Fresenius Kabi
Link Agilia and Link+ Agilia Data Sheet
Data Sheet
2 Pages
Preview
Page 1
MEDICAL DEVICES
CC10342-0_BW_SPEC_SYS_data_sheet_Link_Link_Plus_Agilia_eng
Link Agilia Link+ Agilia
Fresenius Kabi Australia Pty Limited Level 2, 2 Woodland Way Mount Kuring-gai NSW 2080 Australia Phone: 02 9391 5555 Customer Service: 1300 732 001 www.fresenius-kabi.com.au PM2016.034v4
Fresenius Kabi New Zealand Limited 60 Pavilion Drive Airport Oaks Auckland 2022 New Zealand www.fresenius.com.au Customer Service: 0800 144 892 TAPSMR5022
DATA SHEET
Link Agilia & Link+ Agilia Link Agilia and Link+ Agilia are compatible with Agilia IV Pumps and Large Volume Pumps from Fresenius Kabi.
FUNCTION
Data communication with the Agilia IV infusion pumps.
MODE
Wireless optical communication using infrared light.
COMPATIBILITY
Asynchronous Serial Infrared (SIR) physical layer irPHY 1.0, baseband no carrier.
TRANSPORT PROTOCOL
Proprietary.
SPEED
115.2 kb/s max.
WAVELENGTH
880 nm to 900 nm infrared band with 45 nm spectral bandwidth.
EYES SAFETY
< 500 mW/sr maximum intensity for class 1 of IEC/EN 60625-1 - January 2001.
Link+ Agilia
INFRARED (IR) CONNECTOR SPECIFICATIONS
BATTERY SPECIFICATIONS FUNCTION
Backup battery for mains power failures and transport of the rack.
ACCESS
Internal battery holder accessible with specific tools by qualified engineer.
CHARACTERISTICS
7.2 V / 2.2 Ah - Lithium Ion rechargeable battery.
BATTERY AUTONOMY
1 H min.
SELF PROTECTIONS
Over-current, over-discharge, over-voltage and overtemperature.
BATTERY LIFE
Preventive replacement after 3 years.
USB PC CONNECTOR SPECIFICATIONS FUNCTION
External PC connection for maintenance and configuration.
CONNECTOR
External standard Mini B slave connector.
INDICATOR
Blue LED.
OUTPUT POWER
No power output available and remote host power not used.
COMPATIBILITY
USB 2.0 OTG Slave role only.
INSULATION
Not insulated from USB external device, but provides double insulation (4 kV) from mains.
SPEED
480 Mb/s max. (theoretical).
RS232 SERIAL COMMUNICATION CONNECTOR SPECIFICATIONS FUNCTION
System management, PC, PDMS, or maintenance connection.
CONNECTOR
ADAM TECH shielded circular female connector, 3 pins.
INDICATOR
Blue LED.
COMPATIBILITY
RS232 levels, serial asynchronous, half duplex.
INSULATION
1.5 kV insulation from external system + provides double insulation (4 kV) from mains.
SPEED
115.2 kb/s max. (theoretical).
DIMENSIONS - WEIGHT
Link Agilia & Link+ Agilia
Link+ Agilia
DATA COMMUNICATION CONNECTOR
DIMENSIONS - WEIGHT
Link 4+: H/W/D 720 mm (28.35 in) x 195 mm (7.68 in) x 163 mm (6.42 in) - Weight 4.1 kg (9.04 lb). Link 6+: H/W/D 1010 mm (39.76 in) x 195 mm (7.68 in) x 163 mm (6.42 in) - Weight 5.2 kg (11.46 lb). Link 8+: H/W/D 1290 mm (50.79 in) x 195 mm (7.68 in) x 163 mm (6.42 in) - Weight 6.4 kg (14.11 lb).
MAINS POWER INPUT SPECIFICATIONS FUNCTION
Primary power source for Agilia power outlets and for internal electronics of Link Agilia and Link+.
CONNECTOR
Standard appliance inlet IEC type C14 - male - 3 poles.
POWER SUPPLY
100 V to 240 V AC / 50 - 60 Hz.
MAXIMUM POWER (with pumps)
Link 4: 60 VA. Link 6: 90 VA. Link 8: 120 VA.
WIRED NETWORK CONNECTOR SPECIFICATIONS FUNCTION
System management, PC, or maintenance connection.
