GB... Digital Scale model RS
Item no.: 560602 & 560604 1.00... General provisions... 4 2.00... Intended use... 5 2.01... Safety... 6 2.02... Use... 8 2.03... Configuration table... 10 2.04... Accessories... 12 2.05... EMC - Information... 12 3.00... Technical properties... 16 4.00... Control panel... 16 5.00... Weighing mode... 17 6.00... Settings and functions... 18 7.00... Choice / change of automatic shut-off and auditory signal... 18 8.00... Settings for the digital scale step-by-step... 19 9.00... Replacement of battery... 19 10.00... Error messages... 20 11.00... Maintenance and service... 21 12.00... Disposal of digital scale including batteries... 22
© Guldmann GB-02/2022 • # 560642_101.01
13.00... Identification labels... 23 14.00... Warranty and service conditions... 24 A... Warranty... 24 B... Service or Repair... 25
3
General provisions Read this user manual carefully before using the digital scale, as it supplies important information concerning use, safety and maintenance. The descriptions and illustrations in these instructions are not binding. Guldmann reserves the right to make changes to the product that they deem to be necessary in order to improve the product, without them being obligated to update this publication. Product type: Professional electronic digital scale for mobile lifters and ceiling hoists. Conventions: In this user manual, the following symbols are used 0476 EC type-approved for medical use Metrological certification and type approval Accuracy class ATTENTION! The digital scale must be mounted correctly before it is placed into use. In the event of erroneous installation, damage may be caused to the digital scale and injury to the patient. Must not be disposed of as ordinary household refuse, but must be brought to recycling. Type B in accordance with UL/EN 60601-1 Battery operated Indication of weight function O
Indication on stable weight Interference may arise in the vicinity of the devices Dual insulation (Class II)
4
© Guldmann GB-02/2022 • # 560642_101.01
1.00
2.00
Intended use This equipment must be used for weighing a person lifted in a sling for general diagnostic purposes. Application area: at institutions, hospitals and specialised medical clinics. We recommend that the unit be used in surroundings that are not subjected to magnetic disturbances. Personnel who may use the product: Trained caregivers, specialised operators and physicians, who are familiar with the procedures for proper use. Controls and responsibility: The medical equipment must be used under the supervision of a qualified physician (only for Class III scale) or qualified personnel with responsibility for maintenance and periodic controls, and who are familiar with all the safety procedures.
© Guldmann GB-02/2022 • # 560642_101.01
Limitations of use: This medical equipment must only be used as described in this user manual.
5
2.01
Safety The operators must read this user manual carefully, follow the directions in the manual and learn the correct procedures for the use and maintenance of the digital scale. These user instructions contain important information on the mounting, use and maintenance of the digital scale. The manufacturer assumes no liability for direct or indirect damages, including loss of earnings or other commercial damage, which may be due to use of the product that is not in accordance with what is described in this user manual. Save this user manual and the declaration of conformity for later use, and use them when training personnel. Avoid overloading the digital scale beyond its maximum capacity Avoid sudden loading of the digital scale Do not use sharp or pointed objects to press the keys Do not try to open the digital scale Do not remove the seals on the digital scale The battery poles must not be short-circuited Do not submerge the digital scale in water or other liquids Perform regular maintenance and subsequent measurement checks (see section 11.00) NOTE: The medical equipment requires special precautions concerning electromagnetic compatibility, and must be installed and used with respect to the information in accompanying documents.
ATTENTION Mounting of the digital scale must be performed by qualified personnel, and before use it must be checked that the different parts are mounted correctly, such that the person is weighed in a safe and sensible manner. Incorrect mounting may cause risks of falls and involve serious consequences for the person.
6
© Guldmann GB-02/2022 • # 560642_101.01
• • • • • • • •
After mounting of the digital scale and before weighing of the person, the personnel must ensure that: • The weighing will be performed over “soft” items such as for example a mattress, pillows, bed or other soft elements, in order to minimise the consequences of a possible accident • The model RS digital scale and the sling are in the VERTICAL POSITION • The personnel must ALWAYS follow and keep their focus on the sling and the person during the lift in order to avoid swinging and possible movements of the digital scale • The ceiling hoist or the mobile lifter with digital scale and the person in the sling must not be moving, relocated or turned during the weighing • Once the weighing has been concluded, the sling must be taken off the person and the digital scale dismounted Swinging and turning of the scale may cause incorrect weighing. The personnel must therefore ensure that the person remains calm during the weighing.
