ForeSight Elite Tissue Oximetry System Setup guide  HemoSphere advanced monitor setup 1. S  lide the HemoSphere tissue oximetry Figure 1 module into the monitor. 2. Connect the ForeSight Elite tissue oximetry module cable into the HemoSphere tissue oximetry module (Figure 1). Note: ForeSight Elite module LED illuminates green when connected to port A, and blue when connected to port B. 3. S  elect Continue Same Patient or New Patient, enter patient data. Select corresponding patient mode. Note: Tissue oximetry works with invasive, minimally-invasive, and noninvasive modes. 4. Touch the displayed parameter label located inside of any parameter tile to select StO2 <Ch> as a key parameter. Figure 2  ForeSight Elite sensor setup Note: Refer to the HemoSphere Operator’s Manual, Table 13-2, for sensor selection guidelines based on patient monitoring Figure 3 mode, patient weight, and body location. 1. Open  ForeSight Elite sensor packaging and inspect for damage (Figure 2). 2. Clean patient’s skin and ensure area is dry prior to sensor application. 3. Remove protective liner and apply sensors. 4. Connect sensors to ForeSight Elite tissue oximetry module cable. Ensure each sensor is connected with a tactile click (Figure 3).