Manual
12 Pages
Preview
Page 1
GETINGE Drying cabinet 363 MANUAL 503062800
SEV0850001-
Contents SAFETY INSTRUCTIONS_ ________________________________3 General safety instructions_ __________________________________ 3 Power cut-off device_________________________________________ 3 Warning symbols____________________________________________ 3
DRYING CABINET________________________________________4 Construction________________________________________________ 4 Capacity___________________________________________________ 4
TECHNICAL DATA________________________________________4 INSTALLATION___________________________________________5 Free-standing model_________________________________________ 5
ELECTRICAL INSTALLATION_____________________________5 DIRECTIONS FOR USE___________________________________7 Error code 01____________________________________________8
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
Drying anaesthetic and respiratory hoses and bellows_ __________ 8 Maintenance________________________________________________ 8 Cleaning the filter cassette_ __________________________________ 9 Overheating cut-out_________________________________________ 10
Page 2 of 12
SAFETY INSTRUCTIONS
This drying cabinet has been designed with a number of integral safety devices. To avoid injury to the person it is essential that the safety devices are not bypassed or in any other way put out of action.
General safety instructions
• The drying cabinet should be connected according to the instructions. • The drying cabinet should be used according to the instructions. • The drying cabinet should not be used by minors. • Installation work and servicing should only be undertaken by staff qualified to do so. • Staff involved should study up-to-date handbooks and service manuals before undertaking any repairs or service work. • All leads must be disconnected from all circuit boards of the control system before commencing welding on or near the drying cabinet. • The drying cabinet should not be sprayed with water.
Power cut-off device
The drying cabinet should be provided with a separate main switch in the power supply line, easily accessible on the wall.
Warning symbols
This manual includes advice, and certain warnings and instructions of such importance that they have been specially marked. The configuration and use of the relevant symbols are as follows: This symbol indicates a warning in the text. It gives warning of dangers which could lead to minor or major injury to the person and even to a fatality. It is also used in warnings to avoid damage to the machine.
Page 3 of 12
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
This symbol indicates a warning in the text concerning high voltage (>2000 V). It gives warning of dangers which could lead to minor or major injury to the person and even to a fatality.
DRYING CABINET
Made of stainless steel, SIS 2333, with a base and four adjustable feet. Insulated with 20 mm mineral wool sheets. The cabinet operates with recirculated hot air, approximately 80% recirculation and 20% additional electrostatically-filtered air. Automatic electronic controls with display of the drying time. The drying time can be adjusted from 0-95 minutes. The time is counted down in steps of 1 minute during the process so that the display shows how much drying time is left. When the drying time is complete, the fan runs for an additional 2 minutes with the heater switched off so that the overheating cut-out does not trip. There are two heater settings: 70° C for plastics and 90° C for rubber hoses and instruments. The display alternates between and the drying time when the filter needs cleaning.
Construction
The cabinet is available as a single configuration or a through-put cabinet. The doors are fitted with a cremone bolt, either right-handed or left-handed.
Capacity
36 hoses or 9 DIN baskets or 9 shelves or SPRI I racks or 18 hoses and 8 shelves with SPRI II racks.
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
TECHNICAL DATA
Environmental requirements: Humidity, maximum 80% at 31° C Room temperature 5–40° C Weight approximately 100 kg. Internal dimensions 1600 x 495 x 375 (H x W x D). Fan capacity 200 m3/h Exhaust air, approximately 50 m3/h exhaust stub pipe Ø 100 mm valve regulated. Heat dissipation into the room, approximately 150 kcal/h if the exhaust stub pipe is connected to the exhaust ventilation. Power 3500 W Electrical connection 3N - 400/415 V, 50/60 Hz 3 - 208/240 V, 50/60 Hz 3 - 230 V, 50 Hz Fuse 10 A
Page 4 of 12
INSTALLATION Installation should only be carried out by authorised staff.
Free-standing model
ELECTRICAL INSTALLATION
The drying cabinet has to be connected to a wall mounted main switch, with a 3 mm gap .
Drying cabinets for the Japanese market are fitted with a main switch on the top of the cabinet.
Page 5 of 12
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
Secure the machine firmly to the wall with two screws through the holes A at the top of the machine. A
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
Page 6 of 12
DIRECTIONS FOR USE LED
&=
Display
Increase Decrease Temperature selection
Start/stop
1. Cassettes together with moist hoses should be lifted directly into the drying cabinet, where they are slid in on guides. 2. Respiratory bellows are placed in the cassette designed for the purpose. 3. Racks for instruments etc. are placed on the rack shelves. 4. Select the required drying temperature, 70° C or 90° C by pressing the temperature select button. 5. Set the required drying time on the display. Increase the time by pressing A, decrease the time by pressing B. 6. Start the drying operation by pressing the start/stop button. The LED comes on when the drying process starts. Note! Do not disconnect the power supply during the drying process.
7. Drying is complete when the display shows 0. Note. It is time to clean the filter cassette when the display alternates between
and the drying time.
Page 7 of 12
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
8. Digital display of chamber temperature. By pressing the two arrow keys and the temperature selection button and the start/ stop button simultaneously, the chamber temperature will be indicated on the digital display. After displaying the temperature for 2 minutes the display window will revert to showing the count-down drying time.
Error code 01
Error code 01 indicates that contact with the temperature gauge has been broken.
