GE Healthcare
Marquette Responder 3000 Defib Ver 2.0 Rev C Operators Manual
Operators Manual
90 Pages
Preview
Page 1
Operator's Manual
Version 2 227 490 02 GA (e)
Revision C
Contents
1
Intended Use and Functional Description
5
2
Controls and Indicators
6
3
Putting the Device into Operation and Performance Check
9
4
Manual Defibrillation
18
4.1 4.2 4.3
18 19 26
Defibrillator Application Guidelines Non-Synchronized Defibrillation Synchronized Defibrillation (Cardioversion)
5
Semiautomatic Defibrillation
30
6
Pacemaker
36
7
Displaying and Monitoring the ECG
41
8
12-Lead ECG Analysis Program (12SL
9
Pulse Oximetry (SpO2)
49
10
Capnometry (etCO2)
54
11
Memories of the Marquette Responder® 3000
58
12
Recording
61
13
Defibrillator Setup
63
14
Battery Power Operation
67
15
Test Discharge
70
16
Operation in the Vehicle Mounting Unit, Mounting the AC Power Adapter
73
17
Error and System Messages
74
18
Cleaning, Maintenance
75
19
Technical Specifications
78
20
Order Information
84
TM
)
46
Appendix The Arrhythmia Detection Program EC Declaration of Conformity Index
87 88 89
Revision History This manual is subject to the Marquette Hellige change order service. The revision code, a letter that follows the document part number, changes with every update of the manual.
227 490 02-C
P/N / Index
Date
Comment
227 490 02-A
January 1999
Initial Release
227 490 02-B
October 1999
Version 2
227 490 02-C
January 2000
ECO 064 064
Marquette Responder® 3000
3
General Information
General Information
The product Marquette Responder 3000 bears the CE marking CE-366 indicating its compliance with the provisions of the Council Directive 93/42/EEC about medical devices and fulfills the essential requirements of Annex 1 of this directive.
The Marquette Hellige quality management system complies with the standards DIN EN ISO 9001 and EN 46001.
The safety information given in this manual is classified as follows:
The radio-interference emitted by this device is within the limits specified in CISPR11/EN 55011, class A.
Warning indicates a hazard. If not avoided, the hazard can
hazard will result in death or serious injury.
result in death or serious injury.
The device is designed to comply with IEC 60601 requirements. It is a protection class I device and has an internal power source. It is classified as an MDD class IIb device.
This manual reflects software version 2.
This manual is an integral part of the device. It should always be kept near the device. Close observance of the information given in the manual is a prerequisite for proper device performance and correct operation and ensures patient and operator safety. Please note that information pertinent to several chapters is given only once. Therefore, carefully read the manual once in its entirety.
4
Danger indicates an imminent hazard. If not avoided, the
The CE mark covers only the accessories listed in the "Order Information" chapter .
On request Marquette Hellige will provide a service manual.
The product complies with the electromagnetic immunity requirements of standard IEC 60601-1-2/EN 60601-1-2 "Electromagnetic Compatibility - Medical Electrical Equipment".
®
The symbol means: Consult accompanying documents. It indicates points which are of particular importance in the operation of the device. This manual is in conformity with the device specifications and standards on safety of electromedical equipment valid at the time of printing. All rights are reserved for devices, circuits, techniques, software programs, and names appearing in this manual.
Caution indicates a potential hazard. If not avoided, this hazard may result in minor personal injury and/or product/property damage.
To ensure patient safety, the specified measuring accuracy, and interference-free operation, we recommend to use only original Marquette Hellige components. The user is responsible for application of accessories from other manufacturers.
The warranty does not cover damage resulting from the use of unsuitable accessories and consumables from other manufacturers.
Marquette Hellige is responsible for the effects on safety, reliability, and performance of the device, only if − assembly operations, extensions, readjustments, modifications, or repairs are carried out by Marquette Hellige or by persons authorized by Marquette Hellige, − the device is used in accordance with the instructions given in this operator's manual.
Marquette Hellige GmbH 2000 Postfach 600265 79032 Freiburg, Germany Telephone +49 761 45 43-0
Marquette Responder® 3000
227 490 02-C
Intended Use and Functional Description
1 Intended Use and Functional Description The Marquette Responder® 3000 is a lightweight, portable defibrillator with ECG monitor and integrated recorder. It is perfectly geared both to hospital and to prehospital use; in conjunction with the vehicle mounting unit, it can also be used in an ambulance.
