GE Healthcare
LOGIQ Ultrasound Systems
LOGIQ P5 and LOGIQ A5 and A5 Pro Advanced Reference Manual Rev 3
Advanced Reference Manual
358 Pages
Preview
Page 1
GE Healthcare
Technical Publications
Direction 5394145-100 Rev. 3
LOGIQ P5/LOGIQ A5/LOGIQ A5Pro Advanced Reference Manual
Copyright© 2011 By General Electric Co. Operating Documentation
Regulatory Requirement
This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices
This manual is a reference for the LOGIQ P5/LOGIQ A5/LOGIQ A5Pro. It applies to all versions of the software for the LOGIQ P5/LOGIQ A5/LOGIQ A5Pro ultrasound systems.
GE Healthcare
GE Healthcare: Telex 3797371 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall: TEL: 49 212.28.02.208 Deutschland GmbH & Co. KG: FAX: 49 212.28.02.431 Beethovenstrasse 239 Postfach 11 05 60 D-42655 Solingen GERMANY
Revision History
Table i-1: Reason for Change REV
DATE
REASON FOR CHANGE
Rev.1
April 20, 2010
Initial Release
Rev.2
October 20, 2010
Add A5
Rev.3
May 20, 2011
Probe update
List of Effective Pages Table i-2: List of Effective Pages PAGE NUMBER
REVISION NUMBER
PAGE NUMBER
REVISION NUMBER
Title Page
Rev. 3
2-1 thru 2-10
Rev. 3
Revision History i-1 and i-2
Rev. 3
3-1 thru 3-74
Rev. 3
Table of Contents ii-1 thru ii-2
Rev. 3
Index 1 thru 2
Rev. 3
1-1 thru 1-112
Rev. 3
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on GPC (GE Medical Systems Global Product Configuration). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
LOGIQ P5/A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
i-1
This page intentionally left blank.
i -2
LOGIQ P5/A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Table of Contents
Table of Contents
Chapter 1 - Acoustic Output Bioeffects Concerns Surrounding the Use of Diagnostic Ultrasound - - - - - - - - - - - 1-2
Acoustic Output Data Maximum output summary- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3
Chapter 2 - Measurement Formulas Cardiac Measurement Cardiac measurement abbreviation list - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Measurement Formulas Formulas–Generic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-14 Formulas-Urology - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-15 Formulas-Abdominal/Small Parts - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-16 Formulas–OB - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-17 Formulas–GYN - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-21 Formulas–Vascular - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-23 Formulas-Cardiac - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-24
DICOM SR Measurements Supported parameters- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-51
Chapter 3 - OB Tables OB Tables ASUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2 Berkowitz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-8 Bertagnoli - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-9 Brenner - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-9 Campbell - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-10 CFEF - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-11 Chitty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-16 Eriksen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-22 Erik-Nes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-23 Goldstein - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-25 Hadlock - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-26 Hansmann - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-33 Hellman - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-41 Hill - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-41 Hohler - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-42 Jeanty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-42 JSUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-54 Kurtz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-58 Mayden - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-59
LOGIQ P5/A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
TOC-1
Mercer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-60 Merz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-61 Moore - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-71 Nelson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-71 Osaka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-72 Paris - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-76 Rempen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-79 Robinson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-84 Tokyo - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-85 Tokyo Shinozuka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-87 Williams - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-93 Yarkoni - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-93
Index
TOC-2
LOGIQ P5/A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Chapter 1 Acoustic Output Provides Acoustic Output information and tables, as well as possible bioeffects and prudent use.
