GE Healthcare
Bellows Assembly
Carestation 600 Series Breathing System Cleaning and Sterilization Users Reference Manual
48 Pages
Preview
Page 1
Cleaning, Disinfection and Sterilization User's Reference Manual
User responsibility This product will perform in conformity with the description thereof contained in this User’s Reference Manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instruction provided. This product must be checked periodically. A defective product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, the manufacturer recommends that a telephonic or written request for service advice be made to the nearest manufacturer Customer Service Center. This product or any of its parts should not be repaired other than in accordance with written instructions provided by the manufacturer and by the manufacturer trained personnel. The product must not be altered without the prior written approval of the manufacturer. The user of this product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than the manufacturer.
CAUTION
U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A., check the local laws for any restriction that may apply.
Table of Contents Cleaning, disinfection and sterilization Intended use... 2 ISO 17664 compliance... 3 Cleaning information...3 Cleaning and maintenance schedule... 6 Surface cleaning and disinfection...7 Cleaning instructions... 7 Disinfecting instructions... 7 Compatible chemical agents... 9 Breathing system autoclavable parts... 11 Breathing system removal... 12 Removing the breathing system bag hose...12 Removing absorber canister...12 Removing the breathing system...13 Disassemble the breathing system... 15 Breathing system disassembly...15 Breathing system cleaning... 17 Cleaning cassette kit...17 Automated washing... 20 Manual cleaning... 22 Sterilization... 24 Assemble the breathing system... 26 Breathing system assembly... 26 Install the breathing system... 29 Bellows assembly... 31 Disassemble the bellows assembly... 31 Assemble the bellows assembly... 33 Bellows assembly test... 36 Absorber canister... 38 Filling the reusable absorber canister... 38
2076865-001
i
Circuit O2 cell... 40
Index
ii
2076865-001
Cleaning, disinfection and sterilization Cleaning, disinfection and sterilization
This User’s Reference manual provides information and instructions for the anesthesia breathing system assembly and disassembly, automated and manual cleaning and sterilization, surface cleaning and disinfection. For reliable operation of the anesthesia breathing system, clean and sterilize the system according to the information in this manual. An overview of the reprocessing is described below. Detailed instructions for each step are provided within this manual. 1. 2. 3.
Remove the canister and breathing system. Disassemble the breathing system. Wash the identified parts and cleaning cassette in an automated washer. 4. Remove the parts from the automated washer and inspect parts for damage. 5. Attach the cleaning cassette to the breathing system. Wash the breathing system using the cleaning cassette in the automated washer. 6. Remove the parts from the automated washer and remove the cleaning cassette from the breathing system. Inspect parts for damage. 7. Manually clean parts that are not allowed in automated washer and inspect parts for damage. 8. Remove the bellows base and wrap the parts prior to sterilization. 9. Perform the sterilization. 10. Inspect parts for damage before reassembly. 11. Reassemble the bellows and breathing system parts, install the breathing system, and the canister. 12. The cleaning and sterilization is complete.
2076865-001
1
Intended use This manual contains instructions for cleaning, disinfection and sterilization of the applicable parts of the anesthesia system. Follow these instructions for cleaning, disinfection and sterilization of the system parts. Important
2
Use this manual along with the anesthesia system User’s Reference manual. Always perform a preoperative checkout before using the anesthesia machine after reassembly. The anesthesia system User’s Reference manual contains a preoperative checkout list and preoperative test instructions.
2076865-001
ISO 17664 compliance The process for cleaning, disinfection and sterilization of the applicable parts of the anesthesia system has been validated and found to be in compliance with ISO 17664. These instructions have been validated by the manufacturer of the medical device as being capable of preparing a medical device for reuse. It remains the responsibility of the processor to make sure that the reprocessing is actually performed using equipment, materials, and personnel in the reprocessing facility to achieve the desired result. This requires validation and routine monitoring of the process. Any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequences. The manufacturer recommends the use of filter(s) that protect both the inspiratory and expiratory ports to reduce the risk of crosscontamination. See "Breathing system removal", "Breathing system disassembly", and "Sterilization".
Cleaning information Clean and sterilize the breathing system prior to first use after installation. See "Compatible chemical agents" for a list of the approved cleaning agents. Immediately clean potentially infectious material, for example blood or secretions, from the system by wiping the contaminated area using a disposable cloth and an approved cleaning and disinfectant agent. Parts marked 134°C are autoclavable. All parts in the breathing system, except the circuit O2 cell and flow sensors, can be washed in an automated washer using mild detergent. Only surface clean the circuit O2 cell. Do not autoclave the circuit O2 cell. Do not wash the circuit O2 cell or flow sensors in an automated washer. A Cleaning Cassette kit is available to assist in the cleaning of the breathing system when using an automatic washer. Always perform a preoperative test after reassembly and prior to use. The anesthesia system User’s Reference manual contains a preoperative checkout list and preoperative test instructions. Inspect and replace any damaged parts in the breathing system, if the test failed.
