Supplement Additional alarms
3 Pages
Preview
Page 1
English Supplement
Oxylog family
en
English
Supplement Oxylog 2000 plus with part no.: 5705035, Edition: 1 – 2008-01 and higher, installed software version: SW 1.05 Oxylog 3000 with part no.: 9037171, Edition: 1 – 2001-04 and higher, installed software version: SW 1.23 Oxylog 3000 plus with part no.: 5705305, Edition: 1 – 2009-05 and higher, installed software version: SW 1.06 The information supplements the instructions for use in the following chapter:
Problem solving Alarm – Cause – Remedy Messages in the alarm message field Alarm
Cause
Remedy
Rank
!!
Control knob faulty (FiO2)
Technical defect.
Ventilation is continued with last settings. Check patient's condition and ventilation. Contact DrägerService.
19
!!
Control knob faulty (Pmax)
Technical defect.
Ventilation is continued with last settings. Check patient's condition and ventilation. Contact DrägerService.
17
!!
Control knob faulty (RR)
Technical defect.
Ventilation is continued with last settings. Check patient's condition and ventilation. Contact DrägerService.
16
!!
Control knob faulty (VT)
Technical defect.
Ventilation is continued with last settings. Check patient's condition and ventilation. Contact DrägerService.
18
Supplement Oxylog family
3
Directive 93/42/EEC concerning medical devices
Manufacturer Drägerwerk AG & Co. KGaA Moislinger Allee 53 – 55 D-23542 Lübeck Germany +49 451 8 82-0 FAX +49 451 8 82-20 80 http://www.draeger.com
9055573 – GA 5503.420 me © Drägerwerk AG & Co. KGaA Edition: 1 – 2016-05 Dräger reserves the right to make modifications to the device without prior notice.
Distributed in the US by Draeger, Inc. 3135 Quarry Road Telford, PA 18969-1042 U.S.A. (215) 721-5400 (800) 4DRAGER (800 437-2437) FAX (215) 723-5935 http://www.draeger.com