GE Healthcare
CARESCAPE series Modules and Frames
CARESCAPE Monitor Addendum for E-LoFlo CO2 measurement Users Documentation 2nd edition April 2018
Addendum
32 Pages
Preview
Page 1
CO₂ measurement with E-LoFlo Addendum for CARESCAPE monitors' user documentation
CO₂ measurement with E-LoFlo English 2nd edition 2104101-001 © 2018 General Electric Company. All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice. For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies, Inc. and GE Healthcare Finland Oy.
2
CO₂ measurement with E-LoFlo
2104101-001 2018-04-09
Contents About this addendum ... 7 Intended use of this addendum ... 7 Intended audience of this addendum... 7 Symbols glossary ... 7 Training requirements... 7 Manufacturer responsibility... 7 Trademarks... 7 Naming conventions... 8 Safety precautions ... 8 Safety message signal words... 8 Indications for use caution ... 8 Airway gases warnings... 8 Airway gases cautions ... 9 LoFlo Sidestream Module warnings ... 9 LoFlo Sidestream Module cautions ... 11 Module overview ... 11 E-LoFlo indications for use... 11 CE marking application year ... 11 LoFlo Sidestream Module indications for use... 11 E-LoFlo compatibility... 11 E-LoFlo parameters... 12 E-LoFlo module connector... 12 Exhaust port of the LoFlo Sidestream Module ... 12 E-LoFlo and LoFlo Sidestream Module points to note ... 12 Measurement setup for E-LoFlo with LoFlo Sidestream Module... 13 E-LoFlo and LoFlo Sidestream Module, equipment connections ... 13 Connecting the E-LoFlo to the monitor... 14 Pump on and off functionality... 14 Setting up the measurement... 14 About sidestream cannulas... 14 About single-use sidestream on-airway adapters... 15
2104101-001
CO₂ measurement with E-LoFlo
3
Using Respironics single-use on-airway adapters... 15 Checking the measurement ... 16 Using CO₂ measurement ... 16 Selecting the CO2 scale ... 16 Selecting the CO2 sweep speed... 16 Selecting CO2 average ... 16 Setting CO2 limit alarms ... 16 Selecting the O2 level ... 17 Selecting the N2O level... 17 About zeroing LoFlo Sidestream Module ... 17 Zeroing the LoFlo Sidestream Module ... 17 Troubleshooting ... 18 Messages related to CO2 measurement with E-LoFlo ... 18 Cleaning and care... 20 LoFlo Sidestream Module cleaning-related warnings ... 20 Cleaning instructions for LoFlo Sidestream Module... 21 Cleaning agents for LoFlo Sidestream Module... 21 Cleaning instructions for LoFlo Sidestream Module accessories... 21 Permitted detergents for E-LoFlo ... 21 Maintenance ... 22 Before any planned maintenance or regular checks ... 22 Planned maintenance for E-LoFlo... 22 Physical specifications... 22 E-LoFlo module specifications... 22 LoFlo Sidestream Module specifications ... 22 E-LoFlo alarm specifications ... 23 Alarm priorities and escalation times for CO2 ... 23 E-LoFlo default settings ... 25 Parameter specifications... 26 Airway gases standards compliance... 26 E-LoFlo with LoFlo Sidestream Module, performance specifications... 26 Supplies and accessories ... 28 Ordering information ... 29
4
CO₂ measurement with E-LoFlo
2104101-001
Supplies and accessories list ... 29
2104101-001
CO₂ measurement with E-LoFlo
5
6
CO₂ measurement with E-LoFlo
2104101-001
About this addendum Intended use of this addendum This addendum is to be used with the user documentation of the CARESCAPE monitor. Pay special attention to all generic safety statements and safety symbols listed in the monitor’s user documentation. This addendum is to be used with the following products (brand and model): ●
E-LoFlo MJT101
The model number is given on the device plate of the product. In addition, the measurement requires the use of Respironics LoFlo Sidestream CO2 Module.
Intended audience of this addendum This manual is intended for clinical professionals. Clinical professionals are expected to have a working knowledge of medical procedures, practices and terminology required to provide patient care. Using the device should never replace nor impede the human intervention and required patient care provided by clinical professionals.
