Quick Reference Guide

3 Pages

AquaC UNO H Quick Reference Guide (Acute)  Page 1  Warning: The following quick reference guide will only be used by staff who have been trained and found competent in the operation and use of the AquaC UNO H system. This guide does not supersede the manufacturer’s instruction for use, which must be referred to for the safe and effective operation of the system. Document information is gathered below on the “Operational Data Acquisition/Daily Report AquaC UNO H” unless otherwise indicated or as directed by company policy.  RO Control Legend  Display  Menu key  Mute key Stop key  Start key  Back/Escape key  Visual indicators (green/yellow/red)  Navigation/Enter keys  System Start-Up and Operation 1. Connect RO system to water, drain, and plug electrical cord into an appropriate outlet. 2. Determine last time system was operated and last disinfection (review log). To view last disinfection recorded by the AquaC UNO H: Press  , select “Status,” press  , then select “Hygiene,” press  .  Evaluate as follows: If...  Then...  System operated within last 72 hours and less than 7 days since last disinfection  Continue to step 3.  System not operated within last 72 hours or 7 or more days since last disinfection  Perform a module heat disinfection prior to use. Go to step 3.