Gyrus ACMI, Inc
Electrosurgical Generator
Gyrus ACMI PlasmaKinetic SuperPulse Generator User Manual July 2010
User Manual
52 Pages
Preview
Page 1
PlasmaKinetic SuperPulse Generator USER MANUAL Manufactured For: Gyrus ACMI, Inc. 136 Turnpike Road Southborough MA 01772-2104 USA
Gyrus Medical Ltd. Fortran Road St Mellons Cardiff CF3 0LT United Kingdom
0344 Customer Service USA: Customer Service: 1-800-852-9361 Technical Service: 1-800-621-3739 www.gyrusacmi.com
Rx only-CAUTION: Federal Law (USA) restricts this device to sale by or on the order of a Physician.
Part Number: 144012-MA 2010-07
:744000
TABLE OF CONTENTS SECTION 1
PART 1A 1B 1C 1D 1E 1F 1G
2 2A 2B 2C 2D 2E 2F 2G 3 3A 3B 3C 3D 3E 4 4A 4B 4C 4D 4E 5 5A 5B 5C 6 6A 6B 6C 6D 6E 6F 7 8 8A 8B 8C 8D 9 10 11 12 13 14 15
PAGE INTRODUCTION Overview of the SuperPulse Generator Comparison with Conventional Electrosurgery Indications for Use Contraindications for Use System Description Instrument Description Connector Cable Description PATIENT AND OPERATING ROOM SAFETY General Servicing/Equipment Disposal Fire/Explosion Before Surgery During Surgery After Surgery EMC Classification INSTALLATION Responsibility of the Manufacturer Generator Power Requirements Grounding of the Generator Routine Maintenance of the Superpulse Generator Medical Electrical Systems GENERAL INFORMATION Gyrus ACMI SuperPulse Generator Indicators and Displays Output Mode Selection and Power Controls SuperPulse Generator Indicators, Set-Up and Malfunction Displays Changing the Display Language Enabling Additional Instruments via a PIN code BEFORE SURGERY Power up the Generator Select and Connect the Connector Cable Attach Gyrus ACMI PK Instrument to the PK Connector Cable DURING SURGERY Recommendations During Surgery PK Default Settings Activation: Output Selection and Audible Tones Changing Output Mode and Power Setting during Surgery Changing Instruments During Surgery Changing Accessories Between Procedures TECHNIQUES TO OPTIMIZE PERFORMANCE AFTER SURGERY Following Surgery Cleaning the Footswitch Cleaning the Generator Cleaning and Sterilizing the PK Connector Cable OPERATING ROOM TROUBLESHOOTING PERFORMANCE SPECIFICATIONS ERROR AND FAULT CODES EXPLANATION OF SYMBOLS PERIODIC EQUIPMENT SAFETY CHECKS EMC TABLES LIMITED WARRANTY
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
1-1 1-1 1-1 1-1 1-2 1-2 1-3 2-1 2-2 2-2 2-3 2-4 2-5 2-5 3-1 3-1 3-1 3-1 3-1 4-1 4-3 4-5 4-9 4-10 5-1 5-1 5-2 6-1 6-1 6-1 6-2 6-2 6-3 7-1
TABLE OF CONTENTS
8-1 8-1 8-1 8-2 9-1 10-1 11-1 12-1 13-1 14-1 15-1
INTRODUCTION
SECTION 1
Gyrus Medical Ltd, Gyrus Medical Inc and Gyrus International Ltd are referred to as Gyrus ACMI in this user manual. This user’s manual will familiarize you with the controls and output functions available from your SuperPulse Generator and instruct you on its proper use. 1A.
Overview of the SuperPulse Generator Bipolar electrosurgery is a familiar tool widely employed in surgery. Based on similar principles, the Gyrus ACMI PlasmaKinetic (PK) technology combined with Gyrus ACMI PK instruments provides more effective coagulation and cutting of tissues than when similar instruments are used with other electrosurgical generators. A key feature of the Gyrus ACMI SuperPulse Generator is the capability of the generator to identify the type of Gyrus ACMI PK instrument connected to it. The identification of the instrument causes the generator to operate in the PlasmaKinetic (PK) mode that selects a default output designed to produce the desired electrosurgical effect for that particular instrument. This feature provides additional convenience for the user of the device. The user can change this default to obtain a wider range of PlasmaKinetic outputs from the device.
1B.