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-1
Acoustic Output
Bioeffects
Concerns Surrounding the Use of Diagnostic Ultrasound During a diagnostic ultrasound examination, high frequency sound penetrates and interacts with tissue in and around the area of anatomy to be imaged. Only a small portion of this sound energy is reflected back to the probe for use in constructing the image while the remainder is dissipated within the tissue. The interaction of sound energy with tissue at sufficiently high levels can produce biological effects (aka bioeffects) of either a mechanical or thermal nature. Although the generation of bioeffect is intentional with therapeutic ultrasound, it is generally undesired in diagnostic applications and may be harmful in some conditions. NOTE:
The American Institute of Ultrasound in Medicine (AIUM) has published a document entitled "Medical Ultrasound Safety". This three part document covers Bioeffects and Biophysics, Prudent Use and Implementing ALARA. Ultrasound users should read the AIUM documents to become more familiar with Ultrasound safety. A copy of this document is included as part of the documentation package (Document 2163920-100). To contact the AIUM concerning their publications: •
In the USA, by telephone at 1-800-638-5352.
•
To write them, use the following address: AIUM 14750 Sweitzer Lane Suite 100 Laurel, MD, USA
1-2
20707-5906
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Bioeffects Measurement Precision and Uncertainty Precision and measurement uncertainty of the system used to measure the acoustic output is provided in the following table. These values are determined in accordance with Section 6.4 of the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, rev 2 AIUM/NEMA 2004, and the Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, rev 3 NEMA 2004. Table 1-1: Measurement Precision and Uncertainty Center Frequency Measurement Uncertainty
+ 0.034%
Acoustic Power + 4.24%
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Peak Rarefactional Pressure + 10.07%
Acoustic Intensity + 20.34%
1-3
Acoustic Output
FDA Acoustic Output Data
Maximum Output Summary The following tables list the typical maximum acoustic output levels achievable with LOGIQ P5 for all probes and operational modes. It is intended that this is information be useful in making ALARA decisions and selecting the most appropriate probe for the application. In accordance with US FDA Guidelines, the overall maximum acoustic SPTA intensity for LOGIQ P5 is limited to 720 mW/cm2 and MI is limited to 1.9. Modes for which TI does not exceed 0.1 are indicated by <0.1. Table 1-2 shows all of the probes’ operating modes. The following pages contain probe acoustic output tables for those noted in Table 1-2. Table 1-2: Summary Table Operating Mode
10L
12L
ERBL
I739
T739
8C
BE9C
E8C
ERBC
3S
5S
B-Mode
X
X
X
X
X
X
X
X
X
X
X
M-Mode (includes B-Mode)
X
X
X
X
X
X
X
X
X
X
X
Color Flow
X
X
X
X
X
X
X
X
X
X
X
Pulsed Doppler (includes B-Mode)
X
X
X
X
X
X
X
X
X
X
X
B-Flow
X
X
X X
X
X
X
X
X
Coded Excitation Contrast Continuous Wave (CW) Doppler
1-4
X X
X X
X
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
FDA Acoustic Output Data
Table 1-3: Summary Table (Continued) Operatin g Mode
7S
3CRF
3.5C
9L
4C
E8CS
4D3C-L
8L
UG7C
3.5CS
B-Mode
X
X
X
X
X
X
X
X
X
X
M-Mode (includes B-Mode)
X
X
X
X
X
X
X
X
X
X
Color Flow
X
X
X
X
X
X
X
X
X
X
Pulsed Doppler (includes B-Mode)
X
X
X
X
X
X
X
X
X
X
B-Flow
X
X
Coded Excitation Contrast Continuous Wave (CW) Doppler
X
X
X
X
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-5
Acoustic Output
Table 1-4: Summary Table (Continued) Operatin g Mode
P2D
P6D
11L
4DE7C
5CS
3Sp
5Sp
i12L
4D8C
BE9CS
B-Mode
X
X
X
X
X
X
X
X
M-Mode (includes B-Mode)
X
X
X
X
X
X
X
X
Color Flow
X
X
X
X
X
X
X
X
Pulsed Doppler (includes B-Mode)
X
X
X
X
X
X
X
X
B-Flow
X
Coded Excitation
X
X
Contrast
X
Continuous Wave (CW) Doppler
