GE Healthcare
Marquette Responder 3000 Defib Ver2.1 Operators Manual
Operators Manual
98 Pages
Preview
Page 1
Version 2.1
Operator’s Manual 227 490 02 ENG
Revision H
0459
Contents
1 2 3 4
Intended Use and Functional Description Controls and Indicators Putting the Device into Operation and Performance Check Manual Defibrillation
7 8 11 20
4.1 4.2 4.3
20 21 28
Defibrillator Application Guidelines Non-Synchronized Defibrillation Synchronized Defibrillation (Cardioversion)
5 6 7 8 9 10 11 12 13 14 15 16
Semiautomatic Defibrillation 32 Pacemaker 38 Displaying and Monitoring the ECG 43 TM 12-Lead ECG Analysis Program (12SL ) 48 Pulse Oximetry (SpO2) 51 Capnometry (etCO2) 56 Memories of the Marquette Responder® 3000 60 Recording 63 Defibrillator Setup 65 Battery Power Operation 69 Test Discharge 73 Operation in the Vehicle Mounting Unit, Mounting the AC Power Adapter 76 17 Error and System Messages 78 18 Cleaning, Maintenance 79 19 Technical Specifications 84 20 Order Information 90 Appendix 93 The Arrhythmia Detection Program Index
93 94
Revision History This manual is subject to the GE Medical Systems Information Technologies change order service. The revision code, a letter that follows the document part number, changes with every update of the manual.
227 490 02-H
P/N / Revision
Date
Comment
227 490 02-A
1999-01
Initial Release
227 490 02-B
1999-10
Version 2
227 490 02-C
2000-01
ECO 064064
227 490 02-D
2000-08
ECO 065062
227 490 02-E 227 490 02-F
2000-10 2001-01
ECO 065526 ECO 066023
227 490 02-G
2002-11
ECO 070336
227 490 02-H
2004-01
ECO 075566
Marquette Responder® 3000
3
General Information
General Information ●
The product Marquette Responder® 3000 bears the
●
The GE Medical Systems Information Technologies
CE marking “CE-0459” (notified body GMED)
quality management system complies with the
indicating its compliance with the provisions of the
standards DIN EN ISO 9001 and EN 46001.
Council Directive 93/42/EEC about medical de-
●
vices and fulfills the essential requirements of
The safety information given in this manual is classified as follows:
Annex 1 of this directive. ●
Danger indicates an imminent hazard. If not avoided, the
The product complies with the electromagnetic immunity requirements of standard IEC 60601-1-
hazard will result in death or serious injury.
2/EN 60601-1-2 "Electromagnetic Compatibility Medical Electrical Equipment". ●
The radio-interference emitted by this device is
Warning indicates a hazard. If not avoided, the hazard can
within the limits specified in CISPR11/EN 55011, class A. ●
result in death or serious injury.
The device is designed to comply with IEC 60601 requirements. It is a protection class I device and has an internal power source. It is classified as an MDD
● ●
class IIb device.
Caution indicates a potential hazard. If not avoided, this
The device fulfills the requirements of the standard
hazard may result in minor personal injury and/or
UL 2601-1.
product/property damage.
The CE mark covers only the accessories listed in the "Order Information" chapter.
●
To ensure patient safety, the specified measuring accuracy, and interference-free operation, we rec-
●
This manual reflects software version 2.1.
●
This manual is an integral part of the device. It
Information Technologies components. The user is
should always be kept near the device. Close obser-
responsible for application of accessories from
vance of the information given in the manual is a
other manufacturers.
prerequisite for proper device performance and correct operation and ensures patient and operator
ommend using only original GE Medical Systems
●
safety. Please note that information pertinent to several chapters is given only once. Therefore, carefully read the manual once in its entirety. ●
The symbol
means: Consult accompanying
documents. It indicates points which are of particular importance in the operation of the device. ●
This manual is in conformity with the device specifications and standards on safety of electromedical equipment valid at the time of printing. All rights are reserved for devices, circuits, techniques, software programs, and names appearing in
from other manufacturers. ●
GE Medical Systems Information Technologies is responsible for the effects on safety, reliability, and performance of the device, only if − assembly operations, extensions, readjustments, modifications, or repairs are carried out by GE Medical Systems Information Technologies or by persons authorized by GE Medical Systems Information Technologies, − the device is used in accordance with the instructions given in this operator's manual.
this manual. ●
The warranty does not cover damage resulting from the use of unsuitable accessories and consumables
On request GE Medical Systems Information Technologies will provide a service manual.