Comparison with Conventional Electrosurgery Conventional bipolar electrosurgery outputs are rarely optimized to the performance characteristics of specific instruments. This can reduce the speed of clinical effect, increase the thermal margin around the application site and result in tissue sticking to the instrument. The following describes the features of the PlasmaKinetic Generator when used with Gyrus ACMI PK instruments that distinguishes it from conventional bipolar generators. •
Vapor Pulse Coagulation (VPC) VPC produces controlled coagulation of vascular pedicles using vapor-focused pulses of energy. VPC has been specifically tailored for delivery through Gyrus ACMI PK Instruments. Once tissue to be coagulated is grasped in a PK Instrument, the tissue is uniquely coagulated using the pulses of PlasmaKinetic energy compared to the continuous output employed in conventional bipolar generators. This feature provides controlled and repeatable outcomes under a variety of surgical situations.
•
PlasmaKinetic Tissue Cutting (PK) In the PK output mode, the energy is used to create an ionized plasma around one pole of a Gyrus ACMI PK instrument. The ‘active’ pole provides tissue cutting and a level of hemostasis dependent on the selected PK output. This mode utilizes the conductive properties of fluids, both those released from tissues during application as well as saline commonly used to irrigate the body cavity.
1C.
Indications for Use The SuperPulse Generator is a general surgical electrosurgical device intended for use with bipolar instruments used in open, endoscopic and laparoscopic surgical procedures involving the coagulation and cutting of soft tissue. The device is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
1D.
Contraindications for Use The use of this device is contraindicated in patients with the following conditions: Circumcision procedures Electrosurgery should not be used for circumcision. Patients with Pacemakers Use with caution in the presence of internal or external pacemakers. Interference from an electrosurgical current can cause a pacemaker to enter an asynchronous mode or can block the pacemaker effect entirely. For further information, consult the pacemaker manufacturer or hospital Cardiology Department.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 1-1
INTRODUCTION
SECTION 1
1E.
System Description The SuperPulse Generator (figure 1.1) is designed for use in surgery for the coagulation and cutting of soft tissue. A typical system setup would comprise of the following items: • • • •
Generator Footswitch Re-usable Connector Cable Gyrus ACMI PK surgical instruments and other bipolar surgical instruments.
Fig 1.1 1F.
Instrument Description The instruments below are designed and rated to be used with the Generator. For convenience and to improve safety during use, all the dedicated instruments have an internal classification code, which is interrogated by the Generator when the instrument is attached. Default settings and power set adjustment limits are then set appropriately for that particular instrument. Instruments are supplied sterile and are single use only.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 1-2
INTRODUCTION
SECTION 1 Instrument types: Model No Description PlasmaKinetic Molly Forceps
Connector Cable
3300PK 33 cm 3 mm Forceps
3 Pin
3301PK 33 cm 3 mm Forceps w/Trocar
3 Pin
3330PK 33 cm 5 mm Forceps
3 Pin
3345PK 45 cm 5mm Forceps
3 Pin
3000PK 33 cm 10mm Cutting Forceps
3 Pin
3001PK 33 cm 10mm Cutting Forceps w/cord
3 Pin
3005PK 33 cm Cutting Forceps
3 Pin
3006PK 33 cm Cutting Forceps w/cord
3 Pin
3045PK 45 cm Cutting Forceps
3 Pin
3804PK 33 cm LP Scissors
3 Pin
3844PK 45 cm LP Scissors
3 Pin
PlasmaKinetic Needle Electrode
3400PK 33 cm Needle Electrode
3 Pin
Lyons Dissecting Forceps
3700PK 33 cm Dissecting Forceps
3 Pin
3740PK 45 cm Dissecting Forceps
3 Pin
3600PK 33 cm Macro-Jaw Forceps
3 Pin
3640PK 45 cm Micro-Jaw Forceps
3 Pin
3601PK 33 cm Macro-Jaw Forceps
3 Pin
3641PK 45 cm Micro-Jaw Forceps
3 Pin
PlasmaKinetic L-Hook
3527PK 33 cm L-Hook
3 Pin
PK SEALOpen Forcep, Curved
3103PK 25 cm (9¾”) Seal Open Forcep, Curved 3 Pin
PK SEALOpen Forcep, Straight
3104PK 25 cm (9¾”) Seal Open Forcep, Straight 3 Pin
PK SEALOpen Forcep, Angle
3105PK 25 cm (9¾”) Seal Open Forcep, Angle
3 Pin
PlasmaSpatula
3220PK 24cm Plasma Spatula
5 Pin
PlasmaSpatula
3200PK 33cm Plasma Spatula
5 Pin
PlasmaSpatula
3240PK 45cm Plasma Spatula
5 Pin
PlasmaKinetic Cutting Forceps
PlasmaKinetic LP Scissors
PlasmaKinetic Forceps Macro-Jaw
PlasmaKinetic Forceps Micro-Jaw
®
PlasmaSEAL 1G.