Symbols glossary USA only: For a list of all product-related symbols with their standard references, see the symbols glossary in the host monitor’s user manual.
Training requirements No product-specific training is required for the use of this module.
Manufacturer responsibility GE is responsible for the effects on safety, reliability, and performance of the equipment only if: ●
Assembly operations, extensions, readjustments, modifications, servicing, or repairs are carried out by authorized service personnel.
●
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
●
The equipment is used in accordance with the instructions for use.
WARNING
SAFETY HAZARD. To avoid risks to personnel and patient, or damage to the equipment, only perform maintenance procedures described in this manual. Unauthorized modifications can lead to safety hazards.
Trademarks GE, GE Monogram, and CARESCAPE are trademarks of General Electric Company.
Third party trademarks Respironics and LoFlo are trademarks of Koninklijke Philips Electronics N.V.
2104101-001
CO₂ measurement with E-LoFlo
7
Naming conventions In this addendum, the following naming conventions are used: ●
E-LoFlo MJT101 is referred to as E-LoFlo
●
Respironics LoFlo Sidestream CO2 Module is referred to as LoFlo Sidestream Module
●
CARESCAPE monitors are referred to as monitor
Safety precautions Safety message signal words Safety message signal words designate the severity of a potential hazard. DANGER
Indicates a hazardous situation that, if not avoided, will result in death or serious injury.
WARNING
Indicates a hazardous situation that, if not avoided, could result in death or serious injury.
CAUTION
Indicates a hazardous situation that, if not avoided, could result in minor or moderate injury.
NOTICE
Indicates a hazardous situation not related to personal injury that, if not avoided, could result in property damage.
Indications for use caution CAUTION
U.S. Federal law restricts this device to sale by or on the order of a physician.
Airway gases warnings
8
WARNING
Do not use the device during magnetic resonance imaging (MRI) or in an MRI environment.
WARNING
Always inspect the airway adapter for a tight connection and proper operation before attaching it to the patient.
WARNING
Leaks in the gas sampling circuit may cause inaccurate readings.
WARNING
Remove the airway sampling line from the patient’s airway while nebulized medications are being delivered.
WARNING
Strong scavenging suction may cause excessive sample gas flow and inaccurate gas readings.
WARNING
Route all tubing away from the patient’s throat to avoid strangulation.
WARNING
To avoid the spread of infectious disease, do not allow the exhaust to discharge in the direction of the patient or user.
CO₂ measurement with E-LoFlo
2104101-001
WARNING
EtCO2 values may differ from blood gas readings.
WARNING
A failure in zeroing airway gases may cause inaccurate readings
WARNING
Always ensure the correct size and fit of accessories according to patient type and application, especially when monitoring pediatric and neonatal patients. The size and fit of accessories may impact the measured gas concentration values at low tidal volumes. It is recommended to have the gas sampling port close to the proximal end of the endotracheal tube. Excessive dead space in the circuit, including the accessories, may cause re-breathing of gases. Very low accessory dead space between the breathing circuit Y-piece and the gas sampling site may impact the measured gas concentration due to dilution of the sampled exhaled gas with fresh gas from the ventilator. To confirm accurate correlation with measured gases and blood, check arterial blood gas values to confirm a suitable setup is used.
Airway gases cautions CAUTION
Do not apply pressurized air or gas to any outlet or tubing connected to the monitor. Pressure may destroy sensitive elements
LoFlo Sidestream Module warnings The following warnings apply to the LoFlo Sidestream Module.
2104101-001
WARNING
EXPLOSION HAZARD. Do not use in the presence of flammable anesthetics or other flammable gasses. Use of the equipment in such environment may present an explosion hazard.
WARNING
ELECTRIC SHOCK. Always disconnect the module before cleaning. Do not use if it appears to have been damaged. Refer servicing to qualified service personnel.
WARNING
Device is to be used by licensed practitioner, or other qualified medical personnel properly trained in its use.
