Introduction
Video Endoscope
OPERATION MANUAL
1 Precautions 2 Product Overview 3 Workflow 4 Preparation and Inspection 5 How to Use
Lower Gastrointestinal Endoscopes
EC-760S-G/M EC-760S-G/L
This Operation Manual provides details on how to prepare and operate the endoscope and describes cautions to be observed. Please read this manual thoroughly before operating the endoscope. After reading this manual, store it nearby the endoscope so that you can review it whenever necessary.
897N202977B
6 Troubleshooting 7 Service Appendix
897N202977B.book
iii ページ
2021年2月18日
木曜日
午後6時6分
Contents at a Glance Introduction This chapter explains about this manual. Chapter 1
Precautions Before using this product, read this chapter carefully so that you can operate it correctly. This chapter describes the warnings and cautions for safe operation of the endoscope.
Chapter 2
Product Overview This chapter describes details on the accessories supplied with this product, the nomenclature and functions of the endoscope, and related equipment
Chapter 3
Contents at a Glance
connected to this product.
Workflow This chapter describes the workflow of endoscopy.
Chapter 4
Preparation and Inspection This chapter describes the inspection and preparation methods to be performed before using the endoscope, its accessories and related equipment.
Chapter 5
How to Use This chapter describes the basic operation procedures of this product and precautions to observe.
Chapter 6
Troubleshooting This chapter describes actions which should be taken if problems occur while inspecting or using the endoscope.
Chapter 7
Service This chapter explains the services regarding this product.
Appendix This chapter describes main specifications, related equipment used in combination with this product, electromagnetic compatibility (EMC), etc.
897N202977B
iii
897N202977B.book
iv ページ
2021年2月18日
木曜日
午後6時6分
Contents
Contents Contents at a Glance... iii
Introduction
1
About This Manual... 1 Endoscope Operation Manuals... 2 How to Read This Manual ... 3 Terms ... 3 Conventions Used in This Manual ... 3
Chapter 1
Precautions
1-1
1.1
Intended Use ... 1-1
1.2
Incident Reporting ... 1-1
1.3
Applicability of Endoscopy and Endoscopic Treatment... 1-1
1.4
User Qualifications ... 1-2
1.5
Prohibition of Modification and Improper Repair ... 1-2
1.6
Reprocessing Before the First Use/Reprocessing and Storage After Use... 1-3 1.6.1
Reprocessing Before the First Use/Reprocessing After Use ... 1-3
1.6.2
Storage After Use... 1-3
1.6.3
Disposal ... 1-3
1.7
For Safe Operation... 1-4
1.8
Maintenance... 1-4
1.9
Symbols... 1-5
1.10 Precautions for Transportation ... 1-7 1.11 Precautions Against Electric Shock... 1-7 1.12 Electromagnetic Compatibility (EMC)... 1-8 1.13 General Warnings and Cautions ... 1-10 1.13.1 Precautions for Inspection Before Use... 1-10 1.13.2 Handling Precautions ... 1-11 1.13.3 Abnormalities during Use of This Product... 1-15 1.13.4 Precautions for Flexibility Adjustment Mechanism... 1-16 1.13.5 Precautions for Water Jet Function... 1-17 1.14 Precautions for Equipment Used in Combination... 1-19 1.14.1 Precautions for High-Frequency Treatment ... 1-21 1.14.2 Precautions for Endotherapy Device and Syringe ... 1-22 1.14.3 Precautions for Endoscope Visualizer ... 1-23 1.15 Main Indications ... 1-24 1.16 Main Clinical Benefits ... 1-24 1.17 Main Clinical Performances... 1-24 1.18 Main Adverse Events ... 1-24 1.19 Contraindications... 1-24 1.20 Patient Target Group... 1-24
iv
897N202977B
897N202977B.book
v ページ 2021年2月18日
Chapter 2 2.1