Open Forcep, Curved
®
2103PK PlasmaSEAL Open Forcep, Curved
5 Pin
Connector Cable Description The connector cable is designed and rated for 20 uses with the Generator and Instruments. The Connector Cable is supplied non sterile and should be sterilized before use in accordance with its instructions.
Cable types: Model No
Description
PlasmaKinetic Single Function
3900
3 Pin
PlasmaKinetic Dual Function
3905
5 Pin
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 1-3
SECTION 2
PATIENT AND OPERATING ROOM SAFETY
The safe and effective use of electrosurgery depends to a large degree upon factors and variables solely under the control of the operator. There is no substitute for good surgical technique and properly trained operating room staff. It is important that the operating instructions supplied with this or any electrosurgical equipment be read, understood, and followed. Electrosurgery has been employed safely in numerous procedures. Before starting any surgical procedure, the physician should be familiar with the medical literature, complications and hazards of electrosurgery in that procedure. 2A.
General WARNING Hazardous Electrical Output: This equipment is for use only by qualified personnel. WARNING Do not use a monopolar generator/accessories simultaneously with the SP generator. Activation of a monopolar generator/accessories may cause interference with the SP generator resulting in user message changes on the display. Before proceeding with surgery, confirm proper power settings are displayed on the generator. Ensure the appropriate output setting is enabled for the desired surgical outcome. WARNING Direct contact between activated monopolar accessories and SP generator connected accessories could damage the SP generator. If such damage is suspected, the SP generator should be returned to Gyrus ACMI for inspection. WARNING Use with caution in the presence of internal or external pacemakers. Interference from an electrosurgical current can cause a pacemaker to enter an asynchronous mode or can block the pacemaker effect entirely. For further information, consult the pacemaker manufacturer or hospital Cardiology Department. WARNING Do not use electrosurgical equipment unless properly trained in its use in the specific procedure intended. WARNING Electrodes and probes used with monitoring, stimulation, and imaging devices (or similar equipment) can provide a path for high frequency current even if they are isolated. To reduce the risk of an inadvertent burn at the electrode site, place the electrode and / or probe as far away as possible from the electrosurgical site. CAUTION
If two accessories are connected to the SP generator, ensure the appropriate accessory is selected prior to activation. Activation of the unintended accessory could cause unintentional tissue effect.
CAUTION
Do not activate electrodes while in contact with other instruments as unintended tissue effect may occur.
CAUTION
Do not activate the generator in an open circuit condition. To reduce the risk of unintended effects, activate the generator only when the active accessory is near or touching the target tissue.
CAUTION
Use the lowest appropriate power setting to achieve the desired effect.
CAUTION
This equipment is capable of producing a physiological effect.
CAUTION
Read the instructions, cautions, and warnings provided with Gyrus ACMI PlasmaKinetic System accessories before use. This device is an integral system; only use Gyrus ACMI approved accessories with the Gyrus ACMI Superpulse Generator.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-1
PATIENT AND OPERATING ROOM SAFETY
SECTION 2 CAUTION
If possible, avoid the use of needle style instruments for any physiological monitoring equipment that may be connected to the patient during electrosurgery.
CAUTION
Where practical, only use monitoring equipment that incorporates high frequency current limiting devices during electrosurgical procedures.
CAUTION
The connector cable should be positioned so that it avoids contact with the patient and any other leads.
CAUTION
Studies have shown that electrosurgical smoke generated during electrosurgical procedures can be potentially harmful to surgical personnel.
CAUTION
Examine all accessories and connections to the electrosurgical generator before use. Improper connection may result in arcs and sparks, accessory malfunction, or unintended surgical effects.
CAUTION
Do not insert fingers or objects other than the correct cables into the socket. Only activate the footswitch with an instrument attached.
WARNING The PK or SP system has not been cleared for tubal sterilization. Do not use this system for these procedures. 2B.
2C.
Servicing/Equipment Disposal CAUTION
Electrical Shock Hazard: Do not tamper with the generator housing or attempt to remove the control panel. Refer to authorized personnel for service.
