DGH TECHNOLOGY INC
DGH 555 Pachette 3 Operators Manual Rev 2
Operators Manual
36 Pages

Preview
Page 1
DGH 555 (PACHETTE 3) ULTRASONIC PACHYMETER
PACHETTE DGH 555 OPERATOR’S MANUAL
Equipment Manufactured By
Authorized Representative
DGH TECHNOLOGY, INC.
EMERGO EUROPE
110 SUMMIT DR IVE SUITE B EXTO N, PA 1934 1 USA (6 10) 594 -9100
EC
REP
Molenstraat 15 2513 BH, The Hague The Netherlands Phone: +31.70.345.8570
0120 555-INS-OMENG Rev: 2
i
TABLE OF CONTENTS
I.
INTRODUCTION ... 1
II.
FEATURES ... 2
III.
GENERAL DESCRIPTION... 3
IV.
POWER-UP SEQUENCE ... 9
V.
CONFIGURING THE PACHETTE 3 ... 11 V.A. V.B.
VI.
Continuous Average Mode (Factory Default Mode) ... 11 Mapping Mode ... 14
OBTAINING PACHYMETRY MEASUREMENTS ... 19 VI.A. Obtaining Measurements With Continuous Average Mode ... 19 VI.B. Obtaining Measurements With The Mapping Mode ... 21
VII.
PROBE QUALITY... 23
VIII.
VERIFYING PACHETTE 3 CALIBRATION ... 24
IX.
CARE AND MAINTENANCE... 25 IX.A. IX.B. IX.C. IX.D. IX.E. IX.F. IX.G.
X.
Cleaning and Disinfecting Instructions... 25 Transport and Storage Conditions ... 25 Operating Conditions ... 25 Warranty... 26 Lifetime / Shelf-life ... 26 Instructions for Changing Batteries ... 26 Service ... 27
INDICATIONS FOR USE, PRESCRIPTION DEVICE STATEMENT, PRECAUTIONS, INTENSITIES AND CAPABILITIES ... 29 X.A. X.B. X.C. X.D. X.E. X.F. X.G. X.H.
Indications For Use ... 29 Prescription Device ... 29 Tissue Exposure To Ultrasound Energy ... 29 Ultrasonic Intensities... 29 Biometric Measurement Capabilities ... 30 Classification ... 31 Precautions... 31 EMI / EMC Compliance... 31
ii
LIST OF FIGURES
Figure III-A DGH 555 Pachette 3 Front View. ... 3 Figure III-B DGH 555 Pachette 3 Back View. ... 5 Figure III-C DGH 555 Pachette 3 Side View ... 6 Figure III-D DGH 555 Pachette 3 Back Panel... 7 Figure III-E DGH 555 Pachette 3 Electronic CalBox ... 8
1
I.
INTRODUCTION
Ultrasonic Pachymetry is an integral part of refractive corneal surgery as well as certain screening procedures that are a function of corneal thickness. The DGH 555 Ultrasonic Pachymeter (Pachette 3) is an ultrasonic pachymeter that uses echo spike techniques to measure the thickness of the cornea. This is the recommended method for obtaining corneal thickness measurements because it offers the following advantages: • • • • •
Reproducibility High accuracy Ability to take measurements anywhere on the cornea Measurements are not dependent upon patient fixation Ease of use
The corneal thickness measurement may then be used in many different ways including: Lasik, glaucoma screening and corneal edema check for extended wear contact lens patients. The Pachette 3 was designed to provide a means of obtaining fast, accurate, corneal thickness measurements with a battery operated instrument that is simple to use, economical, and ultra-portable. We invite you to read this manual carefully to discover how quickly and easily the Pachette 3 can be integrated into your practice.
2
II.
