Evita V800-V600-V500-V300 Flow Reduction InService Guide
InService Guide
10 Pages
Preview
Page 1
Flow reduction – Anti-air shower – Manual disconnection
Footnote: Please insert appropriate company
Manual disconnection August 2019, Lübeck Drägerwerk AG & Co KGaA Footnote: Please insert appropriate company
Manual disconnection Evita V800 / Evita V600 Evita V800, Evita V600 When the Manual disconnection function is activated, the device reduces flow delivery in case of disconnection. The patient can be disconnected within the next 10 seconds. When the patient is reconnected, the device resumes ventilation in the set ventilation mode. 1.
To activate manual disconnection touch the Procedures button
2.
Touch the Manual disconnection button
1 2
3. Confirm by pressing the rotary knob
3 Drägerwerk AG & Co KGaA
4
Manual disconnection Evita Infinity V500 Evita Infinity V500 When the Manual disconnection function is activated, Evita V500 reduces flow delivery in case of disconnection. The patient can be disconnected within the next 10 seconds. When the patient is reconnected, the device resumes ventilation in the set ventilation mode. 1. Touch the Special procedures... button in the main menu bar. The device opens the Special procedures dialog window
1
2
2. Touch the Functions button if the page is not already preset 3. Touch the Manual discon. button (B) and
4
4. Confirm by pressing the rotary knob The Manual disconnection maneuver can be configured as a Manual discon. button in the main menu bar to enable direct access. Drägerwerk AG & Co KGaA
3 5
Manual disconnection Evita V300 Evita V300 When the Manual disconnection function is activated, Evita V300 reduces flow delivery in case of disconnection. The patient can be disconnected within the next 10 seconds. When the patient is reconnected, the device resumes ventilation in the set ventilation mode. 1. To activate touch the Special maneuvers button in the main menu bar. Evita V300 opens the Special maneuvers dialog window 1
2. To activating manual disconnection touch the Manual discon. button 3. Confirm by pressing the rotary knob The Manual disconnection maneuver can be configured as a Manual discon. button in the main menu bar to enable direct access.
Drägerwerk AG & Co KGaA
2
4 6
Anti-Air Shower August 2019, Lübeck Drägerwerk AG & Co KGaA Footnote: Please insert appropriate company
Anti-air shower Evita V800 / Evita V600 Evita V800, Evita V600 Flow reduction – Anti-air shower When the Anti-air shower function is activated and a disconnection is detected during ventilation, the flow is reduced until reconnection is detected. Simultaneously, the “Disconnection detected” alarm is displayed.
1 2
For non-invasive ventilation, the time before the alarm is triggered can be delayed with the setting Tdiscon. The minute volume may be reduced due to the already reduced flow. The Anti-air shower function can be configured on the dialog page System setup > Ventilation > General settings The password must be entered. The settings can be activated or deactivated. Ventilation starts with the selected settings.
Drägerwerk AG & Co KGaA
3 8
Anti Air Shower Evita Infinity V500 / Evita V300 Flow reduction Anti Air Shower If a disconnection is detected, flow delivery is reduced until after reconnection.
1
1. To activate manual disconnection touch System setup … 2. Touch the Ventilation page button (A) 3. Touch the General settings tab (B)
2
4. Switch Anti Air Shower (H) on or off 3
5. Confirm by pressing the rotary knob
5
Evita V500 starts with the selected settings. 4 Drägerwerk AG & Co KGaA
9
Many thanks Footnote: Please insert appropriate company