NOTE
1. 2.
CAUTION
The generator contains electronic printed circuit assemblies. At the end of the useful life of the equipment, it should be disposed of in accordance with any applicable policies relating to obsolete electronic equipment.
CAUTION
Dispose of any system accessories according to normal institution practice relating to disposal of biologically contaminated items.
There are no user serviceable parts within the product. For maintenance of the generator, refer to the recommended periodic equipment safety checks in Section 13.
Fire/Explosion DANGER
Explosion Hazard: Do not use in the presence of flammable anesthetics.
WARNING
Explosion Hazard: The following substances will contribute to increased fire and explosion hazards in the operating room:
Flammable substances (such as alcohol based skin prepping agents and tinctures)
Flammable agents used for cleaning or disinfecting, or as solvents of adhesives should be allowed to evaporate before the application of electrosurgery. There is a risk of pooling of flammable solutions under the patient or in body cavities during endoscopic surgery. Any fluid pooled in these areas should be mopped up before electrosurgery is used. Endogenous gases.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-2
PATIENT AND OPERATING ROOM SAFETY
SECTION 2
Flammable anesthetics or oxidizing gases such as nitrous oxide (N2O) and oxygen enriched atmospheres. Some materials, for example cotton, wool and gauze, when saturated with oxygen may be ignited by sparks produced in normal use of electrosurgical equipment.
The sparking and heating associated with electrosurgery can provide an ignition source. Observe fire precautions at all times. WARNING Fire/Explosion Hazard: Verify that all oxygen circuit connections are leak free before and during use of electrosurgery. When using electrosurgery in the same room with any of the above substances or gases, prevent their accumulation or pooling under surgical drapes, or within the area where electrosurgery is being performed. 2D.
Before Surgery Active Accessories WARNING Electric Shock Hazard: Do not connect wet accessories to the generator. WARNING Electric Shock Hazard: Ensure that all accessories are correctly connected and that no metal is exposed. WARNING Do not attempt to re-use instruments labeled for Single Use Only. Heat or chemical Sterilization may render the instrument mechanically or electrically unsafe CAUTION
Read the instructions, warnings and cautions provided with the instrument accessories before using.
CAUTION
Accessories labeled as re-usable must only be processed according to the recommended procedure and, where appropriate, recycled the specified number of times.
CAUTION
Use default power levels to test an accessory.
CAUTION
Always inspect the system accessories for damage prior to use. In particular, check the cables of any re-usable accessory for possible insulation damage.
CAUTION
Use only Gyrus ACMI approved accessories supplied for use with this product. Product damage or accessory failure may otherwise result during use.
SuperPulse Generator WARNING Electric Shock Hazard; Connect the generator power cord to a properly grounded receptacle. Do not use power plug adapters. WARNING Fire Hazard; Do not use extension cords. CAUTION
Provide as much distance as possible between the generator and other electronic equipment (such as monitors) as an activated generator may cause interference with them.
CAUTION
Non-function of the generator may cause interruption of surgery. Ensure that all installation procedures are followed and that all connectors are correctly inserted before use. A backup generator should be available for use.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-3
PATIENT AND OPERATING ROOM SAFETY
SECTION 2
2E.
CAUTION
Do not stack equipment on top of the generator or place the generator on top of electrical equipment.
CAUTION
Do not set the activation tone down to an inaudible level. The activation tone alerts surgical personnel when an accessory is active.
During Surgery Contact With Metal Objects WARNING Use extreme caution when using electrosurgery in close proximity to or in direct contact with any metal objects. Do not activate in contact with another metal object. Localized heating of the instrument and the adjacent metal object may result in product damage or inadvertent injury. WARNING While using electrosurgery during a surgical procedure, the patient should not be allowed to come into direct contact with grounded metal objects (e.g., surgical table frame, instrument table, etc.). If this is not possible, use extreme caution to maximize patient safety. The use of antistatic sheeting is recommended for this purpose SuperPulse Generator Power Settings WARNING Do not simultaneously activate the SuperPulse Generator whilst activating with any other electrosurgical equipment (on the same patient). Failure to observe this may result in the attached instrument being unrecognized by the system. CAUTION
Upon reconnection of an instrument to the SuperPulse Generator, or after navigation using the Mode / Menu button, the power settings for cutting and coagulation may be changed from previously selected values.
