diameter
Synthes
9
Specifications - Attachments for Air Pen Drive
Burring Speed Attachments The burring speed attachments transfer the speed of the Drive Unit, with a maximum speed of 60,000 rpm (at 6.5 bar) and 80,000 rpm (at 12 bar). Function: – Accept Synthes short, medium and long burrs for Pen Drive 05.001.045
Burr Attachment, short
05.001.048
Burr Attachment, short, angled
05.001.046
Burr Attachment, medium
05.001.049
Burr Attachment, medium, angled
05.001.047
Burr Attachment, long
05.001.050
Burr Attachment, long, angled
10
Synthes Air Pen Drive User’s Manual
05.001.063
Burr Attachment, extra long (XL), angled
05.001.055
Burr Attachment, extra extra long (XXL), angled
05.001.123
Drill / Burr Attachment, for 2.35 mm Shafts
05.001.059
Craniotome Attachment
05.001.051 05.001.052 05.001.053
Dura Guards Short Medium Long
Additional attachments 05.001.177
Hudson Coupling Attachment Maximum speed: 620 rpm
05.001.103
Adaptor for Intra-Coupling Maximum speed: 60,000 rpm
Synthes
11
Specifications - Attachments for Air Pen Drive
Speed/Torque values for Air Pen Drive Attachments Operating Pressure 6.5 bar 8.0 bar 12.0 bar
Maximum Speed (rpm) Drill Burr Attachments Attachments 1800 60000 1975 65800 n/a 78000
Hudson Attachment 620 680 810
K-Wire. Attachment 2700 2960 n/a
Screw Insertion Attachments 400 440 n/a
Operating Pressure 6.5 bar 8.0 bar 12.0 bar
Maximum Torque (Nm) Drill Burr Attachments, Attachments 0.970 0.035 1.200 0.042 n/a 0.050
Hudson Attachment 3.120 3.630 4.250
K-Wire. Attachment 0.650 0.800 n/a
Screw Insertion Attachments 4.180 5.100 n/a
Drill Attachments
05.001.030 05.001.031 05.001.032 05.001.033 05.001.035 05.001.036 05.001.042 05.001.043 05.001.044 05.001.110 05.001.120
Burr Attachments
05.001.045 05.001.046 05.001.047 05.001.048 05.001.049 05.001.050 05.001.055 05.001.059 05.001.063 05.001.123
Hudson Coupling Attachment
05.001.177
K-Wire Attachment
05.001.037
Screw Attachments
05.001.028 05.001.029 05.001.034
12
Synthes Air Pen Drive User’s Manual
Operating Instructions - Drive Unit (05.001.080)
Connecting the Air Hose to the Drive Unit Double Air Hoses 05.001.083* 3 meter hose 05.001.084
5 meter hose
Connecting the Air Hose to the Drive Unit Connect the Air Hose by fitting the pins on the hose coupling into the grooves on the coupling for hoses on the Drive Unit and turning the hose coupling clockwise. To remove the Air Hose, turn the air supply off, then run the Drive Unit to release the pressure in the hose. Turn the hose coupling counterclockwise and pull it off the Drive Unit. Angled Coupling (05.001.085) The Angled Coupling connects the Air Pen Drive and the Air Hose and is used to guide the Air Hose away from the Drive Unit at a 45° angle. It allows 360° rotation. Connecting the Angled Coupling Connect the Angled Coupling to the Drive Unit by fitting the pins into the grooves on the coupling for hoses on the Drive Unit and turning the Angled Coupling clockwise. Connect the Air Hose to the Angled Coupling by fitting the pins on the Air Hose coupling into the grooves of the Angled Coupling and turning it clockwise. To disassemble, turn the air supply off, then run the Drive Unit to release the pressure in the hose. Turn the parts counterclockwise and pull the Air Hose off the Angled Coupling, and then the Angled Coupling off the Drive Unit.
* Also available
Synthes
13
Operating Instructions - Drive Unit (05.001.080)
Attaching the Hand Switch to the Drive Unit Instrument 05.001.082
Hand Switch
Position the Hand Switch on the Drive Unit by aligning the positioning arrows on the Hand Switch and the positioning arrows on the Drive Unit above the guide grooves (Figure 1). Press down on the Hand Switch until it clicks into place on the Drive Unit (Figure 2).
Figure 1
Figure 2
14
Synthes Air Pen Drive User’s Manual
Correct position for the Hand Switch: The Synthes logo is in the upright position with the Hand Switch facing up (Figure 3). Incorrect position for the Hand Switch (Figure 4).
Figure 3
Figure 4
Synthes
15
Operating Instructions - Drive Unit (05.001.080)
Removing the Hand Switch from the Drive Unit To remove the Hand Switch, grasp the lever and pull it away from the Drive Unit.
Operating the Hand Switch For safety, slide the switch to the LOCK position during Drive Unit and/or attachment exchange.
16
Synthes Air Pen Drive User’s Manual
To operate the Drive Unit, slide the switch to the ON position. The speed can be continuously adjusted by depressing the Hand Switch.
Adjusting the Hand Switch The pull-out finger rest can be adjusted by sliding it in and out of the Hand Switch.
Synthes
17
Operating Instructions - Drive Unit (05.001.080)
Inserting attachments on the Drive Unit Important: To prevent injuries, the mode switch on the Drive Unit, or the ON / LOCK switch on the Hand Switch, should be in the LOCK position when inserting or removing attachments or instruments. The attachments can be inserted in 8 different positions, in 45° increments. Turn the release ring clockwise until it locks in the open position. The release ring will protrude slightly from the main body of the Drive Unit. Insert the attachment into the coupling, aligning the positioning pins of the attachment with the grooves on the release ring. Press the attachment lightly against the Drive Unit; the release ring will lock automatically. Pull lightly on the attachment to confirm that it is secure. Note: If the release ring closes before the attachment is fully inserted, align the positioning pins of the attachment with the grooves on the release ring and turn the attachment clockwise while applying slight pressure against the Drive Unit until the attachment engages. Pull lightly on the attachment to confirm that it is secure.
18
Synthes Air Pen Drive User’s Manual
Removing attachments from the Drive Unit While holding the Drive Unit with the attachment facing up, turn the release ring clockwise until it locks in the open position (Figure 1). Remove the attachment. The release ring should stay in the open position, ready for the next attachment (Figure 2).
Figure 1
Figure 2
Synthes
19