Quick Guide
2 Pages
Preview
Page 1
LOGIQ 200 OB QUICK CARD Keyboard Reference for OB Measurements BPD HC AC FL AFI CRL GS HR VOL
(Shaded areas indicate keys necessary for measurements described on reverse)
A
S
D
F
G
H
J
K
L
These measurement keys are programmable. To display current programmed settings, press the Calculation/Space Bar keys. New Patient
ID / Name
Report Page
1
1
2 Q
3 W
4 E
5 R
6
7
T
Y
8 U
9
2
0
I
X
C
V
B
Set
N
M
O
,
2
= P
[ : ;
Z
1
.
] ” ’
/
Measurement
ECG Position Ellipse
M Mode
Clear
L
R
Freeze
2286455–100 Rev 0
For help with additional questions, please call the Customer Answer Center at 1–800–682–5327, 1, 5 or 262–524–5698.
LOGIQ 200 OB QUICK CARD %%% For key location, see reverse. &&&
Measuring BPD, FL, CRL 1. Scan and Freeze the image. 2. Press the appropriate key (A* for BPD, F* for FL, H* for CRL). 3. With Trackball, move cursor to starting point of the measurement. Press Set. 4. Trackball second cursor to second position. Press Set.
Measuring AFI Requires 4 measurements
1. Scan first quadrant. Freeze image. 2. Press G* for AFI. 3. Trackball cursor to first position. Press Set. 4. Trackball second cursor to second position. Press Set. 5. Unfreeze, scan and Freeze next quadrant. Press Measurement. 6. Repeat steps 3 and 4 three more times, until all four quadrants have been measured. NOTE: If a quadrant has no visible fluid, image and measure that quadrant as close to zero as possible, and Set that measurement.
Measuring AC, HC As each measurement is completed and set, 1. Scan and Freeze the image. it automatically goes to the Report Page. 2. Press the appropriate key (D* for AC, Measuring Gestational Sac S* for HC). Use Dual B-Mode Display Requires total of 3 measurements in 2 planes, usually 1 in sagittal & 2 in transverse planes.
1. Press Left display key, image GS in first plane, press Right display key. (Left side of display freezes, Right is active.) 2. Image GS in second plane, press Freeze. 3. Press J* for GS. 4. Trackball cursor to first position, press Set. Trackball second cursor to second position, press Set. 5. Repeat step 4 to measure second distance. 6. Repeat step 4 to measure third distance.
Deleting a Measurement 1. Display the Report Page. 2. Press the bottom Ellipse key to display the Measurement Averaging page. 3. Trackball to measurement. Press Clear or Back Space. 4. Press Set to confirm deletion.
3. Trackball cursor to first position. Press Set. 4. Trackball second cursor to the second position. 5. Press the top Ellipse key; a circle opens. 6. Use the Trackball and Ellipse keys to adjust circumference size and position around anatomy. The Measurement key toggles between the fixed and active cursors. 7. Press Set.
Measuring HR 1. Image fetal heart. 2. Press M-Mode. 3. Adjust cursor with Trackball to obtain best fetal heart trace. Freeze image. 4. Press K* for HR. Use Trackball to position first cursor at starting point of the measurement. Press Set. 5. Trackball second cursor across two cardiac cycles, press Set.
For Multi-Gestational OB 1. On the Patient Entry Menu, enter all information. Enter Fetus Number – 2, 3, or 4. Answer question “Are you sure to change number? y/n” with y. NOTE: If Multi-Gestation is found during the exam, use ID/Name key to change Fetus Number. 2. To switch fetuses between measurements, use Ellipse key. The system does not switch fetuses if measurement cursor is active. All measurements are assigned in Report Pages and Graphs for the same Fetus until Ellipse key is pressed. Report With a Report Page displayed, use the Ellipse keys to change pages. To access report pages for other fetuses, Trackball to the left arrow at upper left of Report Page (i.e. ² A/2 ³). Press Set. Pages To view all fetuses’ measurements simultaneously, Trackball to arrow and press Set until the Report Page labelled ALL is displayed.
* These measurement keys are programmable. Those listed are the factory defaults. To display current programmed settings, press the Calculation/Space Bar keys. 2286455–100 Rev 0
For help with additional questions, please call the Customer Answer Center at 1–800–682–5327, 1, 5 or 262–524–5698.