Dräger Medical
Humidifier Nebulizer Instructions for Use Edition 12
Instructions for Use
140 Pages
Preview
Page 1
English Instructions for use
Humidifier Nebulizer en
Instructions for Use Symbols... 12
Fault-Cause-Remedy... 17
For Your Safety and that of Your Patients... 12
Technical Data... 17
Intended Use... 13
Order List... 18
Safety information... 13 Overview... 13 Preparation... 13 Before using for the first time... 13 Assembling humidifier... 13 Assembling humidifier to O2 flowmeter or O2 DigiFlow... 14 Inhalation/Insufflation device... 14 Functional test... 14
Trademarks
NOTE
The trademarks are registered in the following countries:
A NOTE provides additional information intended to avoid inconvenience during operation.
Trademark
Country
O2Star®
Germany, European Union, USA, Canada, Australia, Japan, United Arab Emirates
Shut down... 14 Stripping down... 14
Trademarks owned by third-party manufacturers Trademark
Trademark owner
Reprocessing... 15
Neodisher®
Dr. Weigert
Korsolex®
BODE Chemie
Reprocessing Procedure... 15 Sterilization... 16 Visual inspection... 16 Disposal of the Medical Device... 16 Service... 16 Inspection... 17
A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the medical device or other property.
Trademarks owned by Dräger
Operation... 14
Cleaning and Disinfection... 15
CAUTION
Target groups Duties of the operating organization The tasks described in this document specify the requirements that have to be met by each respective target group. The operating organization of this product must ensure the following: –
The target group has the required qualifications (e.g., has undergone specialist training or acquired specialist knowledge through experience).
–
The target group has been trained to perform the task.
–
The target group has read and understood the chapters required to perform the task.
Safety Information Definitions WARNING A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Repair... 17
Instructions for use Humidifier Nebulizer
11
EnglishEnglish
Description of target groups The target groups may only perform the following tasks if they meet the corresponding requirements. Users Task
Requirement
Use of the product in accordance with the intended use
Specialist medical knowledge in the use of the product
Reprocessing personnel Task
Requirement
Reprocessing
Specialist knowledge in the reprocessing of medical devices
Service personnel Task
Requirement
Installation
Specialist knowledge in electrical engineering and mechanics Experience in the servicing of medical devices
Basic service activities
Dräger recommends arranging a service contract with DrägerService.
12
Symbols Storage temperature limitation
Date of manufacture
Keep away from sunlight
For Your Safety and that of Your Patients WARNING Risk of incorrect operation and of misuse Any use of the medical device requires full understanding and strict observation of all sections of these instructions for use. The medical device must only be used for the purpose specified under Intended use. Strictly observe all WARNING and CAUTION statements throughout these instructions for use and all statements on medical device labels. Failure to observe these safety information statements constitutes a use of the medical device that is inconsistent with its intended use. WARNING Risk due to incompatible accessories The use of incompatible accessories may adversely affect the functional integrity of the product. Personal injury and property damage may occur as a consequence. Use only compatible accessories. The accessories that are compatible with this product are listed in the order list supplied with the product.
WARNING Strictly observe the Instructions for Use of the basic device on which this medical device is used. WARNING Risk due to modifications Modifications to the product may lead to malfunctions and unforeseen risks. This may result in injury to the patient or the user or in property damage. Do not modify this product. WARNING Risk if service is not performed regularly If service is not performed regularly, malfunctions may occur, which can result in personal injury and property damage. Perform the service in accordance with the chapter "Service". Patient safety The design of the medical device, the accompanying documentation, and the labeling on the medical device are based on the assumption that the purchase and the use of the medical device are restricted to persons familiar with the most important inherent characteristics of the medical device. Instructions and WARNING and CAUTION statements are therefore largely limited to the specifics of the Dräger medical device. The instructions for use do not contain any information on the following points: –
Risks that are obvious to users
–
Consequences of obvious improper use of the medical device
Instructions for use Humidifier Nebulizer
English
–
Potentially negative effects on patients with different underlying diseases
Overview
Medical device modification or misuse can be dangerous.
