Periodic Inspection Checklist
5 Pages
![Molift Air 200/350/500 Periodic inspection checklist In accordance with ISO:10535 PI15701 Rev A 2021-12-13 Components, Molift Air 200/350 Trolley Connection point buttons Hoist 4 Rev. Updated according to legal req. Description: Side cover 2021-03 Revised Lifting belt Quick release coupling Emergency stop / Emergency lowering 2 Quick release pin Rev. Updated according to legal req. 2021-03-19 Description: Revised Date: Revised By: Released Date: Released By: ECP ID: 2 Updated according to legal req. Rev. Description: ELJ ELJ 2021-03-19 Sling bar GS1-128 2021-03-21 ELJ 2021-03-21 Revised Date: Revised By: Released Date: Released By: ECP ID: Size 76,5mm x 25,5mm Silver background Size 76,5mm x 25,5mm Silver backgroundHand control QR code URL: http://q.etac.com/004s Product label 200/350 ID label 76,5mm Air 200 Gross weight 216kg / 476lbs SWL 205kg / 452lbs 24V SLA, 25.6V (LiFePO4), IP24 Duty cycle 10% Size 76,5mm x 25,5mm 25,5mm 25,5mm(01)XXXXXXXXXXXXXX(11)XXXXXX(21)XXXXXX ABCDEFGHIJ Silver background XXXXXX QR code URL: http://q.etac.com/004v YYYY-MM-DD 60601-1 SWL value Air 350 76,5mm Serial no. Serial number Gross weight 361kg / 797lbs SWL 350kg / 772lbs 24V SLA, 25.6V (LiFePO4), IP24 Duty cycle 10% Production date 25,5mm 60601-1 76,5mm Article no. Etac A/S Parallelvej 3 DK-8751 Gedved www.etac.com Art. number of marked product](https://public.bioclinicalservices.com.au/library/representations/proxy/eyJfcmFpbHMiOnsiZGF0YSI6MTI3OTMwLCJwdXIiOiJibG9iX2lkIn19--8306bc919823abde873ff465d40f11cc9716a627/eyJfcmFpbHMiOnsiZGF0YSI6eyJyZXNpemUiOiIzMjB4In0sInB1ciI6InZhcmlhdGlvbiJ9fQ==--4214356078d1f110e449cda512a9432562da3e70/Molift%20Air%20200,350,500%20Rev%20A%20Periodic%20inspection%20checklist%20December%202021.pdf.png)
Preview
Page 1
Molift Air 200/350/500
Periodic inspection checklist In accordance with ISO:10535 PI15701 Rev A 2021-12-13
Components, Molift Air 200/350 Trolley Connection point buttons
Hoist 4 Rev.
Updated according to legal req.
Description: Side cover
2021-03 Revised
Lifting belt Quick release coupling
Emergency stop / Emergency lowering 2
Quick release pin Rev.
Updated according to legal req.
2021-03-19
Description:
Revised Date: Revised By: Released Date: Released By: ECP ID:
2
Updated according to legal req.
Rev.
Description:
ELJ
ELJ
2021-03-19
Sling bar
GS1-128 2021-03-21 ELJ 2021-03-21 Revised Date: Revised By: Released Date: Released By: ECP ID: Size 76,5mm x 25,5mm Silver background
Size 76,5mm x 25,5mm Silver backgroundHand control QR code URL: http://q.etac.com/004s Product label 200/350
ID label
76,5mm
Air 200
Gross weight 216kg / 476lbs SWL 205kg / 452lbs 24V SLA, 25.6V (LiFePO4), IP24 Duty cycle 10%
Size 76,5mm x 25,5mm 25,5mm 25,5mm(01)XXXXXXXXXXXXXX(11)XXXXXX(21)XXXXXX ABCDEFGHIJ Silver background XXXXXX QR code URL: http://q.etac.com/004v YYYY-MM-DD
60601-1
SWL value
Air 350
76,5mm
Serial no.
Serial number
Gross weight 361kg / 797lbs SWL 350kg / 772lbs 24V SLA, 25.6V (LiFePO4), IP24 Duty cycle 10%
Production date
25,5mm
60601-1
76,5mm
Article no.
Etac A/S Parallelvej 3 DK-8751 Gedved www.etac.com
Art. number of marked product
Molift Air 200/350/500 / PI15701 Rev A / www.etac.com
Components, Molift Air 500
Trolley Connection point buttons
Hoist Side cover
Lifting belt
4
Updated according to legal req.
2021-03-1
Rev.
Description:
Revised D
Emergency stop / Emergency lowering
Quick release coupling
3
Updated according to legal req.
2021-03-21
Rev.