Drying anaesthetic and respiratory hoses and bellows
The drying cabinet can be used for 6 cassettes at the same time. To enable the drying air to be forced through the tubing, each drying cabinet is also equipped with 4 loose blind flanges. If fewer than 6 cassettes are used during the drying operation, these blind flanges must be slid into the empty guides. If only a single cassette is used, all 4 blind flanges must be slid in, leaving one guide uncovered.
Maintenance
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
If necessary When necessary, the outside and the inside of the drying cabinet can be cleaned and disinfected with an alcohol-based detergent, surgical spirit, or normal detergent. The control panel can be cleaned in the same way. Remove marks with ordinary stain-remover for stainless steel.
Page 8 of 12
Cleaning the filter cassette Only to be carried out by authorised staff Ensure that the power supply is disconnected before starting work on the machine.
1
2
B
A
10
C
Do not use sharp implements when dismantling. Clean the filter once a year.
Page 9 of 12
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
V463
1. Undo screw (A) and remove the cover. 2. Pull out the handle (B) and take out the cassette. 3. Take a suitable plastic wash basin in which the filter cassette can be completely immersed. 4. Fill up with approx 60° C hot water and add about 1% alkaline detergent. 5. Stir the solution and carefully immerse the filter cassette . 6. Take out the filter after 15–30 minutes and rinse it on all sides with clean water under a running water-tap. 7. Place the filter on water-absorbent paper or lint for 24 hours. 8. Do not use compressed air for cleaning or drying. 9. Check that the filter cassette has dried completely. 10.Put the cassette back into place after cleaning it. N.B. Make sure the cassette is not turned upside down. The contacts (C) at the back of the filter should face downwards.
Overheating cut-out Only to be carried out by authorised staff. Ensure that the power supply is disconnected before starting work on the machine
If the heating fails to work, or the fan fails to start, the over-heating cut-out may have tripped. Remove front panel A. Reset the cut-out by pressing red button B.
V651 V428
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
A
Page 10 of 12
B
<Doc_USR><Doc_503062800><Rel_A><Lang_GB>
Page 11 of 12
australia
France
South africa
Getinge Australia Pty ltd PO Box 50, Bulimba QLD 4172 Unit 1, 205 Queensport Rd. Murarrie QLD 4173 [email protected] Phone:+61-7 3399 3311
Getinge France SAS BP 49, avenue du Canada ZA de Courtaboeuf Les Ulis, FR-91942 [email protected] Phone: +33-1 64 86 89 00
Getinge South Africa (Pty) Ltd P O Box 48492 Hercules Pretoria SA 0002 [email protected] Phone: +27-123 721 370
belgium
Germany
Spain
Getinge NV Vosveld 4 B-2 B-2110 Wijnegem [email protected] Phone: +32-33 542 865
canada Getinge Canada Ltd 1575 South Gateway Road, Unit C Mississauga Ontario L4W 5J1 [email protected] Phone: +1-905 629 8777
china Getinge (Suzhou) Co.Ltd No.158,Fang Zhou Road, Suzhou Industrial Park 15021 Suzhou, Jiangsu Province P.R.China [email protected] Phone: +86- 51 262 838 966 Getinge Shanghai Trading Co. Ltd. Rm 1988 Tower B, CityCentr, 100 Zunyi Rd. 200051 Shanghai P.R. [email protected] Phone: +86- 21 623 72 408
Denmark Getinge Danmark A/S Firskovvej 23 DK-2800 Lyngby [email protected] Phone: +45-45 93 27 27
Getinge Vertrieb & Service GmbH Kehler Strasse 31 764 37 RASTATT TYSKLAND Tel: +49-7222 932 306 Fax: +49-7222 932 597 e-mail: [email protected]
Getinge Iberica SL P.E. San Fernando, Avda. Castilla 2, Edif. Francia 1era planta San Fernando de Henares Madrid ES-28830 [email protected] Phone: + 34-916 78 26 26
Italy
Sweden
Getinge S.p.A via Poggio Verde, 34 00148 Roma [email protected] Phone: +39-06 656 631
Netherlands Getinge B.V. Fruiteniersstraat 27, Zwijndrecht Postbus 1004 NL-3330 CA Zwijndrecht [email protected] Phone: +31-78 610 24 33
Norway Getinge Norge A/S Ryenstubben 2 0679 Oslo [email protected] Phone: +47-23 051 180
Poland Getinge Poland Ul. Lirowa 27 02-387 Warszawa [email protected] Phone: +48-22 882 06 26
Finland Getinge Finland AB Ängsgatan 8 FI-02200 Esbo [email protected] Phone: +35-89 6824 120
Singapore Getinge International Far East Pte. Ltd. 20 Bendemeer Road, #06-02, Cyberhub Building Singapore, SG-339914 Phone: + 65- 6396 7298
www.getinge.com
Getinge International AB P O Box 69 SE-310 44 Getinge [email protected] Phone: +46-35 15 55 00
Switzerland Getinge ALFA AG Weidenweg 17 4310 Rheinfelden [email protected] www.getingealfa.ch Phone: +41-61 836 15 15
United Kingdom Getinge UK Ltd Orchard Way Calladine Park Sutton-In-Ashfield Notts NG 17 1JU [email protected] Phone: +44-1623510033
USa Getinge USA Inc. 1777 East Henrietta Road Rochester, NY 14623-3133 [email protected] www.getingeusa.com Phone: +1-5,854,751,400