LOGIQ 500 Quick Start Guide 2276613–100 A Training in Partnership Program
Ti P
TM
This page left blank intentionally.
Regulatory Requirement This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices
This Quick Start Guide is a reference for LOGIQ 500 PRO Series. It applies to all versions of 6.0 software for the LOGIQ 500.
GE Medical Systems GE Medical Systems: Telex 3797371 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. (Asia, Pacific, Latin America, North America) GE Ultrascall: Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstraße 239, Postfach 11 05 60, D–42655 Solingen GERMANY
REV
DATE
REASON FOR CHANGE
0
July 10, 2000
Initial Release
TiP Cover Page Title Page
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
N/A 0
Revision History A and B Quick Start 1 thru Quick Start 40
0 0
Revision History A
Revision History B
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Introduction The Quick Start Guide (TRANSLATED) provides a step-by-step description of the basic features and operation of the LOGIQ 500. It is intended to be used in conjunction with the Basic User Manual in order to provide the information necessary to operate the system safely. The Quick Start Guide takes the user from system familiarization through power on, patient data entry, exam category/preset selection, scan modes/adjustments, basic measurements, report pages, recording images and power off. The LOGIQ 500 manuals are written for users who are familiar with basic ultrasound principals and techniques. They do not include sonography training or clinical procedures.
Prescription Device (for USA only) CAUTION: United States law restricts this device to sale or use by or on the order of a physician.
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 1
Front View The following are major features of the LOGIQ 500 system console. Most features come with the standard configuration, while other features are options to the standard console. 1.
VCR Microphone
2.
Probe Holder
3.
MOD Drive
12
4.
Physio Panel *
11
5.
Probe Connectors (3 ports)
6.
Air Filter
7.
Power Supply Filter
8.
On/Off Switch
9.
B/W Page Printer *
1
10
2
9
3
8
4
10. Keyboard 11. Keyboard Task Light 12. Task Light Switch
5
* = optional feature or accessory
7
Quick Start 2
6
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Back View 7 6 5
1.
Circuit Breaker
2.
Air Filter
3.
Cable Access
4.
Peripheral Cable Access Door
5.
Handle for Pushing
6.
Monitor Arm Swivel Lock
7.
Color Monitor
4 2
Optional Freeze/Record Foot Switch
Freeze
3 1 ON ON
0 OFF
1 Record
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 3
Power On To connect the system to the electrical supply: S
Ensure that the wall outlet is a minimum 20 amp dedicated circuit for 120 VAC (USA) or 7.5 amp dedicated circuit for 220–240 VAC (Europe).
S
Make sure that the power switch is turned off.
S
Unwrap the power cable. Make sure to allow sufficient slack in the cable so that the plug is not pulled out of the wall if the system is moved slightly.
S
Push the power plug securely into the wall outlet.
To power on the system, ensure that the Circuit Breaker, located on the back of the power supply near the bottom of the unit, is turned on.
On
Quick Start 4
Press the Power On Power Off/Stand-by switch, located to the left of the probe connectors, to the on position.
LOGIQt 500 Quick Start Guide 2276613–100 Rev. 0
Power On (cont’d) During power up, the screen display changes as the system runs its self diagnostics. Start of diagnostic
XXXXXXX
XXXXXXX
Version X.XX
Version X.XX
Version X.XX
End of diagnostic
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 5
Control Panel Layout 1
2
3
Keyboard Keyboard controls have been arranged according to function and usage. This helps minimize operator movement while scanning. 1.
Alphanumeric Keys
2.
New Patient
3.
Probe Select
4.
Soft Menu Controls
5.
Doppler and CFM
4
5
↓
6
"yA
6.
User Define
7.
Mode, Display and Record
8.
Measurements/Annotations
9.
TGC and Acoustic Output
7
10. VCR
"yA
10
Quick Start 6
9
8
LOGIQt 500 Quick Start Guide 2276613–100 Rev. 0
Soft Menu Control Panel The Soft Menu Display has 8 Top Menu selections (Mode, Preset, Set Up, ECG, Archive, DICOM, AutoSeq and Cine) and 8 Sub Menu selections (varies depending on choice of Top Menu selection) available. The Top Menu Select key cycles through the top level menu page selections. The far left side top menu is the default selection and its sub-menus are automatically displayed. The Sub-Menu Select rocker switch cycles to the next ( ") or previous (A ) sub-menu page. The Sub-Menu rocker switches allow for the increase/decrease of a parameter value or enabling/disabling of a parameter.
LOGIQt 500 Quick Start Guide 2276613–100 Rev. 0
Top Menu Select
TOP MENU Current Available
1 2
SUB-MENU Current Available
2 4
B
Preset Set Up ECG
Frame Imaging Image Average Freq Softner Color OFF
3 MHz
3D Mode
Sub-Menu Select
Quick Start 7
Probe Controls A probe is activated by pressing the Probe Select key that corresponds to the probe port to which the desired probe is connected. The Single CWD key is used to activate the dedicated continuous wave Doppler (CWD) probe. The active probe key will be brightly lit.