Avoid swinging and sudden movements of the digital scale, which may damage the scale and result in incorrect weighing.
• • • • • •
© Guldmann GB-02/2022 • # 560642_101.01
•
Always follow the applicable rules with the use of electrical components that are subject to increased safety requirements. Incorrect installation will invalidate the warranty. The digital scale is intended for indoor use. Comply with the permissible room temperatures for operation. The digital scale complies with the requirements for electromagnetic compatibility. The maximum values specified in the applicable standards must not be exceeded. Avoid overloading the digital scale. Overloading will be indicated by an audible signal.
7
Use 1 2
3 4
There are many different mounting possibilities with a digital scale, depending upon the situation in which it is being used. The most common manners of mounting the digital scale are described below. A number of the mounting combinations in the configuration table at the end of this chapter are illustrated. 1. Hook from ceiling hoist or mobile lifter 2. Quick release button on the digital scale’s top 3. Quick release button on the digital scale’s bottom 4. Adapter for mounting of lifting hanger 1. A. Mounting of the digital scale on a ceiling hoist GH1, GH3, GHZ or a mobile lifter GL5, GL5.1 Fasten the digital scale to the hook on the ceiling hoist or the mobile lifter by pressing on the quick release button (2) and fastening the hook (1). B. Mounting of the digital scale on a ceiling hoist GH2 or a ”Universal” oval hook Fasten the digital scale to the hook on the ceiling hoist by opening the GH2 snap hook (1) or the oval hook and fasten the hook on the quick release button (2). C. Mounting of the digital scale on a mobile lifter GL Fasten the digital scale to the mobile lifter using an adapter fitted on the mobile lifter’s bar. Fasten the digital scale to the adapter (see page 11) on the mobile lifter by pressing on the quick release button (2) and fastening the adapter.
8
© Guldmann GB-02/2022 • # 560642_101.01
2.02
2. Mounting of adapter hook from lifting hanger to the digital scale Press on the quick release button (3) and fasten the adapter hook (4). Fasten the lifting hanger by inserting the adapter hook (4) on the lifting hanger. 3. Attach the desired sling on the lifting hanger and reset the digital scale (see section 4.00). The display will subsequently show 0.0. 4. Remove the sling from the hanger. Place the sling on the person and attach it on the lifting hanger again. 5. Carefully lift the person. When the sling with the person is at rest and freely suspended, the relevant weight will be shown on the display. Important Always reset the scale before weighing takes place. Note: The power saving function will automatically shut off the display after 30 sec. if the scale has no load. Can be changed (see section 7.00). 6. Carefully lower the person after the scale has been read.
© Guldmann GB-02/2022 • # 560642_101.01
Warning This scale has been produced in order to weigh the person in connection with him/her being relocated, for example between a bed and a chair. In connection with reduced lifting height, it may be necessary to divide the process in 2, weighing of the person and subsequent detachment of the scale, and then performing the final relocation, for example between a bed and chair. The scale has not been produced for use in connection with transport of the person. In the event of transport, the scale must be detached.
9
2.03
Configuration table
Configuration table for Digital scale Lifting capacity, max.
300 kg
300 kg
Length
200 mm
200 mm
Product
Ceiling hoist GH1 / GH3 movable
Ceiling hoist GH2
NA
NA
Lifter interface
Adapter interface top Digital scale RS
Adapter for mounting of lifting hanger
10
© Guldmann GB-02/2022 • # 560642_101.01
Hanger interface
Configuration table for Digital scale Lifting capacity, max.