WARNING Confirm proper power settings are displayed on the generator before proceeding with surgery. Ensure the appropriate output setting is enabled for the desired surgical outcome before activating the instrument and ensure that activation is for the minimum time to achieve the desired surgical effect. CAUTION
Failure of the HF SURGICAL EQUIPMENT could result in an unintended increase or decrease in output power.
CAUTION
Use caution when overriding the default power settings.
CAUTION
Should a power supply interruption occur, the generator will revert to its Standby state. The user should press the standby/on button to restart the generator and then press the Mode / Menu button on the front panel to accept the default instrument settings.
Instrument Accessories WARNING When not in use, place instruments in a clean, dry, non-conductive, and highly visible area not in contact with the patient. Inadvertent activation while in contact with the patient may result in burns. WARNING Do not wrap accessory cords around metal objects. This may induce currents that could lead to injury to the patient or surgical personnel. WARNING Fire Hazard: Do not place active accessories near or in contact with flammable materials (such as gauze or surgical drapes). Electrosurgical accessories, which are activated or hot from use, can cause a fire. Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-4
PATIENT AND OPERATING ROOM SAFETY
SECTION 2
Endoscopic Procedures WARNING As visualization may be impaired during endoscopy for a number of reasons, be particularly alert to these potential hazards:
2F.
Ensure the tip of the instrument is visible before activation.
Inadvertent activation or movement of activated electrodes outside of the field of vision may result in injury to the patient.
Localized burns to the patient or physician may result from electrical currents carried through conductive objects. Electrical current may be generated in conductive objects by direct contact with the active instrument, or by the active or return instrument being in close proximity to the conductive object while activated.
Carefully insert and withdraw instruments from trocars to avoid the possibility of damage to the devices and/or injury to the patient.
The instrument tip may remain hot enough to cause burns after the electrosurgical current is deactivated.
After Surgery WARNING Electric Shock Hazard. Always unplug the generator before cleaning. CAUTION Do not reuse or resterilize accessories labeled “disposable” or “single use only.”
2G
EMC Classification The SuperPulse System has been manufactured and tested to the following requirements: Group 2 Class A as per IEC60601-1-2 (2001) EMC PRECAUTIONS Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information in this document. EMC WARNINGS •
The generator should not be used adjacent to or stacked with other electrical equipment. If adjacent or stacked use is necessary both the generator and other equipment should be observed to verify normal operation in the configuration in which it will be used.
•
The EMC classification of the SuperPulse system (class A) is suitable for use on dedicated supply systems not connected to the public mains network, such as hospitals. NOTE:
Although class A limits have been derived for industrial and commercial establishments, administrations may allow, with whatever additional measures necessary, the installation and use of class A ISM equipment in a domestic establishment or establishment connected directly to domestic electricity power supplies.
•
Portable and mobile RF communications equipment can affect medical electrical equipment.
•
The use of accessories and cables other than those for which the system was designed can significantly degrade emissions and immunity performance.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-5
SECTION 2
PATIENT AND OPERATING ROOM SAFETY
•
Keep the accessory cables away from cables from other electrical equipment. Electrical currents may be induced in the other equipment causing unintended effects.
•
Do not use a monopolar generator/accessories simultaneously with the SuperPulse generator. Activation of a monopolar generator/accessories may cause interference with the SuperPulse generator resulting in user message changes on the display. Before proceeding with surgery, confirm proper power settings are displayed on the generator. Ensure the appropriate output setting is enabled for the desired surgical outcome.
•
Provide as much separation as possible between the generator and other electronic equipment (such as monitors). When activating the generator, unintended electromagnetic coupling may cause interference with the other equipment.
•
Should any unintentional effects appear upon other equipment when using the generator, repositioning the generator, the connecting leads or other equipment may alleviate the problem. It may also help to use different mains supply sockets for any affected equipment
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 2-6
SECTION 3
INSTALLATION
The electrosurgical generator described in this manual, in conjunction with the Gyrus ACMI PK Instruments, is designed to provide advanced electrosurgical effects during endoscopic surgery. 3A.
Responsibility of the Manufacturer The manufacturer is responsible for safety, reliability and performance of the equipment only if:
Installation procedures in this manual are followed.
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by the manufacturer and the electrical installation of the relevant operating room complies with local codes and regulatory requirements.
The equipment is used in accordance with this User Manual and the Instructions For Use that accompany all system accessories. 3B.
