Please reprocess the device according to your hospital’s guidelines. This document is for information purposes only. Please refer to the device instructions for use for complete disassembly, assembly, cleaning and surface disinfection. Only use disinfectants as listed in the instructions for use. Ensure that no liquid enters the device.
Prerequisites: There must be no patient in the device. Allow device to cool for 30 minutes prior to disassembly.
1
Surface disinfection for: • • • •
Disassembly ►
Basic device/trolley All disassembled parts Hood and side walls (interior and exterior) X-ray valve air channels: Pull a cloth soaked in disinfectant through each air channel
Set the incubator to the lowest position.
2
C lose and remove the transfer set.
Machine cleaning and thermal disinfection: • Grommets • Fan impeller • Water tank (option)
Before starting use, ensure that all surfaces are dry and free from disinfectant residues. 3
Remove sterile water bag or water tank (option).
16 Connect water tank
or sterile water bag.
4
Remove upper double wall and continuous seal.
15 Apply continuous seal and
refit upper double wall.
18 Set the incubator to the
desired working height.
5
Remove hood.
14 Refit hood.
17 Connect and open
the transfer set.
6
Cleaning mode.
13
◄ Assembly
Disassembly ► Open side walls.
8
Remove grommets.
9
Open head and foot access ports.
10 Remove mattress.
12 Close side walls.
11 Insert grommets.
10 Close head and foot
9
11 Remove mattress tray.
12 Remove hose holder arm
13 Unlock and remove
14 Remove trough.
8
Insert mattress tray.
15 Remove fan impeller
and connecting piece.
4
Insert fan impeller and connecting piece.
(option).
7
Insert hose holder arm (option).
16 Remove drawer.
3
Insert drawer.
access ports.
x-ray tray.
6
Insert and lock x-ray tray.
17 Clean and disinfect
basic device in full.
2
5
Insert mattress.
Insert trough.
18 Remove display unit rotary
knob and clean and disinfect display unit.
1
Insert display unit rotary knob.
◄ Assembly
91 08 695 | 2004-1 | NW | GT | Subject to modifications | © 2020 Drägerwerk AG & Co. KGaA