Edwards Vigilance II Monitor Controls & Icons
On/Off Press to activate monitor.
Set Time/Date Printer Activation
CO Start/Stop Full/Split Screen Icon - STAT Boxes - Data Relationship Plot - Diagnostic Plot
CCO/ICO Mode Select Patient Data - Cardiac Profile - Oxygenation Profile
Home Icon - Return to Home Configuration
Trend Scroll Alarm Silence, 2 min
Help Icon - Access Help Screens - Alarm/Alert Help Setup Icon - Display Preferences - Cable Test - Language Selection - Serial/Analog Port Setup
Patient CCO Cable Connector
SvO2/ScvO2 Connector
Cable Test Ports
Navigation Knob Rotate to Navigate, Push to Select
05 ED Vigilance II QuickR2
2/21/05
3:29 PM
Page 2
Edwards Vigilance II Monitor Quick Reference See Vigilance II Monitor Operator’s Manual for Detailed Instructions.
Oximetry Setup & Calibration
Continuous Cardiac Output
In Vitro Calibration
Continuous Cardiac Output
1. Connect the Optical Module to the SvO2/ScvO2 color-coded connector on the front of the monitor. Connect catheter to the Optical Module with the “Top” side up. Allow 20 minutes for the module to warm up. 2. Use the Navigation Knob to highlight the SvO2/ScvO2 Large Parameter Frame. Press knob to display calibration menu. Highlight In Vitro Calibration. Press knob to select. 3. Highlight HGB. Press knob to select. Enter hospital lab values by turning the knob. Press knob to confirm setting. 4. Turn knob to highlight Calibrate. Press to select. Upon successful completion of calibration, insert catheter in patient. 5. Highlight Start. Press knob to select.
1. Connect the Patient CCO Cable to the color-coded connector on the front panel. 2. If CEDV is required, connect the ECG interface cable from the bedside monitor to the ECG connector on the Vigilance II monitor rear panel. 3. Connect an Edwards Advanced Technology Swan-Ganz Catheter to the patient cable per catheter instructions. 4. Press CO Start/Stop button to begin continuous cardiac output monitoring. If CEDV monitoring is required: • Highlight one of the Large Parameter Frames. Press knob to display menu. • Highlight and select Parameter, then EDV. Highlight Return. Push knob to select.
In Vivo Calibration 1. Use the Navigation Knob to highlight the SvO2/ScvO2 Large Parameter Frame. Press knob to display calibration menu. 2. Highlight In Vivo Calibration. Press knob to select. 3. After 25 seconds Draw is highlighted. Press knob just prior to drawing waste and lab samples. 4. Use knob to enter Oximetry and HGB or Hct values. 5. Highlight Calibrate. Press knob to select.
Signal Quality Index (SQI) for Oxygen Saturation Monitoring LEVEL 1 2 3 4
SIGNAL Normal Intermediate Poor Unacceptable
DESCRIPTION All aspects of the signal are optimal Indicates a moderately comprised signal Indicates poor signal quality Indicates severe problem with one or more aspect of the signal quality
Activating STAT Boxes 1. Highlight the Full/Split Screen Icon. Press knob to select. 2. Highlight the lower frame of the split screen. Press knob to select. 3. Highlight STAT Boxes. Press knob to select.
Rx only. See instructions for use for full prescribing information. Edwards Lifesciences devices placed on the European market meeting the essential requirements referred to in Article 3 of the Medical Device Directive 93/42/EEC bear the CE marking of conformity. Edwards Lifesciences, Edwards, the stylized E and Vigilance II are trademarks of Edwards Lifesciences Corporation. Swan-Ganz and Vigilance are trademarks of Edwards Lifesciences Corporation and are registered in the United States Patent and Trademark Office. © 2005 Edwards Lifesciences LLC. All rights reserved. AR00802
Edwards Lifesciences LLC · One Edwards Way · Irvine, CA 92614 USA · 949.250.2500 · 800.424.3278 · www.edwards.com Edwards Lifesciences (Canada) Inc. · 1290 Central Pkwy West, Suite 300 · Mississauga, Ontario · Canada L5C 4R3 · Phone 905.566.4220 · 800.268.3993 Edwards Lifesciences S.A. · Ch. du Glapin 6 · 1162 Saint-Prex · Switzerland · Phone 41.21.823.4300 Edwards Lifesciences Japan · 2-8 Rokubancho · Chiyoda-ku, Tokyo 102-0085 · Japan · Phone 81.3.5213.5700