Generator Power Requirements Please refer to section 10-1 for full voltage details. Check the Generator Power Connection The power connector meets all requirements for safe grounding. Its purpose should not be defeated by using extension cords or any form of adaptor. When disconnecting from the mains socket or from the generator, cords should always be grasped by the plug. Do not pull on the cord itself.
3C.
Grounding of the Generator To ensure user safety, the generator must be properly grounded through the inlet plug and power cord. Use only hospital grade power cords. IMPORTANT
Ensure that the electrical installation of the relevant room complies with local codes and regulatory requirements.
3D. Routine Maintenance of the Superpulse Generator It is recommended that the generator be inspected by qualified service personnel in accordance with Section 13, Periodic Equipment Safety Checks. 3E. Medical Electrical Systems When the Superpulse generator forms part of a medical electrical system (as defined in EN 60601-1 2.201) or is used with an endoscope that is compliant with EN 60601 2-18 the following applies:
The Superpulse generator is to be placed outside the patient environment.
The Superpulse generator footswitch accessory is to be placed outside the sterile field.
The instruments intended for use with the Superpulse generator are suitable for use inside the patient environment. WARNING Multiple portable socket outlets shall not be placed on the floor. WARNING
Additional multiple portable socket outlets or extension cords shall NOT be connected to the system.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 3-1
SECTION 4 4A.
GENERAL INFORMATION
Gyrus ACMI SuperPulse Generator Indicators and Displays
Red warning lamp
Gyrus Medical Ltd Superpulse System (c) 2003 Version V2.22
PlasmaKinetic and ThermoKinetic Up/Down
Mode / Menu Button
Standby / On Button
Standby / On indicator
Active instrument indicator Connector Cable sockets
Vapor Pulse and Des Up/Down
Fig 4.1 Keypad: Standby/On, Up, Down Arrows and Mode / Menu Button Standby / On The Standby/On button switches the Generator back and forth between the Standby and Idle / Ready states. The green indicator will change from flashing to continuous when the equipment state changes from Standby to Idle / Ready by pressing the button. To place the generator into Standby press the standby button. When prompted press again to confirm entry to Standby is required. Following an error condition the generator may be reset by pressing the Standby/On button twice. Up/Down Arrows Depressing the up or down arrow when parameter change is permitted increases or decreases the parameter step-wise. Holding the button down will increase or decrease the value in preset steps Mode / Menu This button provides access into the waveform selection and setup menus. Repeated short presses will give access to the frequently used functions, listed below:
Cut waveform selection (PK) Coagulation waveform selection (VP/DES) Volume
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-1
SECTION 4
GENERAL INFORMATION
A long press will give access to the setup menu, giving access to the following functions below with repeated short presses: NOTE
Display intensity Key click on/off Select language Enter PIN Code If there is no user activity for a short period, the generator will exit the menu and return to the previous state.
When a PK connector cable is attached, the symbol below appears on the display.
3 Way Cable attached Insert Device
Fig 4.2 Screen for one 3-Way PK cable installed on the selected socket. Output Displays for PK Instruments The display is split into two halves; the upper portion of the display is used to indicate the type of instrument active, that is the instrument that will provide an output when the Cut or Coag pedal is pressed. The lower half of the display indicates the output waveform type and power selected. The left lower portion displays Plasmakinetic (PK) mode selection PK1, PK2, PK3, and ThermoKinetic mode selection T1 or T2, with the default power setting from 10 to 200 dependent on the type of Gyrus ACMI PK Instrument attached. The right portion displays Vapor Pulse Coagulation (VPC) mode selection VP1, VP2, VP3 and standard desiccate (DES), with the default power setting from 10 to 120 dependent on the type of Gyrus ACMI PK instrument attached. The VPC mode is only available with Gyrus ACMI PK instruments. The appropriate display will flash and an audible alarm will sound when an output is activated.
Needle / Hook PK1 40
VP1 20
Fig 4.3 Screen for Needle or Hook selected NOTE
The PK output is not available for some PK instruments. The lower left hand portion of the display remains blank in this case.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-2
SECTION 4
4B.
GENERAL INFORMATION
Output Mode Selection and Power Controls
When using a Gyrus ACMI PK instrument, the generator default power, and waveforms for that instrument will be selected. Power Up/Down - These buttons adjust the power setting; the yellow arrow buttons for the PlasmaKinetic (PK) and ThermoKinetic (T) outputs, and the blue arrow buttons for the Vapor Pulse Coagulation (VPC) and Desiccate (DES) output. Press the appropriate button once for a power increment or decrement. Holding down the button accelerates the rate of change of setting. NOTE
Power can only be adjusted once an instrument is properly connected to the generator.
