Necessary operating steps 1 Opening the gas bottle
5 Checking the function and seal of the instrument
VWHULOHZDWHURU 1D&OVROXWLRQ DSSUR[Ɠ&
1. Immerse the instrument tip to a depth of approx. 5 cm. 1. Open the bottle valve. 2. Check the seal of all connections. No gas may escape.
2 Switch on the device
2. If bubbles develop during immersion, eliminate the bubbles by moving the instrument tip back and forth quickly in the water. 3. Activate ERBECRYO 2 for approx. 5 seconds. At a sufficient level of freezing performance, a clearly visible, homogeneous ball of ice must form at the tip of the instrument. No gas bubbles should escape. Never use an instrument if the freezing performance is inadequate or if the instrument is not leak-tight. If an instrument is not leak-tight, there is a risk of a gas embolism.
6 Freezing 3 Connecting the instrument
4 Checking the reusability of the instrument
1. Position the instrument. 2. Press the footswitch. The Freeze mode icon is shown on the display. 3. Carry out the required application. 4. Note the timer setting with the optical and acoustic signals in each case.
7 Thawing
Release the footswitch.
CRYOSURGERY
80114-351
07.16
8 Closing the gas bottle after use
11 Changing the timer setting
Optional operating steps 9 Select program The effect and timer are saved in a program.
• No Timer: The freeze interval is not measured. • Count Up: The freeze interval is measured in increments of one second • Count Up & Beep: The freeze interval is measured in increments of one second. A signal sounds once every second. • Count Down & Beep: Enables selection of a countdown interval. A signal sounds once the countdown has been completed.
10 Changing the effect
By selecting the effect you can influence the temperature of the probe tip. The greater the effect, the colder the probe tip will be.