Dräger Medical
Fabius Family Breathing System Cleaning Guide
8 Pages
Preview
Page 1
Fabius Family Breathing System Cleaning Guide
02 I
Fabius Family Breathing System Cleaning Guide
CAUTION To reduce the risk of infection to both hospital staff and patients, clean and disinfect medical devices after each use. Protective clothing, eye protection etc. must be worn.
Warning! This booklet is not a replacement for the Instructions for Use. For full information concerning the performance characteristics of the Dräger devices described in the booklet, each user must carefully read and fully comprehend the Instructions For Use before operating the specfic device.
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
COSY Breathing System CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
•
Hot Steam sterilization is required to dry the breathing system
CLIC Adapter CLEANING - Qty 1 each •
Machine clean and thermal disinfection (preferably anaesthesia program)
•
NB. Place black disc into clic adapter prior to cleaning
COMPONENTS
03 I
04 I
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Inspection Caps CLEANING - Qty 2 pieces •
Machine clean and thermal disinfection (preferably anaesthesia program)
Retaining Nuts CLEANING - Qty 2 •
Machine clean and thermal disinfection (preferably anaesthesia program)
Unidirectional Valve Discs CLEANING - Qty 2 •
Machine clean and thermal disinfection (preferably anaesthesia program)
COMPONENTS
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Unidirectional Valve ‘O’ rings CLEANING - Qty 2 •
Machine clean and thermal disinfection (preferably anaesthesia program)
Expiratory Port CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
APL Valve CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
COMPONENTS
05 I
06 I
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Ventilator Diaphragm CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
Ventilator Lid CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
Flow Sensor CLEANING - Qty 1 • • • •
Disinfect the flow sensor by immersion. Sufficiently rinse the flow sensor under water (preferably drinking-water quality) until no disinfectant residues can be recognized. Inspect flow sensor for visible soiling and damage (measuring wires and corresponding pins). Repeat manual disinfection if necessary. Remove remaining water and allow the flow sensor to air for at least 30 minutes.
COMPONENTS
Fabius Family Breathing System Cleaning Guide
FABIUS FAMILY BREATHING SYSTEM CLEANING
Ventilator Hose and Water Trap CLEANING - Qty 3 pieces •
Machine clean and thermal disinfection (preferably anaesthesia program)
Scavenge Hose CLEANING - Qty 1 •
Machine clean and thermal disinfection (preferably anaesthesia program)
COMPONENTS
07 I
Fabius Family Breathing System cleaning Guide / V1.1 / 20.07.12 / Printed in Australia / Subject to modifications / © 2012 Draeger Medical Australia Pty Ltd
Draeger Medical Australia Pty Ltd Unit 97, Axxess Corporate Park 45 Gilby Road Mt Waverley VIC 3149 Tel: 1800 800 327 Fax: 1800 010 327 E-mail: [email protected] www.draeger.com