Quick Reference Guide
2 Pages
Preview
Page 1
Vigilance II Monitor Quick Reference Guide On/Off Press to Activate Monitor Set Time/Date
CCO Start/Stop
Full/Split Screen Icon • STAT Boxes • Data Relationship Plot • Diagnostic Plot
Bolus Mode Patient Data • Cardiac Profile • Oxygenation Profile
Home Icon • Return to Home Configuration
Trend Scroll
Help Icon • Access Help Screens • Alarm/Alert Help
Navigation Knob Rotate to Navigate, Push to Select
Setup Icon • Display Preferences • Cable Test • Language Selection • Serial/Analog Port Setup
Alarm Silence, 2 min
Patient SvO2 /ScvO2 CCO Cable Connector Connector
Cable Test Ports
Vigilance II Monitor Quick Reference Guide Oximetry Setup and Calibration In-Vitro Calibration 1. Connect the optical module to the SvO2/ScvO2 color-coded connector on the front of the monitor. Connect catheter to the optical module with the “Top” side up. Allow 20 minutes for the module to warm up. 2. U se the navigation knob to highlight the SvO2/ ScvO2 large parameter frame. Press knob to display calibration menu. Highlight In-Vitro Calibration. Press knob to select. 3. Highlight HGB. Press knob to select. Enter hospital lab values by turning the knob. Press knob to confirm setting. 4. Turn knob to highlight Calibrate. Press to select. Upon successful completion of calibration, insert catheter in patient. 5. Highlight Start. Press knob to select. In-Vivo Calibration 1. U se the navigation knob to highlight the SvO2/ ScvO2 large parameter frame. Press knob to display calibration menu. 2. Highlight In-Vivo Calibration. Press knob to select. 3. After 25 seconds Draw is highlighted. Press knob just prior to drawing waste and lab samples. 4. Use knob to enter oximetry and HGB or Hct values. 5. Highlight Calibrate. Press knob to select. To Transport 1. The Swan-Ganz catheter should remain connected to the optical module during transport. Unplug the optical module cable from the Vigilance II monitor. Transport the patient. 2. If reconnecting to a different Vigilance II monitor, press the patient data button, select New Patient, select Yes, and Yes again. 3. Reconnect the optical module cable to the Vigilance II monitor. 4. Highlight and select oximetry large parameter frame. Select Recall OM Data. The calibration data stored in the optical module will be displayed. 5. If stored data is less than 24 hours old a Yes or No option will be displayed. Select Yes to begin oximetry monitoring. If No is selected, In-Vivo calibration will be necessary.
Continuous Cardiac Output 1. C onnect the patient CCO cable to the colorcoded connector on the front panel. 2. If CEDV is required, connect the ECG interface cable from the bedside monitor to the ECG connector on the Vigilance II monitor rear panel. 3. C onnect an Edwards advanced technology Swan-Ganz catheter to the patient cable per catheter instructions. 4. Press CO Start/Stop button to begin continuous cardiac output monitoring. If CEDV monitoring is required: • Highlight one of the large parameter frames. Press knob to display menu. • Highlight and select Parameter, then EDV. Highlight Return. Push knob to select. Activating STAT Boxes 1. H ighlight the Full/Split Screen icon. Press knob to select. 2. H ighlight the lower frame of the split screen. Press knob to select. 3. Highlight STAT Boxes. Press knob to select. Signal Quality Signal Quality May be Compromised by: • Pulsatility • Signal intensity (e.g., kinking of catheter, blood clot, hemodilution) • Intermittent wall contact by the catheter Signal Quality May be Improved by: • Try to aspirate distal lumen; if able to aspirate, flush lumen with extreme caution • Check catheter for kinking and recalibrate; replace catheter if required and recalibrate • Reposition catheter and if SQI > 2, recalibrate monitor by performing in-vivo calibration • Attempt to distance electrocautery equipment and cables from the Vigilance monitor • Plug the power cord into separate AC circuits if possible • Update entered hemoglobin and hematocrit values when there is a physiologic change of 6% or greater in hematocrits or of 1.8 g/dl (1.1 mmol/L) or greater in hemoglobin
See Vigilance II monitor operator’s manual for detailed instructions. For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences devices placed on the European market meeting the essential requirements referred to in Article 3 of the Medical Device Directive 93/42/EEC bear the CE marking of conformity. Edwards, Edwards Lifesciences, the stylized E logo, Swan-Ganz, Vigilance, and Vigilance II are trademarks of Edwards Lifesciences Corporation. © 2012 Edwards Lifesciences Corporation. All rights reserved. AR07822
Edwards Lifesciences Irvine, USA I Nyon, Switzerland I Tokyo, Japan I Singapore, Singapore I São Paulo, Brazil edwards.com