2.2
木曜日
午後6時6分
Product Overview
2-1
Checking Package Contents ... 2-1 2.1.1
Endoscope ... 2-1
2.1.2
Accessories... 2-2
Nomenclature and Functions of Endoscope ... 2-4 Scope Connector (One Step Connector) ... 2-6 LG Flexible Portion ... 2-6 Control Portion (G7 control portion) ... 2-6 Boot... 2-7 Insertion Portion (Applied Part) ... 2-8
2.3
Nomenclature and Functions of Distal End of Endoscope ... 2-9
2.4
Flexibility Adjustment Mechanism ... 2-10
2.5
Location of Each Label... 2-11
2.6
System Configuration ... 2-12 2.6.1
Chapter 3 3.1
Chapter 4 4.1
Location of Labels ... 2-11 System Configuration (Combination with VP-7000 and BL-7000) ... 2-14
Workflow
3-1
When Using Standard Functions... 3-2
Preparation and Inspection
4-1
Preparation of the Equipment... 4-2 Endoscope ... 4-2 Accessories... 4-2 Related Equipment (Essential) ... 4-3 Related Equipment (To be Prepared when Necessary) ... 4-3 Personal Protective Equipment... 4-4 Others ... 4-4
4.2
Transporting Endoscope ... 4-5
4.3
Inspecting Endoscope ... 4-6 4.3.1
Inspecting Control Portion... 4-6
4.3.2
Inspecting Flexibility Adjustment Mechanism... 4-6
4.3.3
Inspecting Insertion Portion... 4-7
4.3.4
Inspecting Distal End ... 4-8
4.3.5
Inspecting Bending Section... 4-9 Inspection for Smooth Operation ... 4-9 Inspecting Up/Down Angulation ... 4-10 Inspecting Left/Right Angulation ... 4-11
4.3.6 4.4
897N202977B
Inspecting Scope Connector ... 4-12
Inspecting and Attaching Accessories... 4-13 4.4.1
Forceps Valve ... 4-13
4.4.2
Suction Valve ... 4-14
4.4.3
Air/Water Valve ... 4-16
4.4.4
J Tube ... 4-18
v
Contents
2.5.1
897N202977B.book
vi ページ
2021年2月18日
木曜日
午後6時6分
Contents
4.5
4.6
4.7
Chapter 5 5.1
5.2
4.5.1
Inspecting Related Equipment ... 4-19
4.5.2
Preparing System... 4-19
Connecting Endoscope to Light Source and Related Equipment ... 4-21 4.6.1
Connecting to Light Source... 4-21
4.6.2
Attaching Water Tank... 4-22
4.6.3
Attaching Suction Unit... 4-23
4.6.4
Attaching J Tube ... 4-24
4.6.5
Attaching Water Pump ... 4-25
Inspecting Functions Used in Combination with Related Equipment ... 4-26 4.7.1
Inspecting Endoscopic Images ... 4-26
4.7.2
Inspecting Scope Switch ... 4-27
4.7.3
Inspecting Air/Water Supply Function ... 4-28
4.7.4
Inspecting Suction Function ... 4-29
4.7.5
Inspecting Instrument Channel ... 4-30
4.7.6
Inspecting Water Jet Channel ... 4-30
4.7.7
Inspecting Functions of Endoscope Visualizer... 4-32
How to Use
5-1
Preparation... 5-3 5.1.1
Preparing Related Equipment ... 5-3
5.1.2
Pretreatment of Patient ... 5-3
Insertion and Observation ... 5-4 5.2.1
Using Flexibility Adjustment Mechanism... 5-4
5.2.2
Insertion ... 5-5
5.2.3
Observing Endoscopic Image ... 5-7
5.2.4
Operating Scope Switch... 5-8
5.2.5
Bending Operation ... 5-8
5.2.6
Operating Air/Water Valve ... 5-10
5.2.7
Operating Suction Valve ... 5-11
5.3
Using Water Jet Function ... 5-12
5.4
Injecting Fluids from Instrument Channel Inlet ... 5-14
5.5
Treatment ... 5-15 5.5.1
Using Endotherapy Devices... 5-15
5.5.2
Use of Non-Flammable Gases... 5-17
5.5.3
High-Frequency Treatment ... 5-17
5.6
Endoscope Withdrawal... 5-19
5.7
Reprocessing Endoscope ... 5-20
Chapter 6 6.1
vi
Preparing Related Equipment ... 4-19
Troubleshooting
6-1
Troubleshooting... 6-2 6.1.1
Problem with Displayed Images... 6-2
6.1.2
Problem with Scope Switch... 6-3
6.1.3
Problem with Bending Section ... 6-3
6.1.4
Problem with Air/Water Supply ... 6-4
897N202977B
897N202977B.book
vii ページ
2021年2月18日
木曜日
午後6時6分
6.1.5
Problem with Suction ... 6-5
6.1.6
Problem with Water Jet Function ... 6-5