WARNING
Follow precautions for electrostatic discharge (ESD) and electromagnetic interference (EMI) to and from other equipment.
WARNING
In the presence of electromagnetic devices (like electrocautery), patient monitoring may be interrupted due to electromagnetic interference. Electromagnetic fields up to 20 V/m will not adversely affect LoFlo Sidestream Module performance.
WARNING
EQUIPMENT FAILURE. If the module fails to respond as described in this manual, do not use it until approved for use by qualified service personnel.
CO₂ measurement with E-LoFlo
9
10
WARNING
Do not position the module cables or tubing in any manner that may cause entanglement or strangulation. Support the Respironics LoFlo on-airway adapter to prevent stress on the ET tube.
WARNING
Reuse, disassembly, cleaning, disinfecting or sterilizing the single use cannulas and on-airway adapters may compromise functionality and system performance leading to a user or patient hazard. Performance is not guaranteed if an item labeled as single use is reused.
WARNING
Inspect the sidestream on-airway adapters and the sidestream nasal and oral or nasal sampling cannulas prior to use. Do not use if they appear to be damaged or broken.
WARNING
Replace the sidestream on-airway adapters and the oral or nasal sidestream sampling cannulas if you notice excessive secretions.
WARNING
Monitor the capnogram (CO2 waveform). If you see changes or abnormal appearance, check the patient and the sampling line. Replace the line if needed.
WARNING
Nitrous oxide, elevated levels of oxygen, Helium, Xenon, halogenated hydrocarbons, and anesthetic gas can influence the CO2 measurement with LoFlo Sidestream Module.
WARNING
Do not use the module if it is wet or has exterior condensation.
WARNING
Do not apply extensive tension to any cable.
WARNING
ELECTRIC SHOCK. No user serviceable parts inside.
WARNING
Do not use the device on patients that cannot tolerate the withdrawal of 50 ml/min ±10 ml/min from the airway or patients that cannot tolerate the added dead space to the airway.
WARNING
MEASUREMENT DRIFT. Always ensure that the LoFlo Sidestream Module is used in a ventilated area. Using it in a confined space like a pouch may result in CO2 entering the module through the inlet. This may lead to a measurement drift
WARNING
Do not connect the exhaust tube to the ventilator circuit. There is a risk of patient cross-infection if the sampled gas is returned to the breathing system.
WARNING
Before use, carefully read all instructions for use.
WARNING
INACCURATE READINGS. If the flow rate falls outside the normal operating range and remains there for 15 seconds or more, the Check Sample Line message appears and the measurements may be inaccurate. To avoid the risk of incorrect readings, ensure that the operating conditions meet the given requirements.
CO₂ measurement with E-LoFlo
2104101-001
LoFlo Sidestream Module cautions The following cautions apply to the LoFlo Sidestream Module and accessories. CAUTION
U.S. Federal law restricts this device to sale by or on the order of a physician.
CAUTION
Use only Respironics Novametrix or GE approved accessories.
CAUTION
Do not sterilize or immerse the module in liquids.
CAUTION
Do not clean the LoFlo Sidestream Module and accessories except as directed in this addendum.
CAUTION
Do not store the LoFlo Sidestream Module at temperatures less than –40 °C (40 °F) or greater than 70 °C (158 °F).
CAUTION
Do not use the LoFlo Sidestream Module at temperatures less than 0 °C (32 °F) or greater than 40 °C (104 °F).
CAUTION
Remove the LoFlo sampling line from the LoFlo Sidestream Module’s inlet port when not in use.
CAUTION
Do not stick appendage into sample receptacle.
CAUTION
Always insert sample cell before inserting the on-airway adapter in the ventilated circuit.
CAUTION
Always remove the on-airway adapter from the ventilated circuit before removing the sample cell.
Module overview E-LoFlo indications for use The E-LoFlo is indicated for use with a host device for monitoring respiratory parameters (CO2 and respiratory rate) of adult, pediatric and neonatal hospitalized patients. These CO2 parameter values and respiratory rate values are provided by Respironics LoFlo Sidestream CO2 Module. The E-LoFlo module is intended for use by qualified medical personnel only.
