Instructions for Use
52 Pages
Preview
Page 1
enUS fr es
Instructions for use 3 Notice d’utilisation 19 Instrucciones de uso 35
Dräger Alcotest 3820 WARNING
!
Strictly follow the Instructions for Use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the Intended use section of this document.
i
6
5
4
4 7
8
3
9
2
1 10 en
For description of device parts, please refer to page 4
fr
Pour obtenir une description des pièces, consultez page 20
es
Para la descripción de las piezas del aparato, consulte pagina 36
2
Safety-related information en
1
Safety-related information Before using this product, carefully read the Instructions for Use. Strictly follow the Instructions for Use. The user must fully understand and strictly observe the intended instructions. Use the product only for the purposes specified in the Intended use section of this document. Do not dispose of the Instructions for Use. Ensure that it is retained and appropriately used by the product user. Comply with all local and national rules and regulations associated with this product. Only trained and competent personnel are permitted to inspect, repair and service the product (contact DrägerService or dealer). Do not use a faulty or incomplete product. Do not modify the product. Notify Dräger in the event of any component fault or failure.
2
Conventions used in this document
2.1
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows: WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
3
Dräger Alcotest 3820
3
en Description 2.2
Typographical conventions This symbol indicates information which facilitates the use of the product.
3
Description
3.1
Product overview
1 2 3 4 5 6 7 8 9
Breath-alcohol analyzer LED ring Button Grip zone on mouthpiece Mouthpiece Safety cap Mouthpiece receptacle Display USB connection (Important: Not for charging the device, for Service only) 10 Loudspeaker (Important: No reset button)
3.2
Intended Use
The Dräger Alcotest® 38201) is used for measuring the alcohol concentration in the breath of a human. The measurement is used as an indication for alcohol intoxication. Based on the measurement results, no therapeutic or legal measures should be initiated.
1)
4
Alcotest is a registered trademark of Dräger.
Use en
4
Use
WARNING Small parts of the device can come loose and be swallowed. ► Keep the device out of reach of children under the age of 3.
4.1
Prerequisites
The instrument is factory-calibrated when delivered. If the calibration date has exceeded 12 months, the instrument should be professionally adjusted, e.g. by DrägerService. Optional: If the calibration date has been exceeded, a warning message is displayed after the startup screen. Confirm the warning message with a short press on the button or wait some seconds to continue. After the last alcohol intake, wait approx. 15 minutes before a measurement is carried out. Residual alcohol that is still in the mouth may cause incorrect measuring values. This may also occur by aromatic drinks (e.g. fruit juices), alcoholic mouth sprays, medicines and drops, as well as by burping and vomiting. Rinsing the mouth with water or non-alcoholic drinks does not replace the wait time!
5
Dräger Alcotest 3820
5
en Use 4.2
Initial start-up
A quick setup process (wizard) starts when the instrument is switched on for the first time. The following settings are accessible using the wizard:
Language Date format Date Time format Time Measuring unit
Language A cursor pre-selects one language in the list. Press briefly on the button to change the position of the cursor in order to select your needed language. If the cursor selects the right language please hold the button for approx. 2 seconds to confirm. Date A cursor pre-selects the day of the date. Press briefly on the button to change the day until the actual day is shown. Please hold the button for approx. 2 seconds to confirm the day. Please continue with the month and the year in the same way: short press to change and long press to confirm. Time format A cursor pre-selects a time format. Press briefly on the button to change the position of the cursor in order to select your needed time format. If the cursor selects the needed time format please hold the button for approx. 2 seconds to confirm. Time A cursor pre-selects the hours of the time. Press briefly on the button to change the hour until the actual hour is shown. Please hold the button for approx. 2 seconds to confirm the hour. Please continue with
6
Use en
the minutes in the same way: short press to change and long press to confirm. Measurement unit A cursor pre-selects a measurement unit in the list. Press briefly on the button to change the position of the cursor in order to select your needed measurement unit. If the cursor selects the needed measurement unit, please hold the button for approx. 2 seconds to confirm. After the setup wizard is finished, the display shows Welcome and continues with READY for tests. The Alcotest® 3820 is now adjusted to your settings and is ready for breath tests. Of course these and further settings can be adjusted in the device menu (see section 4.6 on page 9).
4.3
Carrying out a breath sample A new mouthpiece must be inserted for each person who provides a breath test. For hygiene reasons, hold the mouthpiece only on the grip zones.
Do not allow any high alcohol concentrations on the sensor, e.g. by rinsing the mouth with a high proof alcohol shortly before the measurement, as this will shorten the service life of the sensor. Procedure 1. Remove the safety cap from the top of the device. 2. Insert a new mouthpiece (order no. 83 25 250) in the mouthpiece receptacle (see section 3.1 on page 4). For hygiene reasons, hold the mouthpiece only on the grip zones.
