OCTOPUS 900 Service Manual ©HAAG-STREIT 06/2009 HS No. 7220054 Revision 02
HAAG-STREIT AG, Switzerland, Phone: +41 31 978 01 11, Fax: +41 31 978 02 82, [email protected] HAAG-STREIT DEUTSCHLAND GmbH, Germany, Phone: +49 4103 709 02, Fax: +49 4103 709 370, [email protected] HAAG-STREIT FRANCE, France, Phone +33 4 5009 0033, Fax +33 4 5009 7190, [email protected] HAAG-STREIT UK, United Kingdom, Phone +44 1279 414969, Fax +44 1279 635232, [email protected] HAAG-STREIT USA, INC., USA, Phone: +1 513 336 6858, Fax: +1 513 336 7828, [email protected]
OCTOPUS 900 Service Manual
HAAG-STREIT No. 7220054
TABLE OF CONTENTS 1
ITEMS SUPPLIED ...3
2
SAFETY INSTRUCTIONS...4 2.1 General...4 2.2 Ambient conditions ...4 2.3 Power Connection ...4 2.4 Instrument Transportation ...4 2.5 Special Notices in the Text...5 2.6 Symbols on Display ...5 3 INTRODUCTION ...6 3.1 OCTOPUS Perimeter 900...6 3.2 Octopus 900 control unit / PC ...11 4 WARRANTY AND PRODUCT LIABILITY ...12 4.1 Statutory requirements ...12 4.2 Type plate...13 4.3 Pictogram symbols ...13 4.4 Support stand ...14 4.5 Safe system configuration in accordance with IEC / EN 60601-1 ...15 4.6 Installation ...19 5 INSTALLATION AND USAGE OF EYESUITE PERIMETRY SOFTWARE ...20 5.1 Installation, Exchange, Upgrade ...20 5.2 Releasing Program Options ...21 6 PRINCIPLE OF OPERATION...22 6.1 Stimulus Intensity Control...22 6.2 Current Control...22 6.3 Background Intensity Control ...23 6.4 Reference Point...23 6.5 Optics Module ...24 7 MAINTENANCE & REPAIR ...25 7.1 Cleaning ...25 7.2 Housing ...26 7.3 Power supply unit (1803900)...28 7.4 Processor module (1804005) & Light controller board (1804205) ...30 7.5 Camera module (1803800) ...32 7.6 Stimulus Projector (1803400) & Projector–fixation target (1803600)...36 7.7 Brightness measurement p.c.b. - stimulus (1804421) ...44 7.8 Aperture-/filter-module (1803500) ...45 7.9 LED Board - stimulus (1804525) ...46 7.10 Refractive lens holder module (1803250) ...47 7.11 Mirror Unit (1803300) ...50 7.12 Headrest (1803200) ...51 7.13 Operating unit (1803170) ...52 7.14 IR illumination (1803550) ...53 7.15 Brightness measurement p.c.b. - background (1804421)...53 7.16 Background illumination (1803570)...54 8 SYSTEM MESSAGES...55 8.1 Messages (to be completed in revision 2)...55 8.2 Errors (to be completed in revision 2) ...55 8.3 Resetting the IP address ...59 9 APPENDIX...60 9.1 Technical specifications ...60 9.2 Instrument OCTOPUS 900 ...60 Page 2 / 53
HAAG-STREIT 08.06.09
1 ITEMS SUPPLIED We congratulate you on choosing the OCTOPUS 900 – 90° visual-field perimetry of the latest generation. Depending on your order, the items supplied are as follows: − OCTOPUS 900 − Accessory kit − optional support stand Please check the delivery with care 1 x OCTOPUS Perimeter 900 1803000 1 x accessory kit, consisting of: 1 x patient-response button 1802439 1 x Ethernet switch 100Mbps 1803090 2 x Ethernet cable 1803094 1 x dust cover 1803061 1 x instruction manual 7220031 1 x CD with all Octopus 900 and EyeSuite Perimetry manuals 7220030 1 x eye occluder set 1802349 2 x spare fuses (T3.15 A / 250 V) 3000481 1 x support stand, consisting of: 1 x stand base 1801187 1 x O-900 table top 1803051 1 x set of accessories for support stand 1801194 1 x Visual Field Digest 1800092 Please keep all packaging material, as this can be reused for possible returns or when relocating.