The color concept for the displayed information lets you see at a glance whether − the parameter reading is within the alarm limits (green), − a technical fault is reported (blue),
There are two versions of the Marquette Responder® 3000: − a version for manual defibrillation, − a version for semiautomatic defibrillation which can be switched to manual operation.
− an alarm is reported (red),
Both versions are capable of delivering synchronized and non-synchronized defibrillation shocks.
− an event memory,
The following paddle types can be used with the defibrillator: hard paddles (with integrated contact surfaces for children), adhesive electrodes and internal spoons. The device features can be upgraded with the following options: − a program for ECG measurement and interpretation (12SL), − an etCO2 measurement system (capnometry), − an SpO2 measurement system (pulse oximetry), − a transcutaneous pacemaker.
− the system displays a message (yellow). The defibrillator has the following memories for storage and documentation of the relevant procedure data: − an ECG memory, − a trend memory, and − a memory for the 12SL analysis results. The integrated 3-channel recorder can be started manually and automatically. The defibrillator is powered from − an optional AC power adapter which is permanently attached to the defibrillator, or − 1 or 2 plug-in batteries, or − the vehicle mounting unit / wall mount unit. Batteries are recharged via: − the optional AC power adapter
1 2
− the optional charging unit, or
3
− the optional vehicle mounting system. 1 Pacer On Off
10 Dem Fix
20 30
7 5 2 Autoseq
Start Pause
P/min
+
Biocompatibility The parts of the product described in this operator
2 Sync
-
manual, including all accessories, that come in
Analyse
mA
+
50 100 150 200 300 360
contact with the patient during the intended use,
-
3
fulfill the biocompatibility requirements of the appliPrint
Event
cable standards. If you have questions in this matter, please contact Marquette Hellige GmbH or its representatives.
Figure 1-1.
227 490 02-C
Marquette Responder® 3000
Marquette Responder® 3000
5
Controls and Indicators
2 Controls and Indicators The Device
1
31 1
30 29 28 27 26 25 24 23 22 21 20 19 18 17
2
6
3
3
4
1 Pacer
10
20 30
7 On Off
5 2 Autoseq
Dem Fix Start Pause
+
50 100 150 200 300 360
2
P/min
Sync
- Analyse
mA
+
-
3
16 Figure 2-1.
2
Event
15
5 6 7 8 9 10 11
14 13 12
Controls and indicators of the Marquette Responder® 3000
Marquette Responder® 3000
227 490 02-C
Controls and Indicators
1
Connector for exchange of the defibrillation electrodes (switch off the device before exchanging the electrodes!)
18 Button to start and stop the recorder 19 Button to change the pacer output (current) 20 Button to change the pacer rate
2
APEX paddle
3
Infrared interface
21 Button to pause the pacer (without changing the settings)
4
Battery with "Test" button and charge level indication
22 Button to select the pacer mode (fixed rate, demand)
5
Button to unlock right battery for removal
23 Button to unlock left battery for removal
6
Energy selector, ON/OFF switch
24 Button to turn the pacemaker on and off
7
Indicator, yellow, flashes to the QRS rhythm in synchronized mode
25 Indicator, yellow: blinks with each delivered pacing pulse
8
Button to enable and disable the synchronized mode (cardioversion)
26 Button to open paper compartment
9
Button to initiate ECG analysis in the semiautomatic mode (only on semiautomatic defibrillator models)
10 Button to initiate defibrillator charging and to trigger the shock (together with button 11) when adhesive pads or internal spoons are used 11 Button to trigger the defibrillation shock (together with button 10) when adhesive pads or internal spoons are used 12 1-Volt ECG output 13 etCO2 signal input (optional) 14 SpO2 signal input (optional) 15 Function keys F1 to F5
27 Battery with "Test" button and charge level indication 28 Indicator, green: is lit when defibrillator is powered from an external source (mains, ambulance) 29 Indicator, yellow blinking: left battery charging on: left battery charged off: left battery missing or partially charged, no external power source connected 30 Indicator, yellow blinking: right battery charging on: right battery charged off: right battery missing or partially charged, no external power source connected 31 STERNUM paddle
16 ECG signal input 17 Event marker button Explanation of symbols used on the device
Battery charging
Consult accompanying documents
Caution, High Voltage
+
Type CF signal input: highly insulated, suitable for intracardiac application, defibrillation-proof Type CF signal input: highly insulated, suitable for intracardiac application
227 490 02-C
Housing without battery (to close the battery slot) Standby mode (line power operation) Audio alarm OFF
Marquette Responder® 3000
7
Defibrillation Electrodes
Defibrillation Electrodes
Hard Paddles
Hard paddle
a
c b
CHARGE SHOCK 3 2
APEX
Figure 2-2.
