Neonatal Test Lung, reusable Consumables and Accessories The neonatal test lung is intended for the system test of Dräger ventilators. The ventilator must perform a system test, before it is used on a patient. The innovative design of the test lung helps to improve staff satisfaction by reliable testing and reprocessing.
Machine reprocessing proved1
GS1 Data Matrix Code including the date of manufacture, lot number and revision
Possibility to disassemble before reprocessing
1
Please ind the tested reprocessing method in the technical data.
D-32435-2020
Pressure resistant up to 60 mbar
Standard ISO connector
02 | Neonatal Test Lung, reusable
Benefits Compatibility tested with the following devices ‒ Babylog VN500 / VN600 / VN800 ‒ Babylog 8000 plus ‒ Evita V300 / V500 / V600 / V800 during neonatal mode Please contact your local Dräger sales representative regarding further compatibilities.
Functional design The bellow has a strap to facilitate easy disassembly and assembly. In addition, the loop allows the attachment of an internal identification mark.
Long lasting traceability The cap is lasered with a GS1 Data Matrix Code. This is a two-dimensional code in which the date of manufacture, lot number and revision are encrypted in a tamper-proof way. GS1 is a network of not-for-profit organizations that develop, negotiate and maintain standards for inter-company processes worldwide.
Made in Germany The Neonatal Test Lung is designed and manufactured in Germany. This enables us to keep the emissions low during the transport to our distribution warehouse.
Infection prevention & control The Neonatal Test Lung is a reusable product and it can be reused for multiple self tests of the device. In general, reusable products must be reprocessed according to the instruction for use before first use and after every use, in order to reduce the risk of infection. ‒ Follow the infection prevention policies and reprocessing regulations of the health-care facility and national authorities. ‒ Use a washer-disinfector that meets the requirements of the ISO 158832 standard. ‒ Use validated procedures for reprocessing and disassemble the test lung before reprocessing. ‒ Disassemble the test lung before reprocessing. ‒ Procedure: 1. Ensure components are securely placed within the washer-disinfector. 2. Attach the bellow to a suitable nozzle. 3. Ensure that all surfaces and interiors can be completely rinsed and the water can drain off freely. ²EN ISO 15883 is a standard for washer-disinfectors. It clarifies and describes the equipment requirements and the validation of the processes for reprocessing.
Neonatal Test Lung, reusable | 03
Technical Data Part number
MP20050
Tidal volume
50 ml
Compliance
0.92 ml/min at 26 mbar
ISO connector
15OD
Material Cap
Polyamide
Material Bellow
Silicone
Dimensions
5 cm x 8.5 cm
Ambient conditions During storage Temperature
-20 to 60 °C (-4 to 140 °F)
Relative humidity
10 to 95 % (non-condensing)
Ambient pressure
500 to 1100 hPa (7.3 to 15.9 psi)
During operation Temperature
10 °C to 40 °C (50 °F to 104 °F)
Relative humidity
10 to 95 % (non-condensing)
Ambient pressure
500 to 1100 hPa (7.3 to 15.9 psi)
1Reprocessing
Preliminary cleaning
Min. 1 min
Min. 10 °C (50 °F)
Tap water
Cleaning
Min. 5 min
Min. 55 °C (131 °F)
Dr. Weigert: Neodisher Medi-Clean forte (Min. 0.1 %)
Neutralizing
Min. 1 min
Min. 10 °C (50 °F)
Dr. Weigert: Neodisher Z (Min. 0.1 %)
Flushing
Min. 1 min
Min. 10 °C (50 °F)
Demineralized water
Disinfecting
Min. 5 min
Min. 90 °C (194 °F)
Drying
Drying time depends on the load
04 | Neonatal Test Lung, reusable
91 10 210 | 21.01-1 | HQ | PP | Subject to modifications | © 2021 Drägerwerk AG & Co. KGaA
Notes
Not all products, features, or services are for sale in all countries. Mentioned Trademarks are only registered in certain countries and not necessarily in the country in which this material is released. Go to www.draeger.com/trademarks to find the current status.
CORPORATE HEADQUARTERS
REGION EUROPE
REGION ASIA PACIFIC
REGION NORTH AMERICA
Drägerwerk AG & Co. KGaA Moislinger Allee 53–55 23558 Lübeck, Germany www.draeger.com
Dräger Safety AG & Co. KGaA Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 2080 [email protected] .
Manufacturer:
REGION MIDDLE EAST, AFRICA
Drägerwerk AG & Co. KGaA Moislinger Allee 53-55 23542 Lübeck, Germany
Dräger Safety AG & Co. KGaA Branch Office P.O. Box 505108 Dubai, United Arab Emirates Tel +971 4 4294 600 Fax +971 4 4294 699 [email protected] .
Draeger Singapore Pte. Ltd. 61 Science Park Road The Galen #04-01 Singapore 117525 Tel: +65 6872 9288 Fax: +65 6259 0398 asia.pacifi[email protected] .
Draeger, Inc. 3135 Quarry Road Telford, PA 18969 Tel +1 800 4DRAGER (+1 800 437 2437) Fax +1 215 723 5935 [email protected] .
Locate your Regional Sales Representative at: www.draeger.com/contact
REGION CENTRAL AND SOUTH AMERICA
Draeger Panamá Comercial S. de R.L. 59 East Street, Nuevo Paitilla, House 31, San Francisco Town Panama City, Panama Tel +507 377-9100 / Fax -9130 [email protected] .