Dräger Medical
Fabius GS Technical Service Manual Rev 3.0
Technical Service Manual
76 Pages
Preview
Page 1
Technical Service Manual Fabius GS Inhalation Anesthesia Machine
Revision 3.0 5330.500 9036095
Emergency Care • Perioperative Care • Critical Care • Perinatal Care • Home Care
Because you care
Contents General 1
Notes 1.1
3
Symbols and Definitions ... 4
Function Description 1
General Information about the Fabius GS
7
2
Function diagram of Fabius GS
12
3
Battery backup
14
4
Fabius GS piping diagram
14
5
Function description of gas box
15
6
SORC (Sensitive Oxygen Ratio Controller)
16
7
Cosy 2 breathing system
18
All rights reserved. Copyright reserved.
K5330500IECIVZ.fm 10.11.05
8
9
7.1
Ventilation mode ... 22
7.2
Function description: Manual ventilation ... 23
7.3
Function description: Spontaneous breathing ... 27
7.4
Function description: Volume/pressure control ventilation mode ... 31
7.5
Cosy 2 absorber ... 35
Ventilator
35
8.1
Safety valve ... 38
8.2
Auxiliary air valve ... 38
Pneumatics
39
9.1
PEEP/Pmax valve control ... 39
9.2
APL bypass valve control ... 40
I
Contents 10
Electrical block diagram
41
11
Function description: Control PCB
41
12
Control panel assembly
42
13
FiO2 Measurement
44
14
Respiratory Flow Measurement
45
15
Gas flow rate measurement
46
16
Anesthetic vaporizer(s)
47
17
Leak test
49
17.1 System leak test ... 49 17.2 Patient leak test ... 49
Annex Parts catalog Technical Information
K5330500IECIVZ.fm 10.11.05
All rights reserved. Copyright reserved.
II
General
1
Fabius GS 1
Notes
General This Technical Documentation conforms to the IEC 60601-1 standard. Read each step in every procedure thoroughly before beginning any test. Always use the proper tools and specified test equipment. If you deviate from the instructions and/or recommendations in this Technical Documentation, the equipment may operate improperly or unsafely, or the equipment could be damaged. Use only original Dräger parts and supplies. The maintenance procedures described in this Technical Documentation may be performed by qualified service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer. The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of the manufacturer.
Strictly follow the Instructions for Use manual / Operating Instructions! This Technical Documentation does not replace the Instructions for Use manual / Operating Instructions. Any use of the product requires full understanding and strict observation of the product-specific Instructions for Use manual/ Operating Instructions.
Reference is hereby made to the observance of the relevant safety provisions, for example in Germany, the Medical Product Law (MPG), the Medical Device Operator Ordinance (MPBetreibV), the Pressure Container Ordinance (Druckbehälterverordnung), the Technical Rules for Pressurized Gases (Technische Regeln Druckgase), or the Occupational Health and Safety Provisions (Unfallverhütungsvorschriften). Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany.
Observe protection mark DIN 34. Copyright reserved.
Version 2.0_ Released_Printed on_10.11.05_K5330500_General.fm
Follow the laws and regulations applicable in your country.
5330.500
3
General 1.1
Fabius GS
Symbols and Definitions This symbol indicates a warning.
This symbol indicates tips and useful information.
This symbol is used to alert against unsafe practices when handling electrostatic sensitive devices (ESD).
= examination of actual condition
Maintenance
= measures to maintain specified condition
Repair
= measures to restore specified condition
Servicing
= inspection, maintenance, and repair
Observe protection mark DIN 34. Copyright reserved.
Inspection
4
5330.500
Version 2.0_ Released_Printed on_10.11.05_K5330500_General.fm
Definitions according to German standard DIN 31051:
Function Description
5
Fabius GS 1
General Information about the Fabius GS
Function description The Fabius GS comprises the following assemblies: –
Display and Control Panel
–
Flowmeter assembly
–
Gas box: Gas Inlet Assembly and related items
–
Breathing system
–
Pneumatic assembly
–
Ventilator
–
Vaporizers
–
Trolley
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Monitoring, electrical connections and gas connections as shown in Figure 1, Figure 2, Figure 3, and Figure 4.
Figure 1
5330.500
Front view of Fabius GS anesthesia system, for legend see Table 1
7
Function description
Fabius GS Legend to Figure 1
No.
Name Trolley
2
O2 flush
3
Cosy 2 breathing system
4
Ventilator
5
Oxygen flowmeter (auxiliary)
6
Display
7
Control panel
8
Vaporizers
9
Cylinder Pressure Gauges
10
Pipeline Pressure Gauges
11
Flow Control Valves
12
Total fresh gas flowmeter
All rights reserved. Copyright reserved.
1
8
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Table 1
Function description
Figure 2
Rear view showing interface panel and power entry, for legend see Table 2
Table 2
Legend to Figure 2
No. All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Fabius GS
5330.500
Name
1
Serial communication ports (only one is shown)
2
Tube connection for PEEP valve
3
Tube connection for APL bypass valve
4
O2 sensor connection
5
Airway pressure connection
6
Spirolog sensor connection
7
Power entry
8
Cover for power fuses (2x 2.5 A)
9
Function description
Fabius GS
Name
9
ON/OFF switch
10
Battery fuse
Figure 3
Rear view showing gas pipeline and PIN index cylinder connections, for legend see Table 3
Table 3
Legend to Figure 3
10
Name
1
Pipeline tube connections
2
N2O or AIR PIN index cylinder connections
3
O2 PIN index cylinder connection
4
O2 PIN index cylinder connection
All rights reserved. Copyright reserved.