6.1.7
Problem with Flexibility Adjustment Mechanism ... 6-6
6.1.8
Problem with Related Equipment... 6-7 Problem with Endotherapy Devices ... 6-7 Problem with Image Recorder ... 6-7 Problem with Endoscope Visualizer... 6-8
6.2
6.3
Withdrawal of Endoscope with Abnormality ... 6-9 6.2.1
When Endoscopic Images Appear on the Monitor... 6-10
6.2.2
When Either Normal Observation Mode or Special Light Observation Mode is Not Available ... 6-10
6.2.3
When Endoscopic Images Do Not Appear on the Monitor or a Frozen Image Cannot be Restored ... 6-11
Handling of Endoscope with Abnormality... 6-12
6.4
Chapter 7
When Suction Valve Does Not Return to Its Original Position ... 6-12
Returning Endoscope for Repair ... 6-14
Service
7-1
7.1
Service ... 7-1
7.2
After-Sales Service... 7-1 Repairs during the warranty period ... 7-1 Repairs after the warranty period... 7-1
Appendix
Appendix-1
Main Specification ... Appendix-1 Classification of Medical Electrical Equipment... Appendix-1 Electromagnetic Compatibility (EMC) Related Standard ... Appendix-1 Applied Part ... Appendix-1 Specifications ... Appendix-2 Medical Device Regulation ... Appendix-3 Operating Environment, Transport Environment and Storage Environment ... Appendix-4 Operating Environment ... Appendix-4 Transport Environment ... Appendix-4 Storage Environment ... Appendix-4 Accessories ... Appendix-5 Related Equipment Used in Combination... Appendix-6 Compatible Processor and Light Source ... Appendix-6 Water Tank ... Appendix-6 Suction Unit... Appendix-6 Endoscopic CO2 Regulator and Accessories ... Appendix-6 Water Pump ... Appendix-7 Electrosurgical Unit ... Appendix-7 Ultrasonic Processor ... Appendix-7 Hood ... Appendix-7 Endoscope Visualizer ... Appendix-7 Compatible Endotherapy Devices... Appendix-8 897N202977B
vii
Contents
6.3.1
897N202977B.book
viii ページ
2021年2月18日
木曜日
午後6時6分
Contents
Electromagnetic Compatibility (EMC) Information... Appendix-9 Disposal of Electric and Electronic Equipment ... Appendix-19 Index... Appendix-20 Service Centers
viii
897N202977B
897N202977B.book
1 ページ 2021年2月18日
木曜日
午後6時6分
Introduction
This chapter explains about this manual.
About This Manual
Introduction
This manual provides necessary information for using this product, such as the equipment overview, operation procedures and precautions to observe. In addition, the Reprocessing Manual supplied with this product describes the reprocessing and storage methods for the endoscope. This manual does not provide information about procedures or any aspects of endoscopic techniques. Before using this product, thoroughly read and understand this manual, the Reprocessing Manual and the manual of related equipment and use this product as instructed. Also, after reading this manual, store it close to this product for future reference to keep this product in optimum working condition. If you have any questions or comments about any information in this manual, contact your local FUJIFILM dealer.
This manual may not be reproduced in whole or in part in any form without prior permission. The information contained in this manual may be subject to change without prior notice. Trademarks The company names and product names described in this manual are trademarks or registered trademarks of FUJIFILM Corporation or its subsidiaries. Other holders’ trademarks All other company names and product names described in this manual are trademarks or registered trademarks of their respective owners. Copyright © 2020-2021 FUJIFILM Corporation. All rights reserved.