X
X
X
X
Maximum Probe Temperature Table Release Notes-2: Maximum Probe Temperature (Degrees C) Probe
Max Temp
Probe
With TMM Phan -tom*
In Air
Mode
10L
39.6
37.4
B-Flow
12L
39.1
32.4
ERBL
41.5
I739
Max Temp
Probe
With TMM Phantom*
In Air
ERBC
40.2
41.0
Color Flow
Pulsed Doppler
3S
39.6
33.6
35.5
Color Flow
5S
41.0
40.5
35.1
B-Flow
7S
T739
40.7
35.3
B-Flow
8C
41.0
41.0
Pulsed Doppler
1-6
Max Temp With TMM Phan -tom*
In Air
4D3C-L
40.8
39.2
Color Flow
Pulsed Doppler
8L
36.5
39.8
Pulsed Doppler
41.0
Pulsed Doppler
UG7C
41.9
40.1
Color Flow
38.1
39.2
CW Doppler
3.5CS
40.2
39.2
Pulsed Doppler
3.5C
40.2
39.2
Pulsed Doppler
P2D
38.5
34.5
CW Doppler
9L
36.3
37.5
Color Flow
37.3
36.2
CW Doppler
Mode
P6D
Mode
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
FDA Acoustic Output Data Table Release Notes-2: Maximum Probe Temperature (Degrees C) Probe
BE9C E8C 5CS 3Sp BE9CS
Max Temp With TMM Phan -tom*
In Air
41.3
40.0
41.0 41.8 37.1 40.3
41.0 40.3 41.8 42.1
Probe
Mode Color Flow
4C
Max Temp With TMM Phantom*
In Air
38.6
40.3
42.1
Color Flow
E8CS
Color Flow
3CRF
Pulsed Doppler
5Sp
42.5 37.3
40.0 41.5 41.9
Probe
Mode
Color Flow
11L
Harmo nic
4DE7C
Color Flow
i12L
Pulsed Doppler
4D8C
Max Temp With TMM Phan -tom*
In Air
40.9
39.6
Pulsed Doppler
42.6
37.8
Color Flow
38.3
33.0
Color Flow
38.6
41.4
Color Flow
Mode
Pulsed Doppler
*Transesophageal Probes, UG7C probe, are equipped with internal temperature sensor and mechanism to monitor and limit temperature. Notes: Lens temperature measured under following conditions per IEC 60601-2-37: Thermocouple was placed at the geometric center of the lens. Probe placed upright in contact with above thermal phantom. Autofreeze capability disabled. Lens temperature monitored for 30 min. Measurement uncertainty: +-0.5 degC. *Thermal phantom made with tissue-mimicking material as referenced in IEC606012-37: Appendix J.
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-7
Acoustic Output
Acoustic Output
1-8
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Acoustic Output Table Key Table 1-1: FDA
IEC
Table Key
Meaning-IEC 60601-2-37 / FDA & NEMA UD2, UD3
a
α
Acoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz)
Aaprt
Aarpt
-12db Output Beam Area / Active aperture area
CMI
Normalizing Coefficient
Deq
Deq
Equivalent Aperture Diameter / (same)
d-6
d-6
Pulse Beam Width / Beam diameter at –6 dB
deq
deq
Equivalent Beam Diameter
fc
ƒawf
Acoustic Working Frequency / Center frequency
Ipa
Ipa
Pulse-Average Intensity
Ipa.3
Ipa,α
Attenuated Pulse-Average Intensity
PII
Ipi
Pulse-Intensity Integral
PII.3
Ipi,α
Attenuated Pulse-Intensity Integral
ITA
Ita(z)
Temporal-Average Intensity
ITA.3(Z)
Ita,α(z)
Attenuated Temporal-Average Intensity / (at depth z)
ISPTA(Z)
Izpta(z)
Spatial-Peak Temporal-Average Intensity
ISPTA.3(Z)
Izpta,α(z)
Attenuated Spatial-Peak Temporal-Average Intensity
MI
MI
Mechanical Index
Wo
P
Output Power / Time average acoustic power at the source
W.3(Z)
Pα
Attenuated Output Power / Time average acoustic power derated to depth z
Wo1
P1
Bounded Output Power / Power emitted from the central 1cm of aperture
PII
pi
Pulse Pressure Squared Integral / Pulse intensity integral
pr
pr
Peak-Rarefactional Acoustic Pressure / (same)
pr.3
prα
Attenuated Peak-Rarefactional Acoustic Pressure / (same)
PRF
prr
Pulse Repetition Rate / Pulse repetition frequency
TI
TI
Thermal Index / (same)
TIB
TIB
Bone Thermal Index / (same)
TIC
TIC
Cranial-Bone Thermal Index / (same)
TIS
TIS
Soft-Tissue Thermal Index / (same)
PD
td
Pulse Duration / (same)
x-12,y-12
X, Y
-12 dB Output Beam Dimensions / (same)
Z
Z
Distance from the Source to a Specified Point / (same)
Zsp
Zb
Depth for TIB / Depth at which the relevant index is maximum
Zbp
Zbp
Break-Point Depth / (same)
Zsp
Zs
Depth for TIS / Depth at which the relevant index is maximum
-
-
This parameter is not applicable to the specified probe/mode combination.