Probe Ports All imaging probes can be plugged into any of the three probe ports. The CWD probes can be plugged into the Single CWD port which is located above Probe Port Position 3.
.
NOTE: Ensure that the probe port is deactivated and the image is frozen before disconnecting the probe.
Quick Start 8
lock unlock
LOGIQt 500 Quick Start Guide 2276613–100 Rev. 0
Starting an Exam Patient Entry Menu Press the New Patient key at the beginning of each study to reset the system and enter new patient data. When the New Patient key is pressed, the Patient Entry menu appears. Ensure that the Blue Shift key is not active. Use the Trackball to move the highlight cursor to the Exam Category Selection line. Enter the desired exam number and fill in the appropriate patient data. Highlight EXIT and press Return or press New Patient again when the patient entry menu data has been completed. The system takes a few seconds to load new exam category parameters and calibrate the attached transducers before entering the scan mode.
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 9
Application Presets Select the exam category application preset, that best describes the desired study to be performed, from the factory default preset selections displayed on the Soft Menu.
General OB
Fetal Heart
Tech Diff
Newborn
Select the desired probe.
Quick Start 10
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Scan Mode Controls Select the desired display mode or combination of display modes (B-Mode, Pulsed Doppler, Color Flow Mode, M-Mode or Continuous Wave Doppler Mode). CWD Mode is available for sector probe only.
If the dual display mode (split screen) is desired, the L and R keys activate the Left or Right displayed image.
Scan the desired anatomy. Change modes as desired. Use the mode controls to adjust the image as necessary.
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 11
B-Mode Controls There are four Soft-Menu pages for B-Mode imaging adjustments.
B
Preset
Set Up
ECG
Dynamic Range
Gray Map
Focus Number
ATO Create
48
B–2
2
ATO On/Off
Dynamic Range-Controls how echo intensities are converted to shades of gray, thereby creating a range of gray scale that can be adjusted. Gray Map-Determines how the echo intensity levels received are presented as shades of gray. Focus Number-Changes the number of focal zones so that the beam can be tightened or expanded for a specific area. ATO Create-Automatic Tissue Optimization optimizes the image based upon a specified Region of Interest (ROI) or anatomy within the display. ATO On/Off-Highlighted when the ATO map is active on the image; toggle on and off to display that same ATO map again.
B
Preset
Set Up
Frame Average
Imaging Freq
Image Softner
ECG
Frame Average-Averages previous frames of image data with the current frame. Imaging Freq-Allows more echoes to pass through. Image Softner-Adjust for the amount of smoothing applied.
Color OFF
3 MHz
Quick Start 12
3D Mode
Color-Allows for enabling B-Mode image colorization. 3D Mode-Allows for 3D Mode option activation.
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
B-Mode Controls (cont’d) B
Preset
Set Up
Focus Positn
ECG
Focus Positn-Changes the depth at which the selected number of focal zones are optimized.
Tag Positn
Color Tag-Enables the colorization of a specific gray scale level range. Tag Position-Allows for the movement of the specified color tag range throughout the gray scale displayed.
Color Tag
B
Preset
Set Up
ECG
Biopsy Zone
Image Rotatn
Rejectn
Edge Enhance
Biopsy Zone-Enables the electronic guidezone(s) available for the active probe. Image Rotate-Rotates the single real-time or zoomed B-Mode image in 90° increments. Rejectn-Allows for the elimination of low level echoes from the display.
SGL
0DEG
40
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
MID
B Edge Enhance-Brings out subtle tissue differences and boundaries by enhancing the gray scale differences corresponding to the edges of structures.
Quick Start 13
B-Mode Controls (cont’d) The keyboard controls that effect B-Mode are as follows:
8
9
↓ "yA
"yA
1
2
3 4
5
6
7
1.
Acoustic Output-Allows for varying the power output of the transmitted sound wave.
2.
TGC-Balances the image so brightness is consistent throughout the image. Move the slide pot to the right to increase or the left to decrease.
3.
Scan Area-Used to assign trackball control to adjust size/position B-Mode sector display, CFM window, and horizontal size/position of the real-time zoom window.
Quick Start 14
LOGIQt 500 Quick Start Guide 2276613–100 Rev. 0
B-Mode Controls (cont’d) 4.
Depth-To reduce depth (look at a shallower image), turn the Depth rotary encoder clockwise. To increase depth (look at a deeper image), turn the Depth rotary encoder counterclockwise.
5.
Display Format (Dual)- Left/Right split screen display.
6.
Reverse-The GE logo at the top of the sector wedge corresponds to the orientation mark on the probe body. When Reverse is active, the image flips and the GE logo switches to correspond to the probe mark.
7.
B/M Gain-Controls the gain of the displayed echoes. To increase/decrease, turn the control clockwise/ counterclockwise.
8.
Angle-Controls the gain in B/M and CFM-B/M Mode.
9.
Focus Position-While in B-Mode only, Baseline Shift can be used to change Focus Position.
LOGIQ 500 Quick Start Guide 2276613–100 Rev. 0
Quick Start 15