No.
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
No.
Fabius GS
Function description
Figure 4
Rear view showing gas pipeline and cylinder connections (units without PIN index cylinder connections), for legend see Table 4
Table 4
Legend to Figure 4 Name
1
Pipeline tube connections
2
N2O cylinder connection
3
O2 cylinder connection
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
No.
5330.500
11
Function description Function diagram of Fabius GS
Function diagram of Fabius GS - Cosy 2 breathing system
All rights reserved. Copyright reserved.
Figure 5
12
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
2
Fabius GS
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Fabius GS
Figure 6
5330.500
Function description
Function diagram of Fabius GS - Cosy 2.5 (2.6) breathing system
13
Function description 3
Battery backup
Fabius GS Fabius GS backup power is provided by two series-connected 12 V rechargeable batteries. These batteries remain on charge as long as the machine is plugged into an active AC outlet. Should power supply fail while the machine is in operation, the batteries will allow the machine to continue operating for a minimum of 45 minutes, provided that the batteries are fully charged. The batteries are accessible by opening the ventilator compartment. The 3.15A battery fuse is located at the back of the control box.
Figure 7
Figure 8
14
Fabius GS piping diagram
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Fabius GS piping diagram
All rights reserved. Copyright reserved.
4
Battery backup arrangement
Fabius GS 5
Function description of gas box
Function description The supply gases flow through the filters and non-return valves in the gas inlet assembly. Pipeline supply pressures are indicated on pipeline pressure gauges located on the flowmeter assembly. Cylinder pressure gauges are located on the trolley assembly. The pressures of O2 and N2O delivered to the flowmeter assembly are set by regulators on the gas inlet assembly. Should the O2 supply fail or if its pressure decrease below a certain limit, the O2 low alarm switch signals an alarm.
Figure 9
Gas box function diagram, part 1
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
If the O2 flush button is pressed, oxygen is delivered to the fresh-gas outlet. The fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic vaporizer. This avoids an increase in the anesthetic gas concentration.
5330.500
15
Function description
Fabius GS
Figure 10 Gas box function diagram, part 2
6
SORC (Sensitive Oxygen Ratio Controller)
The SORC is a control element that functions like an N2O shut-off device and ensures a vital O2 concentration in the fresh gas. In the event of an O2 shortage, the SORC limits the N2O flow such that the O2 concentration in the fresh gas does not decrease below 21 vol.%. If the O2 flow control valve is closed or if the O2 flow is lower than or equal to 200 mL/min, the SORC interrupts the N2O flow. N2O can be added as of an O2 flow of approx. 300 mL/min. In this case, the SORC also prevents O2 concentrations below 21 vol.%.
16
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
The SORC bypass allows O2 to bypass the restrictor in the SORC when O2 flows above 10 L/min are needed.
Fabius GS
Function description
Figure 11
SORC function diagram, part 1
The O2 and N2O flows are adjusted with the flow control valves. Restrictors located at the outlets of the SORC generate back-pressures. These back-pressures exert a force on the control diaphragms of the SORC. The O2 back-pressure opens the SORC. The N2O back-pressure closes the SORC. The pressure ratio at the control diaphragm affects the N2O flow.
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
The resistors and the spring force are dimensioned such that a minimum concentration of 21 vol.% of O2 is always ensured. The maximum O2 flow is approx. 12 L/min.
Figure 12 SORC function diagram, part 2, for legend see Table 5
5330.500
17
Function description
Fabius GS Table 5
Legend to Figure 12
No.
7
Cosy 2 breathing system
Name
1
Control diaphragms
2
Restrictors
3
N2O non-return valve
4
Operating-point adjusting screw
5
Flow control valves
The Cosy 2 breathing system allows three modes of patient ventilation: –
Manual ventilation and spontaneous breathing
–
Volume controlled ventilation
–
Pressure controlled ventilation
The APL valve (adjustable pressure limiting valve), lever type, has a selector switch which can be used to toggle between “MAN” and “SPONT”. On APL valves with control knob, switching from “IPPV/SPONT” to “MAN” is carried out by turning the knob. In the “MAN” position, the breathing system is closed to atmosphere. This position is used for manual ventilation of the patient. The APL valve opening pressure can be adjusted from 5 to 70 cmH2O (mbar). In the “SPONT” position the APL valve is open to atmosphere. This position is used for spontaneous breathing.
18
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Using the control box and the PEEP Pmax valve, the pressure limit (Pmax) can also be adjusted during volume control from 15 cmH2O (mbar) to 70 cmH2O (mbar) via the membrane keypad.
Fabius GS
Function description
Figure 13 Cosy 2 breathing system, for legend see Table 6
Legend to Figure 13
No.
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
Table 6
5330.500
Name
1
Expiratory connection
2
Flow sensor (Spirolog) (not shown)
3
PEEP/Pmax valve
4
MAN/SPONT APL valve
5
APL Bypass valve
6
Fresh-gas decoupling valve
7
Fresh-gas port
8
Ventilator port
9
Anesthetic gas scavenging port
10
Absorber
19