Data Sheet
2 Pages
Preview
Page 1
Volumat MC Agilia
The advanced volumetric infusion pump. Vigilant® patient protection built in.
Volumat MC Agilia FLOW RATE RANGE
FLOW RATE ACCURACY VOLUME TO INFUSE INFUSION MODES
DURATION OF INFUSION
PRIMING OCS PATENT BOLUS KVO PAUSE GRAPHICAL HISTORY DATA LOG EVENT NIGHT MODE PROGRAMMING MODES
DRUG LABELLING DRUG LIBRARIES
KEYBOARD PROTECTION
PRESSURE MODE DPS PRESSURE MONITORING ANTI-BOLUS SYSTEM
PUMP STATUS
SET INSTALLATION CONTROL INFUSION CONTROL
± 5% and even better in most clinical situations. 1-9999 ml in normal mode, 0.1-999.9 ml in micro mode tml/h modes: Volume + Flow rate, Volume + Time, Flow rate + Time, Volume + Time + Rate, Rate only (if optional drop sensor is used), Ramp-up / Ramp-down, Sequential / Intermittent, Secondary /Piggyback, Drop/min tDose rate modes: ng/h, ng/kg/min, ng/kg/h, ng/m2/24h, µg/min, µg/h, µg/kg/min, µg/kg/h, µg/m2/min, µg/m2/h, mg/min, mg/h, mg/24h, mg/kg/min, mg/kg/h, mg/kg/24h, mg/m2/min, mg/m2/h, g/h, g/kg/min, g/kg/h, g/m2/min, g/m2/h, g/m2/24h, mU/min, mU/h, mU/kg/min, mU/kg/h, mU/m2/min, mU/m2/h, U/min, U/h, U/kg/h, U/kg/24h, U/m2/min, U/m2/h, mEq/min, Eq/h, mEq/kg/min, mEq/kg/h, mEq/m2/min, mEq/m2/h, mEq/m2/24h, mmol/h, kcal/h, kcal/24h, kcal/kg/h, kcal/kg/24h. With or without loading dose.
Injectomat® Agilia Handy
Injectomat® MC Agilia
tDose time mode: calculation of the rate is based on dose and dilution over an adjusted time. From 0 h 01’ to 168 h 00’ with warning when recommended duration of set use is elapsed. Possibility of a delayed start-up. Warning function: a warning message can be activated and adjusted from 0 h 01’ to 96 h 00’. Manual or by using the purge/prime function of the pump. The Occlusivity Check System is an auto-test to check the good working order of the pump in association with its set and thus preventing risk of free flow. 1500 ml/h, adjustable from 50 ml/h to 1500 ml/h by 50 ml/h increments. Manual or programmable. Automatic Keep Vein Open rate of 1 ml/h (adjustable from 1 to 20 ml/h) when volume limit is reached. Programmable from 1 minute to 24 hours, increments from minute to minute. Volumes infused, doses infused, flow rates, pressure. 1500 data log events in real time. The night mode decreases the brightness of the screen and the green lights. The key beep may be turned off. The night mode can be programmed manually or automatically in a variable time range.
Comprehensive
Injectomat® TIVA Agilia Intelligent
Volumat Agilia Smart
No drug names: infusion without any display of the drug names. Drug labelling: infusion with the drug names displayed. Vigilant® Drug’Lib: the secured IV administration mode by soft and hard limits. Display of the drug name based on a list of 50 pre-set drug names. Customised drug library with Vigilant®, the IV Medication Safety Solution. Each drug is defined by: drug name, default rate, min. and max. rates (soft and hard limits), and infusion mode. Up to 960 drugs in total storable in 4 different drug libraries (dependent on character number).
Volumat MC Agilia Advanced
Option to lock the keyboard to prevent accidental change of setting. PRESSURE MANAGEMENT 2 modes available: variable (from 50 to 750 mmHg, 25-50 mmHg increment) or 3 pre-set levels. The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated. Graphic representation of the pressure in the infusion line and of the pressure limit thanks to the pictogram.
Links & Accessories
Significantly reduces risk of bolus after occlusion release (0.2 ml max). ALARMS / PRE-ALARMS / SECURITY Green for infusion in progress, orange for pre-alarm and warnings, red for alarm. Visibility up to 5 metres. All alarms are expressed by means of light indicators, written words, pictograms and beep sounds. Door’s closing, set positioning, SafeClip control, OCS test. End of infusion, imminent end of infusion, downstream occlusion, upstream occlusion, line disconnection, air in line, under flow rate, over flow rate, empty container, unconfirmed setting, end of pause, unauthorised flow rate (hard or soft).
DEVICE CONTROL
Motor rotation check, power source indication, mains power disconnection, low battery, discharged battery, technical fault, watchdog check, communication connection failure.
MAINTENANCE
Preventative maintenance warning.
PUMPING MECHANISM DISPLAY SWINGLOCK CLAMP STACKABILITY DIMENSIONS / WEIGHT BATTERY
TECHNICAL SPECIFICATIONS 2nd generation linear peristaltic pumping system with software adjustments and compensations. Blue graphic LCD monochrome, size 66 mm x 33 mm (128 x 64 pixels). Versatile clamp that allows the fixation on a rail or a pole (Pole: 20-40 mm max. / Rail: 25-35 x 10 mm) Up to 3 devices self-stackable on a pole with a safe handle. H140 x W185 x D165 mm / Weight: ~2000 g. Li-Ion, battery life at least 8 hours at 125 ml/h, remaining battery life and battery charge level available on the display. New Battery Management software (only charging when <7.2V). Battery replacement: separate compartment - no need for electrical security testing.
COMPLIANCE STANDARD
EN/IEC 60601-1 and EN/IEC 60601-2-24 / CE 0459 marking in compliance with the EEC 93/42 European Medical Device Directive / EMC.
WATER RESISTANCE ELECTRICAL COMPLIANCE POWER SUPPLY
IP22. Protection against leakage current: CF type – protection against electric shocks: class II. 100–240Vac / 50-60Hz with standard cable, 10Vdc with specific DC cable.
FREE-FLOW PROTECTION
VOLUMAT LINE ADMINISTRATION SETS All sets integrate the SafeClip that automatically clamps the line to avoid any risk of free flow when the door is opened and when the set is disengaged.
PUMPING SEGMENT MATERIAL NEEDLE-FREE ACCESS
Agilia The Agilia family offers a complete range of infusion pumps, syringe pumps, accessories and disposables to cover all major IV drug administration needs.
INFUSION 1-1500 ml/h in normal mode (1 ml/h increment). 0.1- 99.9 ml/h in micro mode (0.1 ml/h increment). Flow rate can be limited according to drug name (soft and hard limits) with Vigilant®, the IV Medication Safety System.
Administration Sets
Maintenance Tools
Please Note: Due to our policy of continuous product development as well as changes in standards, the features described are subject to change. Please consult us for the current situation.
The silicone pumping segment allows a tight flow rate accuracy as well as long term performances. All Volumat Lines infusion sets are DEHP-free & Latex-free. PVC-free and opaque materials are available. Some sets integrate a needle-free Y injection site to protect the medical staff against the risk of needle-stick injuries.
Fresenius Kabi Australia Pty Limited 964 Pacific Highway, Pymble NSW 2073, Australia Customer Service 1300 732 001 Phone +61 2 9391 5555 Fax +61 2 9391 5505 Web www.fresenius-kabi.com.au PM 2010.060