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-9
Acoustic Output LOGIQ P5 Acoustic Output Tables for TD Board Rev 1 10L Table Release Notes-2: Transducer Mode: 10L Operating Mode: B-Mode Index Label
MI scan
Global Maximum: Index Value
1.67
IEC
FDA
Units
pra
pr.3
(MPa)
P
Wo
(mW)
TIS
TIB
non-scan Aaprt<
Aaprt>1
0.92
-
-
-
1.84
66.90
-
-
66.90
3.80
min of [Pα(zs, Ita,α(zs)] [(W.3(Z1),ITA.3(z1)])
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
ƒawf
fc
(MHz)
Dim of Aaprt
TIC
nonscan
1.30
-
5.47
5.88
-
-
-
5.88
X
(cm)
1.30
-
-
-
1.30
Y
(cm)
0.50
-
-
-
0.50
td
PD
(µsec)
prr
PRF
(Hz)
4417.0
pr at max. Ipi
pr@PIImax
(MPa)
4.47
deq at max. Ipi
deq@PIImax
(cm)
Focal Length
FLX
(cm)
11.20
-
-
11.20
FLY
(cm)
2.60
-
-
2.60
Ipa,αat max. MI Ipi aat max. MI
IPA.3@MImax
0.23
-
(W/cm2)
769.35
(W/cm2)
176.95
Frequency (MHz)
6.67
8.00
-
-
-
8.00
Image Depth (cm)
14.00
12.00
-
-
-
12.00
1-10
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Acoustic Output Table Release Notes-3: Transducer Mode: 10L Operating Mode: M-Mode TIS Index Label
MI
non-scan
scan Global Maximum: Index Value
1.67
IEC
FDA
Units
pra
pr.3
(MPa)
P
Wo
(mW)
TIB
Aaprt<
Aaprt>1
-
0.12
-
0.25
0.18
-
4.59
4.59
4.59
3.80
min of [Pα(zs, Ita,α(zs)] [(W.3(Z1),ITA.3(z1)])
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
ƒawf
fc
(MHz)
Dim of Aaprt
TIC nonscan
1.30
1.30
0.37 5.47
-
5.47
-
5.47
5.47
X
(cm)
-
0.61
-
0.61
0.61
Y
(cm)
-
0.50
-
0.50
0.50
td
PD
(µsec)
0.23
prr
PRF
(Hz)
500.00
pr at max. Ipi
pr@PIImax
(MPa)
4.47
deq at max. Ipi
deq@PIImax
(cm)
Focal Length
FLX
(cm)
-
1.80
-
1.80
FLY
(cm)
-
2.60
-
2.60
Ipa,αat max. MI Ipi aat max. MI
IPA.3@MImax
0.37
(W/cm2)
769.35
(W/cm2)
176.95
Frequency (MHz)
6.67
-
6.67
-
6.67
6.67
Image Depth (cm)
14.00
-
14.00
-
14.00
14.00
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-11
Acoustic Output Table Release Notes-4: Transducer Mode: 10L Operating Mode: Color Flow TIS Index Label
MI
non-scan
scan Global Maximum: Index Value
1.76
IEC
FDA
Units
pra
pr.3
(MPa)
P
Wo
(mW)
TIB
Aaprt<
Aaprt>1
0.92
-
-
-
2.19
59.85
-
-
59.85
3.94
min of [Pα(zs, Ita,α(zs)] [(W.3(Z1),ITA.3(z1)])
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
ƒawf
fc
(MHz)
Dim of Aaprt
TIC nonscan