4
Marquette Responder® 3000
227 490 02-H
General Information
GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223 USA Tel.:+1.414.355.5000 800.558.5120 (USA only) Fax: +1.414.355.3790 The authorized representative for GE Medical Systems Information Technologies Inc. in Europe is: GE Medical Systems Information Technologies GmbH Munzinger Str. 3 D-79111 Freiburg, Germany Tel. +49.761.45.43.0 Fax: +49.761.45.43.233 The country of manufacture appears on the device label. © GE Medical Systems Information Technologies, Inc., 1999 – 2002, 2004. All rights reserved.
227 490 02-H
Marquette Responder® 3000
5
General Information
For your notes
6
Marquette Responder® 3000
227 490 02-H
Intended Use and Functional Description
1 Intended Use and Functional Description The Marquette Responder® 3000 is a lightweight, portable defibrillator with ECG monitor and integrated recorder. It is perfectly geared both to hospital and to prehospital use; in conjunction with the vehicle mounting unit, it can also be used in an ambulance.
The color concept for the displayed information lets you see at a glance whether − the parameter reading is within the alarm limits (green), − a technical fault is reported (blue),
There are two versions of the Marquette Responder® 3000: − a version for manual defibrillation, − a version for semiautomatic defibrillation which can be switched to manual operation.
− an alarm is reported (red),
Both versions are capable of delivering synchronized and non-synchronized defibrillation shocks.
− an event memory,
The following paddle types can be used with the defibrillator: hard paddles (with integrated contact surfaces for children), adhesive electrodes and internal spoons. The device features can be upgraded with the following options: − a program for ECG measurement and interpretation (12SL), − an etCO2 measurement system (capnometry), − an SpO2 measurement system (pulse oximetry), − a transcutaneous pacemaker.
− the system displays a message (yellow). The defibrillator has the following memories for storage and documentation of the relevant procedure data: − an ECG memory, − a trend memory, and − a memory for the 12SL analysis results. The integrated 3-channel recorder can be started manually and automatically. The defibrillator is powered from − an optional AC power adapter which is permanently attached to the defibrillator, or − 1 or 2 plug-in batteries, or − the vehicle mounting unit / wall mount unit. Batteries are recharged via: − the optional AC power adapter
1 2
− the optional charging unit, or
3
− the optional vehicle mounting system. 1 Pacer
10
On Off
Dem Fix
20 30
7 5 2 Autoseq
Start Pause
P/min
+
Biocompatibility The parts of the product described in this operator
2 Sync
-
manual, including all accessories, that come in
Analyse
mA
+
50 100 150 200 300 360
contact with the patient during the intended use,
-
3
fulfill the biocompatibility requirements of the Print
Event
applicable standards. If you have questions in this matter, please contact GE Medical Systems Information Technologies or its representatives.
Figure 1-1.
227 490 02-H
Marquette Responder® 3000
Marquette Responder® 3000
7
Controls and Indicators
2 Controls and Indicators The Device
1
31 30 29
1
28 27
3
26 25 24 23 22 21 20 19 18 17
2
8
3 4
1 Pacer
10
20 30
7 On Off
5 2 Autoseq
Dem Fix Start Pause
+
50 100 150 200 300 360
2
P/min
Sync
- Analyse
mA
+
-
3
5 6 7 8 9 10
Event
11
16 Figure 2-1.
2
15
14 13 12
Controls and indicators of the Marquette Responder® 3000
Marquette Responder® 3000
227 490 02-H
Controls and Indicators
1
Connector for exchange of the defibrillation electrodes (switch off the device before exchanging the electrodes!)
18 Button to start and stop the recorder 19 Button to change the pacer output (current) 20 Button to change the pacer rate
2
APEX paddle
3
Infrared interface
21 Button to pause the pacer (without changing the settings)
4
Battery with "Test" button and charge level indication
22 Button to select the pacer mode (fixed rate, demand)
5
Button to unlock right battery for removal
23 Button to unlock left battery for removal
6
Energy selector, ON/OFF switch
24 Button to turn the pacemaker on and off
7
Indicator, yellow, flashes to the QRS rhythm in synchronized mode
25 Indicator, yellow: blinks with each delivered pacing pulse
8
Button to enable and disable the synchronized mode (cardioversion)
26 Button to open paper compartment
9
Button to initiate ECG analysis in the semiautomatic mode (only on semiautomatic defibrillator models)
10 Button to initiate defibrillator charging and to trigger the shock (together with button 11) when adhesive pads or internal spoons are used 11 Button to trigger the defibrillation shock (together with button 10) when adhesive pads or internal spoons are used 12 1-Volt ECG output 13 etCO2 signal input (optional) 14 SpO2 signal input (optional) 15 Function keys F1 to F5
27 Battery with "Test" button and charge level indication 28 Indicator, green: is lit when defibrillator is powered from an external source (mains, ambulance) 29 Indicator, yellow blinking: left battery charging on: left battery charged off: left battery missing or partially charged, no external power source connected 30 Indicator, yellow blinking: right battery charging on: right battery charged off: right battery missing or partially charged, no external power source connected 31 STERNUM paddle
16 ECG signal input 17 Event marker button Explanation of symbols used on the device
Battery charging
Consult accompanying documents
Caution, High Voltage
+
Type CF signal input: highly insulated, suitable for intracardiac application, defibrillation-proof Type CF signal input: highly insulated, suitable for intracardiac application
227 490 02-H
Housing without battery (to close the battery slot) Standby mode (line power operation) Audio alarm OFF
Marquette Responder® 3000
9
Defibrillation Electrodes
Defibrillation Electrodes
Hard Paddles
Hard paddle
a
c b
CHARGE SHOCK 3 2
APEX
Figure 2-2.
Hard paddles are the electrodes commonly used for external, transchest defibrillation. There is a special Apex paddle and a Sternum paddle. For delivery of the defibrillation shock, the paddles are placed directly on the body surface. Before use, however, an ample amount of electrode gel must be spread onto the paddles. Both paddles have a shock button: The shock button on the Apex paddle is used to initiate defibrillator charging; afterwards the defibrillation shock is triggered by pushing both shock buttons. The paddles can also be used to acquire the ECG signal. A smaller contact surface for defibrillation of children is integrated in the paddles (see "Defibrillation of Children" in section 4.2). Electrodes for Internal Defibrillation
d
Figure 2-3.
Electrode for internal defibrillation
Electrodes for internal defibrillation consist of a contact spoon (a, Figure 2-3), a handle b, and a counter nut c. The spoon must match the size of the heart and have full contact with the myocardium. There is a choice of 3 different spoon sizes. The electrodes as well as their connection cable must be sterilized before each use. An internal defibrillation is either performed with two spoon electrodes or with one spoon electrode and a so-called "external counter electrode" (d, Figure 2-3) which is placed under the patient and in the immediate vicinity of the heart. Defibrillator charging and release of the defibrillation shock are initiated with buttons on the device. Disposable Adhesive Electrodes Disposable adhesive electrodes are used both for defibrillation and for pacing. These electrodes are self-adhesive and pregelled. They are connected by means of a special cable and may remain attached to the patient for a maximum of 24 hours.
Figure 2-4.
10
Disposable adhesive electrode (external defibrillation, pacing)
Marquette Responder® 3000
227 490 02-H
Putting the Device Into Operation and Performance Check
3 Putting the Device into Operation and Performance Check Safety Information Danger
Warning
Explosion Hazard - The device is not designed
Shock Hazard - Observe the following warnings.
for use in areas of medically used rooms where an
Otherwise the lives of the patient, the user and
explosion hazard may occur. An explosion hazard
bystanders are in danger. The Marquette Responder® 3000 is a highvoltage electrotherapy device and must be handled by qualified and specially trained personnel. Improper use of the device can endanger life. Always follow the instructions given in the operator's manual.
may result from the use of flammable anesthetics, skin cleansing agents and disinfectants. Also, it is not permitted to operate the defibrillator in an oxygen-enriched environment or in the presence of flammable substances (gas) or anesthetics. Oxygenation in the vicinity of the defibrillation
-
When equipped with the AC power adapter, do not use the defibrillator outdoors because the power adapter is not splash-proof.