CE marking application year CE marking application year: 2018
LoFlo Sidestream Module indications for use The intended use of the LoFlo Sidestream Module module is to provide carbon dioxide monitoring to a host monitoring system in the intensive care unit (ICU), and in Emergency Medicine/Transport or Respiratory care.
E-LoFlo compatibility E-LoFlo is compatible with the following host devices:
2104101-001
CO₂ measurement with E-LoFlo
11
●
CARESCAPE Monitor B850 software version 2.4
●
CARESCAPE Monitor B650 software version 2.4
●
CARESCAPE Monitor B450 software version 2.4
E-LoFlo cannot be used in the same system in conjunction with CARESCAPE respiratory modules or E-miniC.
E-LoFlo parameters E-LoFlo is used for monitoring CO2 and respiration rate using sidestream sampling method.
E-LoFlo module connector
There is only one connector on the E-LoFlo. It is for connecting the LoFlo Sidestream Module.
Exhaust port of the LoFlo Sidestream Module There is an exhaust port on the rear of the LoFlo Sidestream Module, containing two different-sized barbs for attaching the scavenging tubing. The barb sizes are: ●
Inner diameter 1.524 mm (0.060 in), outer diameter max. 2.54 mm (0.100 in)
●
Inner diameter 3.175 mm (0.125 in), outer diameter max. 9.525 mm (0.375 in)
E-LoFlo and LoFlo Sidestream Module points to note ●
This equipment is suitable for use in the presence of electrosurgery, as tested according to IEC 60601-2-49 clause 202.6.2.101 Electrosurgery interference.
●
For information on materials used in accessories and their biocompatibility, refer to the instructions for use in the accessory package when available.
●
E-LoFlo and LoFlo Sidestream Module cannot be used in the same system simultaneously with CARESCAPE respiratory modules or E-miniC.
●
Always check the tightness of all connections.
●
E-LoFlo module automatically performs barometric pressure compensation to meet the stated accuracy of the LoFlo Sidestream Module.
●
If the message indicating disconnected sampling line keeps appearing even when the sampling line is connected, perform zeroing.
The following points to note apply to LoFlo Sidestream Module:
12
●
LoFlo Sidestream Module and its accessories are latex free.
●
After the life cycle of the LoFlo Sidestream Module and its accessories has been met, disposal should be accomplished following national and/or local requirements.
CO₂ measurement with E-LoFlo
2104101-001
●
Barometric pressure can influence the CO2 measurement with LoFlo Sidestream Module.
●
If nitrous oxide is used, always use the N2O compensation.
Refer to the CARESCAPE monitor’s user documentation for important information about the monitor, including detailed instructions for the clinical use of the product, specifications, and standards.
Measurement setup for E-LoFlo with LoFlo Sidestream Module E-LoFlo and LoFlo Sidestream Module, equipment connections
1.
E-LoFlo module
2.
LoFlo Sidestream Module
3.
Single use accessories (an example)
4.
Mounting bracket for LoFlo Sidestream Module
Always ensure the correct size and fit of accessories according to patient type and application, especially when monitoring pediatric and neonatal patients. The size and fit of accessories may impact the measured gas concentration values at low tidal volumes. It is recommended to have the gas sampling port close to the proximal end of the endotracheal tube. Excessive dead space in the circuit, including the accessories, may cause re-breathing of gases. Very low accessory dead space between the breathing circuit Y-piece and the gas sampling site may impact the measured gas concentration due to dilution of the sampled exhaled gas with fresh gas from the ventilator. To confirm accurate correlation with measured gases and blood, check arterial blood gas values to confirm a suitable setup is used.
2104101-001
CO₂ measurement with E-LoFlo
13
Connecting the E-LoFlo to the monitor 1.
With the module properly oriented (module release latch facing down), align the insertion guide slot in the module with the insertion guide in the module frame.
2.
Push the module into the module frame until clicks.
Pump on and off functionality The pump turns on and off as follows: ●
When you insert a sampling line into the E-LoFlo, the pump turns on automatically. Make sure you hear a click when inserting the sampling line.
