NOTE Keep the date and time of all EV1000 monitors current. If the date and/or time of the EV1000 monitor being transported “from” differs from the EV1000 monitor being transported “to,” then the 24 hour test may not work properly.
PreSep Oximetry Catheter EV1000 Clinical Platform Setup Guide 1. Connect the power adapter and ethernet cable for both EV1000 panel and databox. Press the button on the panel. 2. When the boot up is complete, enter new patient data (patient ID, gender, age, height, and weight) or continue same patient. 3. If entering new patient data, use the touch screen to select and enter values. Press Home to continue.
For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences devices placed on the European market meeting the essential requirements referred to in Article 3 of the Medical Device Directive 93/42/EEC bear the CE marking of conformity. Edwards, Edwards Lifesciences, the stylized E logo, EV1000, and PreSep are trademarks of Edwards Lifesciences Corporation. © 2012 Edwards Lifesciences Corporation. All rights reserved. AR07451
Edwards Lifesciences I edwards.com USA I Switzerland I Japan I Singapore I Brazil
PreSep Oximetry Catheter Setup 1. Connect the optical module to the oximetry color-coded connector on the back of the EV1000 databox. Allow 20 minutes for the optical module to warm up.
2. Open the PreSep catheter lid exposing only the optics connection or entire tray, depending on technique, remove optics connection being careful not to contaminate tray contents. 3. Connect PreSep catheter to the optical module matching “TOP” on both optical module and optics connection.
Calibration 1. Touch Clinical Actions. 2. Touch Oximetry Calibration. 3. Use In vitro calibration if catheter has not been inserted into the patient yet.
In vitro Calibration 1. Select ScvO2. 2. Touch In vitro Calibration. 3. Touch HGB (hemoglobin) or Hct (hematocrit), and enter the lab value if available or else use default values. 4. Select Calibrate. Calibration is complete in 20 seconds. 5. When In vitro Calibration OK message appears, insert catheter. 6. Touch Start. Catheter is ready in 25 seconds. In vivo Calibration (performed after catheter has been inserted) 1. Touch In vivo Calibration. Baseline setup is complete in 25 seconds. 2. Touch Draw, then withdraw the blood sample slowly (2 mL over 30 seconds) and send sample to lab for analysis. 3. Upon receipt of lab values, enter HGB (hemoglobin), Hct (hematocrit), and ScvO2 values. 4. Touch Calibrate. Catheter is ready in 25 seconds.
EV1000 Monitor Recall Optical Module Data Patient data and system calibration information can be transferred with the patient via the optical module. With the catheter connected to the optical module, unplug the optical module from the EV1000 databox. Do not disconnect catheter from optical module. Once the patient has been transferred, reconnect the optical module to the EV1000 databox. 1. Touch Clinical Actions. 2. Touch Oximetry Calibration. 3. Select Recall OM Data. 4. The calibration data in the optical module is displayed if the data is less than 24 hours old. Select Yes to recall data from optical module or select No to perform an in vivo calibration. If the optical module data is more than 24 hours old, the data cannot be recalled and a new calibration must be performed.