PlasmaKinetic PK3 Cut PK2 PK1
High impedance tissue (fatty, vascular tissue) Medium impedance tissue Low impedance tissue (thin tissue) Voltage increasing
ThermoKinetic Cut
T2 T1
High impedance tissue (fatty, vascular tissue) Low impedance tissue (thin tissue) Voltage increasing
Coagulation
DES VP3 VP2 VP1
General Purpose, Non tissue-specific desiccation High impedance tissue (fatty, vascular tissue) Medium impedance tissue Low impedance tissue (thin tissue) Voltage increasing
NOTE
Output mode selection can only be performed with an instrument and connector cable attached to the generator. The range of modes available will depend on the type of Gyrus ACMI PK instrument being used.
NOTE
If the Mode / Menu button on the front panel is quickly pressed and released the generator enters a configuration state, pressing and holding the Mode / Menu button at any time exits this state.
Mode Selection using Gyrus ACMI PK Instruments - There are three PlasmaKinetic Modes shown as PK1, PK2 and PK3 that produce a tissue effect increasing from PK1 to PK3 (PK3 delivers power more effectively to higher impedance tissue than PK1). Two ThermoKinetic outputs are also available, shown as T1 and T2, which continuously switch between a PlasmaKinetic and Desiccate output during activation. A ThermoKinetic output will provide a greater degree of hemostasis during tissue cutting compared to a PK setting. Three VPC mode levels and the desiccate (DES) are available. VP1 is optimal for low impedance tissue, VP2 for intermediate impedance tissue and VP3 for higher impedance tissue or larger diameter instruments. The frequency of pulses in VPC mode will also vary depending on the type of instrument attached: generally, the larger the tissue contact area, the slower the pulses. DES provides a conventional continuous bipolar output. The modes associated with the blue and yellow pedals can be individually assigned. To change to desired mode press the Mode / Menu button until it is seen on the display (yellow pedal mode appear on the LHS and blue pedal mode on the RHS). Using the UP/DOWN buttons as indicated on the screen, will cycle through the available modes for the respective pedal (see Fig 4.4 Screen for coag waveform selection below). Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-3
SECTION 4
GENERAL INFORMATION
When the desired mode is displayed, a further press on the Mode / Menu button will complete the selection
Select
Mode
DES
Fig 4.4 Screen for coag waveform selection Socket Selection The generator permits two instruments to be fitted simultaneously, via the two 5-way connectors. Only one socket can deliver RF at any one time. When the generator is switched on from the standby state, it initially operates an automatic socket selection mechanism, and assists primary connection by activating whichever socket first has an instrument attached. The instrument connected to the active socket is indicated on the display. Thereafter socket selection can only be altered manually, by pressing the black mode footswitch, ensuring that the surgeon always has control over which instrument is activated. Black Mode Footswitch
Fig 4.5 Footswitch Footswitch - Blue Pedal The blue footswitch pedal is used to administer desiccate and VP output waveforms. The output will be present while the footswitch is held down. Footswitch – Yellow Pedal The yellow footswitch pedal is used to administer PK (cut) and ThermoKinetic waveforms. The output will be present while the footswitch is held down. Footswitch – Black Mode Footswitch The black mode footswitch enables a rapid change between PK instruments.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-4
SECTION 4
GENERAL INFORMATION
Generator Switch On/Off The mains power On/Off rocker switch is located on the top left of the rear panel (viewed from front). The Generator is switched on by pressing the side marked l. The generator will display the Serial number of the generator, then the internal tests are performed then the green LED below the Mode / Menu button will come on continuously, then flash after a short time. The generator is then Ready state dependent on whether an instrument has been fitted. The generator display will dim after a period of 30 minutes of not being used and will automatically enter Standby state if it is not used for a period of four hours. It is advisable to switch off the Generator whenever it is not in use for any extended period, by using the rear panel switch. The side marked “0” should be pressed to do this. To enable use of the generator the Standby / On button must be pressed and the generator will then enter the Idle state if no instrument is fitted, or the Ready state if an instrument is fitted. If an instrument is present at switch on then the user has to accept the default powers, by pressing the Mode / Menu button when prompted. 4C.
