Emergency Q-Card  GE OEC® Elite Start the system 1. Ensure that the C-Arm is connected to a compatible Workstation and to electrical power. 2. Turn the keyswitch, located on the C-Arm interface panel, to the 45° X-rays and Motion Enabled position.  X-ray and Motion Enabled position  Workstation power switch  Keyswitch  3. Press the Workstation power switch. The power switch lights up as the system starts up. The system starts up and is fully functional within 175 seconds, or just under 3 minutes.  Shut the system down  Restart the system  Unexpected power loss  Normal shutdown procedure: 1. Press the Workstation power switch, shown in Start the system.  1. Follow the procedure described in Shut the system down.  When the system loses AC power unexpectedly, the system goes into a standby mode with limited functionality. A message displays on the monitor. Unless AC power is restored, the system begins a normal shutdown at the end of the specified time.  2. A shutdown message displays. The power switch blinks until the system has shut down successfully, and then turns off. Shutdown will not take less than 10 seconds. Immediate system shutdown: If an immediate shutdown of the system is required, or if the normal shutdown does not complete properly, use the system disable switch. CAUTION  Do not use the system disable switch for normal system shutdown.  Pressing the system disable switch removes all power from the system. To restore system operation, unplug the power cord for 10 seconds, plug it back in to the power outlet, and press the power switch.  2. After the system shuts down, wait 10 seconds and then press the green Workstation power switch, shown in Start the system, to restart the system. The power switch lights up as the system starts up. 3. Resume normal system use after the system restarts.  When AC power is not present, X-rays and motorized lateral rotation and cephalad caudal tilt movement are disabled. Lift column movement remains enabled. • If power is restored within the 20-second window, the Workstation will recover within 5 seconds. • If the power switch is pressed during a power interruption, the system will shut down, even if power is restored.  System disable switch  Follow the procedure described in Restart the system.  © 2019 GE OEC Medical Systems, Inc. 384 Wright Brothers Drive Salt Lake City, UT 84116 U.S.A. 801-328-9300  5493124-1EN-09 Rev 1 This document contains product information for the GE OEC Elite. It is to be used in conjunction with your system’s Operator’s Manuals. If system errors persist or occur repeatedly, refer to the GE OEC Elite Operator Manual, or call GE OEC service: 1-800-874-7378. It was originally written and verified in English.
File Type: PDF
File Size: 1.29 MB
File Name: GE Healthcare - 5493124-1EN - OEC Elite Emergency Quick Card Rev 1.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions