Quick Guide
2 Pages
Preview
Page 1
QUICK START GUIDE – Seattle PAP plus – Bubble CPAP System Q
A Hose extention B Heated inspiratory hose (blue) C Connector for inspiratory hose
A
J
32 3 2 30 3 0
15 14 13 12 11 10 9 8 7 6
D Patient-side temperature sensor
20 20
5
E Humidifier-side temperature sensor
10 10
4 3 2 1 0
50
60
3,2 3 ,2 2
70 80
40
F Connector for inspiratory hose heater
90 30
O 21
100%
B
P
G O2 hose
B
G
K
H O2
F C
> 80 cm (31,5 in)
D
E
L
I
H Connector for attaching the O2 hose to the blender I
Connection port for a pressure measuring line
J
Water bag port
K Connector with premounted pressure relief valve (Opening pressure at 8 L/min 17+-3 mbar) L Connector for an oxygen measurement device M Humidifier chamber
M O
N Bubble CPAP chamber O Swiveling filling funnel and connector for expiratory hose
10 9 8 7 6 5
Correct Water Level
N Please note: This Quick Reference Guide is not a replacement or substitute for the instructions for use. Any use of the medical device requires full understanding and strict observation of the instruction for use
P Expiratory hose with water trap Q BabyFlow plus
Insert the humidifier chamber in the Fisher & Paykel humidifier MR850 and remove the caps. Puncture the water bag with distilled water and open the vents of the water supply hose. Ensure that the humidifier chamber fills with water.
Use the funnel to fill the water reservoir with max. 0.9 l of distilled water. 10
To adjust the water level, use a syringe to add or remove distilled water until the correct water level line is reached.
9 8
13 12 12 11 1 1 1 0 10 9 8 7 6 5 4 3
50
60
70 80
40 90 30
O 21
Connect the O2 hose to the blender.
100%
Connect the adapter of the O2 hose (H) to the lateral opening of the connector with the pressure relief valve (fixed limit of 17 cm H2O).
H 2
The lateral opening (L) of the connector can be used for oxygen measurement if necessary. The opening (I) can be used for external pressure measurement.
Connect the heated inspiratory hose (B) to the humidifier chamber. Connect the two temperature sensors (D, E) to the inspiratory hose (B).
O2
2 1 0
L
I
D
E B 17 cm H2 H2O
Connect the expiratory hose of the breathing circuit to the filling funnel.
10 9
Set the CPAP value at the bubble CPAP chamber and lock it in place with the safety tube lock.
8
Connect BabyFlow plus directly to the breathing hoses. Performing a leakage test by closing the BabyFlow adapter for prongs or neo masks using the thumb. Set the CPAP value to 10 cm/H2O and the flow to 1 l/min. Bubbles must rise in the bubble CPAP device. Start CPAP therapy.
PDF-9508 | 19.07-2 | AW | LE | Subject to modifications | © 2019 Drägerwerk AG & Co. KGaA
Set the flow and the oxygen concentration to the required values. Permitted flow limits: 4 to 12 l/min. Recommended 8 l/min.