Quick Guide
Agilia® SP MC WiFi Connected Generation
NOTE: This Quick Reference Guide is not intended to be comprehensive instructions for the set-up and operation of the Agilia SP MC and Agilia SP MC WiFi Volumetric Infusion Pumps.
DESCRIPTION Keypad
Syringe Flange Cradle
Disengagement Lever Syringe Barrel Clasp
Plunger Driver
INSTALLATION Secure the pump stability (e.g IV pole fixing) Plug in power cord
The Power Supply Indicator lights up INSTALLING A SYRINGE Connect the extension set to the syringe according to local practices. Prime and check that there is no air in the syringe or extension set. IMPORTANT: Ensure that the patient is not connected. Place the syringe into the cradle, the flanges inserted in the provided slot. Secure the syringe with the syringe barrel clasp. Move the plunger driver gently until it makes contact with the plunger head.
General Operations 1. Selecting a Profile Press <ON/OFF> to power ON the pump. Use arrow keys to select desired profile. Press <OK> to confirm the profile.
2. Selecting a Syringe Press <OK> to confirm the displayed syringe, or <C> to access the syringe list. Press the arrow keys to select a new syringe.
Press <OK> to confirm the new syringe.
3. Selecting a Drug Press the arrow keys to scroll to the drug’s first letter, and press <OK>. NOTE: Drugs are sorted alphabetically by the first letter.
Press the arrow keys to scroll to the drug name, and press <OK>. Press <OK> to acknowledge the clinical advisory message and continue programming, or <C> to change the drug.
Programming an Infusion by Flow Rate Press the arrow keys to program the flow rate. Press <start>.
Programming an Infusion by Dose Selecting the Drug Concentration
Range
Finite values
Press the arrow keys to select the concentration. Press <OK> to confirm. Selecting the Patient’s Characteristics Press the arrow keys to enter the patient’s weight or body surface area. Press <OK> to confirm.
Programming the Infusion Press the arrow keys to program the dose rate value. Press <OK> to confirm.
Starting an Infusion Connect the administration set to the patient’s IV access device using aseptic technique and according to your healthcare facility policy.
Flow rate
Dose
Press <start> to start the infusion, or <C> to modify the infusion settings. Functions During Infusion 1. Stop infusion Press <STOP> key: After 2 minutes, an alarm is generated as a reminder that the infusion is stopped. Confirm or modify settings prior to restarting the infusion.
2. Rate Titration (while infusing)
Flow rate
Dose
Press the arrow keys to modify the flow rate or dose. Press <OK> to confirm. Accessing the Menu Functions Press <MENU> key to access menu items, listed on left side of display. Use arrow keys to select menu item. Press <enter> to activate menu function. Follow prompts on the screen. Powering off Press <STOP> key to stop the infusion. Press and hold <ON/OFF> until the pump powers off.
Administering a Bolus: Direct or Programmed Bolus There are two ways to deliver a bolus during an infusion: 1. Direct Bolus During the infusion, press <BOLUS> key. Press <bolus> to confirm access.
Press and hold <BOLUS> key. To stop the bolus, release the <BOLUS> key: The infusion resumes its previous rate.
2. Programmed Bolus Press <BOLUS> key. Press <prog>.
Flow rate
Dose
Press the arrow keys to program the bolus volume or dose, and press <OK>. Press the arrow keys to program the bolus duration, and press <OK>: The flow rate is calculated automatically.
Flow rate
Dose
Press <start> to start the bolus. After bolus is delivered the infusion resumes its previous rate. Interrupting a Programmed Bolus
Flow rate
Press <STOP> key. Press <no> to return to the infusion. Press <start> to continue the bolus.
Dose
KEYPAD DESCRIPTION
On / Off Increment (fast/low) Decrement (low/fast) Confirm Value / Move to Next Field Stop / Pause Bolus / Prime Pressure Menu/ Menu Cancel Value / Move Back to Previous Field Alarm Silence LIGHT INDICATORS
Power Supply Indicator Battery Charge Status Indicator Infusion in Progress: flashing green
Low-Priority Alarm: constant yellow / Medium-Priority Alarm: flashing yellow
High-Priority Alarm: flashing red
This document may contain inaccuracies or typographical errors. Modifications may thus be made, and included in later editions. Due to the evolution of standards, and of legal texts and materials, the characteristics indicated in the text and images of this document are applicable only to the device with which it is included.