897N202977B
1
897N202977B.book
2 ページ 2021年2月18日
木曜日
午後6時6分
Introduction
Endoscope Operation Manuals Manage and store “Operation Manual” and “Reprocessing Manual” together as a set.
Endoscope Operation Manual
Model: EC-760S-G/M, EC-760S-G/L This manual provides necessary information for using the endoscope such as the equipment overview, operation procedures and precautions to observe.
Endoscope Reprocessing Manual Model: EC-760S-G/M, EC-760S-G/L This manual describes the reprocessing and storage methods of the endoscope.
Note In this manual, the Endoscope Operation Manual is referred to as “this manual”, and the Endoscope Reprocessing Manual as “the Reprocessing Manual.”
2
897N202977B
897N202977B.book
3 ページ 2021年2月18日
木曜日
午後6時6分
How to Read This Manual Terms Term
Description It refers to disinfection or sterilization performed after the manual cleaning of the endoscope and its accessories according to the Reprocessing Manual supplied with this product.
This product
It refers to the endoscope with or without attached accessories.
Standard accessory
It refers to the parts and devices included in the package or supplied with this product.
Accessory
It refers to the parts and devices directly attached to or used with the endoscope.
Related equipment
It refers to the devices directly or indirectly connected to or used with this product during a procedure.
Consumable item
Introduction
Reprocessing
It refers to parts and products whose life expectancy is limited and which require replacement once they show signs of wear or irregularity. Such parts and products cannot be repaired or refurbished and should be replaced after any irregularity (described in “Section 4.4 Inspecting and Attaching Accessories”) is observed. Note Some of the parts and products may be able to be repaired. For details on them, contact your local FUJIFILM dealer.
Conventions Used in This Manual This manual uses the following conventions for easier understanding. Convention
WARNING
Explains dangerous situations that may cause death or serious injury if not avoided.
CAUTION
Explains situations that may cause injury if not avoided. Explains situations that may cause damage to equipment if not avoided.
(1), (2), (3), ...
897N202977B
Description
Indicates consecutive numbers in operating procedures for the order in which successive steps in the procedure should be taken.
Note
Indicates a comment or supplementary information.
Indicates a reference.
3
897N202977B.book
1 ページ 2021年2月18日
Chapter
木曜日
午後6時6分
1 Precautions Before using this product, read this chapter carefully so that you can operate it correctly. This chapter describes the warnings and cautions for safe operation of the endoscope.
1.1
Intended Use Endoscope Models EC-760S-G/M and EC-760S-G/L are intended for the observation, diagnosis and endoscopic treatment of the lower digestive tract, specifically the rectum, sigmoid colon, large intestine and ileocecal region, at medical facilities under the management of physicians. Never use this product for any other purposes.
Never use this product for any other purposes. It may cause severe harm to patient and/or end-users.
1.2
Incident Reporting Regulation (EU) 2017/745 requires that any serious incident that has occurred in relation to the device should be reported to the manufacturer or European Authorized Representative, and the competent authority of your country. In case you become aware of any serious incident, please report immediately to your local FUJIFILM dealer and the competent authority of your country.
1.3
Applicability of Endoscopy and Endoscopic Treatment Proper clinical judgment should be exercised for the applicability of endoscopy and endoscopic treatment. If there are official standards on the applicability of endoscopy and endoscopic treatment that are defined by the hospital’s administrators or other official institutions, such as academic societies on endoscopy, follow those standards. Perform endoscopy and endoscopic treatment only when the benefits outweigh the risks.
