Dräger Medical
CF800 Technical Documentation Rev 2.0
Technical Documentation
40 Pages
Preview
Page 1
Technical Documentation CF 800 for Continuous Flow CPAP Therarpy
Revision 2.0 5682 9036076
Emergency Care • Perioperative Care • Critical Care • Perinatal Care • Home Care
Because you care
Contents General 1
Notes 1.1
3
Symbols and Definitions ... 4
Function Description 1
General
7
2
Intended use
8
3
Monitoring
8
Annex Parts catalog
All rights reserved. Copyright reserved.
K5682XXXbook.IVZ.fm 11.01.06
Test List
I
II
K5682XXXbook.IVZ.fm 11.01.06
All rights reserved. Copyright reserved.
Contents
General
1
CF 800 1
Notes
General This Technical Documentation conforms to the IEC 60601-1 standard. Read each step in every procedure thoroughly before beginning any test. Always use the proper tools and specified test equipment. If you deviate from the instructions and/or recommendations in this Technical Documentation, the equipment may operate improperly or unsafely, or the equipment could be damaged. Use only original Dräger parts and supplies. The maintenance procedures described in this Technical Documentation may be performed by qualified service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer. The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of the manufacturer.
Strictly follow the Instructions for Use manual / Operating Instructions! This Technical Documentation does not replace the Instructions for Use manual / Operating Instructions. Any use of the product requires full understanding and strict observation of the product-specific Instructions for Use manual/ Operating Instructions.
Reference is hereby made to the observance of the relevant safety provisions, for example in Germany, the Medical Product Law (MPG), the Medical Device Operator Ordinance (MPBetreibV), the Pressure Container Ordinance (Druckbehälterverordnung), the Technical Rules for Pressurized Gases (Technische Regeln Druckgase), or the Occupational Health and Safety Provisions (Unfallverhütungsvorschriften). Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany.
Observe protection mark DIN 34. Copyright reserved.
Version 2.0_ Released_Printed on_11.01.06_K5682XXX_XIEC_INT_General.fm
Follow the laws and regulations applicable in your country.
5682
3
General 1.1
CF 800
Symbols and Definitions This symbol indicates a warning.
This symbol indicates tips and useful information.
This symbol is used to alert against unsafe practices when handling electrostatic sensitive devices (ESD).
= examination of actual condition
Maintenance
= measures to maintain specified condition
Repair
= measures to restore specified condition
Servicing
= inspection, maintenance, and repair
Observe protection mark DIN 34. Copyright reserved.
Inspection
4
5682
Version 2.0_ Released_Printed on_11.01.06_K5682XXX_XIEC_INT_General.fm
Definitions according to German standard DIN 31051:
Function Description
5
CF 800 1
General
Functional Description Air and O2 passes via the filter 1 and 2 into the corresponding flowmeters with metering valve 3 and 4. With the aid of the reference list the required O2 concentration is dajusted (with the proportion of the flow amount O2: air). This flow fills the reservoir 5 and passes continuously via the non-return valve 6 to the inspiration output.The airway pressure is measured with the pressure gauge 7. The relief valve 8 limits the airway pressure in the system to max. 30 mbar.The necessary breathing volume for the inspiration is provided prior with the continuous flow. Inspiration flow for the patient comes from the reservoir 5. The non-return valve 6 stops a rebreathing into the reservoir 5.In case the medical compressed gas supply fails, ambient air is sucked in via the bypass air valve 9. A self-locking coupling 10 supplies the medicaments nebulizer with compressed air.
Figure 1
Function diagram
Table 1
Legend Figure 1
All rights reserved. Copyright reserved.
Revision 2.0_ Released_Printed on_11.01.06_F5682000.fm
Pos.