WARNING
2076865-001
Obey applicable safety precautions:
3
Read and understand the material safety data sheet for each cleaning agent.
•
Read and understand the manual for all sterilization equipment to be used.
•
Follow facility guidelines for use of personal protective equipment during cleaning and sterilization procedures.
•
A damaged O2 cell can leak and cause burns. Do not breathe the fumes.
•
Failure to properly install and use bacterial or viral filters in a way that protects the inspiratory and expiratory ports will result in additional required cleaning and sterilization procedures. If bacterial or viral filters are not used, perform the cleaning and sterilization procedures specified in this guide before each patient to avoid crosscontamination between patients.
•
The breathing system should be reprocessed regularly. Contamination is possible. Follow facility guidelines to determine requirements for reprocessing that addresses potential contamination, including an adequate minimum interval between reprocessing which addresses all potential sources of contamination.
•
To help prevent fires: •
All covers used on the system must be made from antistatic (conductive) materials. Static electricity can cause fires.
•
Desiccated (dehydrated) absorbent material may produce dangerous chemical reactions when exposed to inhalation anesthetics. Adequate precautions should be taken to ensure that absorbent does not dry out. Turn off all gases when finished using the system.
•
Obey infection control and safety procedures. Used equipment may contain blood and body fluids.
•
Moveable parts and removable components may present a pinch or a crush hazard. Use care when moving or replacing system parts and components.
CAUTION
4
•
To help prevent damage:
2076865-001
• • •
• • • • • •
2076865-001
Refer to the manufacturer’s data if you have questions about a cleaning agent. Follow facility procedures for cleaning areas of the machine not specified in this section. Do not wipe the surfaces or clean the system using organic, halogenated, or petroleum-based solvents, anesthetic agents, glass cleaners, acetone, or other harsh cleaning agents. Doing so can damage the labels or the system. Do not use abrasive cleaning agents (such as steel wool, silver polish or cleaner). Keep all liquids away from electronic parts. Do not permit liquid to go into the equipment housings. Do not soak parts for more than 15 minutes. Swelling or accelerated aging can occur. Only autoclave parts marked 134°C. Do not exceed 134°C. Cleaning solutions must have a pH of 7.0 to 10.5. Organic, halogenated or petroleum-based solvents, anesthetic agents, glass cleaners, acetone, and other harsh cleaning agents should not be used.
5
Cleaning and maintenance schedule The components process frequency is event-dependent. The following table lists recommended minimum frequency of maintenance. Refer to the anesthesia system’s User’s Reference manual for complete maintenance instructions regarding the anesthesia system. Minimum Frequency During cleaning and setup As necessary
•
Inspect the parts for damage. Replace, repair or reprocess as necessary.
•
Wipe the anesthesia system’s external surfaces. Empty the absorber canister body and replace the absorbent in the canister. Empty the overflow trap on the optional suction regulator. Replace the circuit O2 cell. (Under typical use the cell meets specifications for 2 years). Replace the autoclavable flow sensors. (Under typical use the sensor meets specifications for a minimum of 1 year.)
• • • •
6
2076865-001
Surface cleaning and disinfection External parts and exposed surfaces that are touched while using this device must be cleaned and disinfected when soiled. If necessary, clean the surface parts (especially the display) while the device is in use. Press the Lock Screen button to lock the touchscreen before cleaning. Follow healthcare facility guidelines for cleaning and disinfecting intervals.
CAUTION
Do not use a dripping cloth on surfaces. Make sure there is no liquid pooling on surfaces or entering the device. Damage to the device could occur. Replace the cloth as needed.
WARNING
Make sure that cleaning and disinfecting solutions do not penetrate the inside of the device. Electrical shock could occur.
Cleaning instructions Cleaning solution Bleach solution (1:10 diluted 5.25% sodium hypochlorite solution) 70% Isopropyl Alcohol
•
• •
Thoroughly clean the surfaces of the device with a lint-free cloth pre-soaked with the cleaning solution. Wring excess liquid from the cloth before wiping the surfaces. Clean the surfaces until soil and organic matter is visibly removed. Pay attention to connected surfaces, recessed areas, ridges, and other difficult to clean areas of the device. Inspect the device to confirm the complete removal of soil from surfaces. If soil is still present, clean the device again.
Disinfecting instructions Disinfecting solution
Disinfection time
Bleach solution (1:10 diluted 5.25% sodium hypochlorite solution)
Keep the surface wet for 5 minutes.