-
Before using the device, the operator must ascertain that it is in correct working order and operating condition. In particular, all connectors, electrodes as well as sensors and probes must be checked for signs of damage. Damaged parts must be replaced immediately, before use.
-
When disconnecting the device from the power line, remove the plug from the wall outlet first, before disconnecting the cable from the device. Otherwise there is a risk of coming in contact with line voltage by inadvertently introducing metal parts in the socket of the power cord.
-
As a general rule, utmost caution is advised for intracardiac application of medical technical devices. Great care must be exercised to prevent that conductive parts (connectors, electrodes, transducers) connected to the isolated patient signal input come in contact with other grounded conductive parts, as this could bridge the patient's isolation and cancel the protection provided by the isolated input.
electrodes must be strictly avoided. Temporarily interrupt the oxygen supply.
227 490 02-H
Marquette Responder® 3000
11
Putting the Device Into Operation and Performance Check
-
Electrically conductive contact with parts connected to the heart (pressure transducers, metal tube connections and cocks, guide wires, electrode catheters and the metal parts of syringes) must be avoided at all cost. When using devices intracardially, observe these guidelines:
-
always wear isolating rubber gloves;
-
parts with a conductive connection to the heart must be isolated from ground;
-
do not use tube fittings and stopcocks made of metal, if possible;
-
when connecting the heart catheter, observe these guidelines: - the connection must be isolated - all electrodes must be attached to the patient and secured against inadvertent disconnection or they must be isolated and protected against inadvertent contact (otherwise electrodes that become disconnected could bring the patient in contact with ground).
-
When devices are used intracardially, the annual Technical Inspections are mandatory. During intracardiac application of medical electrical devices, a defibrillator and pacemaker, both checked for proper functioning, must be readily available.
-
12
Ensure that no conductive connection between the patient and bystanders exists during defibrillation.
-
The mains plug must be connected to an appropriate power supply with a non-fused earthed wire. If these requirements cannot be guaranteed, connect the device to the ambulance power supply or operate it on battery power.
-
Do not use multiple portable socket outlets (MPSO) to connect the device to the power line.
-
Devices may be connected to other devices or to parts of systems only when it has been made certain that there is no danger to the patient, the operators, or the environment as a result. In those instances where there is any element of doubt concerning the safety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combination of devices. Standards IEC 60601-1-1/EN60601-1-1 must be complied with in all cases.
-
The device (without AC power adapter) is suitable for application in a humid environment provided the regulations concerning drip-proof equipment of IEC 60601/EN 60601 are strictly observed. However, do not defibrillate patients in a very moist or wet environment, unless absolutely necessary. Always dry the defibrillation electrodes and connection cables prior to defibrillation.
Marquette Responder® 3000
227 490 02-H
Putting the Device Into Operation and Performance Check
-
Warning Equipment Failure - Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason make sure that all external devices operated in the vicinity of the defibrillator comply with the relevant EMC requirements. Xray equipment, MRI devices, radio systems, and cellular telephones are a possible source of interference as they may emit higher levels of electromagnetic radiation. Keep the defibrillator away from these devices and verify the defibrillator performance before use.
-
Equipment Failure - Similarly, the defibrillator may disturb equipment operating in its vicinity when charging or delivering the shock. Verify the performance of these devices before use.
-
Suffocation Hazard - Dispose of the packaging material, observing the applicable wastecontrol regulations. Keep the packaging material out of children's reach.
-
Literature Medical Device Directive of August 2, 1994 EN 60601-1: 1990 + A 1: 1993 + A 2: 1995 Medical electrical equipment. General requirements for safety. EN 60601-1-1: 9/1994 + A1: 12/1995 General requirements for safety. Requirements for the safety of medical electrical systems. IEC-Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
Caution Equipment Damage, Shock Hazard - Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the application site. Wait until all moisture has vaporized before using the device.
-
Equipment Damage - Exercise great care when using HF surgery equipment on the patient at the same as the defibrillator. As a general rule, the distance between the ECG and defibrillation electrodes and the HF surgery electrodes should not be less than 15 cm. If this is not ensured, disconnect the electrodes and transducer leads while using the HF surgery device.
-
Equipment Damage - Avoid defibrillating repeatedly into open air or with the paddles shorted together, because the device temperature may increase to an inadmissible level due to the internal safety discharges.
227 490 02-H
Marquette Responder® 3000
13
Putting the Device Into Operation and Performance Check
Power Supply The defibrillator can be powered − from the power line (requires AC power adapter, P/N 205 108 01, Figure 3-1), − from the ambulance power supply system (requires vehicle mounting unit, P/N 202 317 01), − from the wall mount unit (P/N 202 317 03) − from 1 or 2 rechargeable batteries (mainsindependent). Batteries are recharged by one of the following methods:
Figure 3-1.
Defibrillator with AC power adapter Note
The Marquette Responder® 3000 is switched on and off with the energy selector. Once you are familiar with the device operating routines, this control lets you turn on the device and select the required defibrillation energy in one action. No shock can be delivered in the
position of the
energy selector.
− in the defibrillator, when the defibrillator is connected to the power line or to the ambulance power supply system, or − by means of the separate charging unit ASU 3000 (P/N 701 279 01). If you prefer to operate the device mainsindependent, ensure that the batteries are charged (chapter 14 "Battery Power Operation"). Please refer to chapter 16 for information on operating the device in the vehicle mounting unit and on installing the AC power adapter.
Turning the Defibrillator On • Connect the device to the power supply.
1 10 7 5 2 Autoseq
Figure 3-2.
14
• Switch on the device by turning the energy
20 30
50 100 150 200 300 360
Turning the defibrillator on
(defibrillation shocks cannot be selector to delivered in this position). The test screen appears and the device emits a short audio signal. On the test screen you can see the software version and a message referring to the self-test. The three color blocks in red, green and blue are displayed to verify the correct representation of the colors. After the self-test the standard screen appears (Figure 3-3).
Marquette Responder® 3000
227 490 02-H
Putting the Device Into Operation and Performance Check
defibrillator charging
440-Hz beep emitted during charging
Audio Signals
defibrillator charged
alternating tone (440 Hz / 880 Hz) emitted until shock release
Refer to the table at left for an explanation of the different audio signals and alarm tones emitted by the Marquette Responder® 3000.
audio alarm (HR alarm, lead fail alarm, SpO2 / EtCO2 alarm)
continuous 698-Hz tone emitted until cause of alarm is eliminated
battery alarm
triple 1046-Hz beep emitted at 20-second intervals
QRS beep
1397-Hz beep, or varying with the oxygen saturation level between 512 and 662 Hz
general alert
triple 880-Hz beep
alert to negative event
triple beep (2x880 Hz, 1x 550 Hz
a b Electrode
c
This is the information presented on the standard screen display: − windows for heart / pulse rate, SpO2 and etCO2 readings including the limit values a, b − battery charge level c
f
− ECG lead d
0
− window for ECG, SpO2 and etCO2 waveforms
Paddle
160 / 40
ECG
d e
The Standard Screen Display
bpm
− window indicating operating mode and defibrillation energy f
SpO2 parameter window
− date and time e
etCO2 parameter window
− menu g.
pacemaker window 15.07.1999
09:05:00
semiautom. ECG
SpO2 etCO2
QRSPulse Tone OFF
Next Menu
227 490 02-H
− the parameter reading is within the alarm limits (green), − a technical fault is reported (blue),
g Figure 3-3.
The color concept for the displayed information lets you see at a glance whether
Standard screen display a measured heart rate/pulse rate b selected limit values c battery charge level indication d ECG lead (ECG signal acquired via "Paddle") e date, time f selected operating mode, defibrillation energy g menu
− an alarm is reported (red), − the system displays a message (yellow). If the device does not receive an ECG signal, the HR window is blue (technical fault) and a sawtooth signal is displayed instead of an ECG. The SpO2 and etCO2 parameter windows are also blank and the corresponding waveforms are missing when the required sensors are not connected.
Marquette Responder® 3000
15
Putting the Device Into Operation and Performance Check
Display Flip
↓ ECG
SpO2 etCO2
Figure 3-4.
QRSPulse Tone OFF
Next Menu
Main menu
↓ Filter ON
Assign Channel Waveform
Figure 3-5.
Display
Previous Menu
• In the main menu, select F5 Next Menu (Figure
Main menu, page 2
↓
↓
Contrast
Contrast
Figure 3-6.
Memory
↓ Select Color
Display Flip
The screen display can be rotated 180° to adapt it to the operating position of the defibrillator. The display can be flipped permanently from the setup menu or temporarily as outlined below. You can also set up the system to flip the display automatically when the defibrillator is inserted in the vehicle mounting unit (chapter 13 "Defibrillator Setup"). 3-4).
Previous Menu
Display menu
You will see page 2 of the main menu (Figure 35).
• Display the Display menu with F4 Display (Figure 3-6).
Note The main menu will automatically reappear, when no button is activated for a period of 30 seconds.
• To flip the display, press F4 Display Flip . Contrast adjustment
• Adjust the contrast from the Display menu with F1 and F2. Adjusting Maximum Contrast (Select Color)
• Adjust the maximum contrast from the Display menu with F3.
Note Press F5 for about 2 seconds to return directly to the main menu.
• Press F5 Previous Menu for about 2 seconds to return directly to the main menu.
System Setup
Note Enter your own settings in the column at the far right (with date and signature).
16
The device has a configuration menu which allows you to customize some of the functions to suit your personal requirements. These settings will be retained. The table at right shows all device settings for which customer defaults can be selected, as well as the factory defaults. The information given in this manual is based on a defibrillator with the factory defaults. In chapter 13 "Defibrillator Setup" you will find instructions on setting up the defibrillator. The same chapter explains how to change the language and how to restore the factory defaults.
Marquette Responder® 3000
227 490 02-H
Putting the Device Into Operation and Performance Check
Parameter
Comment
Factory Defaults
Options
HR Limit
40/160
OFF, 15 to 300 increments of 5
SpO2 Limit
---/90
OFF, 60 to 99
etCO2 Limit
---/20
OFF, 5 to 76 increments of 1
User Setup
ALARM LIMITS
ECG Print on Alarm
autom. recorder start on violation of limit value
off
on/off
Lead Fail Alarm
audio signal indicating disconnected electrode
off
30 s/off
Alarm Tone
audio signal indicating violation of an alarm limit
off
on/off
off
low/middle/high / off
QRS Beep Muscle Filter
suppression of motion artifact
on
on/off
Gain
for ECG display
1 cm/mV
0.5; 1; 2 cm/mV
Lead Channel 1
I
standard leads, paddle acquisition
Lead Channel 2
II
standard leads, SpO2
Lead Channel 3
III
standard leads, SpO2, etCO2 on/off
DEFIB Print on shock
automatic recorder start on shock
on
Operating Mode**
choice of the operating mode
semiautomatic/button semiautom./button semiautom./password semiautom. manual
Autosequence
energy selection
200 J, 200 J, 360 J
150 to 360 J per shock
Pacemaker
default pacer rate
60 P/min
30 to 200 P/min
C-LOCK
C-Lock ECG synchronization
off
on/off
SpO2 Integ. Time
SpO2 integration time
8s
4 s, 8 s, 12 s
DATE/TIME**
Change clears all existing settings. day.mon.year
day.mon.year mon/day/year
SpO2
Date Format** Entry of date and time DEVICE Display
screen display (SmartFlip = display flips normal automatically when defibrillator is placed in vehicle mounting unit)
normal, reverse, SmartFlip
Volume
valid for all audio signals
high
high/low/middle
Cont. Printout
continuous recording or 14-second strip
off
on/off
Analysis
continuous ECG analysis
on
on/off
AC Line Filter**
elimination of AC line interference
50 Hz
off/50 Hz/60 Hz
Language**
selection of the language
Factory Default
restores factory defaults
User**
text or name (20 characters)
PASSWORD**
entry of the password
111
000/999
for config**
protects access to configuration menu
off
on/off
EVENT TEXTS**
entry of event texts
BATTERY
battery maintenance program
on/off
OPTIONS
entry of option code to unlock option
** not affected by reactivation of factory defaults
227 490 02-H
Marquette Responder® 3000
17
Putting the Device Into Operation and Performance Test
Performance Check A performance check must be carried out before each use.