Hard paddles are the electrodes commonly used for external, transchest defibrillation. There is a special Apex paddle and a Sternum paddle. For delivery of the defibrillation shock, the paddles are placed directly on the body surface. Before use, however, an ample amount of electrode gel must be spread onto the paddles. Both paddles have a shock button: The shock button on the Apex paddle is used to initiate defibrillator charging; afterwards the defibrillation shock is triggered by pushing both shock buttons. The paddles can also be used to acquire the ECG signal. A smaller contact surface for defibrillation of children is integrated in the paddles (see "Defibrillation of Children" in section 4.2). Electrodes for Internal Defibrillation
d
Figure 2-3.
Electrode for internal defibrillation
Electrodes for internal defibrillation consist of a contact spoon (a, Figure 2-3), a handle b, and a counter nut c. The spoon must match the size of the heart and have full contact with the myocardium. There is a choice of 3 different spoon sizes. The electrodes as well as their connection cable must be sterilized before each use. An internal defibrillation is either performed with two spoon electrodes or with one spoon electrode and a so-called "external counter electrode" (d, Figure 2-3) which is placed under the patient and in the immediate vicinity of the heart. Defibrillator charging and release of the defibrillation shock are initiated with buttons on the device. Disposable Adhesive Electrodes
Figure 2-4.
8
Disposable adhesive electrode (external defibrillation, pacing)
Disposable adhesive electrodes are used both for defibrillation and for pacing. These electrodes are self-adhesive and pregelled. They are connected by means of a special cable and may remain attached to the patient for a maximum of 24 hours.
Marquette Responder® 3000
227 490 02-C
Putting the Device Into Operation and Performance Check
3 Putting the Device into Operation and Performance Check Safety information Danger
Warning
Explosion Hazard – The device is not designed for
Shock Hazard - Observe the following warnings.
use in areas of medically used rooms where an
Otherwise the lives of the patient, the user and
explosion hazard may occur. An explosion hazard
bystanders are in danger. The Marquette Responder® 3000 is a highvoltage electrotherapy device and must be handled by qualified and specially trained personnel. Improper use of the device can endanger life. Always follow the instructions given in the operator's manual.
may result from the use of flammable anesthetics, skin cleansing agents and disinfectants. Also, it is not permitted to operate the defibrillator in an oxygen-enriched environment or in the presence of flammable substances (gas) or anesthetics. Oxygenation in the vicinity of the defibrillation
-
When equipped with the AC power adapter, do not use the defibrillator outdoors because the power adapter is not splash-proof.
-
Before using the device, the operator must ascertain that it is in correct working order and operating condition. In particular, all connectors, electrodes as well as sensors and probes must be checked for signs of damage. Damaged parts must be replaced immediately, before use.
-
When disconnecting the device from the power line, remove the plug from the wall outlet first, before disconnecting the cable from the device. Otherwise there is a risk of coming in contact with line voltage by inadvertently introducing metal parts in the socket of the power cord.
-
As a general rule, utmost caution is advised for intracardiac application of medical technical devices. Great care must be exercised to prevent that conductive parts (connectors, electrodes, transducers) connected to the isolated patient signal input come in contact with other grounded conductive parts, as this could bridge the patient's isolation and cancel the protection provided by the isolated input.
electrodes must be strictly avoided. Temporarily interrupt the oxygen supply.
227 490 02-C
Marquette Responder® 3000
9
Putting the Device Into Operation and Performance Check
-
Electrically conductive contact with parts connected to the heart (pressure transducers, metal tube connections and cocks, guide wires, electrode catheters and the metal parts of syringes) must be avoided at all cost. When using devices intracardially, observe these guidelines:
-
always wear isolating rubber gloves;
-
parts with a conductive connection to the heart must be isolated from ground;
-
do not use tube fittings and stopcocks made of metal, if possible;
-
when connecting the heart catheter, observe these guidelines: - the connection must be isolated - all electrodes must be attached to the patient and secured against inadvertent disconnection or they must be isolated and protected against inadvertent contact (otherwise electrodes that become disconnected could bring the patient in contact with ground).
