How to clean the Dräger Babyleo TN500 Cleaning in 5 steps
!
Please complete all cleaning practices according to your hospitals guidelines. This document is for informational purposes only. Refer to the instructions for use of the device for full disassembly, cleaning and disinfecting instructions.
Prerequisites: There is no patient in the device.
3
Starting the automatic humidifier cleaning: •
Open the hood.
•
On the Main menu touch the Special procedures...
•
Open the Humidifier cleaning tab.
•
Touch the On button and confirm with rotary knob.
•
Wait until the humidifier cleaning is complete. This can take up to 60 minutes. Do not disassemble the device during cleaning.
•
Switch off the device and let it cool before continuing.
Surface disinfection cleaning for:
4
Machine cleaning and thermal disinfection for:
Each part removed during disassembly.
•
Hood and access panels from inside and outside.
•
grommets
X-ray flap air ducts (1): clean by pulling a piece of cloth soaked in disinfectant through each duct.
•
fan impeller
1)
• •
!
Please turn the page for assembling and dissambling.
• •
2
5
Display, main column, trolley, handles and all outer parts. Drawer from outside and inside including insert piece.
Only use disinfectants as listed in the instructions for use. Make sure no liquid penetrates into the device.
Ensure all surfaces are dry before using for therapy.
90 50 645 | 16.07-1 | Marketing Communications | HQ | © 2016 Drägerwerk AG & Co. KGaA
1
2 Disassembly ►
Step 1, 3, 4 and 5, please turn the page
Open the hood and remove frame seals from both sides.
2 Open front and side
3 Remove hose grommets
4 Remove the 2 mini-walls
12 Fasten the seals on the
11 Close front and side
10 Insert hose grommets on
9 Insert the 2 mini-walls
5 Disconnect heated mat-
6 Remove the mattress tray.
7
8 Place mattress on the
7
1
edges of the hood and ensure the seals are correctly positioned.
access panels.
access panels.
on the sensor wall and on the front access panel (6 pieces total).
the sensor wall and front access panel (6 pieces total).
at both ends of the mattress tray.
at both ends of the mattress tray.
Raise the x-ray tray and pull it out.
8 Remove the bed support.
Fit the mattress tray in place.
6 Slide in the x-ray tray and
5 Fit the bed support and
9 Remove the T-bars.
10 Remove the small cover
11 Remove the large cover for
12 Raise and remove the
4 Attach the T-bars.
3 Insert the small cover for
2 Insert the large cover for
1
tress* and scale* cables and remove mattress. Or simply remove SoftBed.
mattress tray. Pull heated mattress* and scale* cables through grommets and plug into sensor wall.
* Option
for the warm air duct.
the warm air duct.
close x-ray tray flaps.
the warm air duct.
the warm air duct.
push it over the T-bars.
fan impeller.
Insert the fan impeller. Ensure that it is tightly pressed down and clicks into position.
◄ Assembly