●
When you remove the sampling line from the E-LoFlo, the pump turns off immediately.
●
In case of a blocked sampling line, the pump turns off after two minutes.
Setting up the measurement NOTE
Always ensure that all cables are properly connected.
1.
Connect the E-LoFlo to the module frame.
2.
Connect the LoFlo Sidestream Module to the E-LoFlo front panel connector.
3.
Connect the sampling line to the LoFlo Sidestream Module.
4.
Connect the sampling line to the patient’s breathing circuit if you are using an airway adapter accessory, or place it on the patient if you are using a nasal cannula.
5.
Turn on the monitor.
About sidestream cannulas CAUTION
The Nasal and Oral/ Nasal/ Cannulas are intended for single use. Do not reuse or sterilize the cannulas as system performance will be compromised.
Sidestream cannulas are intended for monitoring the CO2 of non-intubated patients.
14
●
Select a cannula that has the best fit and most comfort for the patient. As an example, the cannula labeled as Pediatric, may have the best fit and be more comfortable on a small statured adult than the adult cannula. The cannulas are equally functional regardless of which is used.
●
Verify that the cannula is clean, dry and undamaged. Replace the cannula if necessary.
●
Insert the sample cell into the sample cell receptacle. You will hear a click when the sample cell is properly inserted.
●
Zero the LoFlo Sidestream Module.
●
Place the nasal cannula onto the patient:
CO₂ measurement with E-LoFlo
2104101-001
See accessory-specific instructions for use for more detailed information.
About single-use sidestream on-airway adapters CAUTION
The Adult/Pediatric and the Infant/Neonatal on-airway adapters are intended for single use. Do not reuse or sterilize the adapters as system performance will be compromised.
CAUTION
Always insert the sample cell of the LoFlo sampling line into the sample cell receptable on the LoFlo Sidestream Module before inserting the on-airway adapter into the ventilator circuit. Failure to follow this may introduce a leak in the circuit, thereby reducing set minute volume.
CAUTION
Remove the LoFlo airway adapter and sampling line sample cell from the receptacle when not in use.
NOTE
Do not place the on-airway adapter between the ET tube and the elbow as this may allow patient secretions to accumulate in the adapter. If pooling does occur, the on-airway adapter may be removed from the circuit, rinsed with water (or sterile water) and reinserted into the circuit. If rinsed, make sure that no water has entered the sampling tubing. If so, blow out water or replace the adapter. To prevent moisture from draining into the on-airway adapter tubing, always place the on-airway adapter tubing in an up position:
The Adult/Pediatric sidestream on-airway adapters should be used when monitoring intubated patients with Endotracheal Tube diameters greater than 4.0 mm. Use the Infant/Neonatal on-airway adapters when monitoring intubated patients with Endotracheal Tube diameters less than or equal to 4.0mm.
Using Respironics single-use on-airway adapters
2104101-001
1.
Verify that the on-airway adapter is clean, dry and undamaged. Replace the on-airway adapter if necessary.
2.
Insert the sample cell into the sample cell receptacle. Make sure you hear a click.
3.
Perform a sample cell zero.
CO₂ measurement with E-LoFlo
15
4.
Place the on-airway adapter at the proximal end of the airway circuit between the elbow and the ventilator circuit, making sure that the sampling line is pointing up. Make sure the sampling line remains pointed up during use to minimize condensed water from entering the sampling line.
5.
Check that connections have been made correctly by verifying the presence of a proper capnogram on the monitor.
Checking the measurement 1.
Check that all connections are tight.
2.
Occlude the sampling line and check that the Sample line blocked message appears within 30 seconds.
Using CO₂ measurement Selecting the CO2 scale If EtCO2 is above 6% (45 mmHg), change the scale for capnogram. 1.
Select a gas related parameter window.
2.
Select the CO2 tab > Setup.
3.
Select an option from the Scale list.
Selecting the CO2 sweep speed This selection affects the waveform. 1.
Select a gas related parameter window.
2.
Select the CO2 tab > Setup.
3.
Select an option from the CO2 Sweep Speed list. The options are 0.625 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, and 50 mm/s.