-
When devices are used intracardially, the annual Technical Inspections are mandatory. During intracardiac application of medical electrical devices, a defibrillator and pacemaker, both checked for proper functioning, must be readily available.
-
Ensure that no conductive connection between the patient and bystanders exists during defibrillation.
10
-
The mains plug must be connected to an appropriate power supply with a non-fused earthed wire. If these requirements cannot be guaranteed, connect the device to the ambulance power supply or operate it on battery power.
-
Do not use multiple portable socket outlets (MPSO) to connect the device to the power line.
-
Devices may be connected to other devices or to parts of systems only when it has been made certain that there is no danger to the patient, the operators, or the environment as a result. In those instances where there is any element of doubt concerning the safety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combination of devices. Standards IEC 60601-1-1/EN60601-1-1 must be complied with in all cases.
-
The device (without AC power adapter) is suitable for application in a humid environment provided the regulations concerning drip-proof equipment of IEC 60601/EN 60601 are strictly observed. However, do not defibrillate patients in a very moist or wet environment, unless absolutely necessary. Always dry the defibrillation electrodes and connection cables prior to defibrillation.
Marquette Responder® 3000
227 490 02-C
Putting the Device Into Operation and Performance Check
-
Warning Equipment Failure - Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason make sure that all external devices operated in the vicinity of the defibrillator comply with the relevant EMC requirements. Xray equipment, MRI devices, radio systems, and cellular telephones are a possible source of interference as they may emit higher levels of electromagnetic radiation. Keep the defibrillator away from these devices and verify the defibrillator performance before use.
-
Equipment Failure - Similarly, the defibrillator may disturb equipment operating in its vicinity when charging or delivering the shock. Verify the performance of these devices before use.
-
Suffocation Hazard - Dispose of the packaging material, observing the applicable waste-control regulations. Keep the packaging material out of children's reach.
-
Literature Medical Device Directive of August 2, 1994 EN 60601-1: 1990 + A 1: 1993 + A 2: 1995 Medical electrical equipment. General requirements for safety. EN 60601-1-1: 9/1994 + A1: 12/1995 General requirements for safety. Requirements for the safety of medical electrical systems. IEC-Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
Caution Equipment Damage, Shock Hazard - Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the application site. Wait until all moisture has vaporized before using the device.
-
Equipment Damage - Exercise great care when using HF surgery equipment on the patient at the same as the defibrillator. As a general rule, the distance between the ECG and defibrillation electrodes and the HF surgery electrodes should not be less than 15 cm. If this is not ensured, disconnect the electrodes and transducer leads while using the HF surgery device.
-
Equipment Damage - Avoid defibrillating repeatedly into open air or with the paddles shorted together, because the device temperature may increase to an inadmissible level due to the internal safety discharges.
227 490 02-C
Marquette Responder® 3000
11
Putting the Device Into Operation and Performance Check
Power Supply The defibrillator can be powered − from the power line (requires AC power adapter, P/N 205 108 01, Figure 3-1), − from the ambulance power supply system (requires vehicle mounting unit, P/N 202 317 01), − from the wall mount unit (P/N 202 317 03) − from 1 or 2 rechargeable batteries (mainsindependent). Batteries are recharged by one of the following methods:
Figure 3-1.
Defibrillator with AC power adapter
− in the defibrillator, when the defibrillator is connected to the power line or to the ambulance power supply system, or − by means of the separate charging unit ASU 3000 (P/N 701 279 01).
Note The Marquette Responder® 3000 is switched on and off with the energy selector. Once you are familiar with the device operating routines, this
If you prefer to operate the device mainsindependent, ensure that the batteries are charged (chapter 14 "Battery Power Operation").
control lets you turn on the device and select the required defibrillation energy in one action. No shock can be delivered in the
position of the
energy selector.
Turning the Defibrillator On
1 10 7 5 2 Autoseq
Figure 3-2.
Please refer to chapter 16 for information on operating the device in the vehicle mounting unit and on installing the AC power adapter.
