IES 2 Service Manual
Service Manual Art. No. 80116-443 Unit Art. No. 10322-000 All rights to this Service Manual, in particular rights of duplication, dissemination and translation, are reserved. No part of this Service Manual may be reproduced in any form (by photocopying, microfilming or other methods) or processed, duplicated or disseminated by the use of electronic systems without the written consent of ERBE Elektromedizin GmbH. The information contained in this Service Manual may be amended or supplemented without prior notice and represents no obligation on the part of ERBE Elektromedizin GmbH. Printed by ERBE Elektromedizin Printed in Germany Copyright © ERBE Elektromedizin GmbH, Tübingen 2003
INHALT
Table of Contents
Chapter 1
Title
Page
Safety instructions ... 7 Classification of the safety information ...7 Knowledge of the User Manual...7 Protection from the risk of electric shock ...7 Liability and warranty ...8
2
Device structure and function... 9 General view ...9 Description of function ...11 Basic setting ...12 Software functions ...13 Operating mode ... 13 Setting mode... 17 Service mode... 19
Art.-Nr.: 80116-443 11.03
Hardware description ...21 Processor board ... 22 Pressure sensor... 22 Temperature sensor ... 22 Controller supervisor... 22 Power switch... 23 Power supply ... 23 Potential equalization socket ... 23 Power connection with fuses ... 23 Line filter motor ... 23 Motor control... 24 EEProm filter... 24 ECB interface (ECB = ERBE Communication Bus)... 24 Buttons... 24 LED display... 24
3
Replacing components... 25
4
Remedying malfunctions... 27
5
Maintenance and servicing ... 29 Safety check ...29
6
Technical data ... 31
7
Spare parts... 33
5 / 38
Art.-Nr.: 80116-443 11.03
INHALT
6 / 38
1 • Safety instructions
CHAPTER 1
Safety instructions
Classification of the safety information WARNING!
The WARNING! safety indication refers to a risk of personal injury.
CAUTION!
The CAUTION! safety indication refers to a risk of damage to property.
ATTENTION!
The ATTENTION! safety indication refers to a risk which can cause equipment to become unserviceable.
IMPORTANT!
The IMPORTANT! designation indicates application information and other particularly important information.
Knowledge of the User Manual
Art.-Nr.: 80116-443 11.03
The user manual relating to the unit forms part of this Service Manual. Familiarity with the User Manual, in particular the procedures for setting up, commissioning and handling described in the manual, is a prerequisite for the performance of servicing work.
Protection from the risk of electric shock WARNING!
The supply voltage must match the voltage specified on the rating plate. Connect the unit / the equipment cart to a properly installed grounded outlet. Only use the ERBE power cord or an equivalent power cord for this purpose. The power cord must bear the national test symbol. For safety reasons, multiple outlets and extension cords should not be used. If their use is unavoidable, they, too, must be provided with proper grounding.
WARNING!
Unplug the power cord from the outlet before exchanging parts of the unit or cleaning it.
WARNING!
Do not plug a wet power cord into the unit or into an outlet.
WARNING!
Do not touch any unprotected wires or conductive surfaces while the unit is disassembled and is under voltage.
WARNING!
The unit is protected by mains fuses. If one of these fuses blows, the unit must not be used on patients until it has been checked by a properly trained technician. Only replacement fuses of the rating specified on the unit’s name plate may be used.
7 / 38
1 • Safety instructions
Liability and warranty This Service Manual enables your service technician to perform maintenance and repair work to the extent required. Such work must only be conducted by ERBE technicians or by persons who have been expressly authorized to do so. The manufacturer accepts no liability and the right to claim under warranty lapses, if:
Art.-Nr.: 80116-443 11.03
• improper calibration, maintenance, modifications or repairs to the unit or accessories are performed by unauthorized persons, • original spare parts are not used.
8 / 38
2 • Device structure and function
CHAPTER 2
Device structure and function
Art.-Nr.: 80116-443 11.03
General view
Abb. 2-1
1. 2. 3. 4.
User interface Power switch Filter lock Main filter
9 / 38
Art.-Nr.: 80116-443 11.03
2 • Device structure and function
Abb. 2-2
1. 2. 3. 4. 5.
10 / 38
Potential equalization Power connection Power fuse Footswitch socket ECB socket (ECB = ERBE Communication Bus) for connecting the IES 2 to a VIO electrosurgical unit
2 • Device structure and function
Description of function The IES 2 is an electrical medical device which purifies the air of smoke and gas particles which result from the vaporization of human and animal tissue. This smoke typically arises during the use of laser and electrosurgical devices and generally consists of water vapor, aerosols and organic gases. The IES 2 improves the ambient conditions in the OR: • • • •
reducing the amount of dust by removing respirable particles, improving vision, removing unpleasant-smelling gases, evacuating and filtering bioaerosols.