FEATURES
The Pachette 3 is manufactured with high quality components that are designed and built using the latest technological concepts. The result is an advanced and powerful pachymeter that offers practicality and reliability. The following features are just a sample of the characteristics and capabilities of the Pachette 3. Standard Features
• • • • • • • • • • • • • • •
Simple to use. Turn on power and the Pachette 3 is ready to take measurements. Automatic measurement mode for operation without an activation switch. Rapidly obtains and stores up to 25 measurements at a single location. Displays the current measurement, the average, and the standard deviation of all measurements taken. Mapping mode for obtaining and storing up to 33 actual and biased corneal mapped measurements. 16 x 2 LCD character display with LED backlighting allows easy visibility. Proven measurement algorithm which yields accurate, reproducible measurements in a fraction of a second. Ultra-Portable. Battery operated, lightweight and can be easily transported from office to hospital in its custom carrying case. Operator feedback. Audible signal indicates when a valid measurement is complete. Personalized configuration. User-friendly keypad allows selection of number of measurements to be obtained, and selection of delay time between measurements. Configuration memory. Once personally configured for the operator, non-volatile memory provides permanent storage of configuration data, even when the Pachette 3 is turned off and the batteries are removed. Built in IOP Correction Calculation capabilities to assist in glaucoma screening. Unit comes in a protective holster with a built in tilt stand for easy viewing. When not in use the probe cord can be wrapped up and probe can be stored in the custom molded holster cavity for protection while being transported or stored. Probe is detachable and can be easily removed by the operator for cleaning or, if necessary, for replacement.
3
III. GENERAL DESCRIPTION Front View
Figure III-A DGH 555 Pachette 3 Front View. 1
LCD Display 16 x 2 character display used to present measurement data and/or configuration parameters to the operator.
2
Probe Clips Two clips molded into the protective case to place probe between cases.
3
PWR Key Pressing this key turns the Pachette 3 on. When the Pachette 3 is on, pressing and holding this key turns the Pachette 3 off. Also used in conjunction with the DEL key to enter the CalBox mode.
4
4
DEL Key Used to erase a single measurement from a group of measurements. Also used in conjunction with the PWR key to enter the CalBox mode.
5
CLR Key This key is used to erase all obtained measurements whenever it is desired to begin a new measurement sequence.
6
ENT Key In measurement mode, pressing key will display battery status. In configuration mode, the key is used to advance to the next configurable parameter. The key is also used to display the IOP correction calculation when the key is pressed and held (Continuous Avg mode only).
7
CFG Key Used to enter and exit the configuration mode. Also used to display the unit model number, serial number, software version and option number when the key is pressed and held.
8
OS Key Press key to review or take measurements of the LEFT eye. NOTE: Key is used only when the unit is configured for the bilateral mode.
/
9
Keys Used to review measurements or to program options and numerical values presented on the display.
10
OD Key Press key to review or take measurements of the RIGHT eye. NOTE: Key is used only when the unit is configured for the bilateral mode.
5
Back View
Figure III-B DGH 555 Pachette 3 Back View. 1
Probe Holder Used to hold or store probe when not in use or during transport.
2
Probe Connector Holder Used to hold or store probe connector when not in use or during transport.
3
Tilt Stand Used to place unit in tilted position on a flat surface.
6
Side View
Figure III-C DGH 555 Pachette 3 Side View
1
Probe Cord Wrap Place to wrap the probe cord when not in use or during transport.
2
Probe Connector The connector that mates to the connector on the probe cable.
3
Tilt Stand Used to place unit in tilted position on a flat surface.
7
Back Of Unit
Figure III-D DGH 555 Pachette 3 Back Panel
1
Back Panel Label This label contains the model number and serial number for the unit. It also provides contact information for DGH Technology, Inc and the following classification and attention symbols: This symbol indicates the degree of protection against electric shock. The Pachette 3 is classified as type BF equipment.