20 30
50 100 150 200 300 360
Turning the defibrillator on
• Connect the device to the power supply. • Switch on the device by turning the energy (defibrillation shocks cannot be selector to delivered in this position). The test screen appears and the device emits a short audio signal. On the test screen you can see the software version and a message referring to the self-test. The three color blocks in red, green and blue are displayed to verify the correct representation of the colors. After the self-test the standard screen appears (Figure 3-3).
12
Marquette Responder® 3000
227 490 02-C
Putting the Device Into Operation and Performance Check
The Standard Screen Display
a b Electrode
c
d e
f
This is the information presented on the standard screen display:
0 Paddle
− windows for heart / pulse rate, SpO2 and etCO2 readings including the limit values a, b
160 / 40
ECG
bpm
SpO2 parameter window
− battery charge level c − ECG lead d
etCO2 parameter window
− window for ECG, SpO2 and etCO2 waveforms
pacemaker window 15.07.1999
09:05:00
semiautom. ECG
SpO2 etCO2
QRSPulse Tone OFF
Next Menu
The color concept for the displayed information lets you see at a glance whether
Standard screen display
b selected limit values
− the parameter reading is within the alarm limits (green),
c battery charge level indication
− a technical fault is reported (blue),
d ECG lead (ECG signal acquired via "Paddle")
− an alarm is reported (red),
a measured heart rate/pulse rate
e date, time f selected operating mode, defibrillation energy g menu
227 490 02-C
− date and time e − menu g.
g Figure 3-3.
− window indicating operating mode and defibrillation energy f
− the system displays a message (yellow). If the device does not receive an ECG signal, the HR window is blue (technical fault) and a sawtooth signal is displayed instead of an ECG. The SpO2 and etCO2 parameter windows are also blank and the corresponding waveforms are missing when the required sensors are not connected.
Marquette Responder® 3000
13
Putting the Device Into Operation and Performance Check
Display Flip
↓ ECG
SpO2 etCO2
Figure 3-4.
QRSPulse Tone OFF
Next Menu
Main menu
↓ Filter ON
Assign Channel Waveform
Figure 3-5.
Display
Previous Menu
• In the main menu, select F5 Next Menu (Figure
Main menu, page 2
↓
↓
Contrast
Contrast
Figure 3-6.
Memory
3-4).
↓ Select Color
The screen display can be rotated 180° to adapt it to the operating position of the defibrillator. The display can be flipped permanently from the setup menu or temporarily as outlined below. You can also set up the system to flip the display automatically when the defibrillator is inserted in the vehicle mounting unit (chapter 13 "Defibrillator Setup").
Display Flip
Previous Menu
You will see page 2 of the main menu (Figure 35).
• Display the Display menu with F4 Display
Display menu
(Figure 3-6).
• To flip the display, press F4 Display Flip .
Note The main menu will automatically reappear, when
Contrast adjustment
no button is activated for a period of 30 seconds.
• Adjust the contrast from the Display menu with F1 and F2. Adjusting Maximum Contrast (Select Color)
• Adjust the maximum contrast from the Display menu with F3.
• Press F5 Previous Menu for about 2 seconds to
Note Press F5 for about 2 seconds to return directly to the main menu.
System Setup
Note Enter your own settings in the column at the far right (with date and signature).
14
return directly to the main menu.
The device has a configuration menu which allows you to customize some of the functions to suit your personal requirements. These settings will be retained. The table at right shows all device settings for which customer defaults can be selected, as well as the factory defaults. The information given in this manual is based on a defibrillator with the factory defaults. In chapter 13 "Defibrillator Setup" you will find instructions on setting up the defibrillator. The same chapter explains how to change the language and how to restore the factory defaults.