The smaller the value, the slower the sweep speed.
Selecting CO2 average You can select a time interval for averaging the CO2. 1.
Select a gas related parameter window.
2.
Select the CO2 tab > Setup.
3.
Select an option from the CO2 Average list.
Setting CO2 limit alarms
16
1.
Select a gas related parameter window.
2.
Select the CO2 tab > Alarms.
3.
Set high and/or low limit values for EtCO2, FiCO2 and Respiration Rate: select the parameter and then set the limits.
CO₂ measurement with E-LoFlo
2104101-001
Selecting the O2 level NOTE
Compensation must be selected manually when E-LoFlo is used.
The presence of a large concentration of oxygen causes the CO2 level appear lower than the actual value. Use this option to compensate for the presence of O2. 1.
Select a gas related parameter window.
2.
Select the CO2 tab > Setup.
3.
Select an option from the FiO2 level list.
Selecting the N2O level NOTE
Compensation must be selected manually when E-LoFlo is used.
The presence of N2O causes the CO2 value to appear higher than the actual value. Use this option to compensate for the presence of N2O. 1.
Select a gas related parameter window.
2.
Select the CO2 tab > Setup.
3.
Select an option from the N2O level list.
About zeroing LoFlo Sidestream Module During zeroing the LoFlo Sidestream Module adjusts to the optical characteristics of the sample cell. Always perform zeroing when starting to monitor a new patient. Zeroing is also needed when the message Zero required appears, or if the message Sample line disconnected keeps appearing even when the sampling line is connected. Zeroing cannot be performed with the following conditions: ●
Less than 20 seconds since the last detected breath
●
Unstable temperature
●
No sample cell connected to the LoFlo Sidestream Module.
NOTE
To ensure maximum accuracy, it is recommended to perform zeroing each time a sampling line is changed.
Zeroing the LoFlo Sidestream Module When the message Zero required appears, the LoFlo Sidestream Module needs to be zeroed. Zeroing is also needed if the message Sample line disconnected keeps appearing even when the sampling line is connected. NOTE
2104101-001
For best results, allow five minutes for module warm-up before starting zeroing.
1.
Disconnect the patient from CO2 monitoring.
2.
Connect the LoFlo Sidestream Module and, if necessary, wait until the sensor warm-up message disappears.
3.
Connect a LoFlo sampling accessory to the LoFlo Sidestream Module. Ensure that the accessory is exposed to room air and away from all sources of CO2, including the ventilator, the patient’s breath, and your own.
CO₂ measurement with E-LoFlo
17
4.
Select the CO2 parameter window.
5.
Select the Setup tab > Zero.
6.
Select Start Zeroing. Zeroing typically takes 15 to 20 seconds. A status message Zero OK appears when the zero is complete. This message disappears automatically.
7.
Reconnect the patient to resume CO2 monitoring.
Troubleshooting Problem
Solution
CO2 values seem too low
● Check the sampling line and connectors for
leakage. ● Check the sampling line location. ● Check the patient status. ● Check the arterial blood gas values.
CO2 values seem too high
● Check the sampling line for blockage. ● Check the patient status. ● Check the arterial blood gas values.
Module does not work
● Replace the module and have it checked by
qualified service personnel. No CO2 values
● Check all connections. ● Check that CO2 has been selected to the screen.
Messages related to CO2 measurement with E-LoFlo Make sure you are familiar with the generic layout of the screen. This will help you identify where on screen the following messages appear. The message location is indicated with the following abbreviations: ● al. area = alarm area ● param. = parameter window ● wavef. = waveform area
Message
Location
Explanation
What to do
● Apnea (CO2)
● al. area
No breathing detected.
● Check the patient status. ● Check the ventilator and
breathing status. ● Wait. The message
disappears after a breath is detected. ● Apnea deactivated
18
● param.
A new case has just been started/a new patient admitted, or the measurement has just been started and the apnea alarm is not active yet.
CO₂ measurement with E-LoFlo
● Wait. The message
disappears after the monitor detects 3 breaths during the last minute.