7
Dräger Alcotest 3820
7
en Use 3. If the device is switched off, keep the button pressed for approx.1 second until the startup screen is displayed. 4. After approx. 4 seconds, a READY screen is displayed. The instrument is ready to use. 5. Take a deep breath and blow evenly and without interruption into the mouthpiece. Do not block the outlet of the mouthpiece. You have to blow a volume of 1.3 liter and at least 1.6 seconds for a valid test. If blowing normally, this will take 2-3 seconds A continuous tone sounds simultaneously and stops, when the test parameters are fulfilled and the device has taken the sample. The test result is displayed after 3 to 60 seconds (depending on the instrument‘s temperature and the measured concentration). In case of a negative alcohol test, 0,00 will be displayed. In case of a positive alcohol test, the measured alcohol level will be displayed. To prepare a new measurement, press the button briefly. After a short recovery time, (see section 8 on page 16) READY will be displayed and the device is ready to use again.
4.4
Switching off the Instrument The instrument will switch off after pressing the button for approx. 3 seconds. A switch off count down will be shown as well. If the display is off, stop pressing the button. Otherwise the device will be switched on again. The instrument automatically switches off after 120 seconds without any activity. Switching off the instrument does not reduce the waiting period until it is ready to take a measurement!
8
Use en
4.5
Removing the mouthpiece
1. Pull the mouthpiece upwards out of the mouthpiece receptacle by holding it on the grip zones on the left and right side. 2. Dispose of the mouthpiece in accordance with local regulations.
4.6
Settings
To enter the device menu, the device has to be switched on and has to show READY. 1. Briefly press the button. A menu symbol and text Menu are displayed. 2. Briefly press the button again within approx. 3 seconds to enter the menu. If the button is not pressed a second time, the instrument will return to READY. The instrument will exit the menu after 120 seconds, if the button isn‘t pressed. In the main menu all menu options are listed one below the other. To select a menu option, mark the respective menu option with the cursor by briefly pressing the button several times to toggle through the menu until the cursor is at the intended position. Afterwards hold the button for 2 seconds to enter the submenu. The following table provides an overview of the menu options and about how to use them: Menu option
Description
Interaction instruction
Measuring
Option to leave the menu and to go back to the READY screen to perform a test.
Hold the button for 2 seconds to leave the menu.
9
Dräger Alcotest 3820
9
en Use
Menu option
Description
Results
Overview of the last 10 tests performed on the device (incl. result, date, time, test ID). There is one test displayed per screen.
Briefly press the button to navigate through the test result list. Hold the button for 2 seconds to leave the submenu.
Device Info
Serial no and part no of the device
Hold the button for 2 seconds to leave the submenu.
Next Service
Overview, in how many days the device… should get serviced, should be adjusted and should get an accuracy check.
Hold the button for 2 seconds to leave the submenu.
10
Interaction instruction
Use en
Menu option
Description
Date & Time
Date, time as well as date format (yyyy/mm/ dd, yyyy-mm-dd, mm/dd/ yyyy, dd.mm.yyyy) and time format (12h, 24h) can be adjusted. The submenu can be left via the Back option.
Interaction instruction
Contrast
The display contrast can be adjusted within 5 levels.
Speaker
Any acoustical signals of the device can be turned off or to a volume of 50% or 100%.
Select an option by moving the cursor to the option by briefly pressing the button. Afterwards, enter this option by holding the button for 2 seconds. Change a setting or a set number by briefly pressing the button (e.g. shifting the time). Confirm a selection by holding the button for 2 seconds. Select the Back option and hold the button for 2 seconds to leave the submenu. Change the setting by briefly pressing the button. Hold the button for 2 seconds to confirm the setting and leave the submenu. Change the setting by briefly pressing the button. Hold the button for 2 seconds to confirm the setting and leave the submenu.
11
Dräger Alcotest 3820
11
en Use
Menu option
Description
Language
The language of the display can be adjusted.
Interaction instruction
Measurement The measurement unit unit for the alcohol test results can be adjusted within the following options: %, ‰, g/L, mg/ 100 mL, mg%, mg/mL, mg/L, μg/100 mL, μg%, μg/L, g/210 L.
Software Version
12
Software version and part number are displayed.
Change the set language by briefly pressing the button. Hold the button for 2 seconds to confirm the setting and leave the submenu. Change the measurement unit by briefly pressing the button. Hold the button for 2 seconds to confirm the setting and leave the submenu. Hold the button for 2 seconds to leave the submenu.