300 kg
Length
200 mm
260 mm
365 mm
Product
Mobile lifter GL5 / GL5.1
Mobile lifter GL
Universal
Lifter interface
Adapter interface top
NA
NA
Digital scale RS
Adapter for mounting of lifting hanger
© Guldmann GB-02/2022 • # 560642_101.01
Hanger interface
NA
11
2.04
2.05
Accessories Item description
Item No.
Installation length
Oval hook
561628
107.5 mm
GH3 Adapter for Mobile lifter
552532
60 mm
Adapter for mounting of lifting hanger
558898
45 mm
Illustration
EMC - Information
Manufacturer's instructions and declaration - Electromagnetic emissions
Emission test
Conformity
Guidelines for Electromagnetic environment guidance
RF emissions CISPR11
Group 1 Class B
The model RS digital scale uses only RF energy for its internal functionality. Hence these RF emissions are very low, and there is very little probability that they would cause any interference in nearby electronic equipment.
RF emissions CISPR11
Group 1 Class B
Harmonic emissions IEC 61000-3-2
Class A
The model RS digital scale is suitable for use in all healthcare and hospital systems that are connected to the public low voltage distribution grid.
Voltage fluctuations / voltage emissions IEC 61000-3-2
Conformity
12
© Guldmann GB-02/2022 • # 560642_101.01
The model RS electronic digital scale is intended for use in the electromagnetic environments, such as are specified below. The customer or the user must ensure that the scale is used in such an environment.
Manufacturer's instructions and declaration - Electromagnetic immunity
© Guldmann GB-02/2022 • # 560642_101.01
The model RS electronic digital scale is designed to function in electromagnetic environments that are specified below. The customer or the user must ensure that the scale is used in this environment. Immunity test
Conformity
Guidelines for electromagnetic environment
Electrostatic discharge (ESD) IEC/NEKEN61000 - 4 -2
6 kV contact 8 kV air
The floors should be of wood, concrete or ceramic material. If the floors are covered by synthetic material, the relative humidity must be at least 30%.
Electrical rapid transients/burst NEK-IEC/ EN61000 - 4 - 4
+/-2 kV power supply +/-1 kV to inbound/ outbound lines
The power supply quality must be the same as in an ordinary commercial or hospital environment.
Surge IEC/NEKEN61000 - 4 - 5
+/-2 kV differential mode +/-1kV common mode
The power supply quality must be the same as in an ordinary commercial or hospital environment.
Voltage dips, brief interruptions and voltage variation IEC/NEKEN61000 - 4 - 11
<5 % UT in 0.5 cycles 40 % UT in 05 cycles 70 % UT in 25 cycles <5 % UT in 5 sec. Note= UT is the value of the power supply voltage.
The power supply quality must be the same as in an ordinary commercial or hospital environment.
Magnetic field at power frequencies IEC/NEKEN61000 - 4 - 8
3 A/m
The product power frequency magnetic fields should be at levels af a typical location in a typical commercial or hospital environment.
13
Manufacturer's instructions and declaration - Electromagnetic immunity
Immunity test
Conformity
Guidelines for electromagnetic environment
Immunity against conductor-bound disruptions IEC/ NEK-EN61000 -4-6
3 Vrms 150 kHz to 80 MHz (for devices that are not life-supporting)
Immunity radiated IEC/NEKEN61000 - 4 - 3
3 V/m 80 MHz to 2.5 GHz (for devices that are not life-supporting)
Portable and mobile radio frequency-based communications equipment must not be used closer to the device, including conductors, than the recommended distance calculated based on the applicable formula for the transmitter's frequency. Recommended distance between devices d = 1.2√P d = 1.2 √P From 80 MHz to 800 MHz d = 2.3√P From 800 MHz to 2.5 GHz, where P is the maximum rated output power for the transmitter in watts (W) as per the manufacturer of the transmitter, and d is the recommended, seperation distance in metres (m). The field strengths from fixed radio frequency transmitters, as measured in an electromagnetic examination of the location, a ought to be lower than the compliance level in each frequency range b. Interference may occur in the area around devices, which is marked with the following symbol:
At 80 MHz and 800 MHz the higher frequency range applies. It is possible that these guidelines do not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects and people. a) Field strengths from fixed transmitters, such as base stations for radiotelephones (mobile/wireless) and land mobile radios, amateur radio equipment, AM and FM radio transmitters and TV broadcasting, cannot be estimated reliably theoretically. Consider performing an electromagnetic field survey in order to assess the electromagnetic environment that is caused by fixed radio frequency transmitters. If the measured field strength at the location where the device is being used exceeds the compliance level for radio frequency energy that will be used, then you should observe the device in order to check that its operation is normal. If abnormal operation is observed, further precautions may be necessary, for example turning or relocating the device. b) Over the frequency range from 150 kHz to 80 MHz the field strength should be less than 3 V/m.
