Setup Guide
4 Pages
Preview
Page 1
Edwards PediaSat Oximetry Catheter Edwards Oximetry Central Venous Catheter HemoSphere Advanced Monitor Setup guide
Oximetry setup 1. C onnect the HemoSphere oximetry cable to the left side of the HemoSphere advanced monitor (A). to turn 2. Press the power button on the HemoSphere advanced monitor. All functions are accessed through the touch screen.
HemoSphere advanced monitor (A)
3. Select Continue Same Patient or New Patient and enter new patient data (patient ID, gender, age, height, and weight). 4. O pen side 1 of the oximetry catheter tray, exposing only the optical connector (B). Be careful not to contaminate tray contents in side 2. 5. Insert the optical connector of the catheter “TOP” side up into the oximetry cable and snap the enclosure shut (C). 6. T he HemoSphere oximetry cable must be calibrated before each monitoring session.
Optical connector (B)
Oximetry cable enclosure (C)
HemoSphere advanced monitor How to move data during and after patient transfer In vitro Calibration (if catheter has not been inserted yet)
In vivo Calibration (performed after catheter has been inserted)
1. Touch the clinical actions icon > Oximetry Calibration icon.
1. Touch the clinical actions icon > Oximetry Calibration icon.
2. Select Oximetry Type: ScvO2 or SvO2.
2. S elect Oximetry Type: ScvO2 or SvO2.
3. Touch In vitro Calibration button.
3. T ouch In vivo Calibration button. If setup is unsuccessful, see Chapter 10: Help and Troubleshooting in the HemoSphere Advanced Monitor Operator’s Manual.
4. Enter either the patient’s hemoglobin (HGB) or hematocrit (Hct) value. A default value may be used until the patient’s HGB or Hct is available. 5. Touch Calibrate button. 6. When the calibration successfully completes, the following message appears: In vitro Calibration OK, insert catheter. 7. Insert the catheter as described in the catheter instructions for use. 8. Touch Start button.
4. W hen baseline calibration is successful, touch Draw button and then draw the blood sample. 5. D raw the blood sample slowly and send the blood sample to the lab for measured analysis by co-oximeter. 6. Enter HGB or Hct and ScvO2/SvO2 when lab values are received.
7. Touch Calibrate button. 9. If ScvO2/SvO2 are not current key 8. If ScvO2/SvO2 are not current key parameters, touch the displayed parameters, touch the displayed parameter label located outside parameter label located outside of any parameter globe to select of any parameter globe to select ScvO2/SvO2 as a key parameter ScvO2/SvO2 as a key parameter from the parameter popup. from the parameter popup. 10. Touch within the ScvO2/SvO2 9. Touch within the ScvO2/SvO2 parameter globe to adjust parameter globe to adjust Alarms/Targets. Alarms/Targets.
HemoSphere advanced monitor How to move data during and after patient transfer Recall Oximetry Cable Data can be used to recall data from the oximetry cable after a patient has been transported away from the HemoSphere advanced monitor. This allows the patient’s last calibration to be recalled along with the patient’s demographic data for immediate oximetry monitoring. Calibration data within the oximetry cable must be less than 24 hours old to use this function. 1. W ith the catheter connected to the HemoSphere oximetry cable, unplug the cable from the HemoSphere advanced monitor and transport it with the patient. The catheter should not be disconnected from the oximetry cable. 2. If the oximetry cable is being connected to another HemoSphere advanced monitor, make sure that previous patient data is cleared. 3. Once the patient has been transferred, reconnect the oximetry cable to the HemoSphere advanced monitor and turn it on.
5. Touch Recall Oximetry Cable Data button. 6. If the oximetry cable data is less than 24 hours old, touch Yes button to start oximetry monitoring using the recalled calibration information. OR Touch No button and perform an in vivo calibration. 7. From the oximetry calibration menu, touch In vivo Calibration button to recalibrate the cable. To review patient data that was transported with the oximetry cable, touch the settings icon. 8. Touch Patient Data button.
4. Touch the clinical actions icon > Oximetry Calibration icon.
CAUTION Do not disconnect the oximetry cable while calibration or data recall are in process. CAUTION If the oximetry cable is being transferred from a HemoSphere advanced monitor to another HemoSphere advanced monitor, check that the patient height, weight, and BSA are correct prior to beginning monitoring. Re-enter patient data, if necessary.
NOTE: Keep the time and date of all HemoSphere advanced monitors current. If the date and/or time of the HemoSphere advanced monitor being transported “from” differs from the HemoSphere advanced monitor being transported “to” the following message may appear: “Patient data in oximetry cable more than 24 hours old – Recalibrate.” If the system needs to be recalibrated, a 10 minute warm up period for the oximetry cable may be required.
For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences devices placed on the European market meeting the essential requirements referred to in Article 3 of the Medical Device directive 93/42/EEC bear the CE marking of conformity. Edwards, Edwards Lifesciences, the stylized E logo, HemoSphere, and PediaSat are trademarks of Edwards Lifesciences Corporation. All other trademarks are the property of their respective owners. © 2017 Edwards Lifesciences Corporation. All rights reserved. PP--US-1857 v1.0 Edwards Lifesciences One Edwards Way, Irvine CA 92614 USA edwards.com