Troubleshooting en
5
Troubleshooting
An error message is displayed in the event of an instrument fault. The number that appears below the message is used for service functions. If the fault remains after switching the instrument on and off several times, contact DrägerService or your dealer. Fault / message
Cause
Remedy
Instrument cannot be switched on.
Battery empty.
Contact DrägerService or dealer.
Volume too low / sample invalid.
Test subject is blowing Blow more strongly intermittently or not and without hard or long enough. interruptions into the mouthpiece.
Instrument outside Instrument too hot or temperature range too cold.
Allow the instrument to cool down or become warmer.
↑↑↑ (Measuring range Exceeded)
Measuring range exceeded. Residual mouth alcohol may cause incorrect measuring values.
There must be an interval of at least 15 minutes after alcohol has been taken into the mouth (see section 4.1 on page 5). To repeat test, press the button.
Volume too small
Breath sample volume Repeat breath sample. too small.
Volume too large
Measuring range exceeded.
Sample interruption Irregular outward breath, abrupt end or intake
Repeat breath sample. Repeat breath test.
13
Dräger Alcotest 3820
13
en Troubleshooting
Fault / message
Cause
Remedy
Fault number
Instrument fault
Switch the instrument off and on again. If the fault remains, contact DrägerService or dealer.
… DAYS
In … days next service Contact DrägerService is needed. or dealer for device service.
Until next service! ...DAYS Service expired!
14
Since … days service is needed.
Contact DrägerService or dealer for device service.
Maintenance en
6
Maintenance
Calibration Dräger recommends a device calibration every 12 months, because of the sensor drift the uncertainty of measurement increases over months. Instrument calibration and maintenance should only be carried out by authorized persons, e.g. DrägerService. The last 30 days before next service is needed, a day count down will be shown as information in the display after switching on the device. Please confirm this information with the button. It shall remind you that the device needs to be serviced in the next days. After the service date, the device starts to count up how many days it is over the requested service. This information can be confirmed by pressing the button as well. Cleaning Do not immerse in fluids and avoid contact of fluids with the connections. Do not use any cleaning agents containing alcohol or solvents. Wipe down the instrument with a disposable cloth.
7
Disposal
This product must not be disposed of as household waste. This is indicated by the adjacent icon. You can return this product to Dräger free of charge. For information please contact the national marketing organizations and Dräger.
15
Dräger Alcotest 3820
15
en Technical data
8
Technical data
Measuring principle Electrochemical sensor Measuring range Breath alcohol
0.000 to 0.500 g/210L
Blood alcohol concentration
0.000 to 0.500% BAC
Small values will be shown as 0.000
0.000 to 0.0063 g/210L
Ethanol mass per blood volume or mass at 20°C / 68 °F and 1013 hPa. For the above data, a conversion factor of 2100 is used for breath and blood alcohol concentration data. Units of measure %, ‰, g/L, mg/100 mL, mg%, mg/mL, mg/L, μg/100 mL, μg%, μg/L, g/210 L Ambient conditions During operation
-5°C to +50°C / +23°F to +122 °F 10 to 100 % rel. hum. (non-condensing) 600 to 1300 hPa
During storage
-20°C to +70°C / -4°F to +158 °F
Sampling Respiratory volume
>1.3 L (default)
Minimum flow rate
>10 L/min (default)
Minimum blow time
>1.6 s (default)
16
Technical data en
Measurement accuracy Standard deviation of the repeatability with ethyl alcohol standard up to 0.50 mg/L
0.008 mg/L
≥0.50 mg/L
1.7% of the measured value
up to 0.100 % BAC
0.0017 % BAC or
≥ 0.100 % BAC
1.7% of the measured value whichever value is higher
Sensitivity drift Typically 0.4 % of the measured value per month Wait time for measurement readiness following a previous measurement with: 0.25 mg/L
20 s
0.050 % BAC
20 s
0.50 mg/L
40 s
0.100 % BAC
40 s
1.00 mg/L
80 s
0.200 % BAC
80 s
>1.50 mg/L
120 s
>0.300 % BAC 120 s
Calibration every 12 months Dimensions 2 “ x 5.2 “ x 1.1 “(W x H x D) Weight approx. 120g / 4.2 oz
17
Dräger Alcotest 3820
17
en Technical data
Power supply 1 x 3 V CR123A The battery is designed for the service life of the instrument with average use. Protection class IP 52 CE markings Electromagnetic compatibility (Directive 2014/30/EU). Configuration possible, see name plate Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
18
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help.
Manufactured by Dräger Safety AG & Co. KGaA Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 20 80 90 33 598 - GA 4754.540 enUS, fr, es © Dräger Safety AG & Co. KGaA Edition 03 - July 2017 (Edition 01-June 2016) Subject to alteration