! 0120
This symbol cautions the operator to read the operating manual. This mark indicates that Notified Body 0120 (SGS United Kingdom Ltd) has certified the management system of DGH Technology, Inc. meets the requirements of Directive 93/42/EEC Annex II (excluding section 4) for ultrasonic pachymeters. This symbol located on the DGH 555 indicates that the equipment consists of
8
electronic assemblies and other components that may be subject to Directives 2002/96/EC, 2003/108/EC, and 2002/95/EC of the European parliament, which advises that electrical and electronic devices must not be disposed of as normal domestic refuse. In order to prevent environmental risks or endangerments by non-professional disposal, the disposal of this product, including any accessories, must comply with valid practices as outlined in Directives 2002/96/EC, 2003/108/EC, and 2002/95/EC and local regulations. All electronic components and systems should be returned to Orignal Manufacturer for disposal. 2
Battery Door This door provides access to the battery compartment.
Calibration Standard To check Pachette 3 calibration, an electronic Calibration Verification Box “CalBox” (below) is used to simulate the thickness of the cornea. Instructions for using the CalBox are given in section VIII and they are also printed on the CalBox label. It is recommended that calibration verification be performed at least once a day.
TECHNOLOGY, INC.
1. With the Pachette 3 turned off, disconnect the probe and then connect the CalBox cable to the Pachette 3. 2. Enter CalBox mode by holding in the DEL key and turning on the Pachette 3. 3. Press the CalBox POWER key until the LED lights up. 4. Observe that all Pachette 3 measurements are within ±5um. (See user’s manual for further details) 5. Exit CalBox mode by pressing the CLR key.
3
PACHETTE
CalBox
POWER
Model DGH 555
Figure III-E DGH 555 Pachette 3 Electronic CalBox
9
IV. POWER-UP SEQUENCE NOTE: The Pachette 3 is shipped with (2) AA batteries installed and the probe cord wrapped around the cord wrap on the protective holster, with the probe seated in the protective cavity. (When necessary, refer to section IX.E for Instructions for Changing Batteries). 1.
Remove the probe from the cavity by grasping the body of the probe at the access opening in the holster. Do not pull on the probe cord to remove the probe as this can cause damage to the probe. It is recommended that the probe be returned to the cavity for protection when transporting the Pachette 3, or when the unit is not being used.
2.
Unwrap the probe cord. The probe connector may be removed from the holster cavity by gently pulling on the probe cord at the connector strain relief.
3.
Carefully align probe connector for proper mating orientation and insert into the opening on the right side of the holster. Reference Figure IV-C for connector orientation.
4.
Gently press probe into opening until probe is properly mated. Inspect the probe tip to verify it is clean and free of any nicks, scratches or any other defect that could injure the cornea. (Refer to section IX.A. for Cleaning and Disinfecting Instructions)
5.
Flip tilt stand away from holster and place unit on flat surface in tilted position.
6.
Turn on the unit.
7.
The Pachette 3 performs an internal self-test function.
8.
Unit will briefly display battery status as indicated:
BATTERY OK E 9.
F
When the power-up sequence is finished, the display will indicate: OD
Avg 0 = 0µ µ Std Dv = 0.0µ µ
The Pachette 3 is now ready to take corneal measurements. If any default parameters need to be modified, refer to section V. Otherwise, refer to section VI for a detailed de-
10
scription of the proper method for obtaining measurements. During use, the Pachette 3 is designed to conserve battery life. Therefore the unit has three modes: Measurement Mode – Unit will take a measurement when applanated to the cornea. In this mode display will indicate: OD
Avg 0 = 0µ µ Std Dv = 0.0µ µ
Standby Mode – The unit will go into this mode if a measurement is not attempted within one minute from power up or a previous measurement. The one minute factory default delay can be adjusted from 0.5 to 9.5 minutes by accessing the configuration menu as described in section VI. The display will stay active, but a measurement cannot be obtained. In this mode you can still access the configuration menu and review all measurements. When the unit enters this mode you will hear a beep and the display will have a flashing cursor in the bottom left corner as indicated: OD
Avg 0 = 0µ µ Std Dv = 0.0µ µ
To return to measurement mode, press the PWR key. Sleep Mode – If any key is not pressed for three minutes after the unit went into standby mode, the unit will power down into Sleep Mode. In this mode the display will go blank and the unit will appear to be off, but all the measurements that were taken can still be retrieved. Press the PWR key and the unit will perform an internal test, display battery status, and then the display will indicate:
Clear All Meas? ↑=Yes ↓ =No Press the
key to clear all measurements.