2104101-001
Message ●
Check CO2 adapter
● Check adapter
Location
Explanation
What to do
● al. area
Sampling line is disconnected.
● Reconnect the disconnected
sampling line. Make sure that you hear a click when connecting.
● param.
● Perform zeroing. ● Check sample line
● al. area,
param.
Contamination or gas leak in the LoFlo Sidestream Module.
● Reconnect the disconnected
Pressure changes due to strong scavenging suction in the LoFlo Sidestream Module.
● Check that the exhaust port
Out of range pressure changes in the LoFlo Sidestream Module.
sampling line. is not obstructed or blocked. ● Check that the tubing is
not crimped, pinched, or damaged.
The sampling line is disconnected, or of a different type than the last one used. Water, mucus, or other patient contaminate has entered the sample tubing. The sample tubing is crimped or pinched. The exhaust port is obstructed. Sampling line is damaged, cut, or split. ● CO2 pump off
● al. area
● Change sampling line
● wavef.
● CO2 sensor removed
● al. area
● Sensor removed
● param.
● CO2 sensor temp high
● al. area
● Sensor overheating
● param.
LoFlo Sidestream Module’s pump has stopped.
● Change the sampling line to
LoFlo Sidestream Module has been removed.
● Reconnect the sensor or
Sensor temperature is too high.
● Check that the sensor is not
restart the pump. silence the alarm. covered by anything that might lead to overheating. ● Try to cool down the sensor.
● EtCO2 high
● al. area
● EtCO2 low
Measurement values are equal to or outside the alarm limits.
● Check the patient status. ● Adjust alarm limits if
necessary. ● FiCO2 high
● al. area
● FiCO2 low
Measurement values are equal to or outside the alarm limits.
● Check the patient status. ● Adjust alarm limits if
necessary. ● Gas measurements
● al. area
E-LoFlo has been removed.
removed ● Identical gas modules ● Over range
2104101-001
● Reconnect the module to
restart measurement, if necessary. ● al. area ● param.
There are two or more identical modules in the system.
● Remove all but one gas
CO2 value is outside the measurable range and cannot be measured.
● Check the patient.
CO₂ measurement with E-LoFlo
module.
19
Message
Location
Explanation
What to do
● Over scale
● wavef.
Gas signal exceeds the maximum waveform area.
● Check the patient.
● al. area
Measurement values are equal to or outside the alarm limits.
● Check the patient status.
●
RR (CO2) High
● Adjust alarm limits if
necessary. ● Sample line
disconnected
● al. area
Sampling line is disconnected.
● Reconnect the sampling line.
● param.
Sampling line connection is loose.
● Perform zeroing.
● wavef.
Sensor is faulty.
● Contact qualified service
● Line disconnected ● Service CO2
personnel. ● Service CO2 module
● al.area
● Service CO2 sensor
● al.area
● Check sensor
● param.
● Warming up
● param.
Software failure, pressure sensor error, voltage error, or communication error.
● Contact qualified service
Sensor is faulty.
● Contact qualified service
personnel.
personnel. Sensor is warming up.
● Wait until the message
disappears. ● Zero error
● param.
Zeroing has failed.
● Repeat the zeroing.
● Zeroing
● param.
Zeroing is in progress.
● Wait until the zeroing is
completed. ● Zero required
● al. area,
Measurement needs zeroing.
● Perform zeroing.
param.
Cleaning and care LoFlo Sidestream Module cleaning-related warnings The following warnings apply to cleaning the LoFlo Sidestream Module and accessories.
20
WARNING
Unplug the LoFlo Sidestream Module and allow to cool to room temperature before cleaning.
WARNING
Use only approved cleaners and disinfectants.
WARNING
Do not use abrasive cleaners or strong solvents.
WARNING
Small quantities of unapproved cleaners may damage product, shorten lifetime, or cause safety hazards.
WARNING
Keep all cleaning agents away from the LoFlo Sidestream Module’s sample cell receptacle and electrical connections.
WARNING
The manufacturers’ warranty does not cover damage caused by using unapproved cleaners
CO₂ measurement with E-LoFlo
2104101-001