Marquette Responder® 3000
227 490 02-C
Putting the Device Into Operation and Performance Check
Parameter
Comment
Factory Defaults
Options
User Setup
HR Limit
40/160
OFF, 15 to 300 increments of 5
SpO2 Limit
---/90
OFF, 60 to 99
etCO2 Limit
---/20
OFF, 5 to 76 increments of 1
ALARM LIMITS
ECG Print on Alarm
autom. recorder start on violation of limit value
off
on/off
Lead Fail Alarm
audio signal indicating disconnected electrode
off
30 s/off
Alarm Tone
audio signal indicating violation of an alarm limit
off
on/off
off
low/middle/high / off
QRS Beep Muscle Filter
suppression of motion artifact
on
on/off
Gain
for ECG display
1 cm/mV
0.5; 1; 2 cm/mV
Lead Channel 1
I
standard leads, paddle acquisition
Lead Channel 2
II
standard leads, SpO2
Lead Channel 3
III
standard leads, SpO2, etCO2 on/off
DEFIB Print on shock
automatic recorder start on shock
on
Operating Mode**
choice of the operating mode
semiautomatic/button semiautom./button semiautom./password semiautom. manual
Autosequence
energy selection
200 J, 200 J, 360 J
150 to 360 J per shock
Pacemaker
default pacer rate
60 P/min
30 to 200 P/min
C-LOCK
C-Lock ECG synchronization
off
on/off
SpO2 Integ. Time
SpO2 integration time
8s
4 s, 8 s, 12 s
DATE/TIME**
Change clears all existing settings. day.mon.year
day.mon.year mon/day/year
SpO2
Date Format** Entry of date and time DEVICE Display
screen display (SmartFlip = display flips normal automatically when defibrillator is placed in vehicle mounting unit)
normal, reverse, SmartFlip
Volume
valid for all audio signals
high
high/low/middle
Cont. Printout
continuous recording or 14-second strip
off
on/off
Analysis
continuous ECG analysis
on
on/off
AC Line Filter**
elimination of AC line interference
50 Hz
off/50 Hz/60 Hz
Language**
selection of the language
Factory Default
restores factory defaults
User**
text or name (20 characters)
PASSWORD**
entry of the password
111
000/999
for config**
protects access to configuration menu
off
on/off
EVENT TEXTS**
entry of event texts
BATTERY
battery maintenance program
OPTIONS
entry of option code to unlock option
on/off
** not affected by reactivation of factory defaults
227 490 02-C
Marquette Responder® 3000
15
Putting the Device Into Operation and Performance Test
Performance Check A performance check must be carried out before each use.
Electrode 160 / 40
ECG
Paddle
The check includes:
bpm
− a visual inspection of the device, the cables and the electrodes for signs of mechanical damage, − verification of the functional readiness of the device, − delivery of a test discharge. 15.07.1999
09:05:00
semiautom. SpO2 etCO2
ECG
Figure 3-7.
QRSPulse Tone OFF
Next Menu
Standard screen display of a defibrillator ready for operation (no ECG, SpO2 and etCO2 signal available)
After power-up and during operation, the Marquette Responder® 3000 runs automatic selftests. If malfunctions are identified, an error message will be displayed (see chapter 17 "Error and System Messages"). In this situation do not put the device into service. In all other cases you will see the standard screen display (Figure 3-7) and the device is ready for use.
Note A special simulator is required to test the defibrillator performance in the semiautomatic mode.
Now verify that the defibrillation shock is correctly delivered by triggering a test discharge (chapter 15 "Test Discharge"). If the energy of the test discharge is not within the specified limits, a defibrillation is possible all the same (it is the user's decision whether or not to employ the defibrillator). However, the device must be immediately checked and repaired by a service technician.
Event Button
62 160 / 40
ECG
Event
I
bpm
II
III
15.07.1998
Text 1
Figure 3-8.
16
You can use the button to mark specific events (e.g. administration of medications). When you press this key, the corresponding point in time is earmarked in the full-disclosure ECG. Furthermore, you can assign a maximum of 8 "event" texts to the function keys F1 to F4 (e.g. names of medications). When you press these texts appear in the menu line (Figure 3-8). You can press one of the function keys to assign the corresponding text to the event. With F5 Next Menu you can display the next line of 4 texts. (Refer to chapter 13 "Device Setup" for instructions on entering event texts.)
Text 2
Event
09:05:00
Text 3
Event texts
semiautom. Next Text 4 Menu
Marquette Responder® 3000
227 490 02-C
Putting the Device Into Operation and Performance Test
How to toggle the defibrillator from semiautomatic to manual operation 2 Sync
Analyse
Depending on their setup, semiautomatic defibrillators can be switched to manual control. The defibrillator can be set up for four different modes of operation: − semiautom./button (switching to manual mode by activating button)
3
− semiautom./password (switching to manual mode by activating button and entering password) − semiautomatic (manual control not possible) − manual (only manual control possible)
Figure 3-9. Electrode
Buttons to activate the manual mode Operating Mode "semiautomatic/button"
0 Paddle
• To activate the manual mode, simultaneously
160 / 40
ECG
bpm
press F5 and
15.07.1998
Analyse
(Figure 3-9).