897N202977B
1-1
Chapter 1 Precautions
WARNING
897N202977B.book
2 ページ 2021年2月18日
木曜日
午後6時6分
Chapter 1 Precautions
1.4
User Qualifications
WARNING The healthcare facilities owning this product are responsible for the use and maintenance of this product. If this product is not used or maintained properly, it may cause severe harm to patient or end-users. This product is intended for use by medical professionals who have received proper training in endoscopic procedures. This manual does not provide information about procedures or any aspects of endoscopic techniques. Not following the recommendations may cause severe harm to patient or end-users. If there are official standards for user qualifications for performing endoscopy and endoscopic treatment that are defined by the hospital’s medical administrators or other official institutions such as academic societies on endoscopy, follow those standards. The physician should be capable of safely performing the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment.
1.5
Prohibition of Modification and Improper Repair Do not modify this product or its components, and do not disassemble, repair or in any other way reverse-engineer these products. Even if you find a defect, do not attempt to repair these products yourself. FUJIFILM Corporation shall not be liable for any defects, adverse events or device failures caused by such modifications, disassembly, repairs or reverse-engineering. FUJIFILM Corporation shall not be liable for malfunctions, adverse events or damages caused by remodeling, maintenance, and repair using repair parts other than those authorized by FUJIFILM Corporation. FUJIFILM Corporation shall not be liable for malfunctions, adverse events or damages caused by installation, relocation, remodeling, maintenance, and repair not performed by FUJIFILM Corporation or by dealers authorized by FUJIFILM Corporation.
WARNING Do not disassemble or modify this product. Do not perform unauthorized repairs. If any disassembly, modification or improper repair is performed, it may cause severe harm to patient or end-users.
CAUTION Do not disassemble or modify this product. Do not perform unauthorized repairs. If any disassembly, modification or improper repair is performed, it may cause equipment failure.
1-2
897N202977B
897N202977B.book
3 ページ 2021年2月18日
木曜日
午後6時6分
1.6
Reprocessing Before the First Use/Reprocessing and Storage After Use
1.6.1
Reprocessing Before the First Use/Reprocessing After Use This product has not been reprocessed. The endoscope and its accessories must be reprocessed prior to first use as per instructions provided in the Reprocessing Manual. After using the endoscope and its accessories, reprocess and store them according to the instructions provided in the Reprocessing Manual. Note Delayed reprocessing of endoscopes is not recommended.
WARNING
When using the endoscope for a patient with Creutzfeldt-Jakob disease, use it exclusively for a patient with the same disease, or properly discard the endoscope after use. Since the reprocessing methods described in the Reprocessing Manual cannot eliminate the causal agents of Creutzfeldt-Jakob disease, the product could pose a risk of infection. For details on how to deal with Creutzfeldt-Jakob disease, refer to local guidelines. Immediately upon completion of a procedure, it is imperative that pre-cleaning is performed as per instructions provided in the Reprocessing Manual. Otherwise, residual debris may begin to dry and solidify and hinder effective removal and reprocessing efficacy, which may pose an infection control risk.
CAUTION The entire surface and each channel of the endoscope and the accessories must be reprocessed prior to first use, after any servicing, after any subsequent use, and after storage as per instructions provided in the Reprocessing Manual. In addition, store this product as per instructions provided in the Reprocessing Manual. Inadequate reprocessing or storage may cause equipment damage, or reduce performance. Do not forcibly twist or bend the insertion tube of the endoscope. It could damage the endoscope.
1.6.2
Storage After Use Store this product after reprocessing. For details on the reprocessing and storage of the endoscope, refer to the Reprocessing Manual.
1.6.3
Disposal For details on the disposal of the endoscope and accessories, refer to the Reprocessing Manual.
897N202977B
1-3
Chapter 1 Precautions
The entire surface and each channel of the endoscope and the accessories must be reprocessed prior to first use, after any servicing, after any subsequent use, and after storage as per instructions provided in the Reprocessing Manual, even if the accessories were not used during a procedure. In addition, store this product as per instructions provided in the Reprocessing Manual. Inadequate reprocessing or storage may cause infection.