Preparation
B
Before using for the first time A
CAUTION Risk of breakage Check the medical device before use. Do not use the medical device if it is damaged.
Intended Use Device for humidifying oxygen during inhalation or insufflation therapy. Used in combination with an O2 flowmeter or O2-DigiFlow for operation with O2 cylinders or a central supply system for Oxygen.
D
C E
Make sure that the individual parts have been disinfected/sterilized, see Page 15.
F
Only for patients with spontaneous breathing.
Assembling humidifier
Safety information G
WARNING Risk of fire Do not allow O2 connections to come into contact with oil, grease, or flammable liquids. No smoking, no open flames.
H A
WARNING Risk of breakage Make sure to install the breathing circuit without loops and kinks.
C 004
A Housing B Relief valve C Jet with O-rings D O-ring E Sealing ring F
D
Nebulizer
002
CAUTION Risk of personal injury and/or device malfunctions Magnetic fields can impair the correct functioning of the medical device and therefore endanger the patient or user. Do not use the medical device near MRI scanners.
B
G Support ring H Humidifier bottle Instructions for use Humidifier Nebulizer
13
EnglishEnglish
A Check that the sealing ring is in place in the housing, and that it is undamaged. B Push nebulizer onto the jet as far as it will go. C Push support ring over humidifier bottle.
Connect inhalation device or insufflation device to the humidifier.
A Open flow valve on the O2 flowmeter or O2 DigiFlow and set the oxygen flow necessary for the patient on the device.
Functional test
B Only O2 flowmeter: Read flow on upper rim of ball.
Open flow valve on the O2 flowmeter or O2 DigiFlow. At a flow of 5 L/min the nebulizer produces a fine mist.
D Fill humidifier bottle up to mark with distilled water. Screw the humidifier bottle into the housing.
The nebulizer produces a fine mist.
Shut down
Operation
Assembling humidifier to O2 flowmeter or O2 DigiFlow
Close flow valve on the O2 flowmeter or O2 DigiFlow.
NOTE Use the medical device only within the specified pressure range and environmental conditions. Observe section "Technical Data". Otherwise, the correct functioning of the medical device may be compromised.
Observe the relevant Instructions for Use of the O2 flowmeter or O2 DigiFlow. Check that the sealing ring is in place on the O2 flowmeter or O2 DigiFlow. Close the flow valve on the O2 flowmeter or O2 DigiFlow.
Remove inhalation device or insufflation device.
Stripping down
Screw humidifier to O2 flowmeter or O2 DigiFlow and tighten.
D
Inhalation/Insufflation device B C
003
005
006
A
14
Instructions for use Humidifier Nebulizer
English
A Disconnect humidifier bottle and empty. Disconnect O2 flowmeter or O2 DigiFlow from housing. B Remove nebulizer. C Remove support ring from humidifier bottle. D Disconnect jet
Reprocessing WARNING Risk due to inappropriately reprocessed products Reusable products must be reprocessed, otherwise there is an increased risk of infection. – Follow the infection prevention policies and reprocessing regulations of the healthcare facility. – Follow the national infection prevention policies and reprocessing regulations. – Use validated procedures for reprocessing. – Reprocess reusable products before using them for the first time. – Reprocess reusable products after every use. – Observe the manufacturer's instructions for cleaning agents, disinfectants, and reprocessing devices.
CAUTION Risk of stress cracking Do not use disinfectants or cleaning agents containing alcohol.
Reprocessing Procedure
Rinse items under running water until cleaning agent residue is no longer discernible.
4
Check parts for visible dirt and damage. If necessary, repeat manual cleaning.
Manual disinfection
Testing of procedures and agents
Manual disinfection should preferrably be carried out with disinfectants based on aldehydes or quaternary ammonia compounds.
The cleaning and disinfection of medical products has been tested with the following procedures and agents. The following agents showed good material compatibility at the time of the test:
Observe the applicable country-specific listings for disinfectants. The list of the German Association for Applied Hygiene (Verbund für Angewandte Hygiene VAH) applies in German-speaking countries.