5682
Designation
1
Filter
2
Filter
3
Flowmeter with valve for air
4
Flowmeter with valve for oxygen
5
Reservoir (bellows)
6
Nor-return valve
7
Pressure gauge
8
Relief valve
9
Bypass air valve
10
Coupling for medicament nebulizer
7
Functional Description 2
3
Intended use
Monitoring
CF 800 CF800 is a device for the application of respiratory therapeutical measures with continuously positive airway pressure.The device schould only be applied on patients with spontaneous breathing.The CPAP system CF 800 is a compact unit consisting of: –
Aquapor for the heating and humidifiering of the breathing air
–
AWT 01 or thermometer for the temperature measurement
–
Oxydig for the inspiratory O2-measurement
–
Pneumatic medicaments nebulizer for medicaments nebulizing
–
PEEP valve to built up positive airway pressure (CPAP)
The device has no integrated monitoring. By recognition of insufficient spontaneous breathing or malfunction of the device (disconnection, gas failure, breaking of hoses) a regular visual check by authorized medical personnel is necessary.With the Oxydig the inspiratory O2-concentration is monitored. With the AWT 01 the inspiratory breathing air temperature is automatically monitored.
Weight
3,0 kg
Measurements:
Height = max. 550 mm Width = 212 mm
8
CF 800
5682
Revision 2.0_ Released_Printed on_11.01.06_F5682000.fm
Figure 2
All rights reserved. Copyright reserved.
Depth = 260 mm
CF 800
Functional Description
Compliance of the reservoir
400 mL/mbar
Max. pressure variations at flow > 1.5 x volume per minute
+/- 2 mbar
Safety valve
limits the airway pressure to max. 30 mbar
Emergency air valve
opens at a low pressure of 0.4 mbar, makes it possible to go on with spontaneous breathing if the compresed air supply fails
Manometer for measuring the airway pressure
-10 to 30 mbar
error in measuring
2,5% of final value
Operating values
Gas supply O2-pressure
2,7 to 6 bar
Connection thread
M12x1 innen
Air pressure
2,7 to 6 bar
Anschlussgewinde
M20x1,5 außen
Gas consuption
corresponds to the set total flow
Noise during operation sound pressure acc. to DIN45635
max. 45 dB (A)
All rights reserved. Copyright reserved.
Revision 2.0_ Released_Printed on_11.01.06_F5682000.fm
The gases must be dry, free of oir and dust
5682
9
10
5682
Revision 2.0_ Released_Printed on_11.01.06_F5682000.fm
All rights reserved. Copyright reserved.
Functional Description CF 800
Annex
Parts catalog Test List
Parts catalog CF 800
Revision: 2005-06-08 11:19:33 5682.000
Emergency Care - Perioperative Care - Critical Care - Perinatal Care - Home Care
Because you care
Parts catalog
Products Concerned
Item No.
Part No.
Description
Qty.
Qty. unit
8414300
CF 800
1.000
St
8409400
CF 800 (CPAP DEVICE)
1.000
St
8410130
CF 800 P (PAED. CPAP DEVICE)
1.000
St
Remark
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
3
Parts catalog
Accessories/Consumables
Item No.
Part No.
Description
Qty.
Qty. unit
1
8409400
CF 800 (CPAP DEVICE)
1.000
St
2
8409645
ONE-WAY PATIENT CONNECTION
1.000
St
2.1
8407571
HOSE SET, DISPOSABLE
1.000
St
2.2.3
8290274
ANGLE SOCKET ISO
1.000
St
2.3.3
M25647
NOZZLE 22/22
1.000
St
2.4.3
8402041
CORRUGATED HOSE
1.000
St
2.5.3
8403685
SET 12 STAND. CONE A PLASTIC
1.000
St
3
8413146
HOSE SET E2 (BLUE)
1.000
St
7
8409646
PEEP-VALVE WITH SUPPORT
1.000
St
8
8304411
Oxydig complete
1.000
St
9.1
8405791
SENSOR SLEEVE ADULTS
1.000
St
9.3
2M13259
THERMOMETER
1.000
St
10.1
M20650
Rail clamp
1.000
St
10.2
1335812
ALCALI-BATTERY,9 V
1.000
St
Remark
part 2.1; 2.2.3; 2.3.3; 2.4.3 and 2.5.3
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
4
Parts catalog
Accessories/Consumables
Item No.