Press the
key to retrieve all measurements.
11
V.
CONFIGURING THE PACHETTE 3
Configuration Overview When shipped from the factory, the Pachette 3 is ready to take corneal measurements. It is not necessary to setup or configure anything. However, the Pachette 3 has been designed to allow the operator to modify certain default parameters to tailor the instrument to meet one’s needs. Once modified, these parameters are permanently stored in nonvolatile memory and are automatically recalled each time the unit is powered up. To change a parameter, the operator must access the configuration menu. The following procedure explains how to access the configuration menu and modify the default parameters. The Pachette 3 was designed to obtain multiple corneal measurements at a single location and generate an average for those measurements. In addition, the mapping mode permits the operator to take a single measurement at different positions on the cornea. With this mode enabled, the Pachette 3 can be programmed to record measurements at a maximum of 33 different positions. However, the operator must keep track of which measurement belongs to each corneal position. This can be achieved with the help of corneal thickness charts, which are available upon request from DGH Technology, Inc. Changing Parameters On The Configuration Menu A complete list of all the possible parameters on the configuration menu is given in Table 1. The procedures below show how to activate the configuration menu and modify the parameters on the Pachette 3.
V.A. Continuous Average Mode (Factory Default Mode) 1.
Press the CFG key and the display will indicate:
OPERATIONAL MODE Continuous avg 2.
or keys to Press ENT key to continue in Continuous Avg mode or press the change the operational mode from Continuous Avg to Mapping. Refer to section V.B to continue in Mapping Mode. In Continuous Avg mode display will indicate:
12
STD. DEVIATION Enabled 3.
Use the or keys to enable or disable standard deviation. Press the ENT key to confirm your selection and the display will then indicate:
BILATERAL MODE Enabled 4.
or keys to enable or disable bilateral mode. Press the ENT key to conUse the firm your selection and the display will then indicate:
AUTO SWITCH OD/OS Enabled 5.
Use the or keys to enable or disable auto switch mode. Press the ENT key to confirm your selection and the display will then indicate:
OD/OS SWITCH DLY 4.0 sec 6.
Use the or keys to select the time (in seconds) that the unit will wait, after completing measurements on one eye, before switching to the other eye. Press the ENT key to confirm your selection and the display will then indicate:
NUMB OF MEAS 25 7.
Use the or keys to select the total number of measurements, from 1 to 25, that you wish to obtain. Press and release either key to change the current value slowly, one number at a time, or press and hold the key to scroll through the values more quickly. Press the ENT key to confirm your selection and the display will then indicate:
AUTO REP DELAY <50 msec
13
8.
Use the or keys to select the desired value for the auto repeat delay. This is the period of time that the unit will wait between consecutive measurements while the probe is properly applanated to the cornea. Press the ENT key to confirm your selection and the display will then indicate:
DELAY TO STANDBY 1.0 min 9.
Use the or keys to select the desired value for the standby delay. This is the period of time that the unit will wait before entering standby mode if no measurements are attempted. Press the ENT key to confirm your selection and the display will then indicate:
BATTERY TYPE Rechargeable 10. Use the or keys to select battery type, Alkaline or Rechargeable. Press the ENT key to confirm your selection and the display will then indicate:
BACK LIGHT OFF in Standby 11. Use the or keys to select the type of back light desired. The choices are back light always on, always off, or normally on and then shutting off in Standby mode. Press the ENT key to confirm your selection and the display will then indicate:
BRIGHTNESS
12. Use the or keys to select the desired display brightness. Press the ENT key to confirm your selection and the display will then indicate:
CONTRAST
13. Use the or keys to select the desired display contrast. Press the ENT key to confirm your selection and the display will then indicate:
14
VOLUME
14. Use the
or
keys to select the desired alarm volume.