When switched on again, the defibrillator will reactivate the operating mode selected in the setup menu. Date and time of the change of operating modes is stored in the event memory.
09:05:00
semiautom. 0
0
0
ENTER
Operating Mode "semiautomatic/password"
• To activate the manual mode, simultaneously press F5 and Figure 3-10. Buttons for entry of the password
0
(Figure 3-9).
The screen for entry of the password appears (Figure 3-10).
• Enter the password (3-digit number) with F1,
Paddle
160 / 40
ECG
Analyse
bpm
F2, F3. The factory-set password is 111 (also refer to chapter 13, section "Password").
15.07.1998
09:05:00
manual ECG
etCO2 SpO2
QRSPulse Tone OFF
Next Menu
Figure 3-11. Defibrillator set up for manual operation
227 490 02-C
When switched on again, the defibrillator will reactivate the operating mode selected in the setup menu. Date and time of the change of operating modes is stored in the event memory (chapter 11 "Memories of the Marquette Responder® 3000"). If you wish to return to the semiautomatic mode, you will have to turn the device off and on again.
Marquette Responder® 3000
17
Manual Defibrillation / Non-Synchronized Defibrillation
4 Manual Defibrillation 4.1 Defibrillator Application Guidelines Observe the following guidelines to ensure successful and safe defibrillation. Otherwise the lives of the patient, the user and bystanders are in danger.
Warning
-
-
Defibrillating a patient with normal heart rhythm may induce ventricular fibrillation.
-
Position the patient flat on a hard surface where he is electrically insulated. The patient must not be allowed to come into contact with metal parts, e.g., bed or litter, to prevent unwanted pathways for the defibrillation current which may endanger the assistants. For the same reason, do not position the patient on wet ground (rain, accident in swimming pool). Do not allow the defibrillation electrodes to come into contact with other electrodes or metal parts which are in contact with the patient. The patient's chest must be dry, because moisture can cause unwanted pathways for the defibrillation current. After use of flammable skin cleansing agents, wait until they have completely dried.
-
The operator and all assistants must be briefed regarding the preparations for and execution of defibrillation. All tasks must be clearly assigned. Immediately prior to the shock - interrupt heart massage and artificial respiration, - disconnect tube connections, and - warn bystanders.
-
Ensure that no conductive connection between the patient and bystanders exists during defibrillation.
-
Before delivering the shock, verify that the charged and selected energies are the same.
-
Shock Hazard - Always switch off the device before exchanging the defibrillation electrodes.
18
Pacemaker Patients - Defibrillating a patient with an implanted pacemaker is likely to impair the pacemaker function or cause damage to the pacemaker. For this reason - select the smallest energy level possible for the application, - do not apply the defibrillation paddles in the vicinity of the pacemaker electrodes, - have an external pacemaker at hand, - check the implanted pacemaker for proper functioning as soon as possible after the shock.
Caution -
Equipment Damage - Disconnect transducers and devices that are not defibrillationproof from the patient before delivering the shock.
-
Equipment Damage - Do not defibrillate the patient with a second defibrillator, while defibrillation electrodes (paddles, pads) of the first device are applied. If the use of a second defibrillator is inevitable, disconnect the electrodes from the first device or remove them from the patient.
Marquette Responder® 3000
227 490 02-C
Manual Defibrillation / Non-Synchronized Defibrillation
4.2 Non-Synchronized Defibrillation Using Paddles • Remove the paddles from their compartments as shown in Figure 4-1.
• Carefully dry the paddles and the handles in particular, if they are wet.
• Apply an ample amount of electrode cream to Figure 4-1.
Removing the paddles
a
the paddle surfaces.
b
c
• Set the energy selector to "Autoseq" or to the required energy level.
0 Paddle
160 / 40
ECG
bpm
15.07.1999
ECG
09:05:00
QRSPulse Tone OFF
SpO2 etCO2
Figure 4-2.