The IES 2 extracts the smoke-laden air via a device on the surgical applied part or via a hand-held tube which is separate from the device and positioned near the applied part. The evacuated air is then purified of smoke and gas particles by means of a high-performance filter, and recirculated to the ambient air.
Art.-Nr.: 80116-443 11.03
Air purification refers to particles such as aerosols, which are captured by an ULPA high-performance particle filter, and also to organic gases, which are detected by an odor that humans find unpleasant, among other things, and which can be extracted by a special gas filter.
11 / 38
2 • Device structure and function
Basic setting The IES 2 is delivered ex works with the following basic settings: • • • • • • • • • •
Operation suction: 80 % Basic suction: 10 % Run-on time: 2 s Occlusion / system blocked (sensitivity): 5 Occlusion / system blocked (retrigger hold-time): 0 s Display brightness: 10 Sound: on (=1) Basic suction time: 30 s Setting-mode inhibit: off (=0) Basic setting inhibit: off (=0)
Resetting the IES 2 to the basic setting (RESET) The IES 2 can only be reset to the basic setting if the option "Basic setting inhibit" in the service mode is deactivated (=0).
Art.-Nr.: 80116-443 11.03
IMPORTANT!
Fig. 2-3
1. When switching on the unit, press the "Operation suction" button (1) until S.0 appears in the display. 2. Press the "Activation on / off" button (4) within 5 seconds. The unit is reset to the above basic settings if: • P.1 appears in the display, • the "Basic suction" button (2) illuminates and • the display unit (6) flashes. 3. The unit is now in setting mode. In order to go into operating mode, one of the two buttons "Operation suction" (1) or "Run-on time" (3) must be pressed.
12 / 38
2 • Device structure and function
Software functions Operating mode Overview of user display:
Art.-Nr.: 80116-443 11.03
Fig. 2-4
Buttons 1. 2. 3. 4. 5.
Operation suction Basic suction Run-on time Activation on / off Higher / Lower
Indicators 6. Display 7. System blocked / occlusion 8. Filter exhausted 9. Display unit (with percentage and / or seconds display)
13 / 38
2 • Device structure and function
Operation suction:
The operation suction capacity can be changed at any time. To do so, press the "Operation suction" button. The value can be set between 20 % and 100 %. 100 % corresponds to a suction capacity of approx. 650 l/min. By pressing and holding the "Higher/Lower" button the values change quickly, in increments of 5. When the electrosurgical device is activated, operation suction is automatically switched on. The unit cannot be activated in the absence of a filter. Basic suction (and basic suction time):
The basic suction capacity can be changed at any time. To do so, press the "Basic suction" button. The value can be set between 0 % and 30 %.
The basic suction time can be changed in the setting mode. The value can be set to between 0 seconds (= no basic suction) and 99 seconds or to continuous (= on). The basic suction time is only active if the basic suction time is not set to 0 seconds and the basic suction value is not set to 0 %. There is no basic suction in the absence of a filter. Run-on time:
The run-on time can be changed at any time. To do so, press the "Run-on time" button. The value can be set to between 0 and 10 seconds or to continuous (= on). By pressing and holding the "Higher / Lower" button the values change quickly, in increments of 5. Flow indicator:
The flow of the three suction options (operation suction, basic suction, run-on time) is shown in the display in % or seconds.
14 / 38
Art.-Nr.: 80116-443 11.03
By pressing and holding the "Higher / Lower" button the values change quickly, in increments of 5.
2 • Device structure and function
Filter level indicator:
The "Filter exhausted" indicator: • lights up red when the filter is in the reserve range, • flashes red when the filter is exhausted or an exhausted filter has been inserted by mistake. In addition to the visual alert, three warning tones are emitted at one-second intervals. No activation occurs in the absence of a filter. If the filter is removed during operation, the unit automatically switches off. The current filter status can be checked in the setting mode: • New filter: 99 % • Filter almost exhausted: 3 % • Filter exhausted: 0 % Blocked suction detection / occlusion:
If the unit detects possible blocked suction:
Art.-Nr.: 80116-443 11.03
The indicator "System blocked / occlusion" lights up red and an acoustic signal is emitted (1 second on, 1 second off). The "Activation on / off" indicator goes out. The visual and acoustic alert lasts 5 seconds. If blocked suction is detected, the unit switches off. No basic suction occurs. Another activation (=retriggering) is only possible after the retrigger hold time. After retriggering the warning goes off. The sensitivity of the blocked suction detection and the retrigger hold time can be set in the setting mode.