Electrode 160 / 40
ECG
Paddle
The check includes:
bpm
− a visual inspection of the device, the cables and the electrodes for signs of mechanical damage, − verification of the functional readiness of the device, − delivery of a test discharge. 15.07.1999
09:05:00
semiautom. SpO2 etCO2
ECG
Figure 3-7.
QRSPulse Tone OFF
Next Menu
Standard screen display of a defibrillator ready for operation (no ECG, SpO2 and etCO2 signal available)
After power-up and during operation, the Marquette Responder® 3000 runs automatic selftests. If malfunctions are identified, an error message will be displayed (see chapter 17 "Error and System Messages"). In this situation do not put the device into service. In all other cases you will see the standard screen display (Figure 3-7) and the device is ready for use.
Note A special simulator is required to test the defibrillator performance in the semiautomatic mode.
Now verify that the defibrillation shock is correctly delivered by triggering a test discharge (chapter 15 "Test Discharge"). If the energy of the test discharge is not within the specified limits, a defibrillation is possible all the same (it is the user's decision whether or not to employ the defibrillator). However, the device must be immediately checked and repaired by a service technician.
Event Button
62 I
You can use the
160 / 40
ECG
Event
button to mark specific
events (e.g. administration of medications). When
bpm
you press this key, the corresponding point in time II
is earmarked in the full-disclosure ECG. Furthermore, you can assign a maximum of 8 "event"
III
texts to the function keys F1 to F4 (e.g. names of medications). When you press
15.07.1998
Text 1
Text 2
Event
these texts
09:05:00
Text 3
semiautom. Next Text 4 Menu
appear in the menu line (Figure 3-8). You can press one of the function keys to assign the corresponding text to the event. With F5 Next Menu
Figure 3-8.
Event texts
you can display the next line of 4 texts. (Refer to chapter 13 "Device Setup" for instructions on entering event texts.)
18
Marquette Responder® 3000
227 490 02-H
Putting the Device Into Operation and Performance Test
How to toggle the defibrillator from semiautomatic to manual operation 2 Sync
Analyse
Depending on their setup, semiautomatic defibrillators can be switched to manual control. The defibrillator can be set up for four different modes of operation: − semiautom./button (switching to manual mode by activating button)
3
− semiautom./password (switching to manual mode by activating button and entering password) − semiautomatic (manual control not possible) − manual (only manual control possible)
Figure 3-9. Electrode
Buttons to activate the manual mode
0
Operating Mode "semiautomatic/button" Paddle
160 / 40
ECG
• To activate the manual mode, simultaneously
bpm
press F5 and
15.07.1998
09:05:00
semiautom. 0
0
0
Analyse
(Figure 3-9).
When switched on again, the defibrillator will reactivate the operating mode selected in the setup menu. Date and time of the change of operating modes is stored in the event memory.
ENTER
Operating Mode "semiautomatic/password"
• To activate the manual mode, simultaneously Figure 3-10. Buttons for entry of the password
Analyse
(Figure 3-9).
The screen for entry of the password appears (Figure 3-10).
0 Paddle
160 / 40
ECG
press F5 and
bpm
• Enter the password (3-digit number) with F1, F2, F3. The factory-set password is 111 (also refer to chapter 13, section "Password").
15.07.1998
09:05:00
manual ECG
etCO2 SpO2
QRSPulse Tone OFF
Next Menu
Figure 3-11. Defibrillator set up for manual operation
227 490 02-H
When switched on again, the defibrillator will reactivate the operating mode selected in the setup menu. Date and time of the change of operating modes is stored in the event memory (chapter 11 "Memories of the Marquette Responder® 3000"). If you wish to return to the semiautomatic mode, you will have to turn the device off and on again.