OCTOPUS 900 Service Manual
Page 3 / 61
2 SAFETY INSTRUCTIONS 2.1 General •
This instrument should only be used for the purpose as described in this manual.
•
Condensation can occur on the cold surfaces of the instrument if it has been transported from a cold environment to the environment in which it will be used. In such a case, one should wait a couple of hours before turning on the instrument. Once the temperature of the instrument is the same as the environment in which it has been placed and once it is dry you may proceed with powering on the instrument.
•
The examination of the patients, the operation of the instrument and the interpretation of the results can only be performed by persons who are experienced and who have been trained accordingly.
•
The instrument is to be installed on the height-adjustable table and employed in a dimly lit room in a medical area.
•
Keep the service manual there where persons who have been trained to repair the instrument can consult it at all times. Guarantee claims can only be valid when the instructions in the service and user manuals are followed.
•
The doctor or operator is to instruct the patient in the safety instructions and check that they are observed.
•
Make sure that the instrument is connected only to the power source as defined on the perimeter rating plate. The on/off switch does not separate the perimeter from the power lines. Before maintenance or cleaning work is performed, the power cord must be removed from the wall socket.
•
The instrument should not be used in an explosive area. Do not store any flammable and explosive substances (such as Alcohol, Petrol, Gas) near the instrument.
•
Removal of the housings and repairs to the instrument are permitted only by trained and authorized technicians. Considerable danger for the operator and patients can be caused by improper repairs.
•
The dust-cover should be removed before turning on the instrument. If this is not done the light sources may get damaged due to overheating. On the other hand make sure that the instrument has been switched off before covering it with the dust-cover.
•
Only original replacement parts and original accessories are permitted for repairs.
2.2 Ambient conditions •
Transportation
•
Storage
•
Working
Temperature Air pressure Relative Humidity Temperature Air pressure Relative Humidity Temperature Air pressure Relative Humidity
-40°C 500hPa 10% -10C 700hPa 10% +10°C 700hPa 20%
to to to to to to to to to
+70°C 1060hPa 95% +55°C 1060hPa 95% +40°C 1060hPa 75%
Environment • Electric and electronic equipment must be separated from house waste! This equipment has been introduced into the market after the 13th of August 2005. • Disposal via your local collecting point or your Haag-Streit distributor. • Thus it is ensured, that no harmful materials get into the environment and that valuable raw materials can be reused again. 2.3 Power Connection • Only use a conformed power cable. To reduce the risk of an electro shock, ensure that the power cable and plug is connected with an earth lead. • The cable, plug and ground connection on the wall socket must be 100% functional. 2.4 Instrument Transportation Transport the instrument over large distances in the original packing. For short distances the instrument can be lifted by placing the cupola towards ones chest, then place both hands behind the cupola (bottom section) and lift the instrument. See user’s manual section 4.5.2. Page 4 / 61
OCTOPUS 900 Service Manual
2.5 Special Notices in the Text Further safety instructions are put into the text as needed. They vary in importance and should help reduce risks for the patient and operator, avoid damage to the perimeter and insure optimal functioning.
!
ATTENTION
Instruction which must be followed to reduce risks to the patient and operator.
Warning
Instruction which must be followed to avoid damage to the perimeter.
Tip
Tip for insuring optimal functioning of the perimeter.
All OCTOPUS 900 functions are mentioned in this service manual. Certain functions are offered as options and there is a possibility that they are not installed/ supplied with the instrument. 2.6 Symbols on Display 2.6.1 Messages, Errors Error and warning messages
OCTOPUS 900 Service Manual
Page 5 / 61
3 INTRODUCTION 3.1 OCTOPUS Perimeter 900 The OCTOPUS 900 is an automatic projection perimeter for the examination of the whole field of sight (90°). The system is divided into an examination element (OCTOPUS 900) and a control unit (notebook computer or PC). The examination element communicates with the external PC via an Ethernet link. The OCTOPUS 900 is controlled using the software installed on the PC. The perimeter may be controlled, if required, from a fully-lit adjacent room. The integrated, automatic patient-fixation control function guarantees the reliability of diagnosis. The OCTOPUS 900 is suitable for both clinical and investigational purposes, as its flexibility is virtually unlimited. Thanks to its spherical, Goldmann-type cupola design, the OCTOPUS 900 tests the entire field of sight (up to 90° eccentricity). The flexibility of this appliance allows it to deal with all perimetry tasks, both in the 30° and 90° range, using kinetic, static or flicker perimetry. New PC and perimetry software can be downloaded and updated by going to www.octopus.ch.