Description:
Revised Date: Revised By: Released Date: Released By: ECP ID:
ELJ
ELJ
2021-03-21
Quick release pin
Sling bar
Hand control
GS1-128 Size 76,5mm x 25,5mm Silver background
Size 76,5mm x 25,5mm Silver background QR code URL: http://q.etac.com/004w Product label 500
76,5mm
ID label
Air 500
Gross weight 519,4kg / 1145,1lbs SWL 500kg / 1102,3lbs 24V DC, IP24 Duty cycle 10%
25,5mm 25,5mm(01)XXXXXXXXXXXXXX(11)XXXXXX(21)XXXXXX ABCDEFGHIJ
XXXXXX
60601-1
SWL value
YYYY-MM-DD
76,5mm
Serial no.
Serial number
Article no.
Etac A/S Parallelvej 3 DK-8751 Gedved www.etac.com
Production date
Art. number of marked product
2
Molift Air 200/350/500 / PI15701 Rev A / www.etac.com
A copy of this document is available for download at www.etac.com.
The hoist’s software Molift Service tool provides option to enter date of PI and name of inspector along with additional information. Molift recommends adding all observations / remarks concerning the hoist in the Molift Service tool for complete history log of the hoist
Scan QR code on product label for accessing product home page. The user manual can be found in the document section. Article no. (see ID label): ...
Visual examination
Serial no. (see ID label):
Visual examination of load bearing structure to make sure there is no damage, cracks, frays or deformation. All checkpoints must be checked to approve hoist for further use
... Production Year: ... OK
Owner:
Lifting belt
...
Product label
Specify model:
Plastic covers Hand control
Molift Air 200 Molift Air 350 Molift Air 500
Cables Other electronics
Sling shall be inspected separately and is not included in this inspection.
Sling bar Sling bar hooks
Service light indicator
Quick release coupling
Note colour of service light indicator: Status
Not OK
Instruction for use / user manual is present / available
Service light (No light)
Mode Ready for use Perform periodic Yellow inspection Green blink Order service Red Perform service Red (and sound Perform service when activated) immediately
Situation of use Home Hospital Nursing home Other The periodic inspection is to be performed by a person who is suitable and properly qualified and well acquainted with the design, use and care of the hoist 3
Molift Air 200/350/500 / PI15701 Rev A / www.etac.com
Functional examination
OK
Not OK Perform new visual control; Damage and deformations as described above. Any damaged parts must be repaired or replaced and test is repeated once more until the hoist performs correctly
Test FUNCTION and inspect for wear and damage. All checkpoints must be checked to approve hoist for further use OK
Not OK Emergency stop button is fully operational and interrupts power when engaged. Reset switch Check tools for emergency lowering (found behind covers)
Charger is operational Battery. Battery light is green, and service light indicator is OFF Accessories (e.g. scale, propulsion, IRC):
Electrical emergency lowering and control (buttons, wire H sockets, and battery indicator) Hand control functionality Quick release coupling (swivel joint moves freely when turned 360° and without noise) Quick release pin Connection point buttons Angle sensor. Pull lifting belt sideways in all four directions Checked that the hoist functions correctly and runs without load all the way up and down without abnormal noise End stops If all points so far are “OK”, hoist shall be tested with SWL. NOTE!: Ensure SWL for the complete hoist system exceeds or equals hoist SWL before any load testing. Perform one lifting cycle with weight equal to SWL with min. range of 500 mm in the top of lifting range. Brake must not slip when loaded. Movement must be smooth and even, and without unusual noise Move hoist loaded with SWL through entire rail system. Check deflection and suspension is acceptable
... ...
Performed by Full name: ... Date/Place: ... Signature: ... Approved without faults Next inspection (YYYY / MM): .../... Perform “Reset of hoist” Hoist is marked with “Out of order” and sent for repair Hoist is not eligible for repair and taken out of service If periodic inspection reveals any defect, wear or other damage that jeopardises the safety of the patient the hoist may not be used until the deficiency has been eliminated. The owner is notified
4
Molift Air 200/350/500 / PI15701 Rev A / www.etac.com
Due to legal requirements this document or a copy must be filed with the owner’s service log book If approved without faults, apply inspection label on cover inside and mark with month and year for next inspection. When ordering label, use item no. 1100306 Mark label with month and year for next inspection
Reset of hoist
1 Ensure hoist is ON and emergency stop is deactivated.
2
2
3
3
Press and hold both buttons for approx. 4 sec.: First beep when start pressing. Second beep - stop pressing. 2 sec. pause
Pause for approx. 2 sec.
4
2
Press and hold both buttons for approx. 4 sec.: First beep(s) - stop pressing. Inspection indicatorlight is OFF - reset complete.
5