15 / 38
2 • Device structure and function
Activation:
The unit can be activated in three different ways: Activation type 1
via the "Activation on / off" button
on the IES 2
The unit evacuates in operation suction mode for as long as the button is pressed. Afterwards, the unit continues to evacuate until the run-on time expires. After the run-on time, basic suction may occur, depending on the setting. No activation occurs in the absence of a filter. Activation type 2 IMPORTANT!
via ECB bus (automatic via the VIO electrosurgical unit)
The IES 2 must be connected via an ECB cable to a VIO electrosurgical unit Version 1.3.x or higher. If the smoke evacuation device and the electrosurgical unit are connected up properly, the dot on the extreme right of the IES 2 display lights up. The unit extracts in operation suction mode for as long as the electrosurgical unit is activated. Afterwards, the unit continues to extract until the run-on time expires. After the run-on time, basic suction may occur, depending on the setting. No activation occurs in the absence of a filter. via an external connection (footswitch) on the rear of the IES 2
The unit extracts in operation suction mode for as long as the footswitch is pressed. Afterwards, the unit continues to extract until the run-on time expires. After the run-on time, basic suction may occur, depending on the setting. If the run-on time is set to continuous (= on), the unit can be activated by depressing the footswitch once. Pressing the footswitch once more will switch off the unit again. No activation occurs in the absence of a filter.
16 / 38
Art.-Nr.: 80116-443 11.03
Activation type 3
2 • Device structure and function
Setting mode IMPORTANT!
The setting mode can only be called up if the option "Setting-mode inhibit" is deactivated (=0) in the service mode. Calling up setting mode, modifying settings, quitting the setting mode:
Art.-Nr.: 80116-443 11.03
Fig. 2-5
Calling up the setting mode 1. When switching on the unit, press the "Basic suction" button (2) until "P.0" appears in the display. 2. Press the "Activation on / off" button (4) within 5 seconds. The setting mode is called up if: • "P.1" appears in the display, • the "Basic suction" button (2) illuminates and • the indicator unit (6) flashes.
Modifying settings 1. Using the "Higher / Lower" buttons (5), select the setting (P.1 to P.6) which is to be modified. 2. Press the "Activation on / off" button (4). The selected setting can now be modified using the "Higher / Lower" buttons (5). 3. In order to save the modified setting press the "Activation on / off" button (4) once more.
Quitting the setting mode 1. Press either the "Operation suction" (1) or the "Run-on time" (3) button. The unit switches to the operating mode.
17 / 38
2 • Device structure and function
Settings:
Assembly
Parameter
Range
P. 1
General
Display brightness
1 – 10
P. 2
General
Sound (off / on)
0 (=off) 1 (=on)
P. 3
Occlusion
Sensitivity
0 – 10 0 (=off)
P. 4
Occlusion
Retrigger hold time
0 – 10 s
P. 5
Basic suction
Basic suction time
1 – 99 s continuous (=on)
P. 6
Filter
Display of service life
0 – 99 % 99 % (=new)
Art.-Nr.: 80116-443 11.03
No.
18 / 38
2 • Device structure and function
Service mode Calling up service mode, selecting programs, quitting service mode:
Fig. 2-6
Art.-Nr.: 80116-443 11.03
Calling up service mode 1. When switching on the unit, press the "Higher / Lower" buttons (5) until . 0 appears in the display. 2. Press the "Activation on / off" button (4) within 5 seconds. Service mode is called up if: • . 1 appears in the display, •
the buttons
•
the display unit (6) flashes.
light up and
Calling up the service functions 1. Using the "Higher / Lower" buttons (5), call up the desired service function ( . 1 to . 5). 2. Press the "Activation on / off" button (4). The selected service function can now be seen or modified using the "Higher / Lower" buttons (5). 3. In order to save the modified settings, press the "Activation on / off" button (4).
Quitting the service mode 1. Press either the "Operation suction" (1) or the "Run-on time" (3) button. The unit switches to the operating mode.
19 / 38
2 • Device structure and function
Service functions:
No.
Assembly
Parameter
Range
.1
Error
Error list display
See error list table
.2
Error
Delete error list
0 (= no) 1 (= yes)
.3
Parameter
Basic setting inhibit
0 (= off) 1 (= on)
.4
Parameter
Setting-mode inhibit
0 (= off) 1 (= on)
.5
Software
Display version
xx (00 – 99)
Display
Explanation
Range
0.x
last error
x = error code
1.x
last error but one
x = error code
2.x ... 9. x
Error log
x = error code
Display
Explanation
Range
y.0
No error
y = error list item
y.1
ECB error
y = error list item
y.2
Excess temperature
y = error list item
y.3
Excess temperature
y = error list item
y.4
Software error (check total)
y = error list item
y.5
Setup parameter error
y = error list item
y.6
Filter data erroneous
y = error list item
y.7
Pressure sensor error
y = error list item
Error codes:
20 / 38
Art.-Nr.: 80116-443 11.03
Error list:
2 • Device structure and function
Hardware description
Potential equalization socket
Power connection with fuses
Line filter motor
Motor Motor control
Power switch
Processor board
+24VDC Controller
Art.-Nr.: 80116-443 11.03
Power supply
+12VDC Motor +5VDC ECB
Pressure sensor Temperature sensor Controller supervisor
EEProm filter
ECB interface
Buttons LED display
Fig. 2-7 IES 2 block diagram, from 10.04.03
21 / 38