CONNECTOR
RJ45 8 pins.
INDICATOR
Blue LED.
COMPATIBILITY
IEEE 802.3 Ethernet 10baseT or fast Ethernet 100baseT, twisted pairs, Full duplex.
INSULATION
1.5 kV insulation from external system + provides double insulation (4 kV) from mains.
SPEED
Automatic speed selection 10 Mb/s or 100 Mb/s (theoretical).
Link 4: H/W/D 720 mm (28.35 in) x 184 mm (7.24 in) x 163 mm (6.42 in) - Weight 3.6 kg (7.94lb). Link 6: H/W/D 1010 mm (39.76 in) x 184 mm (7.24 in) x 163 mm (6.42 in) - Weight 4.7 kg (10.36 lb). Link 8: H/W/D 1290 mm (50.79 in) x 184 mm (7.24 in) x 163mm (6.42 in) - Weight 5.8 kg (12.79lb).
Link 4+: 75 VA. Link 6+: 105 VA. Link 8+: 135 VA. PROTECTIVE FUSES
2 fuses, 2 A timed with high breaking capacity - T 2A H 250 V. Fuseholder is externally accessible on mains power inlet connector.
ELECTRIC PROTECTION
Class I with protective earth, with 3-wire power cord.
NURSE CALL CONNECTOR SPECIFICATIONS
AGILIA PUMP POWER OUTPUT SPECIFICATIONS
FUNCTION
External alarm NURSE CALL signaling.
FUNCTION
Power source output for Agilia IV infusion pumps.
CONNECTOR
Stereo JACK 3.5 mm connector.
CONNECTOR
Standard line outlet IEC type C13 - female - 3 poles.
INDICATOR
RED LED.
POWER SUPPLY
100 V to 240 V AC / 50 - 60 Hz.
COMPATIBILITY
Relay, dry contacts stand-off / work contact - positive / negative polarity option.
MAXIMUM POWER
15 VA per outlet.
CONTACTS
DC 24 V / 2 A max or AC 48 V / 1 A.
INSULATION
1.5 kV insulation from external system + provides double insulation (4 kV) from mains.
PROTECTIVE FUSES
Report to mains power input fuses.
ON / OFF POWER
No power on empty position - Earth pole is functional earth only.
CENTRAL PROCESSING UNIT (CPU) SPECIFICATIONS FUNCTION
Data processing and communications.
PROCESSOR
32 bits, 400 MHz.
MECHANICAL SPECIFICATIONS FOR LOCKING CLAMPS POLE
Diameter: from 16 to 42 mm Height: from 16 to 42 mm / Depth: 30 mm max.
CENTRAL MEMORY
64 Mbytes, high speed RAM 266 MHz.
RAIL
FLASH MEMORY
Non volatile 256 Mbytes.
COMPLIANCE WITH STANDARDS
OPERATING SYSTEM
Safety and flexible multitasking / multithreading operating system with GPL license.
USER INTERFACE SPECIFICATIONS PUMP CENTRALIZATION DISPLAY
Function: displays infusion pump alarms and pre-alarms. Size: 70 mm x 30 mm - Type: LED - Viewing angle: 360° @ 4 meters. Color: RED (pump alarms - high priority) / YELLOW (pump pre-alarms - Medium and low priority) / RED and YELLOW (system is starting or rebooting after RESET press).
MAIN STATUS INDICATOR BATTERY STATUS INDICATOR SYSTEM STATUS INDICATOR
Function: displays main connected status - Type: LED - Color: GREEN. Function: displays Link+ battery status and alarm - Type: LED - Color: GREEN / YELLOW / RED. Function: displays Link+ internal software status -Type: LED - Color: GREEN / YELLOW / RED.
RESET BUTTON
Function: reboots Link+ internal software - Type: protected push button.
Complies with the 93/42/CEE Medical device directive SAFETY OF ELECTROMEDICAL EQUIPMENTS
Complies with EN/IEC 60601-1. IP22 protection against dust and splashing liquid. Protection against leakage current; type B applied part. Class I with protective earth. Agilia power outlets are wired on internal functional earth to reduce residual current that may disturb ECG or EEG devices.
EMC (electromagnetic compatibility)
Complies with EN/IEC 60601-1-2.