200 J
manual Next Menu
Power-up screen (paddles connected) a ECG signal acquired via "Paddle" b manual mode c selected energy
In the "Autoseq" position of the energy selector, the defibrillator automatically sequences the preset defibrillation energy levels. The level for the 3rd shock is maintained for all subsequent defibrillations. When you set the energy selector again to "Autoseq", the automatic charge sequence starts over. The factory set Autosequence energy levels are the values recommended by AHA/ERC for ventricular fibrillation and pulseless tachycardia. 1st shock with 200 J 2nd shock with 200 J 3rd and all subsequent shocks with 360 J.
• Check that the selected energy is displayed (c, Figure 4-2).
• Apply the paddles on the patient's thorax such that the greatest possible amount of energy flows through the myocardium. The imaginary line connecting the paddle centers should be identical with the cardiac median line (Figure 4-3).
• Press the paddles firmly onto the thorax (the M
NU
ER
ST
K OC SH 3
ECG appears on the monitor screen).
• Do not touch the patient any more and warn all those present.
CH
AR 2 GE SH
AP
EX
Figure 4-3.
227 490 02-C
O 3 CK
Paddle application points (dashed application points for pacemaker patients)
Marquette Responder® 3000
19
Manual Defibrillation / Non-synchronized Defibrillation
• Initiate energy storage with the button on the
a
b
APEX paddle (a). When the selected energy is stored,
SHOCK 3 STERNUM
CHARGE SHOCK 2 3
− the device emits an audio signal − the message "Energy available" appears, − the available energy is displayed (if the available energy drops below a given level, the defibrillator recharges automatically).
APEX
• Now trigger the shock within 30 seconds. To do so, simultaneously press the buttons a and b on the paddles. After the shock Figure 4-4.
Buttons to initiate defibrillator charging (a) and to trigger the shock (a+b)
Warning Risk of Skin Burns / Equipment Damage - Do not apply the paddles over -
sternum or clavicle,
-
nipples
-
implanted pacemaker or defibrillator devices.
Note − If you do not trigger the shock within 30 seconds of charging, the energy will automatically be discharged internally. You will then have to recharge the defibrillator. − When the defibrillator is charged, you can increase and decrease the energy level to any value with the energy selector (without pressing the "Charge" button again). For an internal discharge of the stored energy, set the energy selector to or to "Off".
20
− the audio signal stops and the delivered energy is displayed for approx. 6 seconds (in place of the available energy), − the recorder prints a 14-second ECG strip (4 seconds before the shock, 5 seconds blanked, 10 seconds after the shock) (configurable); the blanked period of time is indicated by a vertical line on the recording; − the shock delivery is annotated on the stored ECG (also refer to chapter 11 "Memories of the Marquette Responder® 3000"). If the defibrillator cannot store the selected energy so that selected and stored energy values differ, a warning will be displayed. The defibrillation pulse can be triggered all the same. In this situation we recommend to check the batteries first. If the batteries are intact, have the defibrillator immediately repaired.
Marquette Responder® 3000
227 490 02-C
Manual Defibrillation / Non-Synchronized Defibrillation
Shock Counter
Defibrillation of Children
The number of delivered shocks is indicated below the energy value. This counter is reset to 0 when you set the energy . selector to The number of delivered shocks is also shown on the recording strip (Figure 5-5, intervention report d). This counter, however, counts all shocks since the device was turned on and is reset to 0 only when the device is turned OFF.
Warning Damage to Myocardium - Please note that children require less energy for successful ventricular defibrillation than adults. For the first defibrillation shock delivered to babies and small children, select an energy of approx. 2 J/kg body weight. For subsequent shocks, the energy may be increased to 4 J/kg body weight. Risk of Skin Burns - The full electrode must be in contact with the skin surface (use the small contact surface of the paddles / pads for children).
Ending Therapy
• Once therapy has ended, set the energy selector to
for continued monitoring of the patient.
• If there is no need to monitor the patient, switch off the defibrillator by setting the energy selector to "Off".
• Clean the paddles and the device as outlined in chapter 18 "Cleaning, Maintenance".
The paddles have two different contact surfaces; a large one (can be removed) for the defibrillation of adults and a smaller one for the defibrillation of children. Remove the large contact surface for pediatric use:
• Press on the lock button 1 (Figure 4-5). • Slide the contact surface 2 towards the front and take it off the paddle.
• When re-installing it, the large contact surface must audibly click into place.
1
2
Figure 4-5.
227 490 02-C
Marquette Responder® 3000
Removing the large contact surface from the paddles
21