Page 6 / 61
OCTOPUS 900 Service Manual
Components 1 2 3
4 5 6
7 8
1 2 3 4 5 6 7 8 9 10 11 12 13
Top cover for stimulus projector Front cover Housing / Cupola Mirror unit cover Forehead rest Display Keypanel Refractive lens holder with IR eye illumination and refractive lenses Chinrest Response button Response button connector Type plate Table height adjustment
9 10 11 12 13
Fig. 3-1
*
14 15
LCD display The high-contrast TFT colour display allows the video image to be viewed from an oblique angle. The following messages are displayed on-screen:
16
14 Indication of a '' during stimulus presentation. 15 Indication of an '', if the patient-response button is pressed. 16 The crosshairs help to centre up the eye, scale = 1mm intervals. 17 Display of left (OS) or right eye (OD) 18 Warning or error message
18 17
OD Fig. 3-2
OCTOPUS 900 Service Manual
Page 7 / 61
19 22
20
Control panel The control panel consists of an easy-to-use keypad made of soft but hard-wearing rubber material. All buttons are backlit with white light to facilitate their use in a darkened room. The backlighting can be disabled, if required, by adjusting the display brightness settings.
21
19 20 21 22
Turning the refractive lens holder in and out Display brightness settings Adjustment of chin support Start examination see Fig. 3-4
24 25 26
Connections
Fig. 3-3
27
23 24 25 26 27
Plug-in mains power unit Mains switch Holder for two T3.15AL/250V fuses Mains connection Ethernet connection
23 Fig. 3-4
DANGER
Page 8 / 61
All externally connected appliances and items of equipment must conform to the relevant safety standards.
OCTOPUS 900 Service Manual
3.1.1 Housing Optical and electronic components are protected from light and dust by five housing covers. They can be removed in just a few moments for the purpose of servicing. Once the four screws in the back panel have been removed, the panel, hood and both IR covers can be lifted out. The optical unit and electronic components of the OCTOPUS 900 are now accessible. Always disconnect the appliance from the mains power supply before opening it. The removal of housing components must only be carried out by suitably-trained specialists who are qualified and authorized to do so.
DANGER
3.1.2 Cupola The cupola of the OCTOPUS 900 has a diameter of 600mm, and thus conforms to the Goldmann standard. Test zones with eccentricity up the following levels can be measured: Nasal Temporal Superior Inferior
• • • •
60° 89° 60° 70°
3.1.3 Head rest A wide, ergonomically-designed forehead rest allows the patient to remain in a comfortable position while examination is taking place. 3.1.4 Chinrest with sensors The four buttons on the control panel are used to adjust the chin support, and thus the position of the head. Fine adjustment can also be carried out from the control unit (PC) using the computer mouse. Sensors in the chin rest detect if the patient’s head is correctly positioned. 3.1.5 Refractive lens holder module An automatic swivel arm allows the refractive lens holder to swing in an out during the examination, without having to alter the position of the patient. This swivel arm can be operated either from the control panel or on-screen with the computer mouse. Once the refractive lens holder has been swung in, it can be fine-adjusted to the correct distance with respect to the eye being examined.
!
CAUTION
Always use the control-panel buttons on the device itself or on-screen control from the PC to swivel the refractive lens holder in an out. Do not attempt to move the refractive lens holder manually.
3.1.6 Refractive lens holder Refractive lenses can be used during examinations within 30° eccentricity. The corresponding lenses are inserted before the test begins. The refractive lens holder can be tipped forward by about 25° to make it easier to load the refractive lenses.