–
Manual cleaning: Neodisher LM2 by Dr. Weigert
–
Manual disinfection: Korsolex extra by Bode Chemie
The composition of the disinfectant is the responsibility of the manufacturer and can change over time.
–
Machine cleaning: Neodisher MediClean by Dr. Weigert
Strictly observe the manufacturer’s information on the disinfectant.
–
Machine disinfection: thermal disinfection at 93 °C (199.4 °F), 10 minutes
Carry out manual disinfection
–
Sterilization: hot steam sterilization at 134 °C (273.2 °F), 5 minutes, fractional vacuum
1
Immerse items in disinfectant. Use suitable brushes.
2
After contact time, rinse items under running water until disinfectant residue is no longer discernible.
3
Check parts for visible dirt and damage. If necessary, repeat manual disinfection.
4
Thoroughly shake out residual water. Allow items to dry thoroughly.
Manual cleaning Manual cleaning should preferrably be carried out under running water or with commercially available cleaning agents based on mild alkaline compounds. Carry out manual cleaning
Cleaning and Disinfection
1
Clean O2 flowmeter or O2 DigiFlow, inhalation device and insufflation device following the relevant Instructions for Use.
Wash off visible soilings under running water. Using an ultrasound cleaner improves cleaning results.
2
Use cleaning agents in accordance with the manufacturer’s instructions. Make sure that all surfaces to be cleaned can be efficiently reached. Use suitable brushes.
Instructions for use Humidifier Nebulizer
3
Machine cleaning and disinfection Use a washer-disinfector in accordance with EN ISO 15883, preferably with a cart for anesthesia and ventilation accessories, for automatic cleaning and disinfection of breathing-gas conducting parts.
15
EnglishEnglish
Carry out machine cleaning and disinfection
Visual inspection
Service
1
Strictly observe Instructions for Use of washerdisinfector.
2
Position items so that all interior spaces are completely flushed and water can drain off freely.
Inspect all items for damage and wear, e.g. cracking, embrittlement or pronounced hardening, and residual soiling.
Safety information
3
Use suitable cleaning agent.
CAUTION
4
Select suitable program (preferably anesthesia program).
Even accessories designed to be reused have a limited service life. Handling and reprocessing can increase wear and markedly shorten service life (e.g., disinfectant residues can attack the material more intensely during autoclaving). If signs of wear become visible, such as cracks, deformation, discoloration, peeling, etc., affected accessories must be replaced.
–
Cleaning must be carried out at 40 °C to 60 °C (104 °F to 140 °F) for at least 5 minutes.
–
Thermal disinfection must be carried out at 80 °C to 95°C (176 °F to 203°F) and with corresponding contact time.
5
Carry out final rinsing with deionized water.
6
Immediately remove items from the washerdisinfector.
7
Check parts for visible dirt and damage. If necessary, repeat program or carry out manual cleaning and disinfection.
8
Allow items to dry thoroughly.
Sterilization During sterilization, living microorganisms are removed from semi-critical medical devices. Residual water inside the components is also dried out.
WARNING
Disposal of the Medical Device When disposing of the medical device: Consult the relevant waste disposal company for appropriate disposal. Observe the applicable laws and regulations.
Risk due to inappropriately reprocessed products The product may be contaminated with infectious agents. Before service is performed and before the product is sent back for repair, reprocess the product in accordance with the chapter "Reprocessing". WARNING Risk if service is not performed regularly Wear and material fatigue of the components may lead to device failure and malfunctions. Perform service at the specified intervals. WARNING Risk if service is not performed properly Personal injury and property damage may occur if service is not performed properly. Service must be performed by those target groups that are assigned to the particular measure.
Only sterilize cleaned and disinfected items. Use a vacuum steam sterilizer (in accordance with DIN EN 285), preferably with fractional vacuum.