Marquette Responder® 3000
19
Manual Defibrillation / Non-Synchronized Defibrillation
4 Manual Defibrillation 4.1 Defibrillator Application Guidelines Observe the following guidelines to ensure successful and safe defibrillation. Otherwise the lives of the patient, the user and bystanders are in danger. -
Warning -
Defibrillating a patient with normal heart rhythm may induce ventricular fibrillation.
-
Position the patient flat on a hard surface where he is electrically insulated. The patient must not be allowed to come into contact with metal parts, e.g., bed or litter, to prevent unwanted pathways for the defibrillation current which may endanger the assistants. For the same reason, do not position the patient on wet ground (rain, accident in swimming pool). Do not allow the defibrillation electrodes to come into contact with other electrodes or metal parts which are in contact with the patient. The patient's chest must be dry, because moisture can cause unwanted pathways for the defibrillation current. After use of flammable skin cleansing agents, wait until they have completely dried.
-
The operator and all assistants must be briefed regarding the preparations for and execution of defibrillation. All tasks must be clearly assigned. Immediately prior to the shock - interrupt heart massage and artificial respiration, - disconnect tube connections, and - warn bystanders.
-
Ensure that no conductive connection between the patient and bystanders exists during defibrillation.
-
Before delivering the shock, verify that the charged and selected energies are the same.
-
Shock Hazard - Always switch off the device before exchanging the defibrillation electrodes.
20
Pacemaker Patients - Defibrillating a patient with an implanted pacemaker is likely to impair the pacemaker function or cause damage to the pacemaker. For this reason - select the smallest energy level possible for the application, - do not apply the defibrillation paddles in the vicinity of the pacemaker, - have an external pacemaker at hand, - check the implanted pacemaker for proper functioning as soon as possible after the shock.
Caution -
Equipment Damage - Disconnect transducers and devices that are not defibrillationproof from the patient before delivering the shock.
-
Equipment Damage - Do not defibrillate the patient with a second defibrillator, while defibrillation electrodes (paddles, pads) of the first device are applied. If the use of a second defibrillator is inevitable, disconnect the electrodes from the first device or remove them from the patient.
Marquette Responder® 3000
227 490 02-H
Manual Defibrillation / Non-Synchronized Defibrillation
4.2 Non-Synchronized Defibrillation Using Paddles • Remove the paddles from their compartments as shown in Figure 4-1.
• Carefully dry the paddles and the handles in particular, if they are wet.
• Apply an ample amount of electrode cream to the paddle surfaces. Figure 4-1.
a
0
b
c
Paddle
160 / 40
ECG
• Set the energy selector to "Autoseq" or to the
Removing the paddles
bpm
15.07.1999
ECG
09:05:00
QRSPulse Tone OFF
SpO2 etCO2
Figure 4-2.
200 J
manual Next Menu
Power-up screen (paddles connected) a ECG signal acquired via "Paddle" b manual mode c selected energy
required energy level. In the "Autoseq" position of the energy selector, the defibrillator automatically sequences the preset defibrillation energy levels. The level for the 3rd shock is maintained for all subsequent defibrillations. When you set the energy selector again to "Autoseq", the automatic charge sequence starts over. The factory set Autosequence energy levels are the values recommended by AHA/ERC for ventricular fibrillation and pulseless tachycardia. 1st shock with 200 J 2nd shock with 200 J 3rd and all subsequent shocks with 360 J.
• Check that the selected energy is displayed (c, Figure 4-2).
• Apply the paddles on the patient's thorax such that the greatest possible amount of energy flows through the myocardium. The imaginary line connecting the paddle centers should be identical with the cardiac median line (Figure 4-3).
• Press the paddles firmly onto the thorax (the ECG appears on the monitor screen). ST ER NU
K OC SH 3
M
• Do not touch the patient any more and warn all those present. CH
AR 2 GE SH
AP EX
Figure 4-3.
227 490 02-H
O 3 CK
Paddle application points (dashed application points for pacemaker patients)
Marquette Responder® 3000
21