Fig. 3-5
OCTOPUS 900 Service Manual
Fig. 3-6
Page 9 / 61
3.1.7 Patient-response button The patient-response button is connected to the underside of the head-support mounting (RJ11 jumper cable). 3.1.8 Mains connection The Ethernet connection socket is located at the back of the appliance. Always use a crossed Ethernet cable for a direct PC connection or a straight Ethernet cable for a connection through a network switch (shielded cable of category 5e), designed to transmit reliably at 100MHz. This network connection is electrically isolated, with electrical strength of 4kV, in conformity with IEC / EN 60601-1. 3.1.9 Light sources LEDs are fitted for the purposes of periphery or background lighting, fixation help and stimulus. LEDs do not emit any heat, so no active cooling system is required in this respect. 3.1.10 Light intensities The intensity of stimulus and periphery light is measured by independent light-sensors and compared to the preset default values every time the perimeter is switched on. 3.1.11 Stimulus The stimulus light is projected indirectly into the cupola via a mirror unit. The user-defined programs offer a selection of five different diaphragm diameters (Goldmann I - V). Stimulus intensity-damping is infinitely adjustable via the electronic control system. Stimulus presentations of 100-500ms are permitted. A mechanical lock und optical damping elements are no longer required. White stimulus for W/W perimetry and optional blue und red stimulus for B/Y and R/W are possible. Stimulus intensity is detected by a light sensor, which also serves as a reference point for the system of coordinates for test locations. The stimulus LED has a service life of >20,000h and is thus maintenance-free. 3.1.12 Periphery or background lighting White background brightness amounts to 31.4 or 4asb for W/W perimetry. You can also select a yellow background with 31.4asb for B/Y perimetry. Background brightness is based on two light sources, each fitted with several LEDs. The background LEDs have a service life of >20,000h and are thus maintenance-free. Background brightness is detected by a separate light sensor. 3.1.13 Fixation marks Three different fixation marks can be selected and their brightness infinitely adjusted by electronic means. The light source is a maintenance-free green LED with a service life of >20,000h.
centre point
Page 10 / 61
cross mark
circle
OCTOPUS 900 Service Manual
3.1.14 Fixation control The eye being examined is illuminated with IR LEDs, photographed by a CMOS camera and shown on the LCD display. The built-in, automatic fixation-control function guarantees the reliability of diagnosis. Precise positioning of the eye being examined is carried out by motorized fine-adjustment of the chin rest. 3.1.15 Diagnosis data All diagnosis data are transferred via the Ethernet connection to the control unit (PC / Laptop) for processing and saving to a database. The use of a second Ethernet connection (LAN card) allows these data to be exported to a server. Diagnosis data can also be printed out using a printer connected to the control unit. 3.2 Octopus 900 control unit / PC A standard PC can be used as a control unit for the Perimeter. The control-system software runs on WINDOWS XP. 3.2.1 Minimum PC requirements • • • • •
Pentium III / 1.7GHz 512 MB RAM 40GB hard disk CD-ROM / DVD T100 (100MHz) Ethernet interface
Power connector socket with fuse holder on the electronics box.
Fig. 3-7 3.2.2 OCTOPUS 900 Handle the instrument using both lower housing halves to lift it out of the packing. Two ribbed grips hinder sideways slipping.
OCTOPUS 900 Service Manual
Page 11 / 61
4 WARRANTY AND PRODUCT LIABILITY •
The appliance and its accessories are made of high-quality materials and manufactured to the latest technical standards. They left our factory in a perfect state of repair. If an incident should occur however, please contact your dealer.
•
The manufacturer supplies the appliance with a one-year warranty, counted from the date of purchase. The warranty covers malfunctions and defects attributable to faulty materials or design. The warranty does not cover malfunctions or defects arising from incorrect use or external factors.
•
Any attempt to carry out unauthorized repairs will immediately void all warranty liability.
•
Continued use of a product that has been damaged by incorrect operation can lead to personal injury. The manufacturer will accept no liability whatsoever in such cases.
•
Before unpacking the appliance, check the outside of the packaging for signs of incorrect handling and possible damage. If you do detect such damage, please notify the carrier who carried out delivery, and unpack the appliance in the presence of a representative from the transport firm concerned. You should then make out a report detailing any damage detected. The report should be signed by you and the carrier’s representative.
4.1 Statutory requirements •
The regulations are according to the valid ‘CE declaration of Conformity’ for this specific instrument.
•
A copy of the declaration of conformity for the product concerned can be requested at any time from Haag-Streit.
•
Statutory requirements regarding accident prevention should be respected.
EMC The OCTOPUS 900 meets the electromagnetic compatibility requirements of IEC / EN 60601-1-2. The appliance is designed to keep the generation and emission of electromagnetic interference to a level that does not affect the normal operation of other items of equipment. The appliance is likewise designed to provide a certain level of resistance to outside sources of electromagnetic interference.
Page 12 / 61
OCTOPUS 900 Service Manual
4.2 Type plate Read through the instruction manual with special care Manufacturer Serial number Appliance number Product classification type B Date of manufacture Do not push
2
3
41
56 7
4.3 Pictogram symbols Ground connection lead. Do not push. The appliance is liable to tip over if pushed from the side. This appliance conforms to European Directive 2002/95/EC (RoHS).
