Student Support Material
178 Pages
Preview
Page 1
How to use the
ARTIS PHYSIO PLUS
Supporting material for the Student SP01745 SW 9.05
Trainee: Mentor:
Important Notice:
This material includes information about the use of the Artis Physio Plus Dialysis System and is provided for facilitating purposes only. This material shall be used conjointly with the Operator’s Manual and the Quick Reference Guide (QRG). The information included in this material does not supersede the information included in the Operator’s Manual and in the QRG. This material does not relieve any user of the Artis Physio Plus Dialysis System from his or her duty to carefully read the full text of the Operator’s Manual before using the System. Reviewing this material does not replace the need for the user to receive the necessary training program before using the System. Code 9033347000 |
2
Table of contents Chapter 1: Chapter 2:
Introduction and Preparation Treatment in practice
Chapter 3: Chapter 4: Chapter 5: Chapter 6: Chapter 7: Chapter 8:
Hygienic maintenance Functions in practice Troubleshooting Special Procedures Report and Service Environment HemoControl therapy
Chapter 9: Chapter 10: Chapter 11: Chapter 12:
HDF on-line treatment HD-SN DP treatment The AFBK therapy Communication
Baxter, Artis, ArtiSet, BiCart, BiCart Select, CleanCart, Diascan, HemoControl, Hemoscan, Hospasol, Safebag, SelectBag, SelectBag One, SelectCart, SoftPac, U 9000, ULTRALINE HD and ULTRALINE HDF are trademarks of Baxter International Inc. or its subsidiaries. Code 9033347000 |
3
Note: In this manual, the Artis screen figures may contain minor differences (position of date and time, commercial name). These differences are irrelevant for the correctness of the information provided to the operator.
Code 9033347000 |
2
4
NOTES
3
Artis Physio Plus SW 9.05 Chapter 1: Introduction and Preparation
Code 903334600 May 2019
Artis in Practice • Touch screen • Intuitive user interface • BiCart Select • Self-loading blood tubing systems • EvaClean • One touch priming • Alarm status and support
Code 9033347000 |
6
$
" !
! ""#
4
NOTES
5
Sub-screens Treatment Type Patient Card Edit Distribution Volume Settings Timer Settings
BP Settings A/V Limit Settings Hemoscan Settings Heparin Settings Priming Settings Rinseback Settings
Fluid Settings Diascan Settings Isolated UF Settings HemoControl Settings UltraControl Settings
Special Procedures Disinfection Hygiene BlackBox System Data Service
Code 9033347000 |
8
Code 9033347000 | 9
Code 9033347000 | 10
6
NOTES
7
Code 9033347000 | 11
Code 9033347000 | 12
Code 9033347000 | 13
8
NOTES
9
Code 9033347000 | 14
Code 9033347000 | 15
Code 9033347000 | 16
10
NOTES
11
Operator Message Guidance
Code 9033347000 |
17
Code 9033347000 |
18
BiCart Select System
Concentrates are prepared on-line during treatment from the: • The BiCart cartridge • The SelectCart cartridge • The SelectBag One container • The SelectBag Citrate container
Concentrate Combinations At connection ensure the correct concentrate combination is set (pre-setable) • BiCart Select System - BiCart cartridge, SelectCart cartridge and SelectBag One container - BiCart cartridge, SelectCart cartridge and SelectBag Citrate container
• BiCart cartridge + A-concentrate - SoftPac container - Canister - Central distribution
After Preparation phase has started, it is no longer possible to change Concentrate Combination Code 9033347000 |
12
19
NOTES
13
Compatibility with SoftPac • New compatibility with SoftPac Citrate containers. • New concentrate combinations are available for the preparation of Dialysis fluid in Bicarbonate mode using SoftPac Citrate containers. • A fixed number of the SoftPac Citrate container codes have been added to the preset list. Preset: Concentrates sub-screen
Available SoftPac Citrate C140
C150 / G150
C165/ G165
C1540/ G1540
C1550/G1550
C1565/G1565
C1575/ G1575
C240/G240
C250/ G250
C265/G265
C275/ G275
C2540
C2550
C2565/ G2565
C340/ G340
C350/G350
C365/ G365
C375/ G375
C3540/ G3540
C3550
C3565
C450
C465
C2040/ G2040
C2050/ G2050
C2065/ G2065
C2075/G2075
C3050/ G3050
C3065/G3065 Code 9033347000 |
20
Fluid Circuit | Treatment
•Single pass fluid system •Ultrafilters after water inlet and dialysis fluid mixing •The third Ultrafilter for on-line treatments, previously included in the ArtiSet accessories, is now replaced by an internal Ultrafilter equal to the first two ones
1 2
•Clearance measurement after the dialyzer
3
Code 9033347000 |
21
Code 9033347000 |
22
ArtiSet HD DNL HC Dialyzer Arterial line
Venous Infusion line Priming line Arterial Infusion lines Heparin line
Dialyzer Venous line
No-return valve Venous Injection port
Arterial Chamber Venous Chamber
Venous Connector
Hemoscan cuvette
Arterial Injection port Venous Patient line
14
Arterial Patient line
NOTES
15
Accessories for On-line Treatments
Ultra line Ultra line
Rinseback connector
Rinseback connector
Fluid Line
Fluid Line
ULTRALINE HD
ULTRALINE HDF Code 9033347000 |
23
Code 9033347000 |
24
Code 9033347000 |
25
Bloodlines & Accessories ARTIS PHYSIO PLUS THERAPY HDF/HF-PRE HDF-POST
ArtiSet Artiset Pre-Post
Accessory ULTRALINE HDF Code: 955599
HDF-POST only
ULTRALINE HD Code: 955600
HD-DN with on-line Priming
HD-SN with on-line Priming
Artiset HD SN HC
ULTRALINE HD Code: 955600
ArtiSet | Venous & Arterial Pathways Pre-pump Pre-pump
Post-pump
16
NOTES
17
Recommended Sequence for Dressing & Priming The all-in-one sequence
The 2-step sequence
Student Version Code 9033346100 |
26
Auto-loading Blood Tubing System Follow operator’s message for: • Auto-loading of the ArtiSet blood tubing system • Disposable set up / connections
All disposables are connected in one single series of action Code 9033347000 |
27
Code 9033347000 |
28
EvaClean Priming fluid draining device i.e. no waste bag is needed • Designed to draw fluid through the blood tubing system – where the A/V patient blood lines are connected • Anti-backflow safety system embedded • Automatically disinfected during the regular disinfection processes
18
NOTES
19
Heparin Syringe Installation
Code 9033347000 |
29
Heparin Administration
•The heparin line is automatically primed if the heparin administration function is activated before Auto-Prime starts •If heparin administration function is not activated before AutoPrime starts, the heparin line can be primed manually using the heparin position key •During heparin administration the light indicator on the pump panel is green lit •In the Linear administration mode, Artis performs a few “kicks” to open the one-way-valve •The Initial Dose is administered when blood enters the arterial dialyzer line (30 ml processed)
Code 9033347000 |
30
Heparin Administration 30cc syringe is standard
Administration types available: • Linear: range 1.5 to 10 mL/h • Intermittent: dose range 0.5-12mL • Manual: range 0.5 to 12mL 10 cc syringe kit (field option)
Administration type available: • Linear: range 0.5 to 4 mL/h • Intermittent: dose range 0.5-4mL • Manual: range 0.5 to 4 mL/h
Extra Bolus button becomes available after Start Treatment
Options available: • Initial Dose • Automatic bolus is administered upon the Operator’s confirmation (Extra Bolus)
Code 9033347000 |
20
31