SuperPulse Generator Indicators, Set-up and Malfunction Displays Impedance Indicator Impedance is measured by the generator to provide assistance in determining the tissue effect endpoint, in addition to the visual and tactile feedback available to the user. The impedance of the tissue rises during the desiccation/coagulation process. Indication of this is provided via audible and visual signals to be used as a guide to determining when treatment is complete. The impedance indication is available as a bargraph with audible indication, or the indications can be removed if desired. The impedance display is in the form of a bar graph.
For the bargraph display thirteen diamonds are displayed initially with the number reducing as the impedance rises.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-5
SECTION 4
GENERAL INFORMATION
The audible tone that accompanies the visual display starts at a high pitch and falls as the impedance increases.
Activation Status Activate - When the generator is activated, the power display of the selected output flashes and an audible tone sounds. If neither occurs, then there is a malfunction. Activate Tones and Impedance Indicator Volume Adjustment Volume Adjustment - The activation tone volume can be adjusted between minimum and maximum using the up control of the desiccate (blue) power control. Depress and release the Mode / Menu button until the symbol “SELECT VOLUME” appears (Fig 4.6 Screen for Alarm volume selection). Press and release the Mode / Menu button once more to accept the setting.
Select
volume
1
10
Fig 4.6 Screen for Alarm volume selection System Failure Displays Non-critical (Soft) Faults Shorting For jawed or grasping Gyrus ACMI PK instruments, if an instrument short occurs an oscillating audible alert is sounded and the display shows the symbol “Regrasp”. The surgeon should release the instrument jaw and reposition the tissue.
T2 40
Regrasp
VP2 30
Fig 4.7 Screen for the selected PK electrode shorted
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-6
SECTION 4
GENERAL INFORMATION
For non-grasping Gyrus ACMI PK instruments, if a Generator “Power Limit” message persists it is indicative of two or more poles of the electrode tip being bridged by a short. This could be caused by contact with other metallic instruments present in the surgical field, or possibly even a build-up of eschar between poles at the electrode tip. If the condition does not resolve itself, the generator after a few seconds, interrupts RF activation with the “OUTPUT SHORTED, RE-APPLY PEDAL” symbol (Fig 4.10 Screen for shorting of a nongrasping electrode), until the pedal is released and pressed again. Power levels will remain the same as those previously in use. This method of resolution is appropriate where the cause of the short is accidental, transitory and understood.
OUTPUT SHORTED RE-APPLY PEDAL
Fig 4.8 Screen for shorting of a non-grasping electrode Storage at low temperatures – The generator will detect where the enclosure temperature is below the specified minimum allowable for use, and will display a ‘WARMING UP” message until an acceptable temperature is reached internally, whereupon normal operation will resume. This condition is possible where the generator has been brought directly into the operating room from a cold storage environment.
WARMING UP
Fig 4.9 Screen for an excessively cold condition Foot pedals stuck on – The generator will detect where one or more the foot pedals appears to be stuck on, and will wait until the condition disappears before resuming normal behaviour. This condition can be inadvertently caused by the inverting the footswitch or standing on a pedal as the generator is switched on.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-7
SECTION 4
GENERAL INFORMATION
Non-critical (Recoverable) Faults Fault - If a fault is detected during set-up or during use a fault code message is displayed as indicated by the “Fault Code XXX REF XX” symbol on the front panel display (Fig 4.10 Screen for recoverable fault display ). Refer to Section 9, Operating Room Troubleshooting.
BUTTON
STUCK
Fault code 400 Ref 15
Fig 4.10 Screen for recoverable fault display Critical (Non-recoverable) Errors Red Warning Symbol - Except during the self-test routine and a non-critical failure, illumination of the red warning symbol on the front panel when accompanied by the “Error Code XXX REF XX” symbol (Fig 4.11 Screen for critical error display) on the front panel display indicates a critical failure. In the case of a critical failure, DO NOT ATTEMPT TO USE THE UNIT. Please refer to Section 11 for information.
INTERNAL
FAILURE
Error code 200 Ref 25
Fig 4.11 Screen for critical error display SuperPulse Generator Connectors and Sockets The PK Single Function connector cable is connected to the generator through either of the two PK connector cable sockets on the front panel of the generator (Fig 4.1). The electrical power cord; footswitch cable and protective earth cable are connected to the generator through fittings on the rear panel (Fig 4.17). A five-way cable may be available to enable use of tri-polar instruments.
Gyrus ACMI PlasmaKinetic SuperPulse Generator Part Number: 144012-MA
USER MANUAL
Page 4-8