OCTOPUS 900 Service Manual
Page 13 / 61
4.4 Support stand An electrically adjustable support stand (available as an option) allows the appliance to be adjusted easily and smoothly to the individual patient. The support stand offers ample legroom and is wheelchair-compatible.
Support stand 1 Table top with mountings for patient-response button 2 Electrical connection box 3 Patient-response button 4 Electrically-actuated lifting column 5 Stand feet with castors
1
2 3 4 5
Fig. 4-1 6 7
Electrical voltage selector in electrical connection box Power socket for connection of the OCTOPUS 900
6 7
DANGER
Check that the mains voltage is correct before connecting to the power supply.
Fig. 4-2 Mains connection with fuse holder on electrical connection box.
Fig. 4-3 ON/OFF switch on electrical connection box. DANGER
Fig. 4-4
Page 14 / 61
Take special care, especially when dealing with patients in wheelchairs, to ensure that the table top does not come into contact with the patient’s legs.
OCTOPUS 900 Service Manual
4.5 Safe system configuration in accordance with IEC / EN 60601-1 4.5.1 OCTOPUS 900 standard configuration (single PC)
Always set a network switch between the Octopus 900 and the PC. This ensures correct data flow
OCTOPUS 900 Service Manual
Page 15 / 61
4.5.2 Octopus 900 in small networks (no dedicated server)
Never attach the Octopus 900.or its control PC to a hub or to a wall plug directly. Always use a local network switch for the connection.
Page 16 / 61
OCTOPUS 900 Service Manual
4.5.3 Octopus 900 in standard networks (with dedicated server)
Never attach the Octopus 900 or its control PC to a hub or to a wall plug directly. Always use a local network switch for the connection. OCTOPUS 900 Service Manual
Page 17 / 61
Observations 1 If the control unit (notebook computer or PC) and printer are farther than 1.5m from the OCTOPUS 900, there is no need for a safety isolating transformer conforming to IEC 60601-1 / EN 60601-1. If a medically-approved control unit or a control unit with network element approved for medical use is in operation without a printer or optional LAN connection, neither a safety isolating transformer nor a distance of > 1.5m from the OCTOPUS 900 are required. For safety reasons, you are however recommended to maintain the minimum distance of 1.5 m wherever possible. All other non-medical appliances and items of equipment must be operated via a safety isolating transformer.
Page 18 / 61
OCTOPUS 900 Service Manual
4.6 Installation 4.6.1 Support stand The support standard is supplied in a separate package. Assemble the support stand according to the instructions supplied with it, and ensure that you select the correct mains voltage before connecting the power supply lead. 4.6.2 OCTOPUS 900 Transporting and relocating the appliance (over short distances only) a) Stand in front of the device and grasp it by the cupola with both hands before lifting (Fig. 4-5). b) Stand to one side of the appliance, with one hand on the front cover and the other at the back. Then take a firm hold and lift (Fig. 4-5; Fig 4-6 & Fig. 4-7).
Fig. 4-5
Fig. 4-7
Fig. 4-6
Push the patient-response button into its connection socket. The connection socket for the response button is located at the bottom of the front cover. The retaining catch on the connection plug of the response button faces forwards CAUTION
! Fig. 4-8
The RJ11 plug is designed for connection of the patient-response button only. Do not use it with any other item.
Fig. 4-9 Front cover Cupola housing Connection plug with retaining catch
− Push in the connection plug until you hear the retaining catch click into place. To remove the response button, push the retaining catch towards the headrest and pull the cable downwards. − Connect the OCTOPUS 900 and PC with the Ethernet cable or integrate the Perimeter appliance into a LAN. You are recommended to use a direct connection, as variations due to network load can affect the diagnostic performance of the appliance. Further information can be found in section 4.4. − An electrically adjustable support stand (available as an option) allows the appliance to be adjusted easily and smoothly to the individual patient. The support stand offers legroom and is wheelchair-compatible.
OCTOPUS 900 Service Manual
Page 19 / 61
5 INSTALLATION AND USAGE OF EYESUITE PERIMETRY SOFTWARE 5.1 Installation, Exchange, Upgrade Instructions on how to use the appliance for diagnostic purposes and a description of the EyeSuiteTM user software are included in a separate manual.
Fig. 5-1
Page 20 / 61
OCTOPUS 900 Service Manual