14
© Guldmann GB-02/2022 • # 560642_101.01
The model RS electronic digital scale is intended for use in the electromagnetic environments as specified below. The customer or the user must ensure that the scale is used in such an enviroment.
Recommended distance between RS Digital scale and portable- and mobile radio frequency-based communications equipment The model RS digital scale is intended for use in an electromagnetic environment, where radiated RF disruptions are under control. The customer or user of device may assist in preventing electromagnetic disruptions by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, with respect to the communications equipment's maximum output power. Maximum rated output effect for transmitter W
Distance with respect to the transmitter's frequency m 150 kHz to 80 MHz d = 1.2 √P
80 MHz to 800 MHz d = 1.2 √P
800 MHz to 2.5 GHz d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
© Guldmann GB-02/2022 • # 560642_101.01
For transmitters with a rated maximum output power not mentioned above, the recommended distance in metres (m) can be estimated using the formula that applies for the transmitter frequency, where P is the transmitter's maximum rated output power in watts (W) according to the manufacturer of the transmitter. Note: At 80 MHz and 800 MHz the higher frequency range applies. It is possible that these guidelines do not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects and people.
15
3.00
Technical properties
Model
RS
Manufacturer
Wunder Sa.Bi. Srl- Trezzo Sull’Adda (MI), Italy
Capacity and division
Max. 300 kg, e=100 g
OIML approval
Class III
Measurement unit
Kg
Display
LCD 25 mm with 5 digits
Dimensions mm
(W) 120 x (L) 70 x (H) 160 mm
Function keys
ON/ZERO/OFF, HOLD, TARE
Power supply
6 alkaline batteries AAA
Operating temperature
0 °C / 40 °C Complies with Directive 2011/65/EU
Protection IP54 (Protects against dust and spray)
Control panel
Function keys On/Off/Zero/Tare Hold (locking of weight)
RS Digital Scale: HOLD = Indicator for locking of weight NET = net weight GROSS = gross weight kg = unit kg
3. HOLD
4. NET
5. GROSS
1. ON/OFF/ZERO/TARE: 1. Push the ON/OFF key to activate the digital scale, and hold the key down for 3 seconds to deactivate the scale. 2. Push the ZERO key in order to reset the digital scale (corresponds to approx. +/-2 % of the maximum capacity). 3. Push the TARE key to find the tare weight.
16
© Guldmann GB-02/2022 • # 560642_101.01
4.00
2. HOLD: Push this key in order to lock the result of the weighing during the weighing phase. Push the HOLD key again to deactivate the HOLD function. 3. NET: Net weight function 4. HOLD: Gross weight function 5.00
Weighing mode Please read the following important guidelines before you read the detailed instructions on how you should use the weighing functions: - You must always make sure that the display shows “Zero” before use. Press the ZERO key if this is not the case. - The professional medical digital scale is designed to perform weighing when the weight is stable. Only first at this point in time can a correct weight be read.
© Guldmann GB-02/2022 • # 560642_101.01
- A tilt of more than 3° of the digital scale will result in an incorrect reading - Swinging and turning of the scale may cause incorrect weighing. The personnel must therefore ensure that the person remains calm during the weighing. - Avoid tilting or rotating the digital scale horizontally
17
6.00
Settings and functions Adjustment of time for automatic shut-off: This function concerns a period of time the digital scale is not in operation. Once the amount of time set has passed, the digital scale will automatically be switched off. Time for automatic shut-off: Programmable 120 sec / 180 sec / 240 sec / 300 sec. Adjustment of auditory signal: This function concerns the choice of the setting for the auditory signal for ON (switched on) or OFF (switched off). This depends upon what the user prefers. Auditory signal: On/Off
Choice / change of automatic shut-off and auditory signal
Press “Hold” for 3 seconds.