16
Instructions for use Humidifier Nebulizer
English
Definition of service terminology
Fault-Cause-Remedy
Concept
Definition
Fault
Cause
Remedy
Operating principle
Venturi nebulizer
Service
All measures (inspection, repair) intended to maintain or restore the functional integrity of a product
No aerosol
No distilled water
Re-fill with distilled water
Drive gas
Oxygen (O2)
Operating pressure
5 ±0.5 bar
No oxygen
Check oxygen supply
Gas consumption
5 to 15 L/min
Push nebulizer onto housing as far as it will go
Average droplet size
1 µm
Nebulizing rate at 15 L/min 36 g/h O2
Inspection
Measures intended to determine and assess the current state of a product
Repair
Measures intended to restore the functional integrity of a product after a failure
Nebulizer not attached properly Leak in humidi- Loose fier humidifier bottle
Inspection .
Measure
Interval
Target group
Functional test and visual inspection
Service personSame as for connected ba- nel sic device (O2 flowmeter, O2 DigiFlow)
Insufficient humidification
Repair Dräger recommends that all repairs are performed by DrägerService and that only authentic Dräger repair parts are used.
Instructions for use Humidifier Nebulizer
"Heavy" droplets
Technical Data
Tighten bottle properly
Sound level at 15 L/min
50 dB (A)
Connecting thread
M 34 x 1.5
Distilled water volume
220 mL
Dimensions
Ø 59 mm Height 198 mm
Faulty sealing ring
Replace sealing ring
Loose hose connection
Connect hose properly
Defective sealing ring on O2 flowmeter
Replace sealing ring, or use new nebulizer.
Materials used
Loose fitting at connection
Weight without distilled water 230 g
Humidifier bottle
Polysulfone
Nebulizer
Polysulfone
Tighten fitting at connection
Housing
Polyamide
Blockage of jet on nebulizer
Replace nebulizer
Ambient conditions
Pressure of drive gas too low
Ensure pressure of 5 bar
Leak in nebulizer
Check O-ring Fit nebulizer properly
During operation Temperature
5 °C to 40 °C (41 °F to 104 °F)
Ambient pressure
700 to 1200 hPa (10.2 to 17.4 psi)
Relative humidity
0 to 95 %, no condensation
17
EnglishEnglish
During storage and transport Temperature
–20 °C to 70 °C (–4 °F to 158 °F)
Ambient pressure Relative humidity
Order List
Designation
Designation
Part No.
800 to 1200 hPa (11.6 to 17.4 psi)
Humidifier nebulizer
2M85835
0 to 95 %, no condensation
O2 flowmeter 16 L/min, hose, DIN ISO
MP04573
O2 flowmeter 16 L/min, rail, NIST
MP04567
O2 flowmeter 16 L/min, central supply, DIN
MP04558
O2 flowmeter 32 L/min, central supply, DIN
MP04571
O2 double flowmeter 16 L/min, central supply, DIN
MP04560
O2 flowmeter 16 L/min, central supply, DIN, long
MP04561
O2 flowmeter 16 L/min, rail, 90°, NIST
MP04570
DigiFlow O2, central supply, 3 L/min, M34x1.5 DIN
MP01211
DigiFlow O2, rail, 3 L/min, M34x1.5 NIST
MP01213
DigiFlow O2, central supply, 16 L/min, M34x1.5 DIN
MP01227
DigiFlow O2, rail, 16 L/min, M34x1.5 NIST
MP01229
Hose (reusable)
8403896
Class II a Classification as per Directive 93/42/EEC Annex IX UMDNS-Code 10 - 046 Universal Medical Device Nomenclature System – nomenclature for medical devices
Accessories required for operation
Part No.
Spare parts Housing
2M85827
Relief valve
2M85829
Jet with O-rings
2M85472
O-ring
E20566
Sealing ring
2M85854
Nebulizer
2M85473
Set of spare parts for the nebulizer
2M85470
Support ring
M12159
Humidifier bottle
2M85675
The humidifier nebulizer is also compatible with the O2Star oxygen therapy accessories of Dräger. For further information contact Dräger.
18
Instructions for use Humidifier Nebulizer