Part No.
Description
Qty.
Qty. unit
10.3
8402868
BACTERIA FILTER
1.000
St
10.4
1190520
HOSE 4X1,5-SI 50 SH A NF
1.000
m
10.5
8400904
TAPER ADAPTER
1.000
St
11
8405000
MEDICAMENT NEBULIZER (PNEU)
1.000
St
12.1
8405029
PATIENT PART (AQUAPOR)
1.000
St
12.2
8403345
SET OF SPARE BRACKETS
1.000
St
12.4
8409450
AQUAPOR SUPPORT FOR RAIL
1.000
St
13
8403201
TEST LUNG
1.000
St
14
8409609
HINGED ARM 240-DEGR.
1.000
St
14.1
8409746
SUPPORT FOR HINGED ARM A
1.000
St
14.2
8409841
HOSE CLAMP FOR HINGED ARM
1.000
St
Remark
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
5
Parts catalog
Concerned Devices
Item No.
Part No.
Description
Qty.
Qty. unit
8414700
Aquapor EL, Humidifier
1.000
St
Remark
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
6
Parts catalog
Accessories Aquapor EL
Item No.
Part No.
Description
Qty.
Qty. unit
1
8414699
Heating Top
1.000
St
1-5
8414700
Aquapor EL, Humidifier
1.000
St
2
8404739
TROUGH
1.000
St
2-5
8405029
PATIENT PART (AQUAPOR)
1.000
St
3
8404738
FLOAT
1.000
St
4
8404741
LIP SEAL
1.000
St
5
8406135
REP.SET COVER (AQUAPOR)
1.000
St
6
1190520
HOSE 4X1,5-SI 50 SH A NF
1.000
m
6-7
8405031
FILLING KIT FOR AQUAPOR
1.000
St
7
8405077
FUELLDORN,UGR.
1.000
St
8
M20650
Rail clamp
1.000
St
8-10
8403345
SET OF SPARE BRACKETS
1.000
St
9
1300849
SPLIT WASHER B5 DIN127-X12CRNI
2.000
St
10
1320440
SCREW AM5X16 DIN84
2.000
St
Remark
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
7
Parts catalog
Hose Set Adults
Item No.
Part No.
Description
Qty.
Qty. unit
1
8414699
Heating Top
1.000
St
2-5
8405029
PATIENT PART (AQUAPOR)
1.000
St
8-10
8403345
SET OF SPARE BRACKETS
1.000
St
14
8405371
TEMPERATURE SENSOR
1.000
St
15
2165627
BREATHING HOSE E 60CM
3.000
St
15-26 8412092
ADULT HOSE SET, AQUAPOR
1.000
St
16
8404985
Water trap
2.000
St
16A
8403976
POT
0.000
St
17
8403367
2.000
St
18
M25647
NOZZLE 22/22
1.000
St
19
M25649
MASKENKRUEMMER
1.000
St
20
8405435
Y-PIECE
1.000
St
21
M23840
STANDARD CONN.12,5 PLASTIC
1.000
St
22
8403566
TUBE CLAMPS
1.000
St
Remark
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
8
Parts catalog
Hose Set Adults
Item No.
Part No.
Description
Qty.
Qty. unit
23
8402041
CORRUGATED HOSE
1.000
St
24
8403685
SET 12 STAND. CONE A PLASTIC
12.000
St
25
8403076
ADAPTER E
1.000
St
26
8402918
SET OF 5 CAPS (8401644)
5.000
St
Remark
different types from size 6 to 12
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5682.000
Revision: 2005-06-08 11:19:33
9