1.75
-
5.02
5.02
-
-
-
5.02
X
(cm)
0.73
-
-
-
0.73
Y
(cm)
0.50
-
-
-
0.50
td
PD
(µsec)
0.47
prr
PRF
(Hz)
1900.0
pr at max. Ipi
pr@PIImax
(MPa)
5.08
deq at max. Ipi
deq@PIImax
(cm)
Focal Length
FLX
(cm)
2.20
-
-
2.20
FLY
(cm)
2.60
-
-
2.60
Ipa,αat max. MI
IPA.3@MImax
(W/cm2)
-
800.15
Ipi aat max. MI
376.07
Frequency (MHz)
5.00
5.00
-
-
-
5.00
Image Depth (cm)
3.00
3.00
-
-
-
3.00
Velocity Scale ( kHz)
1.90
10.40
-
-
-
10.40
Penet
On
On
-
-
-
On
ROI
72.00
37.00
-
-
-
37.00
PS
14.00
14.00
-
-
-
14.00
1-12
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
Acoustic Output Table Release Notes-5: Transducer Mode: 10L Operating Mode: Pulsed Doppler TIS Index Label
MI
non-scan
scan Global Maximum: Index Value
0.53
IEC
FDA
Units
pra
pr.3
(MPa)
P
Wo
(mW)
TIB
Aaprt<
Aaprt>1
-
1.43
-
1.55
1.70
-
59.85
59.85
59.85
1.18
min of [Pα(zs, Ita,α(zs)] [(W.3(Z1),ITA.3(z1)])
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
ƒawf
fc
(MHz)
Dim of Aaprt
TIC nonscan
1.82
3.39
0.52 5.02
-
5.02
-
5.02
5.02
X
(cm)
-
1.22
-
1.22
1.22
Y
(cm)
-
0.50
-
0.50
0.50
td
PD
(µsec)
1.89
prr
PRF
(Hz)
5277.0
pr at max. Ipi
pr@PIImax
(MPa)
1.53
deq at max. Ipi
deq@PIImax
(cm)
Focal Length
FLX
(cm)
-
3.60
-
3.60
FLY
(cm)
-
2.60
-
2.60
Ipa,αat max. MI Ipi aat max. MI
IPA.3@MImax
0.52
(W/cm2)
58.29
(W/cm2)
110.17
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3
1-13
Acoustic Output Table Release Notes-5: Transducer Mode: 10L Operating Mode: Pulsed Doppler TIS Index Label
MI
TIB
non-scan
scan
TIC
Aaprt<
Aaprt>1
nonscan
Frequency (MHz)
5.00
-
5.00
-
5.00
5.00
Image Depth (cm)
4.00
-
4.00
-
4.00
4.00
Velocity Scale (kHz)
5.30
-
10.20
-
10.20
10.20
SV
3.00
-
3.00
-
3.00
3.00
Penet
On
-
On
-
On
On
Table Release Notes-6: Transducer Mode: 10L Operating Mode: B-Flow TIS Index Label
MI
non-scan
scan Global Maximum: Index Value
0.25
IEC
FDA
Units
pra
pr.3
(MPa)
P
Wo
(mW)
TIB
Aaprt<
Aaprt>1
3.03
-
-
-
3.17
107.9
-
-
107.9
0.76
min of [Pα(zs, Ita,α(zs)] [(W.3(Z1),ITA.3(z1)])
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
ƒawf
fc
(MHz)
Dim of Aaprt
1-14
TIC nonscan
0.98
-
6.42
6.07
-
-
-
6.07
X
(cm)
1.14
-
-
-
1.14
Y
(cm)
0.50
-
-
-
0.50
LOGIQ P5 / A5 Advanced Reference Manual Direction 5394145-100 Rev. 3