GAMBRO
Infocard Two-Compartment Bag
Quick User Guide
2 Pages
Preview
Page 1
How to use the Gambro two-compartment PVC-free bag I. How to open the over wrap Immediately before use remove the over wrap from the bag either by peeling apart the two layers of the over wrap starting at the bottom of the bag (A) or by using the tear notches at the top of the bag (B).
B.
A.
or
II. How to mix the solution
1. Mix the solutions in the two different compartments by holding the small compartment with both hands and squeezing it until an opening is created in the peel seal between the two compartments.
2. Push with both hands on the large compartment until the peel seal between the two compartments is entirely open.
3. Secure complete mixing of the solution by shaking the bag gently. The solution is now ready for use and can be hung on the equipment.
The replacement or dialysate line may be connected to either of the two access ports, according to alternatives 1 and 2:
Alternative 1 – Injection port First remove the snap-off cap. Then introduce the spike through the rubber septum. Verify that the fluid is flowing freely.
Alternative 2 – Luer access
1. Remove the cap and connect the male luer lock on the replacement or dialysate line to the female luer receptor on the bag; tighten.
Gambro Lundia AB PO Box 10101 SE-22010 Lund Sweden Phone + 46 46 16 90 00 www.gambro.com
2. Using both hands, break the coloured frangible pin at its base, and move it back and forth. Do not use a tool.
3. Verify that the pin is completely broken and that the fluid is flowing freely. The pin will remain in the luer port during treatment.
HCEN5334_1 © 2008.09. Gambro Lundia AB
III. How to connect the bag to the line-set