Pump Rotor Cleaning Update
English – EN The Pump rotor Cleaning Addendum provides updated cleaning instructions for the:
ACP®215 device
CCS device
MCS®+ 8150 device
MCS®+ 9000 device
PCS®2 device
Please read all information in this addendum before operating the devices. Use this addendum in conjunction with the associated product user manuals.
Routine Pump Rotor Cleaning
Caution: Never use organic solvents (e.g., acetone), acids, bases, chlorine solutions, phenolics, or soap to clean any part of the pump rotor. Caution: Do not use oil, grease, or any material that leaves a deposit when servicing the pump rotor assembly. Caution: Do not immerse the pump rotor assembly in warm/hot water or place directly under a spray of warm/hot water. For routine cleaning, the user should: 1. Unplug the device before performing any cleaning. 2. Visually inspect the pump rotor and remove any residue on a routine basis. Do not clean the pump rotor if there is any visible damage or deterioration, and contact your local Haemonetics service representative. 3. Remove the pump rotor from the housing, using the hexagonal head wrench to remove the pump screw. 4. Immerse a lint-free cloth or gauze sponge in clean, cold, fresh water. Squeeze out excess water so the cloth or sponge is not dripping. Wipe or scrub the rotor to remove any residue from the rollers. Repeat as needed to ensure all visible residue is removed. 5. Dry with a lint-free cloth or gauze sponge (or compressed air, if available). 6. Clean and dry the pump well using the same method. 7. Ensure all of the rollers spin freely and place the pump rotor back into the pump well, aligning the cross pin in the rotor with the pump shaft. 8. Place the pump screw back onto the rotor and tighten with the hexagonal head wrench.
108416-MULTI(AA)
Haemonetics® User Manual Addendum
Pump Rotor Cleaning Update
Disinfecting the Pump Rotor
4
In the event of a fluid spill, the same routine cleaning method should be followed; however, only cleaning solutions containing the following active ingredients should be used: Acceptable Cleaning Solutions Active Ingredient
Example of Trade/Common Name
Quaternary ammonium
Coverage Plus NPD®, Super Sani-Cloth® Wipes
Benzyl ammonium chloride
Med-Chem Pink Germicide, CaviCide™
Glutaraldehyde
MetriCide™
Potassium Peroxymonosulfate
Virkon® S
Cleaning solutions that contain phenolic compounds, such as Sporicidin® and Supergard®, should not be used. In addition, 10% chlorine solutions and very warm water should not be used, as these solutions will damage the pump rotors. Coverage Plus NPD, Super Sani-Cloth, CaviCide, MetriCide, Virkon S, Sporicidin, and Supergard are trademarks or registered trademarks of their respective owners.