70% Isopropyl Alcohol
Keep the surface wet for 5 minutes.
• •
2076865-001
Always clean the surfaces of the device before disinfecting. Wet the surfaces by wiping them with lint-free cloth pre-soaked with disinfecting solution.
7
• •
8
Wiped surfaces must remain wet with the disinfecting solution for the stated disinfection time. If the surface dries before the disinfection time duration, wipe the surface again with another new lint-free cloth pre-soaked with disinfecting solution.
2076865-001
Compatible chemical agents The compatible chemical agents listed have been tested and found not to cause harm to the anesthesia system parts. Read the material safety data sheet (MSDS) for each chemical agent. Use of these chemicals will not damage the components when used as suggested, but efficacy of cleaning or disinfection has not been demonstrated in all cases. The compatible chemical agents listed may not be available or approved for use in all countries. Follow hospital guidelines for cleaning and chemical agent use.
CAUTION
Important Note
Cleaning solutions not shown in the chemical agents list must have a pH of 7.0 to 10.5. Organic, halogenated or petroleum-based solvents, anesthetic agents, glass cleaners, acetone, and other harsh chemical agents are not recommended. Use of any chemical agents other than listed in this table may degrade system performance. The use of chlorinated cleaning and disinfecting agents on metal parts can over time lead to small cosmetic surface oxidation spots on the metal parts. .
Surface cleaning and disinfection
Chemical agent
Concentration
70% Ethanol (Medicinal Ethanol, NOT denatured)
Refer to manufacturer labeling
70% Isopropyl Alcohol
Refer to manufacturer labeling
mikrozid® universal wipespremium
Refer to manufacturer labeling
Bacillol® 30 Sensitive Tissues
Refer to manufacturer labeling
terralin® liquid
Refer to manufacturer labeling
Incidin® OxyWipe
Refer to manufacturer labeling
CLOROX HEALTHCARE®
Refer to manufacturer labeling
Hydrogen Peroxide Cleaner Disinfectant Wipes Oxivir® TB Wipes
Refer to manufacturer labeling
Super Sani-Cloth® GERMICIDAL
Refer to manufacturer labeling
DISPOSABLE WIPE
2076865-001
Premier detergent & disinfectant wipes - Reynard
Refer to manufacturer labeling
Virex® II 256
Refer to manufacturer dilution recommendations
Clinell®UNIVERSAL WIPES
Refer to manufacturer labeling
Clinell® Sporicidal
Refer to manufacturer labeling
9
Chemical agent
Concentration
mikrozid® PAA wipes
Refer to manufacturer labeling
Rely+On™ Virkon™
Refer to manufacturer labeling
Clorox Bleach
13.0 oz to 1 gal of water
Automated washing
Prolystica Ultra concentrate Neutral Dispensed at 0.8 ml per liter of water Detergent, Steris
Manual cleaning
Prolystica 2x Neutral Detergent Neutral, Steris
10
4 ml per liter of water
2076865-001
Breathing system autoclavable parts
1
2
4
3
5
6
1. 2. 3. 4. 5. 6. 7.
AC.24.172
7
Bellows housing and bellows with rim and pressure-relief valve Bellows base Bottom breathing system assembly APL diaphragm assembly Flow sensors Patient port assembly Reusable absorber canister
Figure 1 • Autoclavable assemblies
2076865-001
11
Breathing system removal The breathing system patient circuit, bag hose, and canister must be removed before removing and disassembling the breathing system.
Removing the breathing system bag hose 1.
Disconnect the bag hose from the bag hose port.
1
AC.24.328
2
1. 2.
Bag hose port Bag hose connection
2. 3.
Remove the bag from the holder at the end of the bag arm. Remove the bag hose from the clip and the bag arm base. 2
3
AC.24.325
1
1. 2. 3.
Holder Clip Bag arm base
Removing absorber canister 1.
12
Pull down the canister lifter handle to unlock the canister from the breathing system.
2076865-001
AC.24.071
Lift the canister and remove it from the canister lifter base.
3.
See "Absorber canister" for cleaning and filling instructions.
AC.24.072
2.
Removing the breathing system 1.
2076865-001
Open the breathing system door. Push the release button to unlock the breathing system.
13
AC.24.154 AC.24.155
Grasp the breathing system base, rotate down and pull the assembly away from the anesthesia machine.
AC.24.156
2.
14
2076865-001
Disassemble the breathing system The breathing system assembly can be disassembled for cleaning, sterilization, and part replacement.
Breathing system disassembly Push the patient port assembly tab and pull the assembly from the breathing system.
2.
Pull out both of the flow sensors.
AC.24.158
AC.24.157
1.
2076865-001
15
Remove the APL valve diaphragm assembly.
AC.24.159
3.
16
2076865-001