1 Page
Preview
Page 1
STEP 1
STEP 2
Cleaning the HARMONIC® Hand Piece (HP054) is essential to ensuring optimal device performance. Failure to clean the Hand Piece contact rings (gold rings) properly can result in communication issues between the Hand Piece and the disposable device.
Assemble the Hand Piece Cleaning Tool
Clean the Hand Piece
This step-by-step guide explains how to properly clean the HARMONIC® Hand Piece contact rings manually with the Hand Piece Cleaning Tool or a cotton swab. The Hand Piece (HP054) may be cleaned manually or by machine. Regardless of the method you choose, it is recommended you follow the steps in this guide to clean the upper and lower contact rings on the hand piece (HP054). The Hand Piece can be cleaned with a soft bristle brush or equivalent. Do not use abrasive materials. The Hand Piece Cleaning Tool or cotton swab should be used to clean the Hand Piece prior to sterilization and reuse of the Hand Piece.
1b. Place a large rubber ring on the Hand Piece Cleaning Tool
1c. P lace a small rubber ring on the Hand Piece Cleaning Tool
The HARMONIC® Hand Piece (HP054) is immersible and may be soaked in a pH neutral detergent or pH neutral enzymatic detergent for a period of time up to one hour prior to cleaning.
1d. F ully assembled Hand Piece Cleaning Tool
Manual cleaning can be performed one of the following two ways.
2a. Immerse Hand Piece in pH neutral detergent or pH neutral enzymatic detergent for up to one hour and clean with a soft bristle brush or equivalent
Ensure all detergent and disinfectant residuals are sufficiently removed by rinsing the Hand Piece in purified water. Repeat rinsing as needed until the Hand Piece is thoroughly rinsed and free of all detergents.
Wipe the blade mount surface on the Hand Piece with an alcohol wipe.
Note: After cleaning the Hand Piece, please be sure to clean the contacts (gold rings) following one of the two processes outlined below: Recommended Cleaning Method: Hand Piece Cleaning Tool (HP54CTA)
Alternative Cleaning Method: Cotton Swab
2b. Wet the rubber rings on the Hand Piece Cleaning Tool with pH neutral or enzymatic detergent
2b. Dip a cotton swab in Isopropyl alcohol or a pH neutral solution or detergent and clean around the inside of the upper contact ring until clean.
2c. Insert the Hand Piece Cleaning Tool into the HP054 Hand Piece
2c. Dip a cotton swab in Isopropyl alcohol or a pH neutral solution or detergent and clean around the inside of the lower contact ring until clean.
OR
Small Lower Contact Ring
2d. Rotate the Hand Piece Cleaning Tool back and forth to clean the upper and lower contact rings (gold rings inside the hand piece). Repeat as necessary until clean.
Large Upper Contact Ring
Product Information
Note: Please be sure to clean both the smaller lower contact ring and the larger upper contact ring. Repeat as necessary until clean.
STEP 5
1a. R emove one large and one small rubber ring from the Hand Piece Cleaning Tool package
STEP 3
HARMONIC® Hand Piece (HP054)
STEP 4
HARMONIC® Hand Piece Cleaning
Sterilize the Hand Piece Sterilize the Hand Piece after cleaning following the instructions in the IFU. Note: The rubber rings on the Hand Piece Cleaning Tool should be replaced as needed, when they become dirty or damaged.
How Supplied
Product Code
Description
Quantity/Sales Units
HP54CTA
HARMONIC® Hand Piece Cleaning Tool with 10 large and 10 small rubber rings
The Hand Piece Cleaning Tool is supplied non-sterile. It comes with 10 large rubber rings and 10 small rubber ring replacements.
1
To Order Call JJHCS at 1-800-255-2500 (option 1) between 8:30 a.m. and 8:00 p.m. eastern time, or fax your order to 1-832-562-2212 or 1-800-997-1122. For more information or product support, call 1-877-ETHICON (384-4266).
HARMONIC® Blue Hand Piece (HPBLUE) Note: The Hand Piece Cleaning Tool is only compatible with the HP054. It is recommend you follow the cleaning method outlined on this poster for the HPBLUE Hand Piece. Refer to the instructions to the right on how to clean the HPBLUE Hand Piece:
STEP 1
Inner Contact Ring
Outer Contact Ring
Note: Please be sure to clean both the inner and outer contact rings.
1a. I mmerse Hand Piece in pH neutral detergent or pH neutral enzymatic detergent for up to one hour and clean with a soft bristle brush or equivalent.
1b. Dip a cotton swab in Isopropyl alcohol or a pH neutral solution or detergent and clean around the inner contact ring until clean.
STEP 2
STEP 3
STEP 4
Ensure all detergent and disinfectant residuals are sufficiently removed by rinsing the Hand Piece in purified water. Repeat rinsing as needed until the Hand Piece is thoroughly rinsed and free of all detergents.
Wipe the blade mount surface on the Hand Piece with an alcohol wipe.
Sterilize the Hand Piece after cleaning following the instructions in the IFU.
1c. Dip a cotton swab in Isopropyl alcohol or a pH neutral solution or detergent and clean around the outer contact ring until clean.
©2015 Ethicon US, LLC. All rights reserved. 038231-150806