15. Press the ENT key to return to the beginning of the configuration mode or press the CFG key to exit the configuration mode. If you did not make any changes the display will read “Config Not Changed” and return to measurement mode. If you changed any parameter the display will indicate:
Save New Config? ↑=Yes ↓ =No Press press
to save changes and return to measurement mode using new parameters, or to restore previous configuration and return to measurement mode.
Note: You can press the CFG key at any time to exit the configuration mode. 16. Pachymetry measurements may now be taken. Refer to section VI for a detailed description of the proper method for obtaining measurements.
V.B. Mapping Mode 1. Press the CFG key and the display will indicate:
OPERATIONAL MODE Continuous Avg 2. Use the or keys to change the operational mode from Continuous Avg to Mapping. Press the ENT key to confirm your selection and the display will then indicate:
BILATERAL MODE Enabled 3. Use the or keys to enable or disable bilateral mode. Press the ENT key to confirm your selection and the display will then indicate:
15
AUTO SWITCH OD/OS Enabled 4. Use the or keys to enable or disable auto switch mode. Press the ENT key to confirm your selection and the display will then indicate:
OD/OS SWITCH DLY 4.0 sec or keys to select the time (in seconds) that the unit will wait, after 5. Use the completing measurements on one eye, before switching to the other eye. Press the ENT key to confirm your selection and the display will then indicate:
NUMB OF POSN 33 6.
Use the or keys to select the number of positions that will be measured. Press ENT key to confirm your selection and the display will then indicate:
DISP BIAS MEAS Disabled 7.
Use the or keys to enable or disable the display of biased measurements (If you choose to disable this feature press ENT and go to step 9). Press ENT key to confirm your selection and the display will then indicate:
AMOUNT OF BIAS 100% 8.
Use the or keys to select the percentage that will be used to calculate the biased measurement. Press ENT key to confirm your selection and the display will then indicate:
GOOD MEAS DELAY 1.0 sec 9.
Use the
or
keys to select the time (in seconds) that the unit will wait before
16
storing the current measurement and advancing to the next measurement position. Press ENT key to confirm your selection and the display will then indicate:
POOR APPL DELAY 2.0 sec 10. Use the or keys to select the time (in seconds) that the unit will wait after a poor applanation before advancing to the next measurement position. Press ENT key to confirm your selection and the display will then indicate:
DELAY TO STANDBY 1.0 min or keys to select the desired value for the standby delay. This is the 11. Use the period of time that the unit will wait before entering standby mode if no measurements are attempted. Press the ENT key to confirm your selection and the display will then indicate:
BATTERY TYPE Rechargeable 12. Use the or keys to select battery type, Alkaline or Rechargeable. Press the ENT key to confirm your selection and the display will then indicate:
BACK LIGHT OFF in Standby 13. Use the or keys to select the type of back light desired. The choices are back light always on, always off, or normally on and then shutting off in Standby mode. Press the ENT key to confirm your selection and the display will then indicate:
BRIGHTNESS
14. Use the or keys to select the desired display brightness. Press the ENT key to confirm your selection and the display will then indicate:
17
CONTRAST
15. Use the or keys to select the desired display contrast. Press the ENT key to confirm your selection and the display will then indicate:
VOLUME
16. Use the
or
keys to select the desired alarm volume.
17. Press the ENT key to return to the beginning of the configuration mode or press the CFG key to exit the configuration mode. If you did not make any changes the display will read “Config Not Changed” and return to measurement mode. If you changed any parameter the display will indicate:
Save New Config? ↑=Yes ↓ =No Press press
to save changes and return to measurement mode using new parameters, or to restore previous configuration and return to measurement mode.
Note: You can press the CFG key at any time to exit the configuration mode. 18. Pachymetry measurements may now be taken. Refer to section VI for a detailed description of the proper method for obtaining measurements. Note: When changing operational modes (Continuous Avg to Mapping, or Mapping to Continuous Avg mode) any measurements obtained in the previous mode will be deleted before the selected mode becomes active. This is necessary because the two operational modes are very different and measurements from the previous mode would have no relevance if carried over to the new mode.