Note: “Hold” means “select”
Display showes “Set”.
On/Off/Zero/Tare means “enter”
o to another selecg tion horizontally. o to another g selection vertical.
Setting for time for automatic shut-off -120/180/240/300/off (seconds) Auditory signal ON (beep on) / OFF (beep off) Save the settings
18
© Guldmann GB-02/2022 • # 560642_101.01
7.00
8.00
Settings for the digital scale step-by-step Example: Setting with AUTO-OFF (auto-shut-off) after 180 seconds and with disconnected auditory signal. Step 1. Press HOLD for 3 seconds in order to go to the settings status SETUP. Step 2. Press ZERO in order to go to the status A.OFF, when A.OFF is shown. Step 3. Press HOLD in order to select 180 S = time for automatic shut-off is 180 sec. Step 4. Press ZERO in order to select and go back to the A.OFF status, and press HOLD in order to switch to the setting status BEEP. Step 5. Press ZERO in order to go to the setting status BEEP, and press HOLD again in order to choose between ON and OFF. Step 6. Press ZERO in order to select and go back to the setting status BEEP. Step 7. Press HOLD in order to select END, and press ZERO in order to end selecting the settings.
9.00
Replacement of battery The model RS digital scale uses 6 AAA alkaline batteries. Read the instructions below before you use the digital scale. 1. The cover for the battery case is located on the back side
© Guldmann GB-02/2022 • # 560642_101.01
2. Remove the battery case cover on the digital scale.
3. Take the battery holder out.
19
4. Insert 6 AAA alkaline batteries into the battery holder.
5. Insert the battery holder back into place again.
6. Insert the cover on the battery case.
Error messages 1. Low battery level This message indicated that the voltage in the batteries is too low for the digital scale to be able to function. Replace the batteries 2. High Zero The load is over the limit when the unit is switched on. Reduce the load. 3. Low zero The load is under the limit when the unit is switched on. Increase the load. 4. Overload or counter error The load is over the permitted limit, reduce the load and try again. Contact a Guldmann service center if the problem persists. 5. Memory error The programs in the scale contain errors. We ask that you contact a Guldmann service center if the problem persists.
20
© Guldmann GB-02/2022 • # 560642_101.01
10.00
11.00
Maintenance and service We recommend that regular safety/service inspections is performed at least once a year. This check must be performed by personnel who are qualified to carry out the task. Contact your local Guldmann representative, who will be at your disposition if you need further information. In order to give the digital scale a better and longer life span it is recommended that regular cleaning be performed. Conduct regular (at least once per year) functional checks on the model RS digital scale in the following manner: • • • • •
Checks of mechanical parts, hooks, bolts, screws, etc. Functionality tests of keyboard Check of the ABS enclosure Check of the batteries Scale control (with respect to metrological standards DL N.517 and DM N.182).
Clean the digital scale with a soft cloth, using warm water and a mild soap solution. Avoid the use of solvents or abrasive substances. Do not use large quantities of water when cleaning the scale, since this may lead to damage to the scale’s electronic parts. Always shut the digital scale off before performing cleaning. In the event of long-term inactivity for the digital scale: Remove the batteries from the battery compartment, and cover the digital scale in order to keep it intact. Do not subject the digital scale to shocks or excessive mechanical stress during transport. Contact your local Guldmann representative if you have a need for repairs or assistance. The digital scale is sold type-approved with the first metrological check (sign with M). A later calibration is always necessary if one or more security seals have been damaged, or the display shows abnormal weight indications. In order to maintain the medical approval the digital scale MUST be calibrated/verified according to national regulatiory requirements at an accridited testing institute.
© Guldmann GB-02/2022 • # 560642_101.01
Storage If the digital scale is stored for an extended period of time, make sure that those parts that can be damaged are protected from any possible dust accumulation.