897N202977B.book
4 ページ 2021年2月18日
木曜日
午後6時6分
Chapter 1 Precautions
1.7
For Safe Operation Be sure to prepare a spare endoscope for unexpected events such as the failure of this product. Otherwise, you may not be able to continue the endoscopic procedure. If the spare endoscope is not available, prepare other alternative means such as abdominal surgery. Note It is highly recommended that spare or back-up equipment be available to complete procedures due to unforeseen circumstances including but not limited to interruption or loss of a clear endoscopic image, equipment failure, etc. If an endoscopic image is lost or compromised during a therapeutic procedure, immediately stop treatment and withdraw the endotherapy device together with the endoscope. Use back-up equipment to complete the procedure as necessary.
1.8
Maintenance
WARNING Deterioration or degradation of the endoscope components or its accessories may occur due to factors such as long-term use, procedures, routine handling and repeated reprocessing. Have this product checked by service personnel once every six months. Use of nonfunctional equipment may cause severe harm to patient or end-users. This product should not be subjected to any type of repair or maintenance procedure while it is being clinically used on a patient (or while it is being reprocessed). The more the product is used, the greater the probability of failure of the endoscope and its accessories. Do not use the endoscope that shows any sign of abnormality or irregularity. Take appropriate measures by following “Chapter 6 Troubleshooting.” If the irregularity is still observed after inspection, contact your local FUJIFILM dealer.
1-4
897N202977B
897N202977B.book
5 ページ 2021年2月18日
1.9
木曜日
午後6時6分
Symbols This section explains the safety signs used in this product. Symbol
Description Do not re-use / Single patient use only Lot number Serial number Year of manufacture
Manufacturer Chapter 1 Precautions
Authorised representative in the European Community Refer to instructions for use (Blue)
Temperature limitation
Keep dry
Non-sterile CE marking
Type BF applied part
WEEE marking *1
Humidity limitation
Atmospheric pressure limitation
Minimum diameter of the instrument channel: 3.8 mm
3.8
897N202977B
1-5
897N202977B.book
6 ページ 2021年2月18日
木曜日
午後6時6分
Chapter 1 Precautions
Symbol
Description Medical device Made in Japan
*1 This product shall not be treated as household waste.
1-6
897N202977B
897N202977B.book
7 ページ 2021年2月18日
木曜日
午後6時6分
1.10 Precautions for Transportation WARNING Carry a reprocessed endoscope in a clean manner. If personal protective equipment such as gloves is contaminated, the contaminants adhere to the endoscope and it can be a source of infection. Contact your local FUJIFILM dealer when the endoscope is returned for repair. Be sure to reprocess the endoscope before returning for repair. A returned product which is not reprocessed can create a risk of infection to users, service personnel or other persons in contact with it.
CAUTION
When transporting a reprocessed endoscope, do not coil the insertion tube or the LG flexible portion of the endoscope with a small diameter. Doing so may cause endoscope failure. When transporting the endoscope to the outside of the hospital, store the endoscope in a FUJIFILM-specified carrying case. Not doing so may cause product failure. When transporting the endoscope to the outside of the hospital, make sure that the insertion portion is set to the most flexible condition before storing the endoscope in a FUJIFILM-specified carrying case. Putting the endoscope in the carrying case while the insertion portion is not set to the most flexible condition could damage the endoscope.
1.11 Precautions Against Electric Shock WARNING Connect the power plug of related equipment to be used to the protective earth receptacle. Not doing so may cause an electric shock.
CAUTION Do not use related equipment which is not described in this manual. If the endoscope is used in combination with endoscopic accessories connected to other medical devices, it may cause an electric shock due to an increase in patient leakage current.
897N202977B
1-7
Chapter 1 Precautions
When transporting a reprocessed endoscope, firmly grasp the control portion and scope connector. If only the LG flexible portion or the boot is grasped, it may damage the endoscope.
897N202977B.book
8 ページ 2021年2月18日
木曜日
午後6時6分
Chapter 1 Precautions
1.12 Electromagnetic Compatibility (EMC) This product generates, uses and can radiate electromagnetic energy. To prevent electromagnetic interference within the vicinity of this product, read the following precautions and properly handle this product and other devices in the vicinity. Install and use this product according to “Electromagnetic Compatibility (EMC) Information” in Appendix.
WARNING Do not place any objects that emit strong electromagnetic waves near this product. Otherwise, malfunction of this product may occur. Do not use this product adjacent to other equipment. If such use is necessary, this product and the other equipment should be observed to verify that they are operating normally. Failure to do so could result in improper operation. Do not use portable and mobile RF communications equipment closer than 30 cm to any part of this product. Otherwise, degradation of the performance of this product could result.
CAUTION Use this product in the specified environment and with specified methods. Failure to do so may result in an abnormality of an endoscopic image (rotation or inversion of the viewing image). Noise may appear on the monitor of this product due to the effect of electromagnetic interference. In this case, turn off the device emitting the electromagnetic waves or move the device away from the monitor. This product may receive electromagnetic interference even if related equipment conforming to CISPR 11/EN 55011 is used. Use of this product may cause electromagnetic interference. Depending upon the strength of electromagnetic interference within the vicinity of this product, malfunction of this product or peripherals may occur. If this product does cause harmful electromagnetic interference to other devices, or if this product receives electromagnetic interference from other devices, we recommend that you may try to correct the interference by one or more of the following measures: Change the orientation or position of any affected device. Increase the spacing between devices. Connect the product into an outlet on a circuit different from that to which the other device(s) are connected. Take mitigation measures such as shielding the installation location of any affected device. If the problem cannot be solved with the above measures, stop using this product and consult the manufacturer or your local FUJIFILM dealer for help.
1-8
897N202977B
897N202977B.book
9 ページ 2021年2月18日
木曜日
午後6時6分
Do not use this product near devices, such as MRI systems, that generate strong electromagnetic waves. Doing so may cause malfunction of this product. (If this product is used in combination with an electrosurgical unit, follow the instructions provided in the operation manuals of the electrosurgical unit and high-frequency endotherapy device.)
Chapter 1 Precautions
897N202977B
1-9
897N202977B.book
10 ページ
2021年2月18日
木曜日
午後6時6分
Chapter 1 Precautions
1.13 General Warnings and Cautions Observe the following cautions when handling this product. Also, there are same cautions in each chapter.
1.13.1 Precautions for Inspection Before Use WARNING Make sure to inspect the endoscope and accessories before use according to the procedures provided in this manual. Do not use the equipment that shows any signs of abnormality or irregularity. Use of abnormal equipment may lead to misdiagnosis, cause severe harm or inflammation to the patient and/or pose an infection control risk to the patient. Make sure to check the forceps valve before use. If the inspection result shows any sign of abnormality or irregularity, replace the forceps valve with a new one already reprocessed. Use of an abnormal forceps valve may cause the leakage of body fluid, posing an infection risk. Turn the up/down and left/right angulation knobs slowly in each direction until they stop. Repeat this operation several times to confirm that the bending section angulates smoothly and correctly. If the endoscope with an abnormal angulation knob is used, the bending section does not return to its neutral position, causing patient injury or making it impossible to withdraw the endoscope from the patient. If water leaks from the forceps valve or suction valve during the inspection of the suction function, replace the forceps valve with a new reprocessed forceps valve or the suction valve with a reprocessed spare suction valve. A leaking forceps valve or suction valve may cause backflow of body fluid, posing an infection control risk. If water leaks from the air/water valve during the inspection of the air/water supply function, replace it with a reprocessed spare one. A leaking air/water valve may cause backflow of body fluid, posing an infection control risk. Turn off the light of the light source before inspecting the objective lens. Viewing the light from the light guide directly may damage your eyes.
CAUTION Make sure to inspect the endoscope and accessories before use according to the procedures provided in this manual. Do not use the equipment that shows any signs of abnormality or irregularity. Use of abnormal equipment may cause equipment malfunction. Do not forcibly turn the angulation knob further after turning the knob until it stops. If the angulation knob is forcibly turned, it may cause malfunction of the endoscope. Make sure that no moisture or foreign matter (such as dust, gauze fibers, metallic fragments) adheres to the scope connector before connecting it to the light source. If the scope connector with moisture or foreign matter (such as dust, gauze fibers, metallic fragments